K131449 · Globus Medical, Inc. · OVE · Jul 30, 2013 · Orthopedic
Device Facts
Record ID
K131449
Device Name
COALITION SPACER
Applicant
Globus Medical, Inc.
Product Code
OVE · Orthopedic
Decision Date
Jul 30, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The COALITION® Spacer is a stand-alone interbody fusion device intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) at one level. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of nonoperative treatment. The COALITION® Spacer is to be filled with autogenous bone graft material, and is to be used with two titanium alloy screws which accompany the implant.
Device Story
COALITION® Spacer is a stand-alone cervical interbody fusion device; provides structural stability post-discectomy. Device features radiolucent polymer body with titanium alloy or tantalum markers; superior/inferior surface protrusions grip vertebral endplates to resist expulsion. Implant filled with autogenous bone graft; secured to adjacent vertebrae via two titanium alloy screws inserted through anterior titanium portion. Used in clinical settings by surgeons; provides mechanical stabilization to facilitate fusion. Benefits patients by restoring disc space height and promoting bony fixation in DDD cases.
Clinical Evidence
Bench testing only. Mechanical testing included static and dynamic compression, static and dynamic compression-shear, and expulsion testing conducted in accordance with ASTM F2077 and the Class II Special Controls Guidance Document for Intervertebral Fusion Devices.
Technological Characteristics
Materials: Radiolucent polymer, titanium alloy (ASTM F136, F1295), tantalum markers (ASTM F560). Design: Stand-alone interbody spacer with integrated screw fixation. Dimensions: Various heights/geometries. Sterilization: Not specified.
Indications for Use
Indicated for skeletally mature patients with one-level cervical (C2-T1) degenerative disc disease (DDD) presenting with discogenic pain confirmed by history and radiography, following at least 6 weeks of failed nonoperative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
K153629 — ArcadiusXP C Spinal System · Aesculap Implant Systems, Inc. · May 25, 2016
K170455 — MectaLIF Anterior Stand Alone · Medacta International S.A. · Mar 16, 2017
K203129 — NexGen Standalone Anterior Cervical Discectomy and Fusion (ACDF) System · Precision Spine, Inc. · Jan 12, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
K131449
# 510(k) SUMMARY: COALITION® Spacer
JUL 3 0 2013
| Company: | Globus Medical Inc. |
|----------|-----------------------------|
| | 2560 General Armistead Ave. |
| | Audubon, PA 19403 |
| | (610) 930-1800 |
Christina Kichula Contact: Group Manager, Regulatory Affairs
Date Prepared: May 17, 2013
Device Name: COALITION® Spacer
Classification: Per 21 CFR as follows: §888.3080 Intervertebral Body Fusion Devices. Product Code OVE. Requlatory Class II. Panel Code: 87.
COALITION® Spacer (K083389) Predicate(s): Synthes® Zero-P (K112459)
#### Purpose:
The purpose of this submission is to request clearance for additional sizes of COALITION® Spacers.
#### Device Description:
The COALITION® Spacer is a stand-alone cervical interbody fusion device used to provide structural stability in skeletally mature individuals following discectorny. The spacers are available in various heights and geometric options to fit the anatomical needs of a wide variety of patients. Protrusions on the superior and inferior surfaces of each device grip the endplates of the adjacent vertebrae to aid in expulsion resistance. Screws are inserted through the anterior titanium portion of the implant into adjacent vertebral bodies for bony fixation. The spacer is to be filled with autogenous bone graft material.
The COALITION® Spacer is made from radiolucent polymer, with titanium alloy or tantalum markers, as specified in ASTM F2026, F136, F1295, and F560. The anterior portion of the implant and the mating screws are manufactured from titanium alloy, as specified in ASTM F136 and F1295.
#### Intended Use:
The COALITION® Spacer is a stand-alone interbody fusion device intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) at one level. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These
{1}------------------------------------------------
patients should be skeletally mature and have had at least six (6) weeks of nonoperative treatment. The COALITION® Spacer is to be filled with autogenous bone graft material, and is to be used with two titanium alloy screws which accompany the implant.
#### Technological Characteristics:
The technological characteristics of the additional COALITION® Spacer are similar to the predicate devices in terms of design, dimensions, intended use, materials, and performance characteristics.
#### Performance Data:
Mechanical testing consisting of static and dynamic compression, static and dynamic compression-shear, and expulsion was conducted in accordance with "Class II Special Controls Guidance Document: Intervertebral Fusion Device," June 12, 2007, and ASTM F2077 to demonstrate substantial equivalence to the predicate systems.
#### Basis for Substantial Equivalence:
The COALITION® Spacer additional implants are similar to the predicate systems with respect to technical characteristics, performance and intended use. The information provided within this premarket notification supports substantial equivalence of the subject spacers to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The caduceus is black, and the text is also black. The logo is simple and recognizable, representing the department's mission to protect the health of all Americans.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
July 30, 2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, Mi) 20993-0002
Globus Medical Incorporated % Ms. Christina Kichula Group Manager. Regulatory Affairs 2560 General Armistead Avenue Audubon. Pennsvivania 19403
Re: K131449
Trade/Device Name: COALITION" Spacer Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVE Dated: May 17, 2013 Received: May 20, 2013
Dear Ms. Kichula:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 -- Ms. Christina Kichula
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the requiation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Mark N. Melkerson - S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
## Indications for Use Statement
510(k) Number:
K131449
Device Name:
COALITION® Spacer Commen
### Indications:
The COALITION® Spacer is a stand-alone interbody fusion device intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) at one level. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of nonoperative treatment. The COALITION® Spacer is to be filled with autogenous bone graft material, and is to be used with two titanium alloy screws which accompany the implant.
X_____________________________________________________________________________________________________________________________________________________________________________ Prescription Use (Per 21 CFR §801.109)
OR
Over-The-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anton E. Dmitriev, PhD Division of Orthopedic Devices
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.