Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. 2. Rheumatoid arthritis. 3. Correction of functional deformity. 4. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques. 5. Revision procedures where other treatment or devices have failed. Acetabular shells and femoral stems with porous coatings are indicated for uncemented biological fixation. Non-coated or polyethylene components may be used with mating components that are indicated for either cemented or uncemented use. The Biomet G7™ Freedom™ Constrained Liner is indicated for use as a component of a total hip prosthesis in primary and revision patients at high risk of dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability, and for whom all other options to constrained acetabular components have been considered.
Device Story
Modular acetabular system for total hip arthroplasty; provides surgeons with multiple shell, liner, and head options. Components include titanium alloy shells with porous plasma spray coating, UHMWPE liners (neutral, high wall, or face-changing), and cobalt chrome femoral heads. System includes constrained liner/head options for high-dislocation-risk patients. Used in clinical/OR settings by orthopedic surgeons. Device replaces damaged hip joint surfaces to restore function and reduce pain. System compatibility extends to existing Biomet ceramic heads and femoral stems. Output is a reconstructed hip joint assembly; clinical decision-making relies on surgeon assessment of patient anatomy, bone quality, and stability requirements.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by extensive bench testing, including poly push-out, lever-out, torque-out, rim impingement, fatigue testing, range-of-motion studies, screw pull-through/torsional properties, and metal head fretting/corrosion analysis.
Technological Characteristics
Modular hip prosthesis. Materials: Titanium alloy (ASTM F136) shell, Porous Plasma Spray coating (ASTM F1580), UHMWPE liners (ASTM F648, ArComXL® or E1®), Cobalt chrome femoral heads (ASTM F1537). Features: Solid/limited hole shell designs, neutral/high wall/face-changing liners, constrained liner options. Mechanical fixation via uncemented porous coating or cemented techniques. No software or electronic components.
Indications for Use
Indicated for patients with non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, proximal femur fractures (non-union, femoral neck, trochanteric), or revision needs. Constrained liners indicated for primary/revision patients at high risk of dislocation (prior dislocation, bone loss, laxity, neuromuscular disease, intraoperative instability).
Regulatory Classification
Identification
A hip joint metal/polymer constrained cemented or uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device prevents dislocation in more than one anatomic plane and has components that are linked together. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular component made of ultra-high-molecular-weight polyethylene with or without a metal shell, made of alloys, such as cobalt-chromium-molybdenum and titanium alloys. This generic type of device is intended for use with or without bone cement (§ 888.3027).
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Hip Joint Metal/Polymer Constrained Cemented or Uncemented Prosthesis.”
Predicate Devices
Porous Plasma Spray (PPS®) RingLoc®+ Acetabular System (K093235)
K110087 — CORIN TRINITY ACETBULAR SYSTEM WITH HXLPE LINERS · Corin USA · May 3, 2011
K123705 — TRINITY NON-OCCLUDED TITANIUM PLASMA SPRAYED (TPS) ACETABULAR SHELLS · Corin USA · Feb 12, 2013
K103384 — ESCALADE ACETABULAR CUP SYSTEM · Ortho Development · Mar 16, 2011
K983014 — SUMMIT ACETABULAR SYSTEM · Johnson & Johnson Professionals, Inc. · Oct 27, 1998
K103518 — TRINITY ACETABULAR SYSTEM · Corin USA · Mar 9, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
K121874
Image /page/0/Picture/1 description: The image shows the logo for Biomet Inc. The logo is in black and white and features a stylized font. The letters are bold and outlined, giving them a three-dimensional appearance.
### 510(k) Summary
# NOV 2 3-2012
Preparation Date:
Contact Person:
Proprietary Name:
Classification Name:
Common Name:
Applicant/Sponsor:
August 23, 2012
Biomet Manufacturing Corp. 56 East Bell Drive P.O. Box 587 Warsaw, IN 46581-0587 Establishment Registration Number: 1825034
Becky Earl Regulatory Specialist
G7™ Acetabular System
Acetabular System
LPH- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented (21 CFR 888.3358)
LZO-Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non-Porous, Uncemented (21 CFR 888.3353)
OQG-Prosthesis, hip, semi-constrained, metal/polymer + additive, porous uncemented (21 CFR 888.3358)
KWZ-Prosthesis, hip, constrained, cemented or uncemented, metal/polymer (21 CFR 888.3310) 0
jDI- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented (21 CFR 888.3350)
OQH-Hip, semi-constrained, cemented, metal/polymer + additive, cemented (21 CFR 888.3350)
OQI-Hip, semi-constrained, cemented, metal/ceramic/polymer + additive, porous uncemented (21 CFR 888.3353)
PBI-Prosthesis, hip, constrained, cemented or uncemented, metal/polymer, + additive (21 CFR 888.3310)
Mailing Address: P.O. Box 587 Warsaw, IN 46581-0587 Toll Free: 800.348.9500 Office: 574.267.6639 Main Fax: 574.267.8137 ww.biomet.com
Shipping Address: 56 East Bell Drive Warsaw, IN 46582
p. 1 of 3
{1}------------------------------------------------
# K121874
#### Legally Marketed Devices To Which Substantial Equivalence Is Claimed:
#### Acetabular Shells:
Porous Plasma Spray (PPS®) RingLoc®+ Acetabular System, K093235 (Biomet) Reflection Acetabular Cup System, K033442 (Smith and Nephew) Continuum Acetabular System, K103662 (Zimmer)
#### Acetabular Liners:
Ringloc® E1® Acetabular Liners, K093549 (Biomet) Reflection Acetabular Cup System, K033442 (Smith and Nephew) Continuum Acetabular System, K103662 (Zimmer)
Constrained Heads and Liners: Freedom Constrained Liners, K043537 (Biomet) Cobalt Chrome Femoral Heads: Cobalt Chrome Femoral Components, K911684 (Biomet)
#### Device Description:
The G7™ Acetabular System is a modular system, designed to provide numerous options for surgeons and patients in one compatible system. The titanium alloy (ASTM F136) shell is available in both a solid shell and limited hole design, with an apical hole and plug. The outer diameter of the shells is coated with Biomet's Porous Plasma Spray (ASTM F1580). The acetabular liners are manufactured from UHMWPE (ASTM F648) as ArComXL® or E1® (UHMWPE infused with vitamin E) and available in a Neutral, High Wall, or Ten Degree Face Changing design. Additionally, the system has provided for a constrained liner and head when such an option is required. The Biomet Freedom" product line has added a 32mm Freedom" head (cobalt chrome, ASTM F1537) to its offerings, as well as a new series of constrained liners (E1®, ASTM F648) designed for compatibility with the G7™ shells. A new line of cobalt chrome (ASTM F1537) femoral heads in varying sizes and neck lengths, with either a Type 1 or 12/14 taper, has been added to the system. The system is compatible with Biomet® ceramic heads and femoral stems.
#### Indications for Use:
Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. 1.
- 2. Rheumatoid arthritis.
- 3. Correction of functional deformity.
- Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with 4. head involvement, unmanageable by other techniques.
- ഗ് Revision procedures where other treatment or devices have failed.
Acetabular shells and femoral stems with porous coatings are indicated for uncemented biological fixation. Non-coated or polyethylene components may be used with mating components that are indicated for either cemented or uncemented use.
#### Indications for Biomet G7™ Freedom™ Constrained Liners:
The Biomet G7™ Freedom™ Constrained Liner is indicated for use as a component of a total hip prosthesis in primary and revision patients at high risk of dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability, and for whom all other options to constrained acetabular components have been considered.
{2}------------------------------------------------
Klzs74
#### Summary of Technologies:
The technological characteristics of the proposed device are the same as or similar to the predicates. The G7™ acetabular shells are similar in design, dimensions, and intended use to their Biomet predicate, K093235, but have incorporated a different locking mechanism, which is also similar to legally marketed predicates, K091508 and K103662. The acetabular liners are made of the same material as their predicates, and use previously cleared liner types, sizes and designs. The additional constrained head is a line extension to the Biomet Freedom™ constrained heads, and the constrained liners use the same design principles as the Freedom liner predicate, K030047. The cobalt chrome femoral heads are very similar to the Biomet predicate, K911684.
#### Non-Clinical Testing:
The following testing has been completed to support substantial equivalence: v
- Poly Push Out .
- Poly Lever Out .
- Poly Torque Out .
- Poly Rim Impingement .
- Poly Deformed Cup Push Out .
- Poly Deformed Cup Lever Out
- Poly Fatigue, Lever Out �
- Poly Constrained Liner Rim Impingement and Lever Out Testing .
- Range-of-Motion Study (Poly) .
- Poly Deformed Cup Push In .
- . Freedom Head Pull Out of Constrained Liner
- Screw Pull Through �
- Screw Torsional Properties .
- Torque Curve for MIM Screw Hole Plugs .
- Verification of Insertion Torque and Measurement of Torque Out .
#### Justifications:
- Poly Wear Simulator .
- Metal Head Justification .
- . MRI Justification
- Metal Head Axial Head Pull Off .
- Metal Head Fretting/Corrosion ●
#### Clinical Testing:
None provided as a basis for substantial equivalence.
Testing demonstrates that the modifications made to the G7™ Acetabular System do not introduce any new risks of safety or efficacy, and that the G7™ Acetabular System is substantially equivalent to the predicates.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three wing-like extensions. The bird is positioned within a circle that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference of the circle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
November 23, 2012
Biomet Manufacturing Corporation % Ms. Becky Earl 56 East Bell Drive P.O. Box 587 Warsaw, Indiana 46581-0587 US
Re: K121874
Trade/Device Name: G7™ acetabular system Regulation Number: 21 CFR 888.3310 Regulation Name: Hip Joint Metal/Polymer Constrained Cemented or Uncemented Prosthesis Regulatory Class: Class II . Product Code: PBI, LPH, LZO, OQG, KWZ, JDI, OQH, OQI
- Dated: October 19, 2012
Received: October 22, 2012
Dear Ms. Earl:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
:
{4}------------------------------------------------
## Page 2 - Ms. Becky Earl
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Erin I. Keith
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):_K121894
Device Name: G7™ Acetabular System
Indications For Use:
1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular
- necrosis. ..
- 2. Rheumatoid arthritis.
- 3. Correction of functional deformity.
- 4. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques.
- 5. Revision procedures where other treatment or devices have failed.
Acetabular shells and femoral stems with porous coatings are indicated for uncemented biological fixation. Non-coated or polyethylene components may be used with mating components that are indicated for either cemented or uncemented use.
## Indications for Biomet G7 Freedom™ Constrained Liners:
The Biomet G7 Freedom™ Constrained Liner is indicated for use as a component of a total hip prosthesis in primary and revision patients at high risk of dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability, and for whom all other options to constrained acetabular components have been considered.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use NO (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
K. Asindi
Jor (Division Sign-Off) . Division of Orthopedic Devices 510(k) Number K121874
Page 1 of 1
42
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.