Biomet Fusion System

K163543 · Zimmer Biomet Spine, Inc. · OVD · Feb 28, 2017 · Orthopedic

Device Facts

Record IDK163543
Device NameBiomet Fusion System
ApplicantZimmer Biomet Spine, Inc.
Product CodeOVD · Orthopedic
Decision DateFeb 28, 2017
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Intended Use

When used as a cervical intervertebral body fusion device, the Biomet Intervertebral Body/VBR Fusion System ("Biomet Fusion System") is intended for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients with degenerative disc disease ("DDD") at one spinal level from the C2-C3 disc to the C7-T1 disc. DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radios. These patients should have had at least six weeks of non-operative treatment. The Biomet Cervical Intervertebral Body Fusion System is to be implanted via an anterior approach and is to be combined with supplemental fixation. Approved supplemental fixation systems include the Biomet Anterior Cervical Plate System. When used as a lumbar intervertebral body fusion device, the Biomet Intervertebral Body/VBR Fusion System ("Biomet Fusion System") is intended for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients with degenerative disc disease ("DD") at one or two contiguous spinal levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment. These DDD patients may have had a previous non-fusion spinal surgery at the involved spinal level(s), and may have up to Grade 1 spondylolisthesis at the involved level(s). Implants with 14 degree lordosis or greater are only indicated from levels L2-L5 and are to be used with at least one integrated fixation screw. The Biomet Fusion System is to be combined with supplemental fixation (except as noted below). Approved supplemental fixation systems include the Biomet Spinal Fixation System. The Biomet SA standalone interbody implants, when implanted via an anterior approach and used with the integrated fixation screws, do not require use of supplement fixation. When used as vertebral body replacement, the Biomet Intervertebral Body/VBR Fusion System ("Biomet Fusion System") is indicated for use to replace a vertebral body that has been resected or excised due to tumor or traumalfracture. The device is intended for use as a vertebral body replacement in the thoracolumbar spine (from T1 to L5) The Biomet Vertebral Body Replacement System may also be used in the thoracolumbar spine (i.e. T1- L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Biomet Vertebral Body Replacement System is also indicated for treating fractures of the thoracic and lumbar spine. The Biomet Vertebral Body Replacement System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column. For either indication the system must be used with supplemental internal fixation. Supplemental internal fixation is required to properly utilize this system.

Device Story

Biomet Fusion System is a line of intervertebral body fusion and vertebral body replacement implants; used for spinal fusion or vertebral body reconstruction. Spacers are PEEK-OPTIMA LT1 with tantalum markers; plates and screws are titanium alloy (Ti-6AI-4V ELI). System provides structural support to restore biomechanical integrity of spinal column; used with autogenous bone graft. Implanted by surgeons via anterior approach; requires supplemental internal fixation (except for specific SA standalone configurations). Modifications include additional plate heights (18mm, 20mm), 1-hole/2-hole configurations, and 14-degree lordotic spacers. System components are provided non-sterile; intended to facilitate fusion or replace resected/excised vertebral bodies due to tumor or trauma.

Clinical Evidence

Bench testing only; no clinical data provided. Substantial equivalence demonstrated through design controls, risk analysis (FMEA), and comparison of technological characteristics to the predicate device.

Technological Characteristics

Spacers: PEEK-OPTIMA LT1 (polyaromatic semicrystalline thermoplastic polymer) with tantalum markers. Plates/Screws: Titanium alloy (Ti-6AI-4V ELI) per ASTM F136-08 and ISO 5832-3. Form factor: Intervertebral spacers, plates, and screws in various sizes/configurations. Provided non-sterile.

Indications for Use

Indicated for skeletally mature patients requiring spinal fusion or vertebral body replacement. Cervical DDD (C2-T1) with neck pain; lumbar DDD (L2-S1) with back pain; thoracolumbar (T1-L5) tumor or trauma/fracture. Contraindications include skeletal immaturity.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol is a stylized representation of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, contrasting with the white background. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 28, 2017 Zimmer Biomet Spine, Inc. Megan Fessenden Regulatory Affairs, Senior Specialist 310 Interlocken Parkway, Suite 120 Broomfield, Colorado 80021 Re: K163543 Trade/Device Name: Biomet Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVD, MAX, ODP, MQP Dated: January 31, 2017 Received: February 1, 2017 Dear Ms. Fessenden: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K163543 Device Name Biomet Fusion System #### Indications for Use (Describe) When used as a cervical intervertebral body fusion device, the Biomet Intervertebral Body/VBR Fusion System ("Biomet Fusion System") is intended for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients with degenerative disc disease ("DDD") at one spinal level from the C2-C3 disc to the C7-T1 disc. DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radios. These patients should have had at least six weeks of non-operative treatment. The Biomet Cervical Intervertebral Body Fusion System is to be implanted via an anterior approach and is to be combined with supplemental fixation. Approved supplemental fixation systems include the Biomet Anterior Cervical Plate System. When used as a lumbar intervertebral body fusion device, the Biomet Intervertebral Body/VBR Fusion System ("Biomet Fusion System") is intended for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients with degenerative disc disease ("DD") at one or two contiguous spinal levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment. These DDD patients may have had a previous non-fusion spinal surgery at the involved spinal level(s), and may have up to Grade 1 spondylolisthesis at the involved level(s). Implants with 14 degree lordosis or greater are only indicated from levels L2-L5 and are to be used with at least one integrated fixation screw. The Biomet Fusion System is to be combined with supplemental fixation (except as noted below). Approved supplemental fixation systems include the Biomet Spinal Fixation System. The Biomet SA standalone interbody implants, when implanted via an anterior approach and used with the integrated fixation screws, do not require use of supplement fixation. When used as vertebral body replacement, the Biomet Intervertebral Body/VBR Fusion System ("Biomet Fusion System") is indicated for use to replace a vertebral body that has been resected or excised due to tumor or traumalfracture. The device is intended for use as a vertebral body replacement in the thoracolumbar spine (from T1 to L5) The Biomet Vertebral Body Replacement System may also be used in the thoracolumbar spine (i.e. T1- L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Biomet Vertebral Body Replacement System is also indicated for treating fractures of the thoracic and lumbar spine. The Biomet Vertebral Body Replacement System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column. For either indication the system must be used with supplemental internal fixation. Supplemental internal fixation is required to properly utilize this system. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) {3}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ______________________________________________________________________________________________________________________________________________________________________________ This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 807.92. | Preparation Date: | February 18, 2017 | |--------------------------------------------|------------------------------------------------------------------------------------------------------| | Applicant/Sponsor: | Zimmer Biomet Spine Inc.<br>310 Interlocken Parkway, Suite 120<br>Broomfield, CO 80021 | | Contact Person: | Megan Fessenden<br>Regulatory Affairs, Senior Specialist<br>Phone: 303.465.8994<br>Fax: 303.501.8444 | | Trade Name: | Biomet Fusion System | | Common Name: | Vertebral Body Replacement/Intervertebral Body<br>Fusion Device with Integrated Fixation, Lumbar | | Device Class: | Class II | | Classification Name | Orthosis, spinal intervertebral fusion and/or Spinal<br>intervertebral body fixation orthosis. | | Product Code:<br>Subsequent Product Codes: | OVD<br>MAX, MQP, ODP | | Device Panel - Regulation No.: | Orthopedic - 21 CFR 888.3080 and/or 888.3060 | | Primary Predicate: | Biomet Fusion System K141791 | {5}------------------------------------------------ ### Device Description: The subject Biomet Fusion System has the same or similar intended use, indications for use, technological characteristics and principles of operation as the intervertebral body fusion devices which were reclassified as explained in FDA's guidance document, "Class II Special Controls Guidance Document: Intervertebral Body Fusion Device" (June 12, 2007). The subject Biomet Fusion is substantially equivalent to the previously cleared Biomet Fusion System (K141791). The Biomet Fusion System is intended for use in intervertebral body fusion or vertebral body replacement. However, the implants included in this application are intended for use in lumbar intervertebral body fusion only. The subject implants are a line extension of those currently available in the Biomet Fusion System. Modifications have been made to the currently available Biomet Fusion System implants to add additional plate heights (18mm and 20mm) in both 1-hole and 2-hole configurations as well as to add a new 14degree spacer option. As a result of these new 18mm and 20mm plates and 14 degree spacers, taller cover plates, implant trials, and implant caddies were developed to correspond to the new implant sizes. The Biomet Fusion System implants consist of a spacer body and plates with screws for additional fixation. The spacers are made from PEEK-OPTIMA® LT1, a polyaromatic semicrystalline thermoplastic polymer (-CcH4-O-CcH4-O-CcH4-CO-), with tantalum markers. The plates and screw are made from implant grade titanium alloy (Ti-6AI-4V ELI) meeting ASTM F136-08 and ISO 5832-3. The Biomet Fusion System spacers, plates, and screws are available in a variety of sizes and configurations to accommodate anatomical variation in different vertebral levels and/or patient anatomy. The Biomet Fusion System is provided non-sterile. ### Indications for Use: The indications for use of the Biomet Fusion System are as follows: When used as a cervical intervertebral body fusion device, the Biomet Intervertebral Body/VBR Fusion System ("Biomet Fusion System") is intended for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients with degenerative disc disease ("DDD") at one spinal level from the C2-C3 disc to the C7-T1 disc. DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should have had at least six weeks of non-operative treatment. The Biomet Cervical Intervertebral Body Fusion System is to be implanted via an anterior approach and is to be combined with supplemental fixation. Approved supplemental fixation systems include the Biomet Anterior Cervical Plate System. When used as a lumbar intervertebral body fusion device, the Biomet Intervertebral Body/VBR Fusion System ("Biomet Fusion System") is intended for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients with {6}------------------------------------------------ degenerative disc disease ("DDD") at one or two contiguous spinal levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of nonoperative treatment. These DDD patients may have had a previous non-fusion spinal surgery at the involved spinal level(s), and may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). Implants with 14 degree lordosis or greater are only indicated from levels L2-L5 and are to be used with at least one integrated fixation screw. The Biomet Fusion System is to be combined with supplemental fixation (except as noted below). Approved supplemental fixation systems include the Biomet Spinal Fixation System. The Biomet SA standalone interbody implants, when implanted via an anterior approach and used with the integrated fixation screws, do not require use of supplement fixation. When used as vertebral body replacement, the Biomet Intervertebral Body/VBR Fusion System ("Biomet Fusion System") is indicated for use to replace a vertebral body that has been resected or excised due to tumor or trauma/fracture. The device is intended for use as a vertebral body replacement in the thoracolumbar spine (from T1 to L5) The Biomet Vertebral Body Replacement System may also be used in the thoracolumbar spine (i.e. T1- L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Biomet Vertebral Body Replacement System is also indicated for treating fractures of the thoracic and lumbar spine. The Biomet Vertebral Body Replacement System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column. For either indication the system must be used with supplemental internal fixation. Supplemental fixation is required to properly utilize this system. ### Summary of Technological Characteristics: The subject Biomet Fusion System implants included in this submission have the same intended use, indications for use, technological characteristics, and principles of operation as the previously cleared Biomet Fusion System (K141791). The subject and predicate devices both include the following high-level technological characteristics: - Lumbar intervertebral body fusion device ● - Materials of construction ● - Surgical procedure ● The following technological characteristics are different between the subject and predicate device: - Taller plate flange (plates) - Taller cover plates ● - o Additional lordotic angle (spacers) {7}------------------------------------------------ The minor differences in the new components do not raise any new issues of safety or effectiveness. ## Non-Clinical Testing (Risk Analysis) During the development of the subject Biomet Fusion System modifications a review of the Biomet Fusion System Failure Modes and Effects Analysis (FMEA) was conducted to assess and identify any new risks, including performance risks, associated with the modifications. Recommended risk mitigations described in FDA's guidance document, "Class II Special Controls Guidance Document: Intervertebral Body Fusion Device" (June 12, 2007), Section 6, have been implemented in accordance with this guidance document, as appropriate based on the scope of changes. ### Conclusion: The Biomet Spine Fusion System is substantially equivalent to the predicate systems as a spinal fusion device in regards to: intended use, indications for use, principles of operation, technological characteristics, and sterility. Furthermore, design controls and supporting information sufficiently demonstrate the substantial equivalence of the subject components to the Biomet Spine Fusion System to the predicate system. Based on this information, the subject modifications do not raise any new issues regarding the safety or efficacy when compared to its predicates.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...