Plateau Spacer System

K182470 · Life Spine · MAX · Nov 26, 2018 · Orthopedic

Device Facts

Record IDK182470
Device NamePlateau Spacer System
ApplicantLife Spine
Product CodeMAX · Orthopedic
Decision DateNov 26, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Plateau Spacer System is intended for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). It is to be used in patients who have had at least six months of non-operative treatment. This device is intended to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone and supplemental internal spstem (e.g., pedicle screw or anterolateral plating system) that is cleared for use in the lumbosacral spine.

Device Story

Intervertebral body fusion device; provides structural support to disc space during spinal fusion. Implant features hollow center for bone graft packing; superior/inferior surface teeth prevent rotation/migration. Available in various sizes/footprints to match patient anatomy. Used in lumbosacral spine procedures; requires supplemental internal fixation (e.g., pedicle screws). Operated by surgeons in clinical/OR settings. Benefits patient by promoting fusion in DDD cases.

Clinical Evidence

Bench testing only. Testing per ASTM F2077 and F2267 included static axial compression, static expulsion, static subsidence, and dynamic axial compression to demonstrate mechanical equivalence.

Technological Characteristics

Materials: PEEK-OPTIMA LT1 or Zeniva PEEK ZA-500 (ASTM F2026) or Titanium Ti-6Al-4V ELI (ASTM F136). Features tantalum markers. Design: Hollow interbody spacer with surface teeth for endplate interface. Single-use.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1), including those with up to Grade 1 spondylolisthesis, who have failed at least six months of non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ November 26, 2018 Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text. Life Spine Ms. Angela Batker RA/OA Specialist 13951 S Quality Drive Huntley, Illinois 60142 Re: K182470 Trade/Device Name: Plateau Spacer System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: September 6, 2018 Received: September 10, 2018 Dear Ms. Batker: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Melissa Hall -S For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K182470 Device Name Plateau Spacer System #### Indications for Use (Describe) The Plateau Spacer System is intended for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). It is to be used in patients who have had at least six months of non-operative treatment. This device is intended to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone and supplemental internal spstem (e.g., pedicle screw or anterolateral plating system) that is cleared for use in the lumbosacral spine. Type of Use (Select one or both, as applicable) | <span></span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |--------------------------------------------------------------|-------------------------------------------------------------| |--------------------------------------------------------------|-------------------------------------------------------------| #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## K182470 510(k) Summary Life Spine Plateau Spacer System | Submitted By: | Life Spine, Inc.<br>13951 S. Quality Drive<br>Huntley, IL 60142<br>Telephone: 847-884-6117<br>Fax: 847-884-6118 | |----------------------|----------------------------------------------------------------------------------------------------------------------------------| | 510(k) Contact: | Angela Batker<br>Life Spine, Inc.<br>13951 S. Quality Drive<br>Huntley, IL 60142<br>Telephone: 847-884-6117<br>Fax: 847-884-6118 | | Date Prepared: | September 9, 2018 | | Trade Name: | Plateau Spacer System | | Common Name: | Interbody Fusion Device | | Classification: | MAX, CFR 888.3080, Class II | | Primary Predicate: | Life Spine Plateau (K080411/K111569) | | Secondary Predicate: | Life Spine Plateau Ti (K130630)<br>Life Spine Plateau Lo (K172105)<br>Life Spine Pro Lift (K153400) | ### Device Description: The Plateau-V Spacer System is intended to serve as an intervertebral body fusion device. The implant is available in a range of sizes and footprints to suit the individual pathology and anatomical conditions of the patient. It is fabricated and manufactured from either PEEK-OPTIMA LT1 or Zeniva PEEK ZA-500 with tantalum markers. The implant is hollow to permit packing autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone to help promote intervertebral body fusion. The superior and inferior surfaces have teeth to assist in the interface with the vertebral endplates to prevent rotation and/or migration. {4}------------------------------------------------ All implants are intended for single use only and should not be reused under any circumstances. Do not use any of the Life Spine Plateau Spacer System components with components from any other system or manufacturer. The Life Spine Spacer System components should never be reused under any circumstances. ### Intended Use of the Device: The Plateau Spacer System is intended for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). It is to be used in patients who have had at least six months of non-operative treatment. This device is intended to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone and supplemental internal spine fixation system (e.g., pedicle screw or anterolateral plating system) that is cleared for use in the lumbosacral spine. #### Technological Characteristics: The Life Spine Plateau Spacer System is substantially equivalent to the predicate systems in terms of design, materials, indications for use and sizing. ### Material: This submission seeks clearance of a device made from either PEEK-OPTIMA LT1 or Zeniva PEEK ZA-500 per ASTM F2026 or Titanium (Ti-6A1-4V ELI) per ASTM F136. This this is the same material used in the predicate devices. ### Performance Data: Testing according to ASTM F2077 & F2267 included Static axial compression, static expulsion, static subsidence, and dynamic axial compression tests was presented to demonstrate the substantial equivalency of the Life Spine Plateau (K080411/K111569). ### Substantial Equivalence: The Life Spine Plateau Spacer System was shown to be substantially equivalent to the predicate devices in indications for use, design, function, materials used and mechanical performance. ### Conclusion: The information presented demonstrates the substantial equivalency of The Life Spine Plateau Spacer System.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.