The GMK® UNI knee prosthesis is designed for cemented use in partial knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components. Partial replacement of the articulating surfaces of the knee is indicated when only one side of the joint is affected due to the compartmental primary degenerative or post-traumatic degenerative disease, previous tibial condyle or plateau fractures, deformity or revision of previous arthroplasty.
Device Story
GMK UNI is a unicompartmental knee prosthesis for partial knee arthroplasty. System consists of a cemented cobalt-chromium-molybdenum (Co-Cr-Mo) femoral component, a fixed ultra-high molecular weight polyethylene (UHMWPE) tibial insert, and a metal-backed Co-Cr-Mo tibial tray. Implants are provided sterile for single use. Device is used by orthopedic surgeons in a clinical/surgical setting to replace articulating surfaces of the knee joint. By providing a stable, cemented interface, the device restores joint function and addresses degenerative disease or post-traumatic deformity. Clinical benefit is derived from the mechanical replacement of damaged joint surfaces, facilitating improved mobility and pain relief for the patient.
Clinical Evidence
Bench testing only. Studies included fatigue testing under dynamic physiological loads (ASTM F1800), excessive wear testing (ASTM F2083), sterilization validation (ANSI/AAMI/ISO 11137-1:2006, 11137-2:2006, ISO 11135-1:2007), accelerated and real-time aging, and packaging validation.
Technological Characteristics
Materials: Co-Cr-Mo (ISO 5832-4) for femoral/tibial tray; UHMWPE (ISO 5834-2 Type 1) for insert/plug. Design: Fixed-bearing, cemented, unicompartmental knee prosthesis. Stabilization: Femoral component (2 pegs); Tibial tray (peg and keel). Sterilization: Gamma irradiation or ethylene oxide. Connectivity: None (mechanical implant).
Indications for Use
Indicated for patients requiring partial knee arthroplasty due to compartmental primary degenerative or post-traumatic degenerative disease, previous tibial condyle or plateau fractures, deformity, or revision of previous arthroplasty, provided there is sufficient sound bone to seat and support components.
Regulatory Classification
Identification
A knee joint femorotibial metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral condylar resurfacing component or components made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components made of ultra-high molecular weight polyethylene and are intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight, composed of three overlapping profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 14, 2017
Medacta International SA Stefano Baj Regulatory Affairs Manager Strada Regina 6874 Castel San Pietro (CH) Switzerland
Re: K161741 Trade/Device Name: GMK UNI Regulation Number: 21 CFR 888.3520 Regulation Name: Knee Joint Femorotibial Metal/Polymer Non-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: HSX Dated: March 13, 2017 Received: March 14, 2017
Dear Stefano Baj:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K161741
Device Name GMK(R) UNI
#### Indications for Use (Describe)
The GMK(R) UNI knee prosthesis is designed for cemented use in partial knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components. Partial replacement of the articulating surfaces of the knee is indicated when only one side of the joint is affected due to the compartmental primary degenerative or post-traumatic degenerative disease, previous tibial condyle or plateau fractures, deformity or revision of previous arthroplasty.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Medacta International. The logo consists of two blue triangles, followed by the word "Medacta" in blue. Underneath the word "Medacta" is a yellow line, and below that is the word "International" in blue, followed by a red and white Swiss flag.
#### 3 510(k) Summary
- Applicant/Sponsor: Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
- Stefano Baj Contact Person: Regulatory Affairs Manager Phone: +41 91 696 60 60 Fax: +41 91 696 60 66 Email: baj@medacta.ch
- Date Prepared: June 17, 2016
# DEVICE INFORMATION
| Trade/Proprietary Name: | GMK® UNI |
|-------------------------|-------------------------------------------------------------------------------|
| Common or Usual Name: | Unicompartmental Knee prosthesis |
| Classification Name: | Knee joint femorotibial metal/polymer/ non-constrained cemented<br>prosthesis |
| Product Code: | HSX |
| Regulation Number: | 21 CFR 888.3520 |
| Device Class: | II |
## PREDICATE DEVICE INFORMATION
## Primary predicates:
| 510(k) | Product | 510(k) Holder |
|---------|---------|----------------|
| K073175 | Journey | Smith & Nephew |
## Additional predicates:
| 510(k) | Product | 510(k) Holder |
|---------|---------|----------------|
| K102069 | Journey | Smith & Nephew |
| K033363 | UNI | Zimmer |
{4}------------------------------------------------
# DEVICE DESCRIPTION
The GMK® UNI is a unicompartmental knee prosthesis. The GMK® UNI is comprised of individually packaged femur and tibial components designed for cemented use in partial knee arthroplasty if there is evidence of sufficient sound bone to seat and support the components. The femoral component is made of cobalt-chromium-molybdenum (Co-Cr-Mo per ISO 5832-4), and the tibial component consists of an ultra-high molecular weight polyethylene (UHMWPE per ISO 5834-2 Type 1) insert, and a metal-backed tray component (Co-Cr-Mo per ISO 5832-4). The GMK® UNI system implants are offered sterile (via gamma irradiation or ethylene oxide), are intended for single use only, and may not be re-sterilized.
The Anatomic Femoral Component (cemented) is anatomically shaped and designed with two (2) fixation pegs. Available in six (6) sizes (0 - 5), the femoral component is offered in both Right Medial/Left Lateral (RM-LL) and Left Medial/Right Lateral (LM-RL) options for each size.
The Fixed Tibial Insert has a fixed design and is available in five (5) sizes (1 – 5). Each size is offered in six (6) levels of thickness (8, 9, 10, 11, 12, 14 mm).
The Fixed Plus Tibial Tray (cemented) has a fixed bearing design with one fixation keel and peg to ensure primary stability, and an UHMWPE plug. Available in five (5) sizes (1 – 5), the Fixed Plus Tibial Trays are offered in both Right Medial/Left Lateral (RM-LL) and Left Medial/Right Lateral (LM-RL) options for each size.
# INDICATIONS FOR USE
The GMK® UNI knee prosthesis is designed for cemented use in partial knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components. Partial replacement of the articulating surfaces of the knee is indicated when only one side of the joint is affected due to the compartmental primary degenerative or post-traumatic degenerative disease, previous tibial condyle or plateau fractures, deformity or revision of previous arthroplasty.
## Discussion:
The Indications for Use Statement is similar to the predicate device. The only difference is that the subject device calls out the requirements for "evidence of sufficient sound bone to seat and support the components", and the predicate device cites specific example of degenerative joint disease ("osteoarthritis, traumatic arthritis, or avascular necrosis"). These differences do not create new intended uses for the subject device; the subject and predicate devices are all prostheses intended for use in knee arthroplasty.
{5}------------------------------------------------
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
The design features, and materials of the subject devices are substantially equivalent to those of the predicate devices. The tables below compare characteristics of the subject and predicate devices.
| Feature | GMK UNI - Anatomic<br>Femoral Component<br>(cemented) | Smith & Nephew<br>Journey<br>(K073175) | Zimmer UNI<br>(K033363) |
|-----------------------|-----------------------------------------------------------------|---------------------------------------------------------|-------------------------|
| Sizes | 6 sizes (Size 0 - 5) | 7 sizes (Size 1 -7) | 7 sizes (Size A - G) |
| Configuration | RM-LL and LM-RL | Same | Same |
| Material | CoCrMo - ISO 5832-4 | Same<br>(with an oxinium oxidized<br>zirconium coating) | Same |
| Cemented? | Cemented | Same | Same |
| Device<br>Usage | Single Use | Same | Same |
| Sterility | Provided sterile via Gamma<br>Radiation | Same | Same |
| Shelf Life | 5 Years | Same | > 5 Years |
| Stabilization<br>pegs | Anatomic Femoral<br>component is cemented<br>with two (2) pegs. | Same | Same |
| Feature | GMK UNI - Fixed Tibial Insert | Zimmer UNI<br>(K033363) |
|-----------------|----------------------------------------------------------------------------|-------------------------|
| Sizes | 5 sizes (from 1 to 5) | 6 sizes (from 1 to 6) |
| Thickness | Each size is offered in 6 levels of thickness<br>(8, 9, 10, 11, 12, 14 mm) | Same |
| Material | UHMWPE (ISO 5834 -2) Type 1 | Same |
| Device<br>Usage | Single Use | Same |
| Sterility | Provided sterile via Ethylene Oxide | Same |
| Shelf Life | 5 years | > 5 years |
{6}------------------------------------------------
| Feature | GMK UNI - Fixed Plus<br>Tibial Tray (cemented) | Smith & Nephew Journey<br>(K102069) | Zimmer UNI<br>(K033363) |
|-------------------|-------------------------------------------------------------|-------------------------------------|--------------------------------|
| Sizes | 5 sizes (from 1 to 5) | 6 sizes | 6 sizes (from 1 to 6) |
| Configurations | RM-LL and LM-RL | Same | Same |
| Material | CoCrMo (ISO 5832-4)<br>Plug: UHMWPE (ISO<br>5834 -2) Type 1 | Same | Ti-6Al-4V with PMMA<br>precoat |
| Cemented? | Cemented | Same | Same |
| Device Usage | Single Use | Same | Same |
| Sterility | Provided sterile via<br>Gamma Radiation | Unknown | Same |
| Shelf Life | 5 years | Unknown | > 5 years |
| Stabilization | Stabilized with peg and<br>keel | Same | Same |
| Fixed<br>Bearing? | Fixed bearing | Same | Same |
The fundamental scientific technology of the modified devices has not changed relative to the predicate devices. The safety and effectiveness of the subject devices is adequately supported by the substantial equivalence information, materials information, and analysis data provided within this Premarket Notification.
# PERFORMANCE TESTING
The following studies were performed to support substantial equivalence:
- Fatigue under dynamic physiological loads according to ASTM F1800 ●
- Excessive wear according to ASTM F2083 ●
- Sterilization validation in accordance with ANSI/AAMI/ISO 11137-1:2006, . ANSI/AAMI/ISO 11137-2:2006 and ISO 11135-1:2007
- · Accelerated and real time aging
- Packaging validation
がWWWWWWwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwww
Á½½½½½½½Û´ ![ * ^} ã㺠Áº· çã * ÁçãÁ·Â¡ â[ çã Áº ç¹ º ÁÇ Á£ Áº ǹ º ÍÇ· Íߣ ÍGÁ
# //////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////
AWWWBased on the above information, the GMK® UNI can be considered substantially equivalent AWWW the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.