When used as a cervical intervertebral body fusion device, the Biomet Intervertebral Body/VBR Fusion System ("Biomet Fusion System") is intended for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients with degenerative disc disease ("DDD") at one spinal level from the C2-C3 disc to the C7-T1 disc. DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radios. These patients should have had at least six weeks of non-operative treatment. The Biomet Cervical Intervertebral Body Fusion System is to be implanted via an anterior approach and is to be combined with supplemental fixation. Approved supplemental fixation systems include the Biomet Anterior Cervical Plate System. When used as a lumbar intervertebral body fusion device, the Biomet Intervertebral Body/VBR Fusion System ("Biomet Fusion System") is intended for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients with degenerative disc disease ("DD") at one or two contiguous spinal levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment. These DDD patients may have had a previous non-fusion spinal surgery at the involved spinal level(s), and may have up to Grade 1 spondylolisthesis at the involved level(s). Implants with 20 degree lordosis or greater are only indicated from levels L2-L5 and are to be used with at least one integrated fixation screw. The Biomet Fusion System is to be combined with supplemental fixation (except as noted below). Approved supplemental fixation systems include the Biomet Spinal Fixation System. The Biomet SA standalone interbody implants, when implanted via an anterior approach and used with the integrated fixation screws, do not require use of supplement fixation. When used as vertebral body replacement, the Biomet Intervertebral Body/VBR Fusion System ("Biomet Fusion System") is indicated for use to replace a vertebral body that has been resected or excised due to tumor or traumalfracture. The device is intended for use as a vertebral body replacement in the thoracolumbar spine (from T1 to L5) The Biomet Vertebral Body Replacement System may also be used in the thoracolumbar spine (i.e. T1- L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Biomet Vertebral Body Replacement System is also indicated for treating fractures of the thoracic and lumbar spine. The Biomet Vertebral Body Replacement System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column. For either indication the system must be used with supplemental internal fixation. Supplemental internal fixation is required to properly utilize this system.
Device Story
Biomet Fusion System comprises PEEK-OPTIMA spacer bodies and titanium alloy (Ti-6Al-4V ELI) plates/screws; designed for spinal fusion or vertebral body replacement. Spacers feature hollow centers for autogenous bone graft and transverse grooves for stability. Implanted via anterior approach by surgeons; requires supplemental internal fixation (except specific standalone lumbar configurations). Device restores biomechanical integrity of spinal column; provides structural support during fusion process; aids in decompression and height restoration in tumor/trauma cases.
Clinical Evidence
Bench testing only. Performance data included static compression, static compression shear, static torsion, dynamic compression (per ASTM F2077), and expulsion testing (per ASTM Draft Z8423Z). A cadaveric study was also performed. All tests met acceptance criteria.
Technological Characteristics
Materials: PEEK-OPTIMA LT1 (ASTM F2026), Titanium alloy Ti-6Al-4V ELI (ASTM F136-08/ISO 5832-3), Tantalum markers (ASTM F560). Form factor: Spacer body with optional plates/screws. Non-sterile. Mechanical fixation via transverse grooves and integrated screws.
Indications for Use
Indicated for skeletally mature patients requiring spinal fusion for degenerative disc disease (DDD) at one cervical level (C2-T1) or one/two contiguous lumbar levels (L2-S1); or vertebral body replacement in the thoracolumbar spine (T1-L5) due to tumor, trauma, or fracture. Contraindications include patients without skeletal maturity or those not meeting non-operative treatment duration requirements.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with the first profile being the most prominent and the other two fading into the background. The profiles are connected by a flowing line that resembles a ribbon or wave. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 9, 2014
Biomet Spine Mr. Ted Kuhn Regulatory Affairs Product Manager 310 Interlocken Parkway, Suite 120 Broomfield, Colorado 80021
Re: K141791
Trade/Device Name: Biomet Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVD, MAX, ODP, MOP Dated: September 9, 2014 Received: September 11, 2014
Dear Mr. Kuhn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{2}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Ronaldflygean -S for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K141791
Device Name Biomet Fusion System
#### Indications for Use (Describe)
When used as a cervical intervertebral body fusion device, the Biomet Intervertebral Body/VBR Fusion System ("Biomet Fusion System") is intended for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients with degenerative disc disease ("DDD") at one spinal level from the C2-C3 disc to the C7-T1 disc. DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radios. These patients should have had at least six weeks of non-operative treatment. The Biomet Cervical Intervertebral Body Fusion System is to be implanted via an anterior approach and is to be combined with supplemental fixation. Approved supplemental fixation systems include the Biomet Anterior Cervical Plate System.
When used as a lumbar intervertebral body fusion device, the Biomet Intervertebral Body/VBR Fusion System ("Biomet Fusion System") is intended for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients with degenerative disc disease ("DD") at one or two contiguous spinal levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment. These DDD patients may have had a previous non-fusion spinal surgery at the involved spinal level(s), and may have up to Grade 1 spondylolisthesis at the involved level(s). Implants with 20 degree lordosis or greater are only indicated from levels L2-L5 and are to be used with at least one integrated fixation screw. The Biomet Fusion System is to be combined with supplemental fixation (except as noted below). Approved supplemental fixation systems include the Biomet Spinal Fixation System. The Biomet SA standalone interbody implants, when implanted via an anterior approach and used with the integrated fixation screws, do not require use of supplement fixation.
When used as vertebral body replacement, the Biomet Intervertebral Body/VBR Fusion System ("Biomet Fusion System") is indicated for use to replace a vertebral body that has been resected or excised due to tumor or traumalfracture. The device is intended for use as a vertebral body replacement in the thoracolumbar spine (from T1 to L5) The Biomet Vertebral Body Replacement System may also be used in the thoracolumbar spine (i.e. T1- L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Biomet Vertebral Body Replacement System is also indicated for treating fractures of the thoracic and lumbar spine. The Biomet Vertebral Body Replacement System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column. For either indication the system must be used with supplemental internal fixation. Supplemental internal fixation is required to properly utilize this system.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Biomet Spine. The word "BIOMET" is in large, bold, sans-serif letters. Below it, in smaller letters, is the word "SPINE". The logo is simple and modern, with a focus on the company name.
## 510(k) Summary
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR § 807.92.
| Preparation Date: | September 30, 2014 |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant/Sponsor: | Biomet Spine<br>310Interlocken Parkway, Suite 120<br>Broomfield, CO 80021 |
| Contact Person: | Ted Kuhn<br>Regulatory Affairs Project Manager<br>Phone: 303-501-8549<br>Fax: 303-501-8444 |
| Trade name: | Biomet Fusion System |
| Common Name: | Intervertebral fusion device with bone graft, cervical;<br>Intervertebral fusion device with bone graft, lumbar;<br>Intervertebral Body Fusion Device with Integrated Fixation,<br>Lumbar; Vertebral Body Replacement |
| Classification Name | Orthosis, spinal intervertebral fusion and/or Spinal<br>intervertebral body fixation orthosis. |
| (Product Code): | (ODP, MAX, OVD, MQP) |
| Device Panel - Regulation No.: | Orthopedic - 21 CFR § 888.3080 and/or § 888.3060 |
#### Device Description:
Biomet Fusion System implants consist of a spacer body with screws for additional fixation. The Biomet Fusion System the spacer body is made from PEEK-OPTIMA® LT1, a polyaromatic semicrystalline thermoplastic polymer (-C,Hz-O-C,Hz-O-C,Hz-CO-). The Biomet Fusion System implants include an anterior plate and screws made from an implant grade titanium alloy (Ti-6Al-4V ELI) meeting the requirements of ASTM F136-08 and ISO 5832-3. The spacer body, plates and screws are available in a variety of sizes and configurations to accommodate anatomical variation in different vertebral levels and/or patient anatomy. The Biomet Fusion System is provided non-sterile.
The implants included in this application are intended for us in lumbar intervertebral body fusion. The implants have a hollowed out central area to accommodate bone graft. The upper and lower surfaces of each implant have a series of transverse grooves formed to improve stability and fixation once the device is inserted.
The purpose of this 510(k) Notification is to seek clearance for a modification to the Lanx Fusion System, for additional configurations. The new implants are also referred to as the Biomet Fusion System.
{6}------------------------------------------------
#### Indications for Use:
When used as a cervical intervertebral body fusion device, the Biomet Intervertebral Body/VBR Fusion System ("Biomet Fusion System") is intended for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients with degenerative disc disease ("DDD") at one spinal level from the C2-C3 disc to the C7-T1 disc. DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should have had at least six weeks of non-operative treatment. The Biomet Cervical Intervertebral Body Fusion System is to be implanted via an anterior approach and is to be combined with supplemental fixation. Approved supplemental fixation systems include the Biomet Anterior Cervical Plate System.
When used as a lumbar intervertebral body fusion device, the Biomet Intervertebral Body/VBR Fusion System ("Biomet Fusion System") is intended for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients with degenerative disc disease ("DDD") at one or two contiguous spinal levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment. These DDD patients may have had a previous non-fusion spinal surgery at the involved spinal level(s), and may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). Implants with 20 degree lordosis or greater are only indicated from levels L2-L5 and are to be used with at least one integrated fixation screw. The Biomet Fusion System is to be combined with supplemental fixation (except as noted below). Approved supplemental fixation systems include the Biomet Spinal Fixation System. The Biomet SA standalone interbody implants, when implanted via an anterior approach and used with the integrated fixation screws, do not require use of supplement fixation.
When used as vertebral body replacement, the Biomet Intervertebral Body/VBR Fusion System ("Biomet Fusion System") is indicated for use to replace a vertebral body that has been resected or excised due to tumor or trauma/fracture. The device is intended for use as a vertebral body replacement in the thoracolumbar spine (from T1 to L5) The Biomet Vertebral Body Replacement System may also be used in the thoracolumbar spine (i.e. T1- L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Biomet Vertebral Body Replacement System is also indicated for treating fractures of the thoracic and lumbar spine. The Biomet Vertebral Body Replacement System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column. For either indication the system must be used with supplemental internal fixation. Supplemental internal fixation is required to properly utilize this system.
#### Summary of Technologies:
This submission is intended to seek clearance for a product line extension to the Biomet Fusion System. The product line extension includes new configurations of a PEEK spacer body similar to the existing PEEK spacer body of the previously cleared Biomet Fusion System.
All components in the Biomet Fusion System are made of PEEK (OPTIMA®) per ASTM F2026 and Titanium alloy (Ti-6A1-4V) per ASTM F136. The PEEK components include Tantalum markers per ASTM F560. The Biomet Fusion System has a hollowed out area to accommodate autogenous bone graft, and transverse grooves to improve fixation and stability. It is available in a variety of different sizes to accommodate anatomical variation in different vertebral levels and/or patient anatomy. It
{7}------------------------------------------------
comes available with plates and screws to provide additional fixation; however, these are intended to be used with supplemental fixation. The Biomet Fusion System is provided non-sterile.
The additional devices added to the product line have or similar intended use and indications, principles of operation, and technological characteristics as the current Biomet Fusion System. The minor differences between the current devices added to the product line do not raise any new questions of safety or effectiveness. Mechanical testing and engineering analysis demonstrated comparable mechanical properties to the predicate device.
#### Performance Data:
Performance testing performed to demonstrate substantial equivalence:
- Static Compression per ASTM F2077 ●
- Static Compression Shear per ASTM F2077 ●
- Static Torsion per ASTM F2077
- Dynamic Compression per ASTM F2077
- Expulsion Testing per ASTM Draft Z8423Z ●
- Cadaveric Study
In all instances, the Biomet Fusion System met acceptance criteria and functioned as intended.
#### Substantial Equivalence:
The additional devices included in the product line extension have the same or similar intended uses, indications, technological characteristics, and principles of operation as previously cleared primary predicate Lanx Fusion System (K131547). Performance data demonstrate that these additions to the Biomet Fusion System do not raise new issues of safety or effectiveness; hence it is as safe and effective as its predicate devices. Thus, the additional devices are substantially equivalent.
#### Conclusion:
In summary, the modified Biomet Fusion System has the same or similar: intended use, indications for use, technological characteristics, principles of operation and performances as the previously cleared primary predicate Lanx Fusion System (K131547) and (K123767). Performance testing demonstrates that the modified implants are stronger than or as strong as the predicate implants and provide equivalent performance. Thus, the addition of the new implant models do not raise new questions of safety and efficacy and is substantially equivalent to previous cleared primary predicate Lanx Fusion System (K131547) and K123767).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.