When used with the bone screws, the Theken Spine Vu aPOD Intervertebral Body Fusion Device is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graft material. When used with the SpinPlate, the Theken Spine Vu aPOD Intervertebral Body Fusion Device is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graft material. When used with the SpinPlate, the Theken Spine Vu aPOD Intervertebral Body Fusion Device is intended for use with supplemental fixation such as the Coral Spinal System or the BodyForm Thoracolumbar Fixation System. The Theken Spine Vu aPOD Intervertebral Body Fusion Device, when used with bone screws, is a stand alone device. If the Theken Spine Vu aPod Intervertebral Body Fusion Device is used with the SpinPlate then additional supplemental fixation, which has been cleared by the FDA for use in the lumbar spine, must be used to augment stability. The SpinPlate and bone screws are not intended to be used together. This device is intended to be used with autogenous bone graft. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device.
Device Story
Lumbar intervertebral body fusion device; consists of PEEK OPTIMA LT1 polymer spacers and Titanium alloy (Ti-6Al-4V ELI) bone screws or SpinPlate; includes tantalum radiographic markers. Implanted by surgeons in clinical settings to facilitate spinal fusion in DDD patients. Used with autogenous bone graft. When used with bone screws, functions as stand-alone device; when used with SpinPlate, requires additional cleared supplemental fixation. Provides structural support to intervertebral space; promotes fusion; stabilizes spinal segment. Benefits patients by addressing discogenic back pain.
Clinical Evidence
Bench testing only. No clinical data presented. Performance verified via mechanical testing per ASTM F2077 and ASTM F2267 standards.
Technological Characteristics
Spacer: PEEK OPTIMA LT1 (ASTM F2026). Screws/SpinPlate: Titanium alloy Ti-6Al-4V ELI (ASTM F136). Markers: Tantalum (ASTM F560). Mechanical fixation via screws or plate. Lumbar interbody application.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1, including those with up to Grade 1 spondylolisthesis, who have failed at least 6 months of non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
# 510(k) Summary
| Manufacturer: | Theken Spine, LLC | SEP 02 2010 | |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|--|
| | 1800 Triplett Blvd<br>Akron, OH 44306 | 14101310 | |
| Device Trade Name: | Theken Spine Vu aPOD Intervertebral Body Fusion Device | | |
| Contact: | Glenn Stiegman<br>Vice President, Regulatory Affairs<br>Musculoskeletal Clinical Regulatory Advisers, LLC<br>1331 H Street NW, 12th Floor<br>Washington, DC 20005<br>Office: 202.552.5800<br>Fax: 202.552.5798 | | |
| Date Prepared: | August 5, 2010 | | |
| Classification: | §888.3080, Intervertebral body fusion device | | |
| Class: | II | | |
| Product Code: | MAX | | |
# Indications For Use:
When used with the bone screws, the Theken Spine Vu aPOD Intervertebral Body Fusion Device is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graft material.
When used with the SpinPlate, the Theken Spine Vu aPOD Intervertebral Body Fusion Device is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graft material. When used with the SpinPlate, the Theken Spine Vu aPOD Intervertebral Body Fusion Device is intended for use with supplemental fixation such as the Coral Spinal System or the BodyForm Thoracolumbar Fixation System.
{1}------------------------------------------------
The Theken Spine Vu aPOD Intervertebral Body Fusion Device, when used with bone screws, is a stand alone device. If the Theken Spine Vu aPod Intervertebral Body Fusion Device is used with the SpinPlate then additional supplemental fixation, which has been cleared by the FIDA for use in the lumbar spine, must be used to augment stability. The SpinPlate and bone screws are not intended to be used together. This device is intended to be used with autogenous bone graft. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device.
## Device Description:
The Vu aPOD Intervertebral Body Fusion Device consists of lumbar spinal interbody fusion devices as well as instrumentation designed specifically for the implantation of these devices. The Vu aPOD spacers are manufactured from PEEK OPTIMA LT1 polymer per ASTM F2026 while the bone screws and SpinMate are comprised of Titanium alloy (Ti-6Al-4V ELI) per ASTM F136. Radiographic markers present with the Vu aPOD spacers are comprised of tantalum per ASTM F560. The Vu aPOD Intervertebral Body Fusion Device is for lumbar spinal use at one or two contiguous levels from the L2-L3 to L5-S1 disc levels.
# Predicate Device(s):
The Vu aPOD Intervertebral Body Fusion Device was shown to be substantially equivalent to previously cleared devices, including the Theken Spine Vu aPOD System (K080822), Titan Spine Endoskeleton TA (K080615), Spinal Elements Lucent Magnum + (K083475), and I.DR Spine ROI-A (K082262), and has the same indications for use, design, and function.
#### Performance Standards:
Preclinical testing has been performed per ASTM F2077 (static axial compression, static compression-shear, static torsion, dynamic axial compression, dynamic compressionshear, expulsion) and ASTM F2267 (static subsidence) indicating that the Vu alPOD Intervertebral Body Fusion Device is substantially equivalent to predicate devices.
# Conclusion:
Sufficient information, including extensive testing, has been presented to demonstrate the Vu aPOD Intervertebral Body Fusion Device is substantially equivalent to predicate devices with the same indications, intended use, and technological features.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is enclosed in a circle, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
SEP 12 2011
Theken Spine, LLC % Musculoskeletal Clinical Regulatory Advisers, LLC Mr. Glenn Stiegman Vice President, Regulatory Affairs 1331 H Street NW, 12th floor Washington, DC 20005
Re: K101310
Trade/Device Name: Vu a·POD Intervertebral Body Fusion Device Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVD Dated: August 12, 2010 Received: August 13, 2010
Dear Mr. Stiegman:
This letter corrects our substantially equivalent letter of September 2, 2010.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other
{3}------------------------------------------------
requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, by ont rot limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions. (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket photifics: " now (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
**Sincerely yours,**
Mark A. Millbern
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
SEP 02 2010
### 510(k) Number (if known): K101310
#### Vu aPOD Intervertebral Body Fusion Device Device Name:_
Indications For Use:
When used with the bone screws, the Theken Spine Vu aPOD Intervertebral Body Fusion Device is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graft material.
When used with the SpinPlate, the Theken Spine Vu aPOD Intervertebral Body Fusion Device is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graft material. When used with the SpinPlate, the Theken Spine Vu aPOD Intervertebral Body Fusion Device is intended for use with supplemental fixation such as the Coral Spinal System or the BodyForm Thoracolumbar Fixation System.
The Theken Spine Vu aPOD Intervertebral Body Fusion Device, when used with bone screws, is a stand alone device. If the Theken Spine Vu aPod Intervertebral Body Fusion Device is used with the SpinPlate then additional supplemental fixation, which has been cleared by the FDA for use in the lumbar spine, must be used to augment stability. The SpinPlate and bone screws are not intended to be used together. This device is intended to be used with autogenous bone graft. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device.
AND/OR -Prescription Use x (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|---------------------|--|
|---------------------|--|
Division of Surgical. Orthopedic. and Restorative Devices
Page 1 of 1 -
KIDI310 510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.