K122773 · Stelkast Company · OQG · Oct 9, 2012 · Orthopedic
Device Facts
Record ID
K122773
Device Name
CROSS-OVER ACETABULAR SHELL & LINER
Applicant
Stelkast Company
Product Code
OQG · Orthopedic
Decision Date
Oct 9, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Cross-Over Acetabular Shell and Liner are intended for use in reconstruction of the articulating surface of the acetabular portion of the hip that is severely disabled and/or very painful resulting from: 1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. 2. Rheumatoid arthritis. 3. Correction of functional deformity. 4. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques. 5. Revision of previously failed total hip Arthroplasty. The Cross-Over Acetabular Shell and Liner are used as part of the Provident Hip Systems. The components of this hip system are intended for cementless fixation.
Device Story
Acetabular component for total hip arthroplasty; consists of titanium alloy shell with plasma spray coating and Vitamin E-blended polyethylene liner. Used as part of Provident Hip System for cementless fixation. Implanted by orthopedic surgeons in clinical/OR setting to replace articulating acetabular surface. Provides stable joint reconstruction; restores function; reduces pain. Non-clinical testing evaluated interconnection mechanism (disassembly force, lever-out torque, rotational failure torque).
Clinical Evidence
Bench testing only. No clinical data presented.
Technological Characteristics
Titanium alloy shell with commercially pure titanium plasma spray coating; Vitamin E-blended polyethylene liner. Available in hooded and non-hooded configurations. Designed for cementless fixation. Mechanical testing performed on shell-liner interconnection.
Indications for Use
Indicated for patients with severe hip disability or pain due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, proximal femur fractures (non-union, femoral neck, trochanteric with head involvement), or revision of failed total hip arthroplasty.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Predicate Devices
Acetabular Shells in the Stelkast Provident Hip System (K001745)
K090998 — DEPUY PINNACLE 100 WITH GRIPTION ACETABULAR CUPS · DePuy Orthopaedics, Inc. · Jun 12, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
K122773 - Page 1 of 2
ОСТ
9 2012
# 510(k) Summary
Device Trade Name:
Manufacturer:
Contact:
Cross-Over Acetabular Shell and Liner
StelKast, Inc. 200 Hidden Valley Road McMurray, PA 15317
Mr. David Stumpo Vice President of Product Development Phone: 724.731.2208 Fax: 724.941.5987 dstumpo@stelkast.com
1331 H Street NW, 12th Floor Washington, DC 20005 Phone: (202) 552-5800 Fax: (202) 552-5798 .
Musculoskeletal Clinical Regulatory Advisers, LLC
Prepared by:
Date Prepared:
September 7, 2012
Classification:
21 CFR 888.3358, Hip joint metal/polymer/metal semiconstrained porous-coated uncemented prosthesis
Class:
Product Codes:
OOG, OOH, OQI, LPH, LWJ, JDI, MAY, LZO
#### Indications For Use:
The Cross-Over Acetabular Shell and Liner are intended for use in reconstruction of the articulating surface of the acetabular portion of the hip that is severely disabled and/or very painful resulting from:
- 1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- 2. Rheumatoid arthritis.
- 3. Correction of functional deformity.
- Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal 4. femur with head involvement, unmanageable using other techniques.
- 5. Revision of previously failed total hip Arthroplasty.
II
The Cross-Over Acetabular Shell and Liner are used as part of the Provident Hip Systems. The components of this hip system are intended for cementless fixation.
{1}------------------------------------------------
#### Device Description:
The Cross-Over Acetabular Shell and Liner will be used as part of a complete total hip system in conjunction with a femoral head and femoral stem in total hip arthroplasty.
The Cross-Over Acetabular Shell is made of titanium alloy with a commercially pure titanium plasma spray coating. The Cross-Over Liner is made of polyethylene blended with Vitamin E. The liner is available in both non-hooded and hooded options.
### Predicate Devices:
Comparative information presented in the 510(k) supports the substantial equivalence of the Cross-Over Acetabular Shell and Liner with respect to their indications for use, design, materials, and performance.
This 510(k) demonstrates the substantial equivalence of the Cross-Over Acetabular Shell and Liner to the following predicate devices:
- Acetabular Shells in the Stelkast Provident Hip System (K001745), and .
- . EXp Acetabular Liner (K094035)
These acetabular shells and liners have the same intended use and general design, are available in a similar size range, and are made of the same materials.
### Preclinical Testing:
Non-clinical testing was performed on the Cross-Over Acetabular Shell and Liner to assess the interconnection mechanism between the shell and liner (i.e., disassembly force, lever-out torque, and rotational failure torque). The results of the performed tests demonstrate that the Cross-Over Acetabular Shell and Liner are substantially equivalent to tegally marketed predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, rendered in a stylized, minimalist design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -- WO66-G609 Silver Spring, MD 20993-0002
ОСТ
2012
StelKast Company
% Mr. David Stumpo
Vice President of Product Development
200 Hidden Valley Road
McMurray, Pennsylvania 15317
Re: K122773 Trade/Device Name: Cross-Over Acetabular Shell and Liner Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: OQG. OQH, OQI, LPH, LWJ, JDI, MAY, LZO Dated: September 7, 2012 Received: September 13, 2012
Dear Mr. Stumpo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{3}------------------------------------------------
### Page 2 - Mr. David Stumpo
CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Sincerely yours,
Mark N. Melkerson
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K122773
Device Name: Cross-Over Acetabular Shell and Liner
The Cross-Over Acetabular Shell and Liner are intended for use in reconstruction of the articulating surface of the acetabular portion of the hip that is severely disabled and/or very painful resulting from:
- 1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. .
- 2. Rheumatoid arthritis.
- 3. Correction of functional deformity.
- 4. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques.
- 5. Revision of previously failed total hip Arthroplasty.
The Cross-Over Acetabular Shell and Liner are used as part of the Provident Hip Systems. The components of this hip system are intended for cementless fixation.
V Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number_K122777
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.