Endoskeleton TA Interbody Fusion Device, Endoskeleton TAS and TAS Hyperlordotic Interbody Fusion Device, Endoskeleton TO Interbody Fusion Device, Endoskeleton TT Interbody Fusion Device, Endoskeleton TC Interbody Fusion Device, Endoskeleton TCS Interbody Fusion Device, Endoskeleton TL Interbody Fusion Device
K192018 · Titan Spine, Inc. · ODP · Aug 16, 2019 · Orthopedic
Device Facts
Record ID
K192018
Device Name
Endoskeleton TA Interbody Fusion Device, Endoskeleton TAS and TAS Hyperlordotic Interbody Fusion Device, Endoskeleton TO Interbody Fusion Device, Endoskeleton TT Interbody Fusion Device, Endoskeleton TC Interbody Fusion Device, Endoskeleton TCS Interbody Fusion Device, Endoskeleton TL Interbody Fusion Device
Applicant
Titan Spine, Inc.
Product Code
ODP · Orthopedic
Decision Date
Aug 16, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ENDOSKELETON® TA Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogence back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the device may be used with supplemental fixation that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). ENDOSKELETON® TA is indicated to be used with autograft bone and/or allograft bone comprised of cancellous bone and/or demineralized allograft bone with bone marrow aspirate. The ENDOSKELETON® TAS Interbody Fusion device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogence back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients should have received 6 months of nonoperative treatment prior to treatment with the device is a standalone system that is intended to be used with the bone screws provided and requres no additional supplementary fixation. The Device is indicated to be used with autograft bone and/or allograft bone comprised of cancellous bone and/or demineralized allograft bone with bone marrow aspirate. Hyperlordotic Devices ≥16°: The ENDOSKELETON® TAS Hyperlordotic Interbody Fusion Device (≥ 16°) is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). Patients should have received 6 months of non-operative treatment with the devices. The device is indicated to be used with autograft bone and/or allograft bone comprised of cancellous bone and/ or demineralized allograft bone with bone marrow aspirate.The ENDOSKELETON® TAS Hyperlordotic Interbody Fusion Device must be used with a posterior supplemental internal spinal fixation that has been cleared by the FDA for use in the lumbar spine must be used with a posterior supplemental inxation that has been cleared by the FDA for use in the lumbar spine. The ENDOSKELETON® TC is indicated for use for anterior cervical interbody fusion in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine at one disc level from C-3 to C-7. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have received 6 weeks of non-operative treatment prior to treatment with the device. The ENDOSKELETON® TC is indicated to be used with supplemental fixation that has been cleared by the FDA for use in the cervical spine and be used with autograft bone and/ or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow aspirate. The ENDOSKELETON® TCS System is an anterior cervical intervertebral body fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) (defined as pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) at one disc level from C2 to T1. Patients should have received 6 weeks of non-operative treatment with the device. The device is indicated to be used with autograft bone and/or allograft bone comprised of cancellous bone and/or demineralized allograft bone with bone marrow aspirate. The device is a stand-alone system when used with ENDOSKELETON® TCS integrated screws and when used without the integrated screws it requires additional supplemental fixation cleared for the cervical spine. The ENDOSKELETON® TL Interbody Fusion Device is indicated for use in spinal fusion procedures in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment with the devices. The device must be used with supplemental fixation systems that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients with previous non-fusion spinal surgery at the involved level(s) may be treated with the device. It is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow aspirate. The ENDOSKELETON® TO Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogence back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the devices. The device must be used with supplemental fixation that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). It is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone marrow aspirate. The ENDOSKELETON® TT Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogence back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non- operative treatment prior to treatment with the devices. The device must be used with supplemental fixation that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). It is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone marrow aspirate.
Device Story
Endoskeleton® System consists of interbody and vertebral body fusion devices of varying sizes/geometries; manufactured from Ti6Al4V-ELI titanium alloy. Implants feature hollow centers for bone graft material (autograft, allograft, or demineralized allograft with bone marrow aspirate) and windows for graft visualization/bone formation. Superior/inferior surfaces utilize Chemtex® or nanoLOCK® (MMN™) surface treatments to enhance fixation. Used in spinal fusion procedures; implanted by surgeons. Modification expands indications to include use with demineralized allograft bone with bone marrow aspirate. No changes to design, materials, or surgical technique.
Clinical Evidence
No clinical data or performance testing was performed for this submission; substantial equivalence is based on literature review and similarity to predicate devices.
Technological Characteristics
Medical grade titanium alloy (Ti6Al4V-ELI) per ASTM F-136/ISO 5832-3 and ASTM F-3001. Features hollow geometry for bone graft, visualization windows, and Chemtex® or nanoLOCK® (MMN™) surface treatments. Implanted for >30 days. Sterilization via gamma irradiation.
Indications for Use
Indicated for skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels (lumbar L2-S1 or cervical C2-T1). DDD defined as discogenic pain with degeneration confirmed by history/radiography. Patients must have failed 6 months (lumbar) or 6 weeks (cervical) of non-operative treatment. May include Grade I spondylolisthesis or retrolisthesis (lumbar).
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
Medtronic Sofamor Danek USA, Inc. ARTiC-L™ 3D and ARTiC-XL™ 3D Ti Spinal System (K190959)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out in a sans-serif font.
August 16, 2019
Titan Spine, Inc Christine Scifert 6140 West Executive Dr., Suite A Mequon, Wisconsin 53092
Re: K192018
Trade/Device Name: Endoskeleton TA Interbody Fusion Device, Endoskeleton TAS and TAS Hyperlordotic Interbody Fusion Device, Endoskeleton TO Interbody Fusion Device, Endoskeleton TT Interbody Fusion Device, Endoskeleton TC Interbody Fusion Device, Endoskeleton TCS Interbody Fusion Device, Endoskeleton TL Interbody Fusion Device
Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP, OVD, MAX, OVE Dated: July 26, 2019 Received: July 29, 2019
Dear Christine Scifert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@tda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Brent Showalter -S
for Melissa Hall Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
#### K192018
#### Device Name
ENDOSKELETON® TA Interbody Fusion Device
#### Indications for Use (Describe)
The ENDOSKELETON® TA Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogence back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the device may be used with supplemental fixation that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). ENDOSKELETON® TA is indicated to be used with autograft bone and/or allograft bone comprised of cancellous bone and/or demineralized allograft bone with bone marrow aspirate.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{3}------------------------------------------------
510(k) Number (if known)
K192018
#### Device Name
ENDOSKELETON® TAS Interbody Fusion device
#### Indications for Use (Describe)
The ENDOSKELETON® TAS Interbody Fusion device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogence back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients should have received 6 months of nonoperative treatment prior to treatment with the device is a standalone system that is intended to be used with the bone screws provided and requres no additional supplementary fixation. The Device is indicated to be used with autograft bone and/or allograft bone comprised of cancellous bone and/or demineralized allograft bone with bone marrow aspirate.
Hyperlordotic Devices ≥16°: The ENDOSKELETON® TAS Hyperlordotic Interbody Fusion Device (≥ 16°) is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). Patients should have received 6 months of non-operative treatment with the devices. The device is indicated to be used with autograft bone and/or allograft bone comprised of cancellous bone and/ or demineralized allograft bone with bone marrow aspirate.The ENDOSKELETON® TAS Hyperlordotic Interbody Fusion Device must be used with a posterior supplemental internal spinal fixation that has been cleared by the FDA for use in the lumbar spine must be used with a posterior supplemental inxation that has been cleared by the FDA for use in the lumbar spine.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| <div><span style="text-decoration: overline;">区</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{4}------------------------------------------------
510(k) Number (if known)
K192018
Device Name ENDOSKELETON® TC
#### Indications for Use (Describe)
The ENDOSKELETON® TC is indicated for use for anterior cervical interbody fusion in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine at one disc level from C-3 to C-7. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have received 6 weeks of non-operative treatment prior to treatment with the device. The ENDOSKELETON® TC is indicated to be used with supplemental fixation that has been cleared by the FDA for use in the cervical spine and be used with autograft bone and/ or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow aspirate.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------|-----------------------------------------|
| <span style="text-decoration: underline;">Prescription Use</span> (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span style="font-size: 20px;">☐</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{5}------------------------------------------------
510(k) Number (if known)
K192018
Device Name ENDOSKELETON® TCS
#### Indications for Use (Describe)
The ENDOSKELETON® TCS System is an anterior cervical intervertebral body fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) (defined as pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) at one disc level from C2 to T1. Patients should have received 6 weeks of non-operative treatment with the device. The device is indicated to be used with autograft bone and/or allograft bone comprised of cancellous bone and/or demineralized allograft bone with bone marrow aspirate. The device is a stand-alone system when used with ENDOSKELETON® TCS integrated screws and when used without the integrated screws it requires additional supplemental fixation cleared for the cervical spine.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{6}------------------------------------------------
510(k) Number (if known)
K192018
Device Name ENDOSKELETON® TL
#### Indications for Use (Describe)
The ENDOSKELETON® TL Interbody Fusion Device is indicated for use in spinal fusion procedures in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment with the devices. The device must be used with supplemental fixation systems that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients with previous non-fusion spinal surgery at the involved level(s) may be treated with the device. It is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow aspirate.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{7}------------------------------------------------
510(k) Number (if known)
K192018
Device Name ENDOSKELETON® TO
#### Indications for Use (Describe)
The ENDOSKELETON® TO Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogence back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the devices. The device must be used with supplemental fixation that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). It is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone marrow aspirate.
Type of Use (*Select one or both, as applicable*)> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{8}------------------------------------------------
510(k) Number (if known)
K192018
Device Name ENDOSKELETON® TT
#### Indications for Use (Describe)
The ENDOSKELETON® TT Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogence back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non- operative treatment prior to treatment with the devices. The device must be used with supplemental fixation that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). It is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone marrow aspirate.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{9}------------------------------------------------
| Company: | Titan Spine, Inc.<br>6140 West Executive Drive, Suite A<br>Mequon, WI 53092, USA |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>Registration: | 3006340236 |
| Primary Contact: | Christine Scifert<br>Phone: 901-831-8053 |
| Company Contact: | Jane Rodd<br>Phone: 866-822-7800<br>Fax: 262-242-7802 |
| Trade Name: | Endoskeleton® TA Interbody Fusion Device<br>Endoskeleton® TAS and TAS Hyperlordotic Interbody Fusion Devices<br>Endoskeleton® TO Interbody Fusion Device<br>Endoskeleton® TT Interbody Fusion Device<br>Endoskeleton® TC Interbody Fusion Device<br>Endoskeleton® TCS Interbody Fusion Device<br>Endoskeleton® TL Interbody Fusion Device |
| Common Name: | Intervertebral Fusion Device With Integrated Fixation, Lumbar<br>Intervertebral Fusion Device With Bone Graft, Cervical<br>Intervertebral Fusion Device With Integrated Fixation, Cervical<br>Intervertebral Fusion Device With Bone Graft, Lumbar<br>Spinal Vertebral Body Replacement Device |
| Classification: | Class II |
| Regulation Number: | 21 CFR 888.3080 (Intervertebral body fusion device) |
| Panel: | 87- Orthopedic |
| Product Code: | ODP, OVD, OVE, MAX |
| Predicate Devices: | Primary Predicate:<br>Medtronic Sofamor Danek USA, Inc. ARTiC-L™ 3D and ARTiC-XL™<br>3D Ti Spinal System – K190959 |
| Additional Predicates: | Endoskeleton System – K183557<br>Endoskeleton® TAS and TCS - K173535<br>Endoskeleton® TO - K170399 |
{10}------------------------------------------------
Endoskeleton® System (TA IBD and VBR, TO, TT, TAS, TC, TL) - K141953 Endoskeleton® TAS and TCS - K173535 Endoskeleton® TO – K170399 Endoskeleton TCS nanoLOCK - K153122 K141953
#### Device Description:
The current Endoskeleton® System is an interbody and vertebral body family comprised of a variety of sizes and geometries to accommodate various patient anatomy and pathology. All implantable components are manufactured from medical grade titanium alloy (Ti6Al4V-ELI). The center of the implant is hollow and is to be filled with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone marrow aspirate. The design incorporates "windows" through the implant to permit visualization of the graft material and, over time, formation of new bone. The superior and inferior surfaces include either the Chemtex® surface treatment or nanoLOCK® surface treatment (MMNTM) designed to improve fixation to the adjacent bone. The nanoLOCK® surface technology (MMN™) provides a microscopic roughened surface with nano-scale features. This submission seeks to expand the indications of these devices to include use with demineralized allograft bone with bone marrow aspirate.
The predicate Endoskeleton® System (K173535. K170399. K153122. and K141953) is provided either non-sterile or sterile via gamma irradiation.
#### Indication for Use:
The ENDOSKELETON® TA Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment with the devices. The device may be used with supplemental fixation that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). ENDOSKELETON® TA is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow aspirate.
The ENDOSKELETON® TAS Interbody Fusion device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). Patients should have received 6 months of non-operative treatment with the devices. The device is a standalone system that is intended to be used with the bone screws provided and requires no additional supplementary fixation. The Device is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow aspirate.
Hyperlordotic Devices ≥16°: The ENDOSKELETON® TAS Hyperlordotic Interbody Fusion Device (≥ 16°) is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc
{11}------------------------------------------------
confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients should have received 6 months of nonoperative treatment prior to treatment with the device is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow aspirate.The ENDOSKELETON® TAS Hyperlordotic Interbody Fusion Device must be used with a posterior supplemental internal spinal fixation that has been cleared by the FDA for use in the lumbar spine must be used with a posterior supplemental internal spinal fixation that has been cleared by the FDA for use in the lumbar spine.
The ENDOSKELETON® TO Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment with the devices. The device must be used with supplemental fixation that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). It is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow asprate.
The ENDOSKELETON® TT Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non- operative treatment with the devices. The device must be used with supplemental fixation that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). It is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow aspirate.
The ENDOSKELETON® TC is indicated for use for anterior cervical interbody fusion in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine at one disc level from C-3 to C-7. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have received 6 weeks of non-operative treatment with the device. The ENDOSKELETON® TC is indicated to be used with supplemental fixation that has been cleared by the FDA for use in the cervical spine and be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow aspirate.
The ENDOSKELETON® TCS System is an anterior cervical intervertebral body fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) (defined as pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) at one disc level from C2 to T1. Patients should have received 6 weeks of non-operative treatment with the device. The device is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow asprate. The device is a stand-alone system when used with ENDOSKELETON® TCS integrated screws and when used without the integrated screws it requires additional supplemental fixation cleared for the cervical spine.
{12}------------------------------------------------
The ENDOSKELETON® TL Interbody Fusion Device is indicated for use in spinal fusion procedures in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the devices. The device must be used with supplemental fixation that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients with previous non-fusion spinal surgery at the involved level(s) may be treated with the device. It is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow aspirate.
### Substantial Equivalence:
Titan Spine has 14 previous clearances related to the predicate Endoskeleton® systems. The most recent clearances for each sub-family are provided below for the primary and secondary predicates:
| Primary Predicate: | Medtronic Sofamor Danek USA, Inc. ARTiC-L TM 3D and ARTiC-XLTM<br>3D Ti Spinal System – K190959 |
|-----------------------|-------------------------------------------------------------------------------------------------------------|
| Secondary Predicates: | Endoskeleton® TAS and TCS – K173535 |
| | Endoskeleton® TO – K170399 |
| | Endoskeleton® System (TA IBD and VBR, TO, TT, TAS, TC, TL) – K141953 |
| | Endoskeleton TCS nanoLOCK – K153122 |
The modification for this Special 510(k) relates only to indications for use. The intended use, material, surgical technique, surface treatment, sterility, and design of the subject devices are the same as predicate devices. The only change that is the subject of this submission is to change the indication for use with demineralized allograft bone with bone marrow aspirate.
The purpose of this Special 510(k) is to modify the original indications of the Ensoskeleton® System. This submission seeks to clarify that demineralized HCT/Ps (e.g. Grafton™ DBM DBF) combined with bone marrow aspirate may be used as an alternative to autograft and/or allograft bone comprised of cancellous and/or corticocancellous bone. The review of the literature concluded that there were no additional risks due to the modification of the indications of these devices to include use with demineralized allograft bone with bone marrow aspirate and that these devices are substantially equivalent to the predicate devices.
### Technological Characteristics:
The modification for this Special 510(k) relates only to the indications for use. The intended use, material, surgical technique, surface treatment, sterility, and design of the subject devices are the same as predicate devices.
#### Performance Testing:
No performance testing was required or performed, as this modification for this Special 510(k) relates only to the indications for use.
{13}------------------------------------------------
#### Biocompatibility:
There are no changes to the Endoskeleton® System materials. The devices are patient contacting and are implanted for durations greater than 30 days. All implant components of the Endoskeleton® devices are made from medical grade titanium (Ti6A14V-ELI) alloy per ASTM F-136/ISO 5832-3 and ASTM F-3001.
#### Conclusion:
Based upon the clinical literature contained in this submission and the similarities of the subject and predicate devices, the subject Endoskeleton® system is substantially equivalent to the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.