ProLift Lateral Fixated
K200338 · Life Spine, Inc. · OVD · Apr 17, 2020 · Orthopedic
Device Facts
| Record ID | K200338 |
| Device Name | ProLift Lateral Fixated |
| Applicant | Life Spine, Inc. |
| Product Code | OVD · Orthopedic |
| Decision Date | Apr 17, 2020 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3080 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The ProLift® Lateral Fixated is an interbody fusion device intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The ProLift Lateral Fixated System is to be filled with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone, may be used with two bone screws, and is to be used with supplemental fixation.
Device Story
Intervertebral body fusion device; titanium alloy (Ti-6Al-4V ELI) implant; available in various sizes/footprints. Implant packed with autograft/allogenic bone graft to promote fusion; superior/inferior surface teeth prevent rotation/migration. Includes variable and fixed configuration bone screws (5.5mm/6.5mm diameter; 25-60mm length). Used in lumbosacral spine (L2-S1) for DDD; requires supplemental fixation. Surgeon selects implant size based on patient anatomy/pathology. Benefits patient by facilitating spinal fusion in symptomatic DDD cases.
Clinical Evidence
Bench testing only. Static compression, dynamic compression, static shear, and dynamic shear testing performed according to ASTM F2077 to demonstrate substantial equivalence.
Technological Characteristics
Titanium alloy (Ti-6Al-4V ELI) per ASTM F136; interbody fusion implant with integrated bone screws; variable and fixed screw configurations; surface teeth for fixation; mechanical performance per ASTM F2077.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1), including up to Grade 1 spondylolisthesis or retrolisthesis, who have failed at least 6 months of non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
- Life Spine ProLift Expandable System (K191005)
- Life Spine ProLift Expandable System (K173182)
- Globus Elsa (K161379)
- Life Spine Sentry (K180166)
Related Devices
- K201500 — Life Spine Plateau-A Ti Anterior Lumbar Spacer & Plateau-X Ti Lateral Lumbar Spacer · Life Spine, Inc. · Aug 4, 2020
- K242826 — ProLift Wedge Expandable Spacer System · Life Spine, Inc. · Oct 8, 2024
- K171329 — Altus Spine Titanium Interbody Fusion System · Altus Partners, LLC · Aug 29, 2017
- K182470 — Plateau Spacer System · Life Spine · Nov 26, 2018
- K130630 — PLATEAU SPACER SYSTEM TITANIUM · Life Spine, Inc. · Jun 13, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
April 17, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Life Spine Inc. Angela Batker RA/QA Manager 13951 S Quality Drive Huntley, Illinois 60142
Re: K200338
Trade/Device Name: ProLift® Lateral Fixated Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVD, MAX Dated: February 7, 2020 Received: February 11, 2020
Dear Angela Batker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Brent Showalter, Ph.D. Assistant Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200338
Device Name ProLift® Lateral Fixated
#### Indications for Use (Describe)
The ProLift® Lateral Fixated is an interbody fusion device intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The ProLift® Lateral Fixated is to be filled with autograft and/ or allogenic bone graft comprised of cancellous and/or corticocancellous bone, may be used with two bone screws, and is to be used with supplemental fixation.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary ProLift® Lateral Fixated
| Submitted By: | Life Spine, Inc.<br>13951 S. Quality Drive<br>Huntley, IL 60142<br>Telephone: 847-884-6117<br>Fax: 847-884-6118 |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact: | Angela Batker<br>Life Spine, Inc.<br>13951 S. Quality Drive<br>Huntley, IL 60142<br>Telephone: 847-884-6117<br>Fax: 847-884-6118 |
| Date Prepared: | April 13th, 2020 |
| Trade Name: | ProLift® Lateral Fixated |
| Common Name: | Intervertebral Fusion Device With Integrated Fixation, Lumbar |
| Classification: | OVD, CFR 888.3080, Class II<br>MAX, CFR 888.3080, Class II |
| Primary Predicate: | Life Spine ProLift Expandable System (K191005) |
| Additional Predicate: | Life Spine ProLift Expandable System (K173182)<br>Globus Elsa (K161379)<br>Life Spine Sentry (K180166) |
## Device Description:
The ProLift® Lateral Fixated is intended to serve as an intervertebral body fusion device. The implant is available in a range of sizes and footprints to suit the individual pathology and anatomical conditions of the patient. It is fabricated and manufactured from titanium alloy (Ti-6AI-4V ELI) as described by ASTM F136. The implant allows packing of autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone to help promote fusion. The superior and inferior surfaces have teeth to assist in the interface with the vertebral bodies to prevent rotation and/or migration. The screws are manufactured in variable and fixed configurations with diameters of 5.5mm and 6.5mm and lengths of 25mm-60mm. The responsible surgeon will determine the correct size of the implant in accordance with the size of the individual patient.
{4}------------------------------------------------
## Intended Use of the Device:
The ProLift® Lateral Fixated is an interbody fusion device intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The ProLift Lateral Fixated System is to be filled with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone, may be used with two bone screws, and is to be used with supplemental fixation.
# Technological Characteristics:
The ProLift® Lateral Fixated is substantially equivalent to the predicate systems in terms of design, materials, indications for use and sizing.
#### Material:
This submission seeks clearance of a device made from titanium alloy (Ti-6Al-4V ELI) as described by ASTM F136. This this is the same material used in the predicate devices.
#### Performance Data:
Static compression, dynamic compression, static shear and dynamic shear testing according to ASTM F2077, was presented to demonstrate the substantial equivalency of the Pro-Lift Expandable System (K191005).
#### Substantial Equivalence:
The ProLift® Lateral Fixated was shown to be substantially equivalent to the predicate devices in indications for use, design, function, materials used and mechanical performance.
#### Conclusion:
The information presented demonstrates the substantial equivalency of The ProLift® Lateral Fixated.