The CORE Shoulder System is a handheld surgical instrument with computer-assisted instrument tracking and is intended to assist the surgeon with placement of the K-wire (central guide pin) used in the preparation of the glenoid and the positioning of the glenoid component during primary Anatomic or Reverse total shoulder arthroplasty. The CORE Shoulder System tracks the live position of the instruments relative to an untracked virtual anatomical model. It does not track the patient anatomy. The CORE Shoulder System is designed for use with the following LINK Implant systems: - LINK Embrace Shoulder System Reverse Configuration (K200368, K212992, K231445) - LINK Embrace Shoulder System Anatomic Configuration (K210899)
Device Story
Handheld surgical instrument system for computer-assisted K-wire placement during shoulder arthroplasty. Inputs: pre-operative CT-based virtual 3D anatomical model; live instrument position data. Operation: surgeon uses handheld COREmote (sensor unit) and stainless-steel probes to track instrument trajectory relative to virtual model; system displays real-time comparison between planned and actual K-wire position on a tablet workstation. Used in OR by surgeons. Benefits: improved accuracy of K-wire placement relative to pre-operative plan. Output: visual guidance on tablet screen. Does not track patient anatomy directly.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical performance testing, including battery life, sensor lifecycle, patient motion risk assessment, comparative K-wire placement testing in analogue glenoid models, cadaver testing, accuracy testing, usability evaluation, biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1), EMC, cybersecurity, software validation, and sterilization validation.
Technological Characteristics
System includes tablet workstation, handheld COREmote (single-use power unit, reusable sensor unit), and reusable stainless-steel probes. Features computer-assisted instrument tracking. Biocompatibility per ISO 10993-1; electrical safety per IEC 60601-1; EMC compliant. Wireless connectivity between handheld unit and tablet. Sterilization validation performed.
Indications for Use
Indicated for surgeons performing primary Anatomic or Reverse total shoulder arthroplasty to assist in K-wire (central guide pin) placement and glenoid component positioning. Designed for use with LINK Embrace Shoulder System (Reverse and Anatomic configurations).
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
Glenoid Intelligent Reusable Instrument System (Glenoid IRIS) (K123122)
Virtual Implant Positioning (VIP) System (K222007)
K984298 — Z KAT FLUROTACTIC GUIDANCE SYSTEM MKI · Z-Kat, Inc. · Jun 23, 1999
K202151 — Smart SPACE Shoulder System · Techmah Medical, LLC · Feb 3, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
February 20, 2025
LinkBio Corp. Deniz Kortan Sr. Quality Manager - Digital Surgery 69 King Street Dover, New Jersey 07801
Re: K241470
Trade/Device Name: CORE Shoulder System Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: OHE, KWS, PHX Dated: January 21, 2025 Received: January 21, 2025
Dear Deniz Kortan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA
guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/2/Picture/3 description: The image shows a digital signature. The signature is by Farzana Sharmin -S. The date of the signature is 2025.02.20, and the time is 14:03:17 -05'00'.
Farzana Sharmin, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241470
Device Name CORE Shoulder System
#### Indications for Use (Describe)
The CORE Shoulder System is a handheld surgical instrument with computer-assisted instrument tracking and is intended to assist the surgeon with placement of the K-wire (central guide pin) used in the preparation of the glenoid and the positioning of the glenoid component during primary Anatomic or Reverse total shoulder arthroplasty. The CORE Shoulder System tracks the live position of the instruments relative to an untracked virtual anatomical model. It does not track the patient anatomy.
The CORE Shoulder System is designed for use with the following LINK Implant systems:
- · LINK Embrace Shoulder System Reverse Configuration (K200368, K212992, K231445)
- · LINK Embrace Shoulder System Anatomic Configuration (K210899)
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
| 510(k)<br>Submitter: | LinkBio Corp.<br>69 King Street<br>Dover, NJ 07801, USA<br>Registration Number: 3006721341 |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Deniz Kortan<br>LinkBio Corp.<br>69 King Street<br>Dover, NJ 07801, USA<br>Phone: 1-800-932-0616<br>Email: d.kortan@linkbio.com |
| Date Prepared: | February 20, 2025 |
| Trade Name: | CORE Shoulder System |
| Common<br>Name: | Shoulder Arthroplasty Implantation System |
| Classification<br>Name: | 21 CFR §888.3660: Shoulder joint metal/polymer semi-constrained<br>cemented prosthesis |
| | QHE (Shoulder Arthroplasty Implantation System) |
| | KWS (Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer<br>Cemented) |
| | PHX (Shoulder Prosthesis, Reverse Configuration) |
| Classification<br>and Panel: | Class II, Orthopedic |
| Predicate<br>Device: | Glenoid Intelligent Reusable Instrument System (Glenoid IRIS), by<br>Cleveland Clinic, K123122 (Primary Predicate), and Virtual Implant<br>Positioning (VIP) System, by Arthrex, K222007 (Additional Predicate). |
| Reason for<br>Submission | New device |
| Device<br>Description: | The CORE Shoulder System is a handheld surgical instrument with<br>computer-assisted instrument tracking and is intended to assist the<br>surgeon with placement of the K-wire (central guide pin) used in the<br>preparation of the glenoid and the positioning of the glenoid component<br>during primary Anatomic or Reverse total shoulder arthroplasty.<br>CORE Shoulder assists the surgeon in placing the K-wire according to the<br>preoperatively planned location. The CORE Shoulder System tracks the live |
{5}------------------------------------------------
position of the instruments relative to an untracked virtual anatomical model. It does not track the patient anatomy.
It allows the surgeon to visually compare the planned and placed position/trajectory of the guide pin (K-wire) by referencing a virtual 3D model of the pre-operative plan and the measured location of the K-wire. The system components include the Workstation (tablet, AC adapter, stand), the handheld COREmote (single-use Power Unit and reusable Sensor Unit), and reusable stainless-steel probes of different sizes.
- Indications for The CORE Shoulder System is a handheld surgical instrument with computer-assisted instrument tracking and is intended to assist the Use: surgeon with placement of the K-wire (central guide pin) used in the preparation of the glenoid and the positioning of the glenoid component during primary Anatomic or Reverse total shoulder arthroplasty. The CORE Shoulder System tracks the live position of the instruments relative to an untracked virtual anatomical model. It does not track the patient anatomy.
The CORE Shoulder System is designed for use with the following LINK Implant systems:
- LINK Embrace Shoulder System - Reverse Configuration (K200368, K212992, K231445)
- LINK Embrace Shoulder System – Anatomic Configuration (K210899)
Comparison to The predicate device for the subject CORE Shoulder System is the Glenoid Predicate IRIS system.
Device:
Both the subject and predicate systems are intended to aid the surgeon in setting K-wire position (entry point and orientation) according to a patientspecific pre-operative plan for the intended glenoid implant position.
Both the predicate and subject devices rely on pre-operative planning software that uses CT data to create a virtual model of the patient's scapula and determine the K-wire position associated with the desired implant position. The predicate Glenoid IRIS uses its own proprietary software for pre-operative planning, while the CORE Shoulder uses a commercially available software legally marketed for this purpose (K170702).
Both systems assist the surgeon in visually comparing the planned and placed K-wire position by referencing a 3D model of the pre-operative plan. In the case of the predicate device, the 3D model is a physical model, and in the case of the subject CORE Shoulder device, the 3D model is a virtual model. The CORE Shoulder also provides computer-assisted instrument tracking to help target the K-wire by tracking the position of the instrument relative to an untracked virtual model. The CORE Shoulder device interacts wirelessly with the computer tablet running the CORE software to display this information. This is a technological difference from the predicate Glenoid IRIS which does not provide computer-assisted instrument tracking.
{6}------------------------------------------------
| | The subject and predicate devices both feature reusable Stainless Steel<br>components for contact with the glenoid which include a guide for drilling<br>the K-wires. |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Both the subject and predicate devices have the same intended purpose.<br>There are some differences in technological characteristics, but both<br>systems provide a method for comparing the planned and placed K-wire<br>position compared to the pre-operative plan, and scientific methods exist<br>for evaluating the effect of the different technological characteristics on<br>safety and effectiveness. The performance testing provided supports the<br>Substantial Equivalence of the subject device for its intended purpose. |
| Performance<br>Testing: | Non-clinical performance testing and analysis were provided, including:<br>- Battery Life<br>- Sensor Unit Lifecycle<br>- Patient Motion Risk Assessment<br>- Comparative testing of K-wire placement versus pre-op plan in<br>analogue glenoid models<br>- Cadaver testing comparing both K-wire placement and implant<br>position versus pre-op plan<br>- Accuracy testing for K-wire Placement Check feature<br>- Usability evaluation<br>- Biocompatibility Evaluation per ISO 10993-1<br>- Electrical safety testing per IEC 60601-1<br>- EMC testing<br>- Cybersecurity Evaluation<br>- Software validation testing<br>- Sterilization validation testing<br><br>The results of non-clinical performance testing and evaluations<br>demonstrate that the device is suitable for its intended purpose and<br>Substantially Equivalent to the predicate. |
| Clinical Testing: | Clinical performance testing was not required to demonstrate the<br>substantial equivalence of this device. |
| Conclusion: | The subject device is substantially equivalent to the predicate device<br>identified in this premarket notification. |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.