FlexBand; FlexPatch; FlexBand Plus

K230316 · International Life Sciences · OWW · May 26, 2023 · General, Plastic Surgery

Device Facts

Record IDK230316
Device NameFlexBand; FlexPatch; FlexBand Plus
ApplicantInternational Life Sciences
Product CodeOWW · General, Plastic Surgery
Decision DateMay 26, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3300
Device ClassClass 2
AttributesTherapeutic

Intended Use

FlexBand®, FlexPatch®, and FlexBand® Plus are intended for use in surgical procedures for reinforcement of soft tissue where weakness exists. FlexBand®, FlexPatch®, and FlexBand® Plus are also intended for reinforcement of soft tissues that are repaired by suture or other fixation devices during tendon and ligament repair surgery including reinforcement of rotator cuff, patellar, Achilles, biceps, quadriceps tendons, medial collateral ligament, lateral ligament, spring ligament, deltoid ligament, ulnar collateral ligament or other tendons or extra-articular ligaments. FlexBand®, FlexPatch®, and FlexBand® Plus is not intended to replace normal body structure or provide the full mechanical strength to support the rotator cuff, patellar, Achilles, biceps tendons, medial collateral ligament, lateral collateral ligament, spring ligament, deltoid ligament or other tendons or extra-articular ligaments. Sutures, used to repair the tear, and sutures or other fixation devices, used to the bone, provide mechanical strength for the tendon repair. The products reinforce soft tissue and provides a degradable scaffold that is incorporated into the patient's own tissue.

Device Story

Knitted surgical mesh made from ARTELON® (degradable polycaprolactone-based polyurethane urea) fibers; used for soft tissue reinforcement in orthopedic surgeries; provides degradable scaffold for tissue incorporation; does not replace native tissue mechanical strength. FlexBand® Plus includes pre-loaded sutures for surgical usability. Supplied sterile; cut to size by surgeon; implanted during tendon/ligament repair procedures. Acts as adjunct to primary suture/fixation repair; shares load on primary repair; supports healing via scaffold integration. Used in clinical operating room settings by surgeons.

Clinical Evidence

Clinical literature review provided regarding ligament reinforcement procedures. Literature indicates device effectiveness for proposed ligament indications. Bench testing included suture retention and tensile strength testing. Routine endotoxin (LAL) testing performed per lot.

Technological Characteristics

Knitted mesh; material: ARTELON® (degradable polycaprolactone-based polyurethane urea); form factor: mesh strip/patch; sterilization: E-Beam (min 25 kGy); non-electronic; no software.

Indications for Use

Indicated for reinforcement of soft tissue where weakness exists and reinforcement of soft tissues repaired by suture/fixation during tendon and ligament repair surgery (e.g., rotator cuff, patellar, Achilles, biceps, quadriceps, collateral, spring, deltoid, ulnar ligaments). Not intended to replace normal body structure or provide full mechanical strength for tendon/ligament repair.

Regulatory Classification

Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ May 26, 2023 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. International Life Sciences Tiffini Wittwer Vice President Quality and Regulatory Affairs 2150 Northwest Parkway SE Suite G Marietta, Georgia 30076 ## Re: K230316 Trade/Device Name: FlexBand®; FlexPatch®; FlexBand® Plus Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: QWJ, FTL, OWW Dated: April 26, 2023 Received: April 27, 2023 ## Dear Tiffini Wittwer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Yu-chieh Chiu -S Yu-Chieh Chiu, Ph.D. Acting Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K230316 ### Device Name FlexBand®, FlexPatch®, and FlexBand® Plus ## Indications for Use (Describe) FlexBand®, FlexPatch®, and FlexBand® Plus are intended for use in surgical procedures for reinforcement of soft tissue where weakness exists. FlexBand®, FlexPatch®, and FlexBand® Plus are also intended for reinforcement of soft tissues that are repaired by suture or other fixation devices during tendon and ligament repair surgery including reinforcement of rotator cuff, patellar, Achilles, biceps, quadriceps tendons, medial collateral ligament, lateral ligament, spring ligament, deltoid ligament, ulnar collateral ligament or other tendons or extra-articular ligaments. FlexBand®, FlexPatch®, and FlexBand® Plus is not intended to replace normal body structure or provide the full mechanical strength to support the rotator cuff, patellar, Achilles, biceps tendons, medial collateral ligament, lateral collateral ligament, spring ligament, deltoid ligament or other tendons or extra-articular ligaments. Sutures, used to repair the tear, and sutures or other fixation devices, used to the bone, provide mechanical strength for the tendon repair. The products reinforce soft tissue and provides a degradable scaffold that is incorporated into the patient's own tissue. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(K) Summary: K230316 # FlexBand®/ FlexPatch® / FlexBand® Plus #### I. Submitter Information | Submitter: | International Life Sciences (dba Artelon) | | |----------------|--------------------------------------------------------|--| | Address: | 2150 Northwest Pkwy SE<br>Suite G<br>Marietta GA 30076 | | | Telephone: | +1 (800) 610-3446 | | | Contact: | Tiffini Wittwer, MPH | | | Date Prepared: | May 23, 2023 | | #### Device Information II. | Trade Name(s): | FlexBand® / FlexPatch® / FlexBand® Plus | |-----------------------|-------------------------------------------------------------------| | Common Name: | Surgical Mesh | | Classification: | Class II | | Regulation: | 21 CFR 878.3300 | | Classification Name: | Mesh, Surgical, Absorbable, Orthopedics, Reinforcement of Ligamen | | Classification Panel: | Orthopedic | QWJ, OWW, FTL Product Codes: #### Predicate Device Information III. The FlexBand® / FlexPatch® / FlexBand® Plus devices described in this submission are substantially equivalent to the following predicate devices: The FlexBand® / FlexPatch® / FlexBand® Plus (K192112) BioBridge™ Collagen Matrix (K151083) #### IV. Device Description The FlexBand® / FlexPatch® / FlexBand® Plus products are knitted mesh made from ARTELON® fibers. ARTELON® fiber is made of degradable polycaprolactone-based polyurethane urea. The construction permits the mesh to be cut into any desired shape or size without unraveling. FlexBand® Plus devices have suture attached to each end of the knitted mesh strip. The pre-loaded suture is intended to aid in usability in the operating room. The devices are supplied sterile, one product per package in double layer peel pouch packaging. #### V. Indication for Use {4}------------------------------------------------ FlexBand®, FlexPatch®, and FlexBand Plus® are intended for use in surgical procedures for reinforcement of soft tissue where weakness exists. FlexBand®, FlexPatch®, and FlexBand Plus® are also intended for reinforcement of soft tissues that are repaired by suture or other fixation devices during tendon and ligament repair surgery including reinforcement of rotator cuff, patellar, Achilles, biceps, quadriceps tendons, medial collateral ligament, lateral collateral ligament, spring ligament, deltoid ligament, ulnar collateral ligament or other tendons or extra-articular ligaments. FlexBand®, FlexPatch®, and FlexBand Plus® is not intended to replace normal body structure or provide the full mechanical strength to support the rotator cuff, patellar, Achilles, biceps, quadriceps tendons, medial collateral ligament, lateral ligament, spring ligament, deltoid ligament, ulnar collateral ligament or other tendons or extra-articular ligaments. Sutures, used to repair the tear, and sutures or other fixation devices, used to attach the tissue to the bone, provide mechanical strength for the tendon repair. The products reinforce soft tissue and provides a degradable scaffold that is incorporated into the patient's own tissue. #### Comparison of Technological Characteristics With the Predicate Devices VI. FlexBand®, FlexPatch®, and FlexBand Plus® devices have the same principles of operation as the predicate devices (K192112 and K151083). They are all implants intended for the reinforcement of soft tissues that are repaired by suture or other fixation devices. The devices are used in various surgical procedures where soft tissue reinforcement is needed. FlexBand®, FlexPatch®, and FlexBand Plus® products are similar to the BioBridge predicate (K151083) in the following ways: - Intended use, to share load placed on the primary repair and provide structural scaffold . for torn or damaged soft tissue; - Indication for Use - Radiation sterilization (E-Beam method, minimum 25 kGy) ● - Biodegradable - May be cut to size to meet surgeons' preference ● There have been no changes to the FlexBand®, FlexPatch®, and FlexBand Plus® products since the prior 510(k) clearance. Therefore, FlexBand®, FlexPatch®, and FlexBand Plus® products are the same as the FlexBand®, FlexPatch®, and FlexBand Plus® predicate (K192112) in the following ways: - . Material - Knit patterns and design. . - Product sizes and packaging configuration #### VII. Performance Data {5}------------------------------------------------ ## Mechanical testing - Suture Retention Testing ● - Tensile Strength Testing ● The tests listed above provide objective evidence that the FlexBand®, FlexBatch®, and FlexBand Plus® devices provide adequate mechanical properties for use in ligament soft tissue reinforcement. The performance data risk / benefit analysis concluded that the differences do not affect the safety and effectiveness of the FlexBand®, FlexPatch®, and FlexBand Plus® device in relation to the predicate. Routine endotoxin (LAL) testing is performed on each production lot to monitor endotoxin levels. # Clinical Literature Clinical literature was provided on the subject device involving ligament reinforcement procedures. The reviewed literature shows the device is effective when used in the proposed ligament indication. # VIII. Conclusion Based on the above information the FlexBand®, FlexPatch®, and FlexBand Plus® products are substantially equivalent to the soft tissue reinforcement predicate devices: The FlexBand® / FlexPatch® / FlexBand® Plus (K192112) ``` BioBridge™ Collagen Matrix (K151083) ```
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...