T-Rex Standalone ALIF

K170855 · Siats, LLC · OVD · Sep 7, 2017 · Orthopedic

Device Facts

Record IDK170855
Device NameT-Rex Standalone ALIF
ApplicantSiats, LLC
Product CodeOVD · Orthopedic
Decision DateSep 7, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Intended Use

When used with the bone screws, the T-Rex Intervertebral Body Fusion Device is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the interior of the interior of the interbody spacer component may be packed with autogenous bone graft and/or allogeneic bone graft material composed of cancellous and/ or corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device.

Device Story

T-Rex Standalone ALIF is a lumbar intervertebral body fusion device with integrated fixation; used as an adjunct to spinal fusion. Device consists of HA-enhanced PEEK-OPTIMA spacer and titanium bone screws; includes tantalum radiographic markers. Implanted by surgeons in clinical settings to stabilize spinal segments in patients with degenerative disc disease. Spacer interior packed with autogenous or allogeneic bone graft to promote fusion. Various heights, widths, lengths, and lordotic angles available to match patient anatomy. Provides mechanical stability to the intervertebral space; facilitates fusion process; reduces back pain associated with discogenic origin. Provided sterile; surgical instruments provided non-sterile.

Clinical Evidence

Bench testing only. No clinical data provided. Performance demonstrated via mechanical testing per ASTM F2077 (static/dynamic compression, compression shear) and ASTM F2267 (subsidence), plus expulsion and screw push-out testing.

Technological Characteristics

Materials: HA-enhanced PEEK-OPTIMA, titanium screws, tantalum markers (ASTM F560). Form factor: Intervertebral spacer with integrated fixation screws. Energy source: None (mechanical). Sterilization: Provided sterile. Connectivity: None.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1, including those with up to Grade 1 spondylolisthesis. Requires at least 6 months of prior non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 7, 2017 SIATS, LLC % Allison C. Komivama. Ph.D., R.A.C. Principal Consultant AcKnowledge Regulatory Strategies, LLC 2834 Hawthorn Street San Diego, California 92104 Re: K170855 Trade/Device Name: T-Rex Standalone ALIF Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVD Dated: July 25, 2017 Received: July 26, 2017 Dear Dr. Komiyama: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Image /page/0/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. {1}------------------------------------------------ Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. # Katherine D. Kavlock -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K170855 Device Name T-Rex Standalone ALIF #### Indications for Use (Describe) When used with the bone screws, the T-Rex Intervertebral Body Fusion Device is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the interior of the interior of the interbody spacer component may be packed with autogenous bone graft and/or allogeneic bone graft material composed of cancellous and/ or corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. #### Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/2 description: The image shows a logo with the word "SIATS" in orange letters. The "S" is stylized to resemble the head of a dinosaur with its mouth open, revealing sharp teeth. A blue and purple swoosh design wraps around the letters, adding a dynamic element to the logo. # 510(k) Summary K170855 DATE PREPARED July 25, 2017 #### MANUFACTURER AND 510(k) OWNER SIATS, LLC 6316 E 102™ Street, Tulsa, OK 74137, USA Telephone: Official Contact: Robert Compton, CEO #### REPRESENTATIVE/CONSULTANT Allison C. Komiyama, Ph.D., R.A.C. AcKnowledge Regulatory Strategies, LLC Telephone: +1 (619) 208-7888 Email: akomiyama@acknowledge-rs.com #### PROPRIETARY NAME OF SUBJECT DEVICE T-Rex Standalone ALIF #### COMMON NAME Intervertebral Fusion Device with Integrated Fixation, Lumbar #### DEVICE CLASSIFICATION Intervertebral body fusion device (21 CFR 888.3080, Product Code OVD Class II) #### PREMARKET REVIEW ODE/DOD/ASDB Orthopedic Panel #### INDICATIONS FOR USE When used with the bone screws, the T-Rex Intervertebral Body Fusion Device is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the interbody spacer component may be packed with autogenous bone graft and/or allogeneic bone graft material composed of cancellous and/or corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. {4}------------------------------------------------ Image /page/4/Picture/2 description: The image shows the word "SIATS" in orange letters with a dinosaur head as the "S". The letters are in a gradient from orange to yellow. The dinosaur head is orange with white teeth. A blue swoosh surrounds the word and dinosaur head. # DEVICE DESCRIPTION The T-Rex™ Standalone ALIF is a lumbar Intervertebral Body Fusion Device with integrated fixation made from HA Enhanced PEEK-OPTIMA (HAPEEK) incorporating titanium bone screws. The implants are available in various heights, widths, lengths, and angles of lordosis to accommodate patients' anatomy. The implants are provided sterile with surgical instruments provided non-sterile to facilitate implantation. Radiographic markers made of tantalum (ASTM F-560), are included in each implant to allow radiographic visualization. ### PREDICATE DEVICE IDENTIFICATION The T-Rex Standalone ALIF is substantially equivalent to the following predicates: | 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate | |---------------|-------------------------------------------------------------------------|-------------------| | K141942 | Midline II / Midline II / Centinel Spine Inc. | ✓ | | K151785 | Px HA PEEK IBF System / Innovasis, Inc. | | | K162351 | SeaSpine® Vu a•POD™ Prime NanoMetalene® /<br>SeaSpine Orthopedics Corp. | | | K170511 | NeoFuse HA Enhanced PLIF/TLIF / HT Medical, LLC | | # SUMMARY OF NON-CLINICAL TESTING No FDA performance standards have been established for the T-Rex Standalone ALF. The following tests were performed in order to demonstrate safety based on current industry standards: - . Static and dynamic compression and compression shear (per ASTM F2077) - Subsidence (per ASTM F2267) . - . Expulsion - o Push out test on the screws The results of these tests indicate that the T-Rex Standalone ALIF is substantially equivalent to the predicate devices. #### EQUIVALENCE TO PREDICATE DEVICES SIATS, LLC believes that the T-Rex Standalone ALIF is substantially equivalent to the predicate devices based on the information summarized here: The subject device has similar dimensions, similar materials and the same indications for use as the devices cleared in K141942 and K162351. It uses the same materials as the device cleared in K151785 and similar manufacturing processes as the device cleared in K170511. The subject device has a similar design and similar technological characteristics to the devices cleared in K141942 and K162351. # CONCLUSION The T-Rex Standalone ALIF is considered substantially equivalent to the predicate devices based on the testing performed, the similar indications for use, and equivalent technological characteristics. Based on the testing performed, it can be concluded that the subject device is as safe and as effective compared to the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...