K213005 · Glw Medical, Inc. · HRS · Mar 11, 2022 · Orthopedic
Device Facts
Record ID
K213005
Device Name
Apollo Ankle Fracture Plating System
Applicant
Glw Medical, Inc.
Product Code
HRS · Orthopedic
Decision Date
Mar 11, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Apollo™ Ankle Fracture Plating System is intended for fixation of fractures, osteotomies, and non-unions of the distal tibia and fibula such as: - Lateral Malleolar Fractures - Syndesmosis Injuries - Medial Malleolar Fractures - Bi-Malleolar Fractures - Tri-Malleolar Fractures - Posterior Malleolar Fractures - Distal Anterior Tibia Fractures - Vertical Shear Fractures of the Medial Malleolus - Pilon Fractures - Distal Tibia Shaft Fractures - Distal Fibula Shaft Fractures - Distal Tibia Periarticular Fractures - Medial Malleolar Avulsion Fractures - Lateral Malleolar Avulsion Fractures Apollo Locking Screws are intended for use with Apollo's Plating Systems. Apollo non-Locking Screws are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device. Apollo washer is intended to prevent a screw head from breaking through the cortex of the bone by distributing the forces/ load over a large area when used for fracture fixation of bone fragments. Apollo 1/3 tubular plates are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repar, and fracture fixation, appropriate for the size of the device.
Device Story
Apollo Ankle Fracture Plating System consists of titanium alloy (Ti-6AL-4V ELI) and PEEK plates, locking/non-locking screws, and washers. Used by orthopedic surgeons for internal fixation of distal tibia/fibula fractures and osteotomies. Implants provide mechanical stabilization of bone fragments to facilitate healing. Instrumentation included for delivery. Provided sterile via Gamma irradiation. Output is physical stabilization of bone; clinical decision-making based on radiographic assessment and surgical need for fracture reduction.
Clinical Evidence
Bench testing only. Plate components tested per ASTM F382. Screws tested for torsional strength, driving torque, and axial pullout per ASTM F543 and FDA Guidance for Bone Screws and Washers. Static and dynamic 3-point bending performed per ASTM F1264. Results demonstrate performance at least equivalent to predicate devices.
Indicated for fixation of fractures, osteotomies, and non-unions of the distal tibia and fibula, including malleolar fractures, syndesmosis injuries, pilon fractures, and avulsion fractures. Also indicated for bone reconstruction, arthrodesis, and joint fusion.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Wright Medical Ortholoc 3Di Ankle Fracture System (K1603044)
{0}------------------------------------------------
March 11, 2022
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. The FDA logo is on the right, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The logos are placed side by side.
GLW Medical Inc % Cheryl Wagoner Consultant Wagoner Consulting LLC 5215 Crosswinds Drive Wilmington, North Carolina 28409
Re: K213005
Trade/Device Name: Apollo Ankle Fracture Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: February 8, 2022 Received: February 9, 2022
Dear Cheryl Wagoner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For: Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K213005
Device Name Apollo Ankle Fracture Plating System
Indications for Use (Describe)
Apollo™ Ankle Fracture Plating System is intended for fixation of fractures, osteotomies, and non-unions of the distal tibia and fibula such as:
- · Lateral Malleolar Fractures
- Syndesmosis Injuries
- · Medial Malleolar Fractures
- Bi-Malleolar Fractures
- Tri-Malleolar Fractures
- Posterior Malleolar Fractures
- Distal Anterior Tibia Fractures
- · Vertical Shear Fractures of the Medial Malleolus
- Pilon Fractures
- Distal Tibia Shaft Fractures
- Distal Fibula Shaft Fractures
- Distal Tibia Periarticular Fractures
- Medial Malleolar Avulsion Fractures
- · Lateral Malleolar Avulsion Fractures
Apollo Locking Screws are intended for use with Apollo's Plating Systems.
Apollo non-Locking Screws are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.
Apollo washer is intended to prevent a screw head from breaking through the cortex of the bone by distributing the forces/ load over a large area when used for fracture fixation of bone fragments.
Apollo 1/3 tubular plates are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repar, and fracture fixation, appropriate for the size of the device.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## Traditional 510(k) Premarket Notification Apollo™ Ankle Fracture Plating System
## 510(k) Summary (as required by 21 CFR 807.92)
| Date Prepared | September 17, 2021 |
|----------------|---------------------------------------------|
| Manufacturer | GLW, Inc. |
| Address | 300 Sylvan Ave<br>Englewood Cliff, NJ 07632 |
| Telephone | 917-794-2583 |
| Contact Person | Arundhati Radhakrishnan |
| Address | 300 Sylvan Ave<br>Englewood Cliff, NJ 07632 |
| Telephone | 201-268-3281 |
| Email | Arundhati.radhakrishnan@glwmed.com |
| Trade Name | Apollo™ Ankle Fracture Plating System |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Plate, Fixation, Bone<br>Screw, Fixation, Bone |
| Panel Code | Orthopaedics/87 |
| Classification Name | Single/multiple component metallic bone fixation appliances<br>and accessories.<br>Smooth Or Threaded Metallic Bone Fixation Fastener |
| Class | Class II |
| Regulation Number | 21 CFR 888.3030<br>21CFR 888.3040 |
| Product Code | HRS<br>HWC |
| Name of Primary Predicate<br>Device | 510(k) # | Manufacturer |
|-------------------------------------|----------|--------------|
| | | |
| Name of Reference<br>Device(s) | 510(k) # | Manufacturer |
| | | |
| CREED™ Cannulated<br>Screws | K200291 | GLW, Inc |
| Description | ApolloTM Ankle Fracture Plating System consists of<br>implantable components that will be include an array of<br>Titanium alloy Ti-6AL-4V ELI (ASTM F3001) / PEEK plates<br>and locking and non-locking screws.<br><br>The screws are offered in configurations that include a<br>range of Titanium alloy Ti-6AL-4V ELI (ASTM F136) screws.<br>A variety of instrumentation is offered as part of the kit to |
|-------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{5}------------------------------------------------
# Traditional 510(k) Premarket Notification
Apollo™ Ankle Fracture Plating System
| | facilitate delivery of the implants. The implantable devices<br>are provided sterile via Gamma irradiation. |
|--|-------------------------------------------------------------------------------------------------------------|
|--|-------------------------------------------------------------------------------------------------------------|
| Indications and<br>Intended Use | Apollo™ Ankle Fracture Plating System is intended for<br>fixation of fractures, osteotomies, and non-unions of the<br>distal tibia and fibula such as:<br>Lateral Malleolar Fractures Syndesmosis Injuries Medial Malleolar Fractures Bi-Malleolar Fractures Tri-Malleolar Fractures Posterior Malleolar Fractures Distal Anterior Tibia Fractures Vertical Shear Fractures of the Medial Malleolus Pilon Fractures Distal Tibia Shaft Fractures Distal Fibula Shaft Fractures Distal Tibia Periarticular Fractures Medial Malleolar Avulsion Fractures Lateral Malleolar Avulsion Fractures Apollo Locking Screws are intended for use with Apollo's<br>Plating Systems.<br><br>Apollo non-Locking Screws are indicated for use in bone<br>reconstruction, osteotomy, arthrodesis, joint fusion, fracture<br>repair, and fracture fixation, appropriate for the size of the<br>device.<br><br>Apollo washer is intended to prevent a screw head from<br>breaking through the cortex of the bone by distributing the<br>forces/load over a large area when used for fracture fixation<br>of bone fragments.<br><br>Apollo 1/3 tubular plates are indicated for use in bone<br>reconstruction, osteotomy, arthrodesis, joint fusion, fracture<br>repair, and fracture fixation, appropriate for the size of the<br>device. |
|--------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristics<br>and Substantial<br>Equivalence | Documentation was provided to demonstrate that the<br>Subject device is substantially equivalent to the primary<br>predicate Wright Medical Ortholoc 3Di Ankle Fracture<br>System (K1603044). The Subject device is substantially<br>equivalent to the predicate devices in intended use,<br>indications for use, materials, technological characteristics,<br>and labeling. |
{6}------------------------------------------------
г
| Performance<br>Data | The plate components were tested via ASTM F382 and were<br>shown to be at least equivalent to the predicate devices. |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Torsional strength, driving torque and axial pullout testing (per<br>ASTM F543 and FDA Guidance for Bone Screws and<br>Washers, December 2020) confirmed that the Subject device<br>screws performed as intended and are at least equivalent to<br>the predicate devices. Static 3-point bending, and dynamic 3-<br>point bending per ASTM F1264 further confirmed the<br>performance and substantial equivalence of the Subject<br>device screws. |
| Conclusion | Based on the intended use, indications for use, technological<br>characteristics, materials, and comparison to predicate<br>reference devices, the Subject device has been shown to be<br>substantially equivalent to legally marketed predicate<br>devices. |
|------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.