OrthoPediatrics Locking Proximal Femur System Cortical Screws
Applicant
OrthoPediatrics Corp.
Product Code
HRS · Orthopedic
Decision Date
Sep 27, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Intended Use
The OrthoPediatrics Locking Proximal Femur System Cortical Screws are intended for temporary internal fixation and stabilization of long bone fractures and osteotomies, mal-unions, and non-unions, in pediatric and small stature adults. Specific indications include: intertrochanteric derotation and Varus osteotomies, femoral neck and pertrochanteric fractures, and intertrochanteric valgus osteotomies.
Device Story
Device consists of metallic cortical screws designed as a line extension for the OrthoPediatrics Locking Proximal Femur System. Screws provide temporary internal fixation and stabilization for long bone fractures and osteotomies in pediatric and small stature adult patients. Operated by surgeons in clinical settings; screws are implanted to provide immediate stability during the natural bone healing process. Output is mechanical fixation of bone segments. Benefits include stabilization of fractures and osteotomies to facilitate healing.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical engineering analysis.
Indicated for temporary internal fixation and stabilization of long bone fractures, osteotomies, mal-unions, and non-unions in pediatric and small stature adult patients. Specific applications include intertrochanteric derotation, Varus/valgus osteotomies, femoral neck fractures, and pertrochanteric fractures.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
OrthoPediatrics Locking Proximal Femur System (K111086)
Related Devices
K083286 — ORTHOPEDIATRICS PEDILOC LOCKING PLATE SYSTEM · OrthoPediatrics Corp. · Jan 30, 2009
K011603 — VHS PEDIATRIC HIP SCREW SYSTEM · Walter Abendschein, M.D. · Aug 15, 2001
K180736 — WishBone Medical Plate and Screw System · Wishbone Medical, Inc. · Oct 4, 2018
K111086 — ORTHOPEDIATRICS FRACTURE AND OSTEOTOMY BONE PLATE SYSTEM · OrthoPediatrics Corp. · Jun 17, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, stacked on top of each other. The profiles are rendered in a simple, abstract manner.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 27, 2016
OrthoPediatrics, Corp. Mr. Adam Cargill Regulatory Affairs Associate Manager 2850 Frontier Drive Warsaw, Indiana 46582
Re: K162307
Trade/Device Name: OrthoPediatrics Locking Proximal Femur System Cortical Screws Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories. Regulatory Class: Class II Product Code: HRS, HWC Dated: September 1, 2016 Received: September 2, 2016
Dear Mr. Cargill:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K162307
#### Device Name
OrthoPediatrics Locking Proximal Femur System Cortical Screws
Indications for Use (Describe)
The OrthoPediatrics Locking Proximal Femur System Cortical Screws are intended for temporary internal fixation and stabilization of long bone fractures and osteotomies, and non-unions, in pediatric and small stature adults. Specific indications include: intertrochanteric derotation and Varus osteotomies, femoral neck and pertrochanteric fractures, and intertrochanteric valgus osteotomies.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Locking Proximal Femur System Screws Special 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on August 12, 2005.
| Sponsor: | OrthoPediatrics, Corp.<br>2850 Frontier Drive<br>Warsaw, IN 46582<br>Establishment Registration Number: 9102640<br>Phone: (574) 267-6379<br>Fax: (574) 269-3692 |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Adam Cargill<br>Regulatory Affairs Associate Manager |
| Date: | September 26, 2016 |
| Subject Device: | Trade Name: OrthoPediatrics Locking Proximal Femur System Cortical Screws |
| | Regulation Number: 888.3030<br>Regulation Name: Single/multiple component metallic bone fixation appliance<br>and accessories<br>Product Code: HRS |
| | Regulation Number: 888.3040<br>Regulation Name: Smooth or threaded metallic bone fixation fastener<br>Product Code: HWC |
| | Common Name(s): Plate, Fixation, Bone<br>Screw, Fixation, Bone |
### Legally marketed devices to which substantial equivalence is claimed:
- K111086 – OrthoPediatrics Locking Proximal Femur System
# Device Description
The OrthoPediatrics Locking Proximal Femur System Cortical Screws are a line extension to expand the current offering to allow for longer cortical screws. The screws are part of one system that combines implants (plates and screws) and instruments. The additional cortical screw sizes include the following: 55mm, 60mm, 65mm, 70mm, 75mm, and 80mm.
- Materials: The cannulated screws are manufactured from medical grade BIODUR stainless steel ● which meet ASTM 2229-07 standard.
- . Function: The system, which includes the cortical screws, functions to provide immediate stability and temporary fixation during the natural healing process.
{4}------------------------------------------------
# Intended Use and Indications for Use
The OrthoPediatrics Locking Proximal Femur System Cortical Screws are intended for temporary internal fixation and stabilization of long bone fractures and osteotomies, mal-unions, and non-unions, in pediatric and small stature adults. Specific indications include: intertrochanteric derotation and Varus osteotomies, femoral neck and pertrochanteric fractures, and intertrochanteric valgus osteotomies.
# Summary of Technological Characteristics
The technological characteristics (materials, design, sizing, and indications) of the Locking Proximal Femur System Cortical Screws are similar to the predicate system's screws.
The rationale for substantial equivalence is based on consideration of the following characteristics:
Intended Use: The proposed Locking Proximal Femur System Cortical Screws are intended for temporary internal fixation and stabilization of long bone fractures and osteotomies, malunions, and non-unions, in pediatric and small stature adults which is the same as the predicate device.
Indications for Use: The proposed Locking Proximal Femur System Cortical Screws Indications for Use are identical to the predicate devices.
Materials: The proposed Locking Proximal Femur System Cortical Screws use the identical implant material as the predicate system's screws.
Design Features: The proposed Locking Proximal Femur System Cortical Screws incorporate similar design features as the predicate system's screws with the difference being the length of the screw and the change from cannulated to solid.
Sterilization: The proposed Locking Proximal Femur System Cortical Screws are provided nonsterile, which is identical to the predicate device.
### Summary of Performance Data (Nonclinical and/or Clinical)
- . Non-Clinical Tests
- Engineering analysis was performed to show substantial equivalence between the o subject device and predicate device.
- . Clinical Tests - None provided as a basis for substantial equivalence.
### Substantial Equivalence Conclusion
OrthoPediatrics believes that the Locking Proximal Femur System Cortical Screws are substantially equivalent to the legally marketed predical screws (K111086) based on the similarities of design, intended use, materials, and the results of verification activities conducted. No new risks have been identified and it is expected that the Locking Proximal Femur System Cortical Screws will perform substantially equivalent to the legally marketed predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.