K033351 · Global Orthopaedic Technology, USA, Inc. · HRY · Mar 11, 2004 · Orthopedic
Device Facts
Record ID
K033351
Device Name
GLOBAL RESURFACING UNICOMPARTMENTAL (GRU) KNEE
Applicant
Global Orthopaedic Technology, USA, Inc.
Product Code
HRY · Orthopedic
Decision Date
Mar 11, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3530
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Global Resurfacing Unicompartmental (GRU) Knee System is intended for partial replacement of the articulating surfaces of the knee when only one side of the joint is affected due to compartmental primary degenerative or post-traumatic degenerative disease, previous tibial conclyle or plateau fractures, deformity or revision of previous arthroplasty. The device is a single-use implant that is intended for use with bone cement.
Device Story
Global Resurfacing Unicompartmental (GRU) Knee System is a semi-constrained, cemented, metal/polymer knee prosthesis. System comprises anatomic femoral component (cobalt chrome alloy) and semi-lunar tibial component (UHMWPE). Femoral component features central keel and post for cement fixation and rotational stability; requires left/right specific geometries. Tibial component features universal geometry. Device is single-use implant for surgical replacement of articulating knee surfaces. Intended for use by orthopedic surgeons in clinical settings to address compartmental degenerative disease or trauma. Benefits include restoration of joint function and reduction of pain associated with localized knee joint pathology.
Indicated for patients requiring partial knee replacement due to compartmental primary degenerative disease, post-traumatic degenerative disease, previous tibial condyle or plateau fractures, joint deformity, or revision of previous arthroplasty where only one side of the joint is affected.
Regulatory Classification
Identification
A knee joint femorotibial metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that consist of a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Related Devices
K161741 — GMK UNI · Medacta International S.A. · Apr 14, 2017
K020741 — UNICONDYLAR KNEE · Encore Orthopedics, Inc. · Jun 4, 2002
K050764 — UNIGLIDE UNICONDYLAR KNEE PROSTHESIS · Corin USA · Aug 24, 2005
K982859 — UC-PLUS SOLUTION UNICONDYLAR KNEE SYSTEM · Plus Orthopedics · Jul 30, 1999
Submission Summary (Full Text)
{0}------------------------------------------------
K03351
MAR 1 1 2004
## SUMMARY OF SAFETY AND EFFECTIVENESS
- Global Orthopaedic Technology USA, Inc. Sponsor: 5349 Fled Leaf Court Oviedo, Florida 32765
Global Resurfacing Unicompartmental Knee System Device:
Classification Name: Knee Joint, Femorotibial, Metal/Polymer, Semi-constrained, Cemented Prosthesis (21 CFR 888.3530)
Intended Use: Partial replacement of the articulating surfaces of the knee when only one side of the joint is affected due to compartmental primary degenerative or post-traumatic degenerative disease, previous tibial condyle or plateau fractures, deformity or revision of previous arthroplasty. The device is a single-use implant intended for implantation with bone cement.
Device Description: The Global Resurfacing Unicompartmental Knee System consists of femoral and tibial components.
The femoral component is anatomic in design, to provide coverage of the condyle from posterior to anterior. The anatomic shape of the femoral component necessitates separate left and right geometries. A central keel and post on the back of the femoral component assists in cement fixation and rotational stability. The device is manufactured from cobalt chrome alloy that conforms to ASTM F-75-01.
The tibial component is semi-lunar in configuration and manufactured from compression molded ultra-high molecular weight polyethylene (UHMWPE) that conforms to ASTM F-648-00. The tibial component is universal in geometry.
Potential Risks: The potential risks associated with this device are the same as with any joint replacement device. These include, but are not limited to:
Reaction to bone cernent Deformity of the joint Cardiovascular disorders Fracture of bone cement Implant loosening/migration Nerve damage
Blood vessel damage Soft tissue imbalance Delayed wound healing Metal sensitivity Fracture of the components Bone fracture Infection Hematoma Dislocation Excessive wear
Page 1 of 1
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 1 2004
Mr. Carl Knobloch Chief Operating Officer Global Orthopaedic Technology, USA, Inc. 5349 Red Leaf Court Oviedo, Florida 32765
Re: K033351
Trade/Device Name: Global Resurfacing Unicompartmental (GRU) Knee System Regulation Number: 21 CFR 888.3530 Regulation Name: Knee joint femorotibial metal/polymer semi-constrained cemented prosthesis Regulatory Class: II Product Code: HRY Dated: February 16, 2004 Received: February 17, 2004
Dear Mr. Knobloch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 - Mr. Carl Knobloch
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Mark N. Millman
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Confidential
## Indications for Use
510(k) Number (if known): K033351
Device Name: Global Resurfacing Unicompartmental (GRU) Knee System
Indications For Use: The Global Resurfacing Unicompartmental (GRU) Knee System is intended for partial replacement of the articulating surfaces of the knee when only one side of the joint is affected due to compartmental primary degenerative or post-traumatic degenerative disease, previous tibial conclyle or plateau fractures, deformity or revision of previous arthroplasty. The device is a single-use implant that is intended for use with bone cement.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Do
Mullins
Division of General, Restorative, and Neurological Devices
**510(k) Number** K033351
Page 1 of _1_
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.