SYNTHES DORSAL DISTAL RADIUS PLATE (DDRP) SYSTEM
Device Facts
| Record ID | K962616 |
|---|---|
| Device Name | SYNTHES DORSAL DISTAL RADIUS PLATE (DDRP) SYSTEM |
| Applicant | Synthes (Usa) |
| Product Code | HRS · Orthopedic |
| Decision Date | Sep 3, 1996 |
| Decision | SN |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3030 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
Synthes Dorsal Distal Radius Plate is intended for fixation of fractures, osteotomies and carpal fusions involving the distal radius, applied to the dorsal aspect.
Device Story
Synthes Dorsal Distal Radius Plate; PI-shaped titanium alloy plate; precontoured for dorsal distal radius; features two proximal shaft arms and one distal head arm; proximal arms include four compression holes and one round hole for 2.7 mm cortex screws; distal head includes seven internally threaded holes for 1.8 mm threaded head buttress pins or 2.4 mm cortex screws; plate can be cut to size; available in left and right versions; used by orthopedic surgeons for internal fixation of distal radius fractures, osteotomies, and carpal fusions.
Clinical Evidence
Bench testing only; mechanical test results demonstrate equivalence to predicate device.
Technological Characteristics
Material: Titanium alloy. Design: PI-shaped, precontoured, left/right versions. Fixation: 2.7 mm cortex screws (proximal), 1.8 mm threaded head buttress pins or 2.4 mm cortex screws (distal).
Indications for Use
Indicated for patients requiring fixation of fractures, osteotomies, and carpal fusions of the distal radius.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
- Zimmer Small T-Plate, Oblique