VISUALIF CHALLENGING ACCESS PLATE

K122168 · Spine Smith Partners L.P. · OVD · Oct 11, 2012 · Orthopedic

Device Facts

Record IDK122168
Device NameVISUALIF CHALLENGING ACCESS PLATE
ApplicantSpine Smith Partners L.P.
Product CodeOVD · Orthopedic
Decision DateOct 11, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Intended Use

The VisuALIF System is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with autograft to facilitate fusion. The VisuALIF is a stand-alone device intended to be used with an anterior cover plate and a minimum of two provided bone screws angled both cephalad and caudal with a minimum of one screw into each vertebral body. Should the physician choose to use fewer than the two screws provided, additional supplemental fixation cleared by the FDA for use in the lumbar spine must be used. The anterior cover plate must be utilized whenever the device is implanted using the bone screws provided.

Device Story

Lumbar intervertebral body fusion device; implanted via open anterior approach. System comprises PEEK Optima LT1 'U'-shaped implant, titanium alloy screws, and anterior cover plate. Radiopaque tantalum markers assist surgeon with placement. Used in clinical settings by surgeons to treat DDD and spondylolisthesis; requires autograft for fusion. Stand-alone device; requires minimum two bone screws (cephalad/caudal) and cover plate; supplemental fixation required if fewer screws used. Provides structural support to vertebral bodies to facilitate spinal fusion; benefits patients by stabilizing affected spinal segments.

Clinical Evidence

Bench testing only. Mechanical performance evaluated via static and dynamic compression (ASTM F2077-03), subsidence (ASTM F2267-04), and expulsion (ASTM Draft Standard F-04.25.02.02). No clinical data presented.

Technological Characteristics

Implant: PEEK Optima LT1 (ASTM F2026) with tantalum markers (ASTM F-560-05). Screws/cover plate: Titanium alloy (ASTM F-136). 'U'-shaped geometry. Stand-alone, non-powered, mechanical fixation. Sterilization method not specified.

Indications for Use

Indicated for intervertebral body fusion of the lumbar spine (L2-S1) in skeletally mature patients with degenerative disc disease (DDD) and up to Grade I spondylolisthesis, following six months of failed non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K122168 OCT 1 1 2012 ## SpineSmith Cynch Spinal System – VisuALIF Interbody Fusion Implant System | 510(k) Summary of Safety and Effectiveness | | | OC | |--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|--|----| | SUBMITTED BY | SpineSmith Partners, LP<br>93 Red River<br>Austin, TX 78701 | | | | ESTABLISHMENT<br>REGISTRATION NUMBER | 3006404071 | | | | CONTACT PERSON | Clifton (Chris) Naivar<br>Director - Quality and Regulatory Affairs<br>Phone: 512-637-2068<br>Fax: 512-637-2096<br>Email: cnaivar@spinesmithusa.com | | | | SUBMISSION PREPARED BY | Clifton (Chris) Naivar<br>Director - Quality and Regulatory Affairs<br>Phone: 512-637-2068 | | | | DATE PREPARED | October 10, 2012 | | | | CLASSIFICATION | OVD 888.3080- Intervertebral Fusion Device<br>with Bone Graft, Lumbar | | | | COMMON NAME | Intervertebral Body Fusion Device | | | | PROPRIETARY NAME | SpineSmith Cynch Spinal System - VisuALIF<br>Challenging Access Interbody Fusion Implant<br>System | | | #### IDENTIFICATION OF PREDICATE DEVICES: The SpineSmith VisuALIF Challenging Access System was determined to be substantially equivalent to the previously cleared VisuALIF System (K102090, SpineSmith; Cleared 09/09/2010). #### DEVICE DESCRIPTION: The VisuALIF Challenging Access System an extension to the VisuALIF System and both are part of the Cynch Spinal System. It is available in various sizes to accommodate individual patient anatomy. The VisuALIF Challenging Access implant is a lumbar intervertebral body fusion device that is intended to be implanted singularly via an open anterior approach. VisuALIF Challenging Access is intended to be used with the two (2) or more bone screws provided with the accompanying anterior cover plate assembly. {1}------------------------------------------------ Addition of additional screw holes and positions to the existing VisuALIF System is intended to provide surgeons with additional surgical approach options. There are no changes with respect to indications or intended use as compared to the VisuALIF Spinal System cleared previously via K102090. ### INDICATIONS: The VisuALIF System is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment.The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with autograft to facilitate fusion. The VisuALIF is a stand-alone device intended to be used with an anterior cover plate and a minimum of two provided bone screws angled both cephalad and caudal with a minimum of one screw into each vertebral body. Should the physician choose to use fewer than the two screws provided, additional supplemental fixation cleared by the FDA for use in the lumbar spine must be used. The anterior cover plate must be utilized whenever the device is implanted using the bone screws provided. ## SUMMARY OF TECHNOLOGICAL CHARACTERISTICS: The purpose of this submission is to add the VisuALIF Challenging Access device, which may be implanted via an open anterior approach. The Cynch System implants are manufactured from PEEK Optima LT1 and contain three (3) radiopaque tantalum markers to assist the surgeon with proper placement of the device. The subject device (VisuALIF Challenging Access - ALIF device) has similar technological characteristics as the predicate devices identified above (SpineSmith's Cynch System per K102090). Specifically, the following characteristics support this conclusion: - Intended for stand-alone use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. - 'U' shaped implant design to allow for placement of autograft bone - - Implant material: PEEK Optima LT1 per ASTM F2026 with radiopaque tantalum marker bar per ASTM F-560-05, and Titanium alloy screws/cover plate assembly per ASTM F-136 - Substantially equivalent results of non-clinical testing relative to static and dynamic testing (per ASTM F2077-03), subsidence (per ASTM F2267-04), and expulsion (per ASTM Draft Standard F-04.25.02.02) Page 2 of 3 {2}------------------------------------------------ ## DISCUSSION OF NON-CLINICAL TESTING: The following non-clinical tests were conducted: - Static and dynamic compression testing, conducted in accordance with । ASTM F2077-03 - Static and dynamic compression-shear testing, conducted in accordance . with ASTM F2077-03 - Subsidence testing, conducted in accordance with ASTM F2267-04 - - Expulsion testing, conducted in accordance with ASTM Draft Standard F--04.25.02.02 ## CONCLUSIONS: The subject and predicate device share the same intended use, primary implant design and material of manufacture. The non-clinical mechanical test results demonstrate that any minor differences do not impact device performance as compared to the predicates and demonstrate that the VisuALIF Challenging Access System is substantially equivalent to the predicate device. Page 3 of 3 {3}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines representing the wings and a wavy line representing the body. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 OCT 11 2012 Spine Smith Partners, LLC % Mr. Clifton (Chris) Naviar Director, Quality and Regulatory Affairs 93 Red River Austin, Texas 78701 Re: K122168 Trade/Device Name: SpineSmith Cynch Spinal System - VisuALIF Challenging Access Plate Interbody Fusion Implant System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVD Dated: September 12, 2012 Received: September 14, 2012 Dear Mr. Naviar: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {4}------------------------------------------------ Page 2 - Mr. Clifton (Chris) Naviar comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Eiriel Keith Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # INDICATIONS FOR USE # 510(k) Number (if known): K122168 Device Name: ## SpineSmith Cynch Spinal System - VisuALIF Challenging Access Plate Interbody FisciAL Implant System Indications for Use: The VisuALIF System is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolistics a DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with autografic studi device system is designed for use with autografit to facilitate fusion. The VisuALIF is a stand-alone device intended to be used with an anterior cover plate and a minimum of two provided bone screws angled both cephalad and caudal with a minimum of one screw into each vertebral both cephalad and physician choose to use fewer than the two screws provided, additional supplemental fixation cleared by the FDA for use in the lumbar spine must be used. The anterior cover plate must be utilized whenever the device is implanted using the bone screws provided. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Carl-R (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number. K122168
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...