The Infinity™ OCT System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7), and the thoracic spine from T1-T3: - Traumatic spinal fractures and/or traumatic dislocations. - Instability or deformity. - Failed previous fusions (e.g. pseudarthrosis). - Tumors involving the cervical spine. - Degenerative disease, including intractable radiculopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The Infinity™ OCT System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. The Infinity™ OCT System may be used with PASS OCT Patient Specific UNiD OCT Rods. In order to achieve additional levels of fixation, the Infinity™ OCT System may be connected to the CD Horizon™ Spinal System and VertexTM Reconstruction System rods with the Infinity™ OCT System rod connectors. Transition rods with differing diameters may also be used to connect the Infinity™ OCT System to the CD Horizon™ Spinal System. Refer to the CD Horizon™ Spinal System package insert and Vertex™ Reconstruction System package insert for a list of the indications of use. Note: The 3.0mm multi axial screw (MAS) reguires the use of MAS Crosslink™ at each level in which the 3.0mm screw is intended to be used. The lateral offset connectors and MAS extension connectors are intended to be used with 3.5mm and larger diameter multi axial screws. The lateral offset connectors and MAS extension connectors are not intended to be used with 3.5mm and larger diameter multi axial screws. The lateral offset connectors and MAS extension connectors are not intended to be used with 3.0mm screws. Note: Segmental fixation is recommended for these constructs.
Device Story
Posterior spinal fixation system consisting of rods, hooks, polyaxial screws, occipital plates, and connectors; provides immobilization of spinal segments. Subject device adds titanium alloy and cobalt-chrome transition rods (3.5/4.75 mm) to UNiD™ OCT Patient-Specific Rod line. Rods are pre-bent industrially based on patient-specific anatomy prior to surgery; surgeon-implanted to stabilize spine. Compatible with Infinity™ OCT, PASS OCT, and CD Horizon™ systems. Benefits include rigid fixation for spinal stabilization in trauma, tumor, or degenerative cases.
Clinical Evidence
Bench testing only. Performed ASTM F1717 static and dynamic compression bending and static torsion testing to validate compatibility of the 4.75 mm rod portion with the CD Horizon Spinal System.
Technological Characteristics
Materials: Titanium alloy Ti-6Al-4V ELI (ISO 5832-3, ASTM F136) and Cobalt-chromium molybdenum alloy (ISO 5832-12, ASTM F1537). Components: Rods, hooks, polyaxial screws, occipital plates, connectors. Connectivity: Mechanical coupling to other spinal systems. Sterilization: Not specified.
Indications for Use
Indicated for patients with acute/chronic instabilities of craniocervical junction, cervical spine (C1-C7), and thoracic spine (T1-T3) requiring immobilization as adjunct to fusion. Conditions include traumatic fractures/dislocations, instability/deformity, failed fusions, cervical tumors, and degenerative disease (radiculopathy, discogenic pain, facet instability). Also indicated for spinal column integrity restoration in patients with advanced cervical tumors where fusion is not feasible due to limited life expectancy.
Regulatory Classification
Identification
Posterior cervical screw systems are comprised of multiple, interconnecting components, made from a variety of materials that allow an implant system to be built from the occiput to the upper thoracic spine to fit the patient's anatomical and physiological requirements, as determined by preoperative cross-sectional imaging. Such a spinal assembly consists of a combination of bone anchors via screws (i.e., occipital screws, cervical lateral mass screws, cervical pedicle screws, C2 pars screws, C2 translaminar screws, C2 transarticular screws), longitudinal members (e.g., plates, rods, including dual diameter rods, plate/rod combinations), transverse or cross connectors, interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors), and closure mechanisms (e.g., set screws, nuts). Posterior cervical screw systems are rigidly fixed devices that do not contain dynamic features, including but not limited to: non-uniform longitudinal elements or features that allow more motion or flexibility compared to rigid systems.Posterior cervical screw systems are intended to provide immobilization and stabilization of spinal segments in patients as an adjunct to fusion for acute and chronic instabilities of the cervical spine and/or craniocervical junction and/or cervicothoracic junction such as: (1) Traumatic spinal fractures and/or traumatic dislocations; (2) deformities; (3) instabilities; (4) failed previous fusions (e.g., pseudarthrosis); (5) tumors; (6) inflammatory disorders; (7) spinal degeneration, including neck and/or arm pain of discogenic origin as confirmed by imaging studies (radiographs, CT, MRI); (8) degeneration of the facets with instability; and (9) reconstruction following decompression to treat radiculopathy and/or myelopathy. These systems are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
Special Controls
*Classification.* Class II (special controls). The special controls for posterior cervical screw systems are:(1) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(2) Nonclinical performance testing must demonstrate the mechanical function and durability of the implant.
(3) Device components must be demonstrated to be biocompatible.
(4) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(5) Labeling must include the following:
(i) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(ii) Intended use and indications for use including levels of fixation;
(iii) Device specific warnings, precautions, and contraindications that include the following statements:
(A) “Precaution: Preoperative planning prior to implantation of posterior cervical screw systems should include review of cross-sectional imaging studies (
*e.g.,* CT and/or MRI) to evaluate the patient's cervical anatomy including the transverse foramen, neurologic structures, and the course of the vertebral arteries. If any findings would compromise the placement of these screws, other surgical methods should be considered. In addition, use of intraoperative imaging should be considered to guide and/or verify device placement, as necessary.”(B) “Precaution: Use of posterior cervical pedicle screw fixation at the C3 through C6 spinal levels requires careful consideration and planning beyond that required for lateral mass screws placed at these spinal levels, given the proximity of the vertebral arteries and neurologic structures in relation to the cervical pedicles at these levels.”
(iv) Identification of magnetic resonance (MR) compatibility status;
(v) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user, and;
(vi) Detailed instructions of each surgical step, including device removal.
K210449 — Infinity OCT System, PASS OCT® Spinal System · Medtronic · Mar 18, 2021
K153169 — PASS OCT Patient Specific Rods · Medicrea International · Jan 29, 2016
K151031 — Virage OCT Spinal Fixation System · Zimmer Spine, Inc. · Jun 24, 2015
K140738 — PASS LP SPINAL SYSTEM · Medicrea International · Nov 4, 2014
K171444 — YUKON OCT Spinal System · K2m, Inc. · Aug 8, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
April 19, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The FDA part of the logo is in blue, with the acronym "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION" next to it.
Medicrea International (Medtronic) Ms. Karine Trogneux Regulatory Affairs Manager 5389 Route de Strasbourg - Vancia Rillieux-La-Pape, 69140 France
Re: K220810
Trade/Device Name: Infinity™ OCT System and PASS OCT Spinal System Regulation Number: 21 CFR 888.3075 Regulation Name: Posterior Cervical Screw System Regulatory Class: Class II Product Code: NKG Dated: March 15, 2022 Received: March 21, 2022
Dear Ms. Trogneux:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220810
Device Name Infinity™ OCT System
## Indications for Use (Describe)
The Infinity™ OCT System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7), and the thoracic spine from T1-T3:
- · Traumatic spinal fractures and/or traumatic dislocations.
- Instability or deformity.
- = Failed previous fusions (e.g. pseudarthrosis).
- · Tumors involving the cervical spine.
· Degenerative disease, including intractable radiculopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability.
The Infinity™ OCT System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
The Infinity™ OCT System may be used with PASS OCT Patient Specific UNiD OCT Rods. In order to achieve additional levels of fixation, the Infinity™ OCT System may be connected to the CD Horizon™ Spinal System and VertexTM Reconstruction System rods with the Infinity™ OCT System rod connectors. Transition rods with differing diameters may also be used to connect the Infinity™ OCT System to the CD Horizon™ Spinal System. Refer to the CD Horizon™ Spinal System package insert and Vertex™ Reconstruction System package insert for a list of the indications of use.
Note: The 3.0mm multi axial screw (MAS) reguires the use of MAS Crosslink™ at each level in which the 3.0mm screw is intended to be used.
The lateral offset connectors and MAS extension connectors are intended to be used with 3.5mm and larger diameter multi axial screws. The lateral offiset connectors and MAS extension connectors are not intended to be used with 3.0mm screws.
Note: Segmental fixation is recommended for these constructs.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220810
Device Name PASS OCT Spinal System
### Indications for Use (Describe)
The PASS OCT spinal system is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine from T1-T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The PASS OCT spinal system is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. In order to achieve additional levels of fixation, the PASS OCT spinal system may be connected to the PASS LP spinal system with the dual diameter rods. Refer to the PASS LP spinal system package insert for a list of the PASS LP spinal system indications of use.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-size:100%;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Medicrea. The logo consists of a stylized "M" on the left, followed by the word "MEDICREA" in black and red letters. The letters "MEDIC" are in black, while "REA" is in red, and there is a trademark symbol in the upper right corner of the logo.
MEDICREA INTERNATIONAL Special 510(k) Application UNiD™ OCT Patient Specific Rods
# 510(k) SUMMARY
# MEDICREA INTERNATIONAL S.A. Infinity™ OCT System and PASS OCT Spinal System
# Date Prepared: April 15, 2022
| I. Submitter and contact Person information | | |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| Submitter | MEDICREA INTERNATIONAL (Medtronic)<br>5389 route de Strasbourg - Vancia<br>RILLIEUX-LA-PAPE 69140<br>FR | |
| Contact Person | Karine Trogneux<br>MEDICREA INTERNATIONAL S.A. (Medtronic)<br>5389 route de Strasbourg - Vancia<br>RILLIEUX-LA-PAPE 69140<br>FR<br>Phone: 00 33 4 69 85 95 39 | |
| II. Subject Device Information | | |
| Device Name | Infinity™ OCT System and PASS OCT Spinal System | |
| Regulatory<br>Identification/<br>Classification | Product Codes: | NKG |
| | Common Name: | Posterior Cervical Screw System |
| | Regulation Number: | 21 CFR§ 888.3075 |
| | Classification Name: | Posterior Cervical Screw System |
| | Device Classification: | Class II |
| III. Predicate and Reference Devices | | |
| Primary Predicate<br>Device | | Infinity™ OCT System (K163375; cleared the 08/21/2017) |
| | Product Codes: | NKG |
| | Regulation Number: | 21 CFR 888.3075 |
| | Classification Name: | Posterior Cervical Screw System |
| Additional<br>Predicate devices | | PASS OCT Patient Specific Rods (K153169; cleared the 01/29/2016) |
| | Product Codes: | NKG |
| | Regulation Number: | 21 CFR 888.3075 |
| | Classification Name: | Posterior Cervical Screw System |
| | | Infinity™ OCT System, PASS OCT Spinal System (K210449; cleared the<br>03/18/2021) |
| | Product Codes: | NKG |
| | Regulation Number: | 21 CFR 888.3075 |
| | Classification Name: | Posterior Cervical Screw System |
| Reference Device | | CD Horizon™ Spinal System (K113174; cleared the 11/21/2011) |
| | Product Codes: | NKB |
| | Regulation Number: | 21 CFR 888.3070 |
| | Classification Name: | Thoracolumbosacral Pedicle Screw System |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Medicrea. The logo consists of a stylized "M" made up of two overlapping shapes in gray and black, followed by the word "MEDICREA" in black and red. The letters "MEDIC" are in black, while "REA" is in red, and there is a trademark symbol next to the letter A.
MEDICREA INTERNATIONAL Special 510(k) Application UNiD™ OCT Patient Specific Rods
### IV. Description of Subject Device
The Infinity™ OCT System and PASS OCT Spinal System (including UNID™ OCT Patient-Specific Rods) are posterior systems, which consist of a variety of shapes and sizes of rods, hooks, polyaxial screws, occipital plates, occipital bone screws, and connection components, which can be rigidly locked to the rod in a variety of configurations.
The implants are manufactured in titanium alloy Ti-6Al-4V ELI conforming to ISO 5832-3 specifications and ASTM F136 specifications, Cobalt-chromium molybdenum alloy Co-Cr28Mo6 according to ISO 5832-12 and ASTM F1537.
The purpose of this submission is to add titanium alloy and cobalt chrome 3.5/4.75 mm transition rods to the line of UNID™ OCT Patient-Specific Rods. The subject rods are designed and manufactured for one specific patient and are compatible with Infinity™ OCT System and PASS OCT Spinal Systems (cranially) and the CD Horizon Spinal System (caudally). The UNID™ OCT Patient-Specific Rods must be used during surgery for this patient only and must not be reused (single use only).
### V. Indications for Use
Infinity™ OCT System
The Infinity™ OCT System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7), and the thoracic spine from T1-T3:
▪ Traumatic spinal fractures and/or traumatic dislocations.
- Instability or deformity.
- Failed previous fusions (e.g. pseudarthrosis).
- Tumors involving the cervical spine.
▪ Degenerative disease, including intractable radiculopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability.
The Infinity™ OCT System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
The Infinity™ OCT System may be used with PASS OCT Patient Specific UNiD OCT Rods. In order to achieve additional levels of fixation, the Infinity™ OCT System may be connected to the CD Horizon™ Spinal System and Vertex™ Reconstruction System rods with the Infinity™ OCT System rod connectors. Transition rods with differing diameters may also be used to connect the Infinity™ OCT System to the CD Horizon™ Spinal System. Refer to the CD Horizon™ Spinal System package insert and Vertex™ Reconstruction System package insert for a list of the indications of use.
Note: The 3.0mm multi axial screw (MAS) requires the use of MAS Crosslink™ at each level in which the 3.0mm screw is intended to be used.
The lateral offset connectors and MAS extension connectors are intended to be used with 3.5mm and larger diameter multi axial screws. The lateral offset connectors and MAS extension connectors are not intended to be used with 3.0mm screws.
Note: Segmental fixation is recommended for these constructs.
## PASS OCT spinal system
The PASS OCT spinal system is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine from T1-T3: traumatic spinal fractures and/or traumatic
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Medicrea. The logo consists of the letters "M", "E", "D", "I", "C", "R", "E", and "A". The first "M" is gray, the next four letters are black, and the last three letters are red. There is a trademark symbol in the upper right corner of the logo.
MEDICREA INTERNATIONAL Special 510(k) Application UNiD™ OCT Patient Specific Rods
dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical spine; and degenerative disease, including
intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The PASS OCT spinal system is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
In order to achieve additional levels of fixation, the PASS OCT spinal system may be connected to the PASS LP spinal system with the dual diameter rods. Refer to the PASS LP spinal system package insert for a list of the PASS LP spinal system indications of use.
### Comparison of Technological Characteristics with the Predicate and Reference Devices VI.
The subject UNiD™ OCT Patient Specific Rods do not change to the fundamental scientific technology, indications for use, intended use, materials, and levels of attachment as compared to previously cleared UNiD™ OCT Patient Specific Rods. The subject and predicate UNID™ OCT Patient Specific Rods are directly adapted to a unique patient by means of an industrial bending prior to surgery, which differ from the other rods in the Infinity™ OCT System and the PASS OCT Spinal System which are cut and bent by the surgeon based on the need of the individual case.
The compatibility between the Infinity™ OCT System and 3.5 mm section of the subject UNiD™ OCT Patient Specific Rods has been previously established in K210449.The compatibility between the CD Horizon Spinal System and the 4.75 mm section of the subject UNiD™ OCT Patient Specific Rods has been previously established in K220724.
### VII. Performance Data (Non-Clinical Test Summary)
In accordance with the Guidance for Industry and FDA Staff – Spinal System 510(k)'s, Medicrea has evaluated the subject devices to demonstrate substantial equivalence to the predicate devices. Rationales were provided confirming that the use of the UNID™ OCT patient specific Rods did not introduce a new worst case compared to the predicate Infinity™ OCT Rods. However, to confirm the compatibility of the caudal 4.75 mm portion of the subject rods with the CD Horizon Spinal System, the following tests were performed per ASTM F1717: static and dynamic compression bending, and static torsion.
### VIII. Conclusion
Based on the test results and supporting information provided in this premarket notification, the subject 3.5/4.75 mm UNID™ OCT Patient Specific Rods are substantially equivalent to the identified predicate devices
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.