K151083 · Fibralign Corporation · OWY · Jan 8, 2016 · General, Plastic Surgery
Device Facts
Record ID
K151083
Device Name
BioBridge Collagen Matrix
Applicant
Fibralign Corporation
Product Code
OWY · General, Plastic Surgery
Decision Date
Jan 8, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The BioBridge™ Collagen Matrix is intended to reinforce soft tissue where weakness and deficiencies exist, specifically, for reinforcement of soft tissue repaired by sutures or suture anchors in tendon repair, including small tendons, ligaments, and general surgical procedures for tissue repair where weakness exists, including muscle flap reinforcement. BioBridge is not intended to replace normal body structure or provide the full mechanical strength to support tendon repair.
Device Story
BioBridge is a sterile, biocompatible, biodegradable surgical mesh ribbon made of highly purified porcine Type I collagen. It features a collapsed, multiluminal structure with aligned collagen fibrils to provide tensile strength for soft tissue support. Used in surgical settings by physicians, the device is implanted to reinforce soft tissue weaknesses or bridge healthy tissues. Surgeons may use single or multiple ribbons, potentially bundling or braiding them to suit specific patient needs. The device acts as a mechanical scaffold to support repair sites; it does not replace normal body structure or provide full mechanical strength for tendon repair. It is intended to aid in surgical outcomes for tendon, ligament, and muscle flap procedures.
Clinical Evidence
No clinical data. Evidence consists of non-clinical bench testing (tensile pull, suture pull, knot tensile, degradation rate, characterization) and in vivo animal studies (rabbit tendon repair model, porcine model). Biocompatibility testing performed per ISO 10993.
Technological Characteristics
Material: Highly purified porcine-derived Type I collagen. Structure: Narrow, thin, multiluminal ribbon with aligned collagen fibrils. Crosslinking: Chemical crosslinking agent used during manufacturing; residuals removed via rinsing. Sterilization: Terminal sterilization. Form factor: Ribbon-like mesh. Connectivity: None.
Indications for Use
Indicated for reinforcement of soft tissue with weakness or deficiencies, including small tendons, ligaments, muscle flaps, and general surgical tissue repair sites reinforced by sutures or suture anchors.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a human figure, represented by three overlapping profiles, suggesting a sense of community and care. The logo is presented in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 8, 2016
Fibralign Corporation c/o Mr. Ronald S. Warren Experien Group, LLC 755 N. Mathilda Avenue, Suite 100 Sunnyvale, CA 94085
Re: K151083
Trade/Device Name: BioBridge Collagen Matrix Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: OWY Dated: December 7, 2015 Received: December 8, 2015
Dear Mr. Warren:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
Radiological Health
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K151083
Device Name BioBridge™ Collagen Matrix
## Indications for Use (Describe)
The BioBridge™ Collagen Matrix is intended to reinforce soft tissue where weakness and deficiencies exist, specifically, for reinforcement of soft tissue repaired by suture anchors in tendon repair, including small tendons, ligaments, and general surgical procedures for tissue repair where weakness exists, including muscle flap reinforcement.
BioBridge is not intended to replace normal body structure or provide the full mechanical strength to support tendon repair.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!"
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Notification K151083
### I. GENERAL INFORMATION [807.92(a)(1)]
## Applicant:
Fibralign Corporation 32930 Alvarado-Niles Road, Suite 350 Union City, CA 94587 U.S.A. Phone: 510-489-2199 FAX: 831-335-8348
## Contact Person:
Ronald S. Warren Regulatory Consultant for Fibralign Experien Group, LLC 755 N. Mathilda Avenue, Suite 100 Sunnyvale, CA 94085 U.S.A. Phone: (408) 505-3926 FAX: (408) 400-0865 Email: Ron@experiengroup.com
Date Prepared: January 8, 2016
## II. DEVICE INFORMATION [807.92(a)(2)]
## Trade/Proprietary Name:
BioBridge™ Collagen Matrix
# Generic/Common Name:
Surgical mesh
## Classification Name:
Mesh, Surgical, Collagen, Plastic and Reconstructive Surgery
# Regulatory Classification:
Class II per 21 CFR§878.3300
# Product Code:
OWY
{4}------------------------------------------------
### III. PREDICATE DEVICES [807.92(a)(3)]
- Primary predicate device: TRELLIS™ Collagen Ribbon Wright Medical . Technology, Inc. (K131143; K120019)
- Additional predicate device: HP Tissue Matrix (HPTM)- LifeCell Corporation . (K142326; K08)
- Reference device: Ossix™ Plus ColBar LifeSciences Ltd. (K053260) ●
The ColBar Ossix Plus device is included as a reference device as it is sourced from Datum Dental, the same supplier of collagen used in BioBridge, and is a similar device made of purified pepsin-treated Collagen Type I.
### IV. DEVICE DESCRIPTION [807.92(a)(4)]
BioBridge is a sterile implantable biocompatible and biodegradable surgical mesh ribbon comprised of highly purified porcine collagen that is designed to provide mechanical support for weaknesses and deficiencies in soft tissue. The collapsed ribbon structure contains multiple folds of a thin membrane with aligned collagen fibrils in the same lengthwise direction. This structure provides mechanical properties that contribute to strong tensile strength, that which provides support to weaknesses and deficiencies in soft tissue and aids in bridging a connection between two healthy soft tissues.
BioBridge is fabricated using a proprietary manufacturing process that produces a narrow and very thin ribbon-like membrane comprised of collagen fibrils, creating a multiluminal structure that provides mechanical properties for support of soft tissue repair. This approach presents the opportunity to enable the use of highly purified, pepsin treated collagen, in a defined structure that has mechanical properties similar to those of the predicates. The smaller ribbon-like form factor gives the surgeon greater flexibility to tailor their procedures compared to the predicates without changing the intended use. One or more BioBridge devices can be implanted depending on the surgeon's discretion, including arrangement of the individual devices in bundles or braided together by the physician, to better address a specific patient need. It is particularly well suited, for example, in supporting repair of small tendons and ligaments. Fibralign sources the highly purified porcine-derived Type I collagen from an FDA-registered and ISOqualified supplier. Fibralign employs a proprietary manufacturing process that takes highly purified porcine-derived Type I collagen and produces narrow, ribbon-like membranes comprised of highly aligned collagen fibrils that provides mechanical strength for use in supporting soft tissue. A chemical crosslinking agent is used during the manufacturing process to promote the crosslinking of the collagen but the crosslinker itself is not added to nor bound to the collagen matrix. The crosslinker residuals are water soluble and are removed by product rinsing at the end of the production process. The final product is packaged into individual storage containers and then sealed within an outer tray and terminally sterilized.
### INDICATIONS FOR USE [807.92(a)(5)] V.
The Indication for Use statement for the Biobridge™ Collagen Matrix is as follows:
{5}------------------------------------------------
The BioBridge™ Collagen Matrix is intended to reinforce soft tissue where weakness and deficiencies exist, specifically, for reinforcement of soft tissue repaired by sutures or suture anchors in tendon repair, including small tendons, ligaments, and general surgical procedures for tissue repair where weakness exists, including muscle flap reinforcement.
BioBridge is not intended to replace normal body structure or provide the full mechanical strength to support tendon repair.
The Indications for Use statement for the BioBridge are semantically similar to the predicate device; however, the differences are not critical to the intended surgical use of the device nor do they affect the safety and effectiveness of the device relative to the predicate. Both the subject and predicate devices have intended surgical use for the reinforcement and repair of soft tissue where weakness exists.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES [807.92(a)(6)]
The primary predicate device TRELLIS received 510(k) clearance (K131143; K120019) as a narrow, ribbon-like version of its predicate, the WMT Dermal Matrix (K073219). The new dimensions of TRELLIS allowed for new surgical techniques (i.e. support of soft tissue in orthopedic applications) as compared to its predicate (Wright Dermal Matrix) but did not change the intended use of the device from its predicate. Similarly, the proposed BioBridge device is a thinner collagen ribbon than the primary predicate device, which provides greater flexibility to the surgeon and allows for further expansion of the potential surgical techniques for soft tissue repair, but does not change the intended use of the device as compared to this predicate.
In addition, BioBridge is comparable with respect to intended use to the LifeCell HPTM device (K142326; K082176) which is intended for use as a soft tissue patch to reinforce soft tissue where weakness exists and for the surgical repair of damaged or ruptured soft tissue membranes which require the use of reinforcing or bridging material to obtain the desired surgical outcome. The HPTM device is a surgical mesh derived from porcine dermis and the implant is intended for reinforcement of soft tissue in plastic and reconstructive surgery.
#### VII. PERFORMANCE DATA [807.92(b)]
All necessary bench and non-clinical testing were conducted on the BioBridge to support a determination of substantial equivalence to the predicate device.
# [807.92(b)(1)]
## Nonclinical Testing Summary:
The nonclinical bench testing and in vivo testing included:
- Dry Tensile Pull Test
- Wet Tensile Pull Test .
{6}------------------------------------------------
- Suture Pull Test
- Surgical Knot Tensile Test ●
- Length and Diameter measurements ●
- Characterization of collagen, water, and crosslinking ●
- Enzymatic degradation rate ●
- Biocompatibility testing ●
- In vivo studies including a rabbit tendon repair model which evaluated the ● degradation rate as well as the mechanical strength of the repair over time
Biocompatibility testing was completed on the finished BioBridge device and on the porcine collagen raw material per ISO 10993 requirements for a permanent implant.
In addition to the nonclinical bench testing, the performance of the BioBridge device was supported by animal studies. A porcine model and rabbit tendon model were employed to support the intended use of the device.
The collective results of the nonclinical testing demonstrate that the materials chosen and the design of the BioBridge meet the established specifications necessary for consistent performance during its intended use. In addition, the collective bench and animal testing demonstrates that BioBridge does not raise new questions of safety or effectiveness for soft tissue reconstruction when compared to the predicate devices.
# [807.92(b)(2)]
## Clinical Testing Summarv:
This section is not applicable as the 510(k) clearance of this device was not supported by clinical testing.
## [807.92(b)(3)]
The collective results of the nonclinical testing demonstrate that the BioBridge materials and design meet the established specifications necessary for consistent performance during its intended use. In addition, the collective bench testing demonstrates that BioBridge does not raise new questions of safety or effectiveness for surgical use for the reinforcement and repair of soft tissue where weakness exists when compared to the predicate devices.
#### VIII. CONCLUSIONS
The BioBridge device subject to this 510(k) is substantially equivalent to the predicate devices, namely the Wright Medical Technology, Inc. TRELLIS"" Collagen Ribbon (K131143; K120019), as well as the LifeCell HPTM device (K142326).
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.