The MacroPore OS Trauma System is intended to maintain the relative position of weak bony tissue such as bone graft substitutes, or bone fragments from comminuted fractures. The MacroPore OS Trauma System is also indicated for cement restriction in total joint arthroplasty procedures. Only when used in conjunction with traditional rigid fixation, the MacroPore OS Trauma System is intended to maintain the relative position weak bony tissue in trauma and reconstructive orthopedic procedures involving: - Long bones - Flat bones - Short bones - Irregular bones - Appendicular skeleton - Thorax When used alone (without traditional rigid fixation), the MacroPore OS Trauma System is intended to maintain the relative position of bone graft substitutes in reconstructive orthopedic procedures involving: - Tumor resections where bone strength has not been compromised - Iliac crest harvests - Ribs This device is not intended for use in the spine. The device is not intended for load bearing indications unless used in conjunction with traditional rigid fixation.
Device Story
Resorbable graft containment system; composed of porous/non-porous poly(L-lactide) (PLA) sheets, sleeves, and fixation screws. Sheets/sleeves malleable when heated to ~55°C in sterile water; cut to size by surgeon. Used in orthopedic trauma/reconstructive procedures; applied to bone fragments or graft sites. Provides containment/relative positioning of bone graft substitutes or fragments. Can be used alone or with traditional rigid fixation (plates/screws). Prevents dislocation of graft material. Benefits patient by maintaining graft/fragment position during healing. Operated by orthopedic surgeons in clinical/OR settings.
Clinical Evidence
Bench testing only. Aging studies confirmed rigidity and strength equivalent to predicate after 6 months. Mechanical testing verified tensile strength and spring constant. DSC analysis confirmed amorphous, non-crystalline material structure. Thermal stability testing in 60°C saline confirmed viscosity remains within range for surgical shaping.
Technological Characteristics
Material: Poly(L-lactide) (PLA) 70:30. Form: Porous/non-porous sheets (20x20mm to 120x120mm) and sleeves (150mm length, 10-25mm diameter). Thickness: 0.75-3.0mm. Pore size: 500-3000 microns. Principle: Malleable thermoplastic; heat-activated (55°C) for anatomical contouring. Fixation: PLA screws (2.0-4.8mm diameter). Sterilization: Not specified.
Indications for Use
Indicated for reinforcing weak bony tissues in orthopedic procedures, including comminuted fracture fragments and bone graft substitutes. Used with rigid fixation for long, flat, short, irregular, appendicular, and thoracic bones. Used alone for tumor resections, iliac crest harvests, and ribs. Contraindicated for spine use and load-bearing applications without rigid fixation.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
KO2//64
#### MacroPore OS Trauma System 510(k) SUMMARY
Page 1 of 4
# ADMINISTRATIVE INFORMATION
Manufacturer Name:
Official Contact:
MacroPore, Inc. 6740 Top Gun Street San Diego, CA 92121
Kenneth K. Kleinhenz Director of Regulatory Affairs Telephone (858) 458-0900 Fax (858) 458-0994
DEVICE NAME
Classification Name: Trade/Proprietary Name: Plate, Fixation, Bone MacroPore OS Trauma
### ESTABLISHMENT REGISTRATION NUMBER 2031733
# DEVICE CLASSIFICATION AND PRODUCT CODE
As shown in 21CFR 888.3030 bone fixation appliances intended for use in orthopedic procedures are classified as Class II. They have been assigned Product Code HRS.
{1}------------------------------------------------
#### INTENDED USE
The MacroPore OS Trauma System is intended to maintain the relative position of weak bony tissue such as bone graft substitutes, or bone fragments from comminuted fractures. The MacroPore OS Trauma System is also indicated for cement restriction in total joint arthroplasty procedures.
Only when used in conjunction with traditional rigid fixation, the MacroPore OS Trauma System is intended to maintain the relative position weak bony tissue in trauma and reconstructive orthopedic procedures involving:
- Long bones
- Flat bones
- · Short bones
- · Irregular bones
- · Appendicular skeleton
- Thorax
When used alone (without traditional rigid fixation), the MacroPore OS Trauma System is intended to maintain the relative position of bone graft substitutes in reconstructive orthopedic procedures involving:
- · Tumor resections where bone strength has not been compromised
- Iliac crest harvests
- Ribs
This device is not intended for use in the spine. The device is not intended for load bearing indications unless used in conjunction with traditional rigid fixation.
#### Design Characteristics
MacroPore OS Trauma System is a resorbable graft containment system composed of various sized porous sheet and sleeves, non-porous sheets and associated fixation screws manufactured from poly (L-lactide) 70:30 (PLA). The MacroPore OS Trauma System is composed of MacroPore OS Trauma Protective Sheets, MacroPore OS Trauma Protective Sleeves, and MacroPore OS Trauma Screws.
The MacroPore OS Trauma Protective Sheets and Protective Sleeves can be cut with scissors to the desired shape and size. The MacroPore Power Pen can also be used to cut or shape the MacroPore OS Trauma to the desired shape or size. MacroPore OS Trauma Protective Sheets and Protective Sleeves are fully malleable when heated to approximately 55℃ (for example, by the use of sterile hot water), and thus can be conformed three dimensionally to most any anatomical orientation. The MacroPore Power Pen can also be used to cut or shape the MacroPore OS Trauma Protective Sheets and Protective Sleeves to the desired shape or size.
The MacroPore OS Trauma Protective Sheets can be rolled into a tube or used as a flat sheet. The MacroPore OS Trauma Protective Sheets and Protective Sleeves can be used either alone or in conjunction with internal bone fixation devices such as plates and screws, which also can serve to fixate the MacroPore OS Trauma and prevent dislocation.
{2}------------------------------------------------
#### 510(k) SUMMARY
The MacroPore OS Trauma Screws range in size from 2.0mm to 4.8mm in diameter. The MacroPore OS Trauma Protective Sheet is provided in sheets of 20 x 20 mm to 120 x 120 mm and will be provided in other sizes as needed for particular surgical procedures. The MacroPore OS Trauma Protective Sleeves are provided in lengths of 150mm to 5mm with diameters that range from 10mm up to 25mm. The MacroPore OS Trauma Protective Sheets and Protective Sleeves are provided with and without macroporous holes. The pore size ranges from 500 microns to 3000 microns in diameter, with pores distributed randomly or uniformly throughout the sheet/sleeve in an offset or aligned pattern. The thickness of the MacroPore OS Trauma Protective Sheets and Protective Sleeves ranges from 0.75 mm to 3.0 mm according to the orthopedic region to be treated.
#### Material Composition
The MacroPoreOS Protective Sheet is fabricated from polylactic acid (PLA).
### In Vitro Testing
Because the MacroPore OS Trauma Protective Sheet is intended to be heated in the surgical suite to temperatures above the material's glass transition temperature to facilitate shaping to anatomic structures, testing was performed to determine the effect of prolonged heating in saline at 60°C on inherent viscosity. The testing demonstrates that viscosity stayed within an appropriate range over 120 minutes. Therefore, the relatively brief exposure anticipated during the surgical preparation of MacroPore OS Protective Sheet is not expected to have a significant effect on its mechanical properties.
Aging studies were performed on MacroPore OS Trauma System components. Testing demonstrated that the MacroPore OS Trauma Protective Sheet is as rigid and as strong as the predicate after 6 month of exposure. Mechanical testing was performed on the MacroPore OS Trauma Protective Sheets and MacroPore OS Trauma Screws. Testing determined the MacroPore OS Trauma System to be substantially equivalent to the mechanical strengths of the predicate devices under indication for use conditions.
Crystallinity was tested for by DSC (differential scanning calorimetery). This test measures the amount of heat energy that is absorbed by a material. A crystalline material will require more energy once it reaches it's melting point. This release of heat energy can be seen on a graph as a sharp spike and is referred to as a "melting endotherm". The tests ran on the sterile and nonsterile samples revealed no endothermic spikes, indicating that the implants are amorphous and non-crystalline.
{3}------------------------------------------------
#### EQUIVALENCE TO MARKETED PRODUCT
The MacroPore OS Trauma System shares materials, indications, and design principles with the following predicate devices which have been determined by FDA to be substantially equivalent to pre-amendment devices: MacroPore OS Protective Sheet, Synthes Resorbable Meshes and Sheets, and the MacroPore OS Protective Sheet.
#### Indications For Use
The MacroPore OS Trauma System shares substantially equivalent indications for use with the predicate devices.
#### Design and Materials
The design and materials of MacroPore OS Trauma System and the predicate devices (MacroProe OS, Synthes Resorbable Sheet and Mesh, and MacroPore OS Spine) are nearly identical as they all are made from resorbable polylactide (PLA) material and are provided in sheets and fixation screws of similar shapes and sizes. Both the predicate device and the MacroPore OS Trauma System Sheet have a semi-rigid construction with pores of similar diameter and spacing. The pore size and spacing of the predicate device is within the pore size and spacing specifications of the MacroPore OS Trauma System Protective Sheets. The dimensions of the predicate device are also comparable to the MacroPore OS Trauma System sheet as both devices are provided in rectangular sheets that are several centimeters in size. The mechanical characteristics of the MacroPore OS Trauma System are substantially equivalent to the predicate device with respect to tensile strength, and rigidity as measured by the materials spring constant. In addition to physical characteristics, both the predicate device and the MacroPore OS Trauma Sheets can be shaped with warm water and cut to specific shapes and sizes by the end user.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services. The logo consists of two main elements: the department's name arranged in a circular fashion and an abstract symbol representing the human form. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES" is written in a circular arc around the left side of the logo. To the right of the text is a symbol composed of three stylized human profiles stacked on top of each other.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Kenneth K. Kleinhenz Director of Regulatory Affairs MacroPore Biosurgery 6740 Top Gun Street San Diego, California 92121
JUL · 1 2002
Re: K021164
Trade/Device Name: MacroPore OS Trauma Regulation Number: 21 CFR 888.3030
Regulation Name: Single/Multiple Component Metallic Bone Fixation appliances and
Accessories Regulatory Class: Class II Product Code: HRS Dated: April 10, 2002 Received: April 11, 2002
Dear Mr. Kleinhenz :
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
# Page 2 -- Mr. Kenneth K. Kleinhenz
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and بندر : additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Stypt Rhodes
M. Witten, Ph.D., M.D. Celia Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K 0.21164
Indications for Use: چھون جي سنڌ جي جي جي جي جي جي جي جي جي جي جي ان جي ان جي ايڇ ايو والي ا The MacroPore OS Trauma and the predicate device share substantially equivalent indications for use as they both are indicated for the reinforcing weak bony tissues in orthopeadic procedures and when used in conjuction with traditional rigid fixation. The MacroPore OS Trauma shares identical indications for use wording with the predicate device, MacroPore OS, with the exception of indications for rib bones. Specifically, the MacroPore OS Trauma is indicated for:
The MacroPore OS Trauma System is intended to maintain the relative position of weak bony tissue such as bone grafts, bone graft substitutes, or bone fragments from comminuted fractures. The MacroPore OS Trauma System is also indicated for cement restriction in total joint arthroplasty procedures.
Only when used in conjunction with traditional rigid fixation, the MacroPore OS Trauma System is intended to maintain the relative position weak bony tissue in trauma and reconstructive orthopedic procedures involving:
- · Long bones
- Flat bones
- · Short bones
- · Irregular bones
- · Appendicular skeleton
- Thorax
When used alone (without traditional rigid fixation), the MacroPore OS Trauma System is intended to maintain the relative position of bone graft substitutes in reconstructive orthopedic procedures involving:
- · Tumor resections where bone strength has not been compromised
- · Iliac crest harvests
- Ribs
This device is not intended for use in the spine. The device is not intended for load bearing indications unless used in conjunction with traditional rigid fixation.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | <i>Yes</i> |
|------------------|------------|
|------------------|------------|
(Division Sign-Off)
Division of General, Restorative and Neurological Devices
| 510(k) Number | KO21164 |
|---------------|---------|
|---------------|---------|
| Over-The-Counter Use | <i>No</i> |
|----------------------|-----------|
|----------------------|-----------|
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.