SPINAL ELEMENTS POSTERIOR CERVICAL/THORACIC SPINAL SYSTEM

K120467 · Spinal Elements, Inc. · NKB · Jul 30, 2012 · Orthopedic

Device Facts

Record IDK120467
Device NameSPINAL ELEMENTS POSTERIOR CERVICAL/THORACIC SPINAL SYSTEM
ApplicantSpinal Elements, Inc.
Product CodeNKB · Orthopedic
Decision DateJul 30, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Spinal Elements Posterior Cervical/Thoracic Spinal System is intended for posterior fixation of the cervical and thoracic spine (CI-T3) for the following conditions: degenerative disc disease (neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies); spondylolisthesis; spinal stenosis; trauma (fracture/dislocation); failed previous fusion; tumors. The use of polyaxial pedicle screws is limited to placement in T1-T3 for treating thoracic conditions only. The screws are not intended to be placed in the cervical spine.

Device Story

System comprises titanium alloy fixation components: screws, hooks, rods, and transverse connectors. Surgeon implants components into posterior elements or pedicles of cervical/thoracic vertebrae (C1-T3) to stabilize spine. Mechanical fixation achieved by joining rods to screws/hooks/connectors. Used in surgical settings by orthopedic or neurosurgeons. Provides structural support to facilitate fusion; addresses spinal instability caused by degenerative disease, trauma, or tumors. Polyaxial pedicle screws limited to T1-T3 thoracic application. Benefits include spinal stabilization and correction of deformity.

Clinical Evidence

Bench testing only. Mechanical performance evaluated per ASTM F1717, including static compression, dynamic compression, and torsion testing. No clinical data provided.

Technological Characteristics

Materials: Titanium alloy (Ti-6Al-4V) conforming to ASTM F 136 or ISO 5832-3. Components: Screws, hooks, rods, transverse connectors. Fixation: Mechanical joining of components. Connectivity: None. Sterilization: Not specified.

Indications for Use

Indicated for patients requiring posterior cervical and thoracic (C1-T3) spinal fixation for degenerative disc disease, spondylolisthesis, spinal stenosis, trauma (fracture/dislocation), failed fusion, or tumors. Polyaxial pedicle screws restricted to T1-T3 thoracic use; contraindicated for cervical spine placement.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K120467 Page 1 of 2 Spinal Elements, Inc. Premarket Notification - Spinal Elements Posterior Cervical / Thoracic Spinal System J # 510(k) Summary Spinal Elements Posterior Cervical / Thoracic Spinal System JUL 30 2012 # 510(k) Number: K120467 Manufacturer Identification Submitted by: Spinal Elements, Inc. 2744 Loker Ave. W., Suite 100 Carlsbad, CA 92010 760-607-0121 761-607-0125 (fax) Contact Information: Benjamin A. Kimball Regulatory Affairs Manager Spinal Elements, Inc. 2744 Loker Ave. W., Suite 100 Carlsbad, CA ·92010 760-607-1816 Bkimball@spinalelements.com Date Prepared: July 26, 2012 Device Indentification Proprietary Name Spinal Elements Posterior Cervical / Thoracic Spinal System Spinal Interlaminal Fixation and Spinal Intervetebral Common Name Device Classification Classification Name Fixation Orthosis and/or Pedicle Screw System 21 CFR 888.3050 and / or 21 CFR 888.3070 Spinal interlaminal fixation orthosis, Spinal intervertebral body fixation orthosis. Pedicle screw spinal system Class III Proposed Regulatory Class Device Product Code: MNI, KWP, NKB Purpose of this 510(k) This 510(k) seeks clearance for a new system. {1}------------------------------------------------ ## Device Description The Spinal Elements Posterior Cervical/Thoracic System consists of a variety of fixation devices that are attached to the spine by means of screws and hooks placed in/or on the pedicles or posterior elements of the various vertebrae, rods that span the distance between the screws/hooks, and transverse connectors. Screws are intended for attachment to the thoracic (T1-T3) spine only. All components of the system are manufactured from titanium alloy (Ti-6A1-4V conforming to ASTM F 136 or ISO 5832-3). The system achieves fixation by the mechanical joining of the rods, screws, hooks, and connectors. A variety of constructs may be assembled to suit the individual pathology and anatomy of the patient. #### Intended Use of the Device The Spinal Elements Posterior Cervical/Thoracic Spinal System is intended for posterior fixation of the cervical and thoracic spine (CI-T3) for the following conditions: degenerative disc disease (neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies); spondylolisthesis; spinal stenosis; trauma (fracture/dislocation); failed previous fusion; tumors. The use of polyaxial pedicle screws is limited to placement in T1-T3 for treating thoracic conditions only. The screws are not intended to be placed in the cervical spine. #### Performance Data Non-clinical, mechanical testing was performed in accordance with ASTM F1717 to determine the performance profile of the device. Testing included: - > Static compression testing - > Dynamic compression testing - > Torsion testing All test results indicate the system will perform as intended. ## Substantial Equivalence Spinal Elements Posterior Cervical/Thoracic Spinal System is substantially equivalent to Alphatec's Solanas Posterior Stabilization System (K052201 and K071380), Depuy Spine's Summit® Ocipito-Cervico-Thoracic (OCT) Spinal System (K002733 and K010681), and SeaSpine's Sierra System (K062934) in indications, general design features, performance, function, and materials. {2}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 JUL 30 2012 Spinal Elements, Incorporated % Mr. Benjamin A. Kimball Regulatory Affairs Manager 2744 Loker Avenue West, Suite 100 Carlsbad, California 92010 Re: K120467 Trade/Device Name: Spinal Elements Posterior Cervical/Thoracic Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNI, KWP Dated: July 18, 2012 Received: July 19, 2012 Dear Mr. Kimball: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability , warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ # Page 2 - Mr. Benjamin A. Kimball Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Mark N. Mel Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # · Indications for Use 510(k) K:120467 Spinal Elements Posterior Cervical / Thoracic Spinal System Device Name: ## Indications for Use: The Spinal Elements Posterior Cervical/Thoracic Spinal System is intended for posterior fixation of the cervical and thoracic spine (CI-T3) for the following conditions: degenerative disc disease (neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies); spondylolisthesis: spinal stenosis; trauma (fracture/dislocation); failed previous fusion; tumors The use of polyaxial pedicle screws is limited to placement in T1-T3 for treating thoracic conditions only. The screws are not intended to be placed in the cervical spine. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) signature Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices Page 1 of 1 K120467 510(k) Number_
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...