K210470 · Medicrea International SA · OVE · Sep 3, 2021 · Orthopedic
Device Facts
Record ID
K210470
Device Name
C-CURVE Interbody Fusion Device
Applicant
Medicrea International SA
Product Code
OVE · Orthopedic
Decision Date
Sep 3, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The C-CURVE® Device is an anterior cervical intervertebral body fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) (defined as pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) at one disc level from C2 to T1. Patients should have received 6 weeks of non-operative treatment prior to treatment with the device is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow aspirate. The device is a stand-alone system when used C-Curve screws and when used without the integrated screws it requires additional supplemental fixation cleared for the cervical spine.
Device Story
C-CURVE is an anterior cervical intervertebral body fusion system; consists of titanium alloy cage and two expansion screws. Implanted via anterior approach between adjacent cervical vertebrae after discectomy. Cage features two pre-assembled components (body and expansion screw) allowing expansion and lordotic angle adaptation. Used as stand-alone system when secured with integrated screws; otherwise requires supplemental fixation. Intended to restore interbody height, stabilize spine, and promote bony fusion. Operated by surgeons in clinical/OR settings. Output is physical stabilization of cervical spine segment. Benefits include restoration of disc height and spinal stability in DDD patients.
Clinical Evidence
Bench testing only. No clinical data provided. Mechanical performance validated via ASTM F2077 (static/dynamic axial compression, torsion, compression-shear) and ASTM F2267 (subsidence).
Technological Characteristics
Materials: Ti-6Al-4V ELI titanium alloy (ASTM F136, ISO 5832-3). Components: Expandable cage and two screws. Principle: Mechanical interbody fusion with integrated fixation. Sterilization: Provided sterile. Connectivity: None (mechanical implant).
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one cervical level (C2-T1) who have failed 6 weeks of non-operative treatment. Contraindications include patients not skeletally mature or those without confirmed discogenic pain.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
K231763 — Stable-C Interbody System · Nexus Spine, LLC · Aug 21, 2023
K240715 — Idys® C ZP 3DTi · Clariance · Mar 22, 2024
K172788 — MONET Anterior Cervical Interbody Fusion Cage System · Ctl Medical Corporation · Jun 25, 2018
K162760 — Spine Wave Anterior Cervical Spine System · Spine Wave, Inc. · Feb 16, 2017
K211417 — F3D-C2 Cervical Stand-Alone System · Corelink, LLC · Feb 4, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 3, 2021
Medicrea International S. A. Karine Trogneux Regulatory Affairs Department Manager 5389 route de Strasbourg - Vancia Rillieux-La-Pape, 69140 France
Re: K210470
Trade/Device Name: C-CURVE Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVE, ODP Dated: July 23, 2021 Received: August 4, 2021
Dear Karine Trogneux:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brent L. Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K210470
Device Name C-CURVE
#### Indications for Use (Describe)
The C-CURVE® Device is an anterior cervical intervertebral body fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) (defined as pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) at one disc level from C2 to T1. Patients should have received 6 weeks of non-operative treatment prior to treatment with the device is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow aspirate. The device is a stand-alone system when used C-Curve screws and when used without the integrated screws it requires additional supplemental fixation cleared for the cervical spine.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
The Company's 510(k) Summary is provided on the following pages.
## 510(k) SUMMARY MEDICREA INTERNATIONAL's C-Curve
In accordance with 21 CFR 807.92 of the Federal Code of Regulations the following 510(k) summary is submitted for the C-Curve:
Date Prepared: February, 15, 2021 Prepared answers: July, 23, 2021
## 1. Submitter:
MEDICREA INTERNATIONAL 5389 route de Strasbourg - Vancia RILLIEUX-LA-PAPE 69140 FR
## Contact Person:
Karine Trogneux MEDICREA INTERNATIONAL 5389 route de Strasbourg - Vancia RILLIEUX-LA-PAPE 69140 FR
## 2. Trade name: C-Curve
## Regulatory Identification/ Classification
Intervertebral Fusion Device With Integrated Fixation, Cervical / Intervertebral Fusion Device With Bone Graft, Cervical Regulation Number: 21CFR 888.3080 Product Code: OVE, ODP Class II
## 3. Predicate or legally marketed devices which are substantially equivalent:
#### Primary predicate device:
- Endoskeleton TCS Interbody Fusion Device (Titan Spine, K192018)
## Additional predicate devices:
- HIJAK SA (Atlas Spine, K192570)
- PEEK Prevail (Medtronic, K073285, K153373)
- Optio-C (ZIMMER SPINE, K132894)
None of those devices have not been subject to a design-related recall.
{4}------------------------------------------------
#### 4. Description of the device:
MEDICREA® INTERNATIONAL S.A. C-CURVE® is composed of a cervical cage range stabilized by C-CURVE® screws. The titanium cage is composed of two pre-assembled components (the cage body and an expansion screw) that enable its expansion and adaptation of its lordotic angle. The cages are delivered individually and the screws are provided separately by two. The cages are available in different footprints and heights. These implants are designed to be inserted between two adjacent cervical vertebrae after removal of the disc. C-CURVE® is implanted by anterior approach. It is important to refer to the surgical technique and to the Instructions For Use (see paragraph «IMPLANT INSERTION», especially to use the two
screws to stabilize the cage). C-CURVE® can be considered as a stand-alone cage if the two screws are used to stabilize the cage. If it is not the case, an additional fixation system must be added. Titanium cages & screws are manufactured in titanium alloy Ti-6Al-4V ELI conforming to ISO 5832-3 specifications and ASTM F136 specifications. Under no circumstances are the implants reusable.
MATERIALS: Components are manufactured from titanium alloy (Ti-6Al-4V ELI) that conforms to ASTM F136 and ISO 5832-3.
#### Function:
C-Curve was developed as an implant:
- To restore and maintain adequate interbody height.
- To stabilize the spine at one level & promote bony fusion.
#### 5. Indication for Use
C-CURVE" is an anterior cervical intervertebral body fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) (defined as pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) at one disc level from C2 to T1. Patients should have received 6 weeks of non-operative treatment with the device. The device is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone marrow aspirate. The device is a stand-alone system when used C-Curve screws and when used without the integrated screws it requires additional supplemental fixation cleared for the cervical spine.
## 6. Substantial equivalence claimed to predicate devices
The C-Curve components are technologically similar to the already cleared predicates: the Titan Spine Endoskeleton TCS (K192018), the Atlas Spine HiJak SA (K192570), the Medtronic PEEK Prevail (K073285, K153373) and the ZIMMER SPINE Optio-C (K132894) in terms of intended use, material used, mechanical safety and performances.
{5}------------------------------------------------
The table below compares the features and characteristics of the submitted C-Curve components to their predicate devices.
| Device | MEDICREA<br>INTERNATIONA<br>L<br>C-Curve | TITAN SPINE<br>Endoskeleton TCS | ATLAS SPINE<br>HIJAK SA | Medtronic<br>PEEK Prevail | ZIMMER SPINE<br>Optio-C |
|--------------------------------------|------------------------------------------------------|----------------------------------------------------|------------------------------------------------------|-------------------------------------------------------------------|------------------------------------------------------------|
| 510(k)<br>number | To be<br>determined | K192018 | K192570 | K073285,<br>K153373 | K132894 |
| Intended<br>use | | | | | |
| Cervical via<br>anterior<br>approach | Yes | Yes | Yes | Yes | Yes |
| Components | | | | | |
| Device<br>component<br>s | Cage & 2<br>screws | Cage & 2 screws | Cage/plate &<br>2 screws | Cage & 2 screws | Cage/plate & 3<br>screws |
| Material | | | | | |
| | Cage: Ti-6Al-<br>4V ELI<br>Screws: Ti-<br>6Al-4V ELI | Cage: Ti-6Al-4V<br>ELI<br>Screws: Ti-6Al-4V<br>ELI | Cage: Ti-6Al-<br>4V ELI<br>Screws: Ti-<br>6Al-4V ELI | Cage: PEEK,<br>Tantalum &<br>Nitinol<br>Screws: Ti-6Al-<br>4V ELI | Cage: PEEK, Ti-<br>6Al-4V ELI<br>Screws: Ti-6Al-<br>4V ELI |
| Sterile<br>state | | | | | |
| | Provided<br>sterile | Provided sterile | Provided non<br>sterile | Provided sterile | Provided sterile |
## 7. Non-clinical Test Summary:
The following performance data was provided in support of the substantial equivalence determination.
#### Biocompatibility Testing
The C-Curve components are made from the same materials as its predicates and the manufacturing processes are similar to the ones of the predicates.
#### Mechanical testing
The tests performed on the C-Curve components (Static & Dynamix Axial Compression, Static & Dynamic Torsion, Static and Dynamic Compression Shear per ASTM F2077 , Subsidence per ASTM F2267 and Postero-anterior expulsion) indicate that the product is mechanically equivalent to its predicates.
{6}------------------------------------------------
#### 8. Conclusions
The subject C-Curve device demonstrated equivalent mechanical performance to the cited predicate device under the same test conditions.
The overall technology characteristics and mechanical performance data lead to the conclusion that the subject device is substantially equivalent to legally marketed predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.