K180700 · Medtronic Sofamor Danek USA, Inc. · NDN · May 18, 2018 · Orthopedic
Device Facts
Record ID
K180700
Device Name
KYPHON HV-R Bone Cement
Applicant
Medtronic Sofamor Danek USA, Inc.
Product Code
NDN · Orthopedic
Decision Date
May 18, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3027
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K180700 · May 18, 2018
KYPHON HV-R Bone Cement
Medtronic Sofamor Danek USA, Inc.
Retrospective clinical literature review
A retrospective review of five clinical articles involving 462 patients was conducted to evaluate the safety and efficacy of PMMA bone cement when used in sacral vertebroplasty or sacroplasty procedures, supporting the expanded indication.
Retrospective literature review; Sacral vertebroplasty; Sacroplasty; PMMA bone cement
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective clinical literature review; Retrospective literature review
Safety and efficacy of PMMA bone cement in the sacrum
Indications for Use
KYPHON™ HV-RTM Bone Cement is indicated for the treatment of pathological fractures of the vertebral body due to osteoporosis, cancer, or benign lesions using a cementoplasty or vertebroplasty procedure. It is also indicated for the fixation of pathological fractures of the sacral vertebral body or ala using sacral vertebroplasty or sacroplasty. Cancer includes multiple myeloma and metastatic lesions, including those arising from breast or lung cancer, or lymphoma. Benign lesions include hemangioma and giant cell tumor. Pathological fracture may include a symptomatic vertebral body microfracture (as documented by appropriate imaging and/or presence of a lytic lesion) without obvious loss of vertebral body height.
Device Story
Two-component PMMA bone cement system; powder (PMMA copolymer, barium sulfate, benzoyl peroxide) and liquid (methylmethacrylate monomer, hydroquinone, N,N-dimethyl-p-toluidine) mixed prior to use. Used by physicians in clinical settings for cementoplasty, vertebroplasty, or sacroplasty procedures. Provides structural fixation for pathological fractures. Benefits include stabilization of vertebral/sacral fractures, potentially reducing pain and restoring structural integrity. Clinical decision-making guided by imaging (e.g., identifying lytic lesions or microfractures).
Clinical Evidence
Retrospective clinical literature review of 462 patients across five articles demonstrated safety and efficacy of PMMA bone cement in sacral vertebroplasty/sacroplasty. Cadaveric study evaluated cement extravasation behavior using long-axis and short-axis surgical techniques; extravasation rates were within acceptable ranges.
Indicated for patients with pathological vertebral body fractures due to osteoporosis, cancer (multiple myeloma, metastatic lesions from breast/lung/lymphoma), or benign lesions (hemangioma, giant cell tumor). Also indicated for fixation of pathological sacral vertebral body or ala fractures via sacral vertebroplasty or sacroplasty. Includes symptomatic microfractures without height loss.
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is blue.
May 18, 2018
Medtronic Sofamor Danek USA, Inc. Laveeda Leflore Regulatory Affairs Specialist 1800 Pyramid Place Memphis. Tennessee 38132
Re: K180700
Trade/Device Name: KYPHON™ HV-R™ Bone Cement Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: NDN Dated: March 15, 2018 Received: March 16, 2018
Dear Ms. Leflore:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{2}------------------------------------------------
K180700
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
## Indications for Use
510(k) Number (if known)
#### K180700
Device Name KYPHON™ HV-R™ Bone Cement
#### Indications for Use (Describe)
KYPHON™ HV-RTM Bone Cement is indicated for the treatment of pathological fractures of the vertebral body due to osteoporosis, cancer, or benign lesions using a cementoplasty or vertebroplasty procedure. It is also indicated for the fixation of pathological fractures of the sacral vertebral body or ala using sacral vertebroplasty or sacroplasty. Cancer includes multiple myeloma and metastatic lesions, including those arising from breast or lung cancer, or lymphoma. Benign lesions include hemangioma and giant cell tumor. Pathological fracture may include a symptomatic vertebral body microfracture (as documented by appropriate imaging and/or presence of a lytic lesion) without obvious loss of vertebral body height.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|-------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) Summary Medtronic Sofamor Danek USA, Inc.
## May 15, 2018
| Submitter | Medtronic Sofamor Danek USA, Inc.<br>1800 Pyramid Place<br>Memphis, Tennessee 38132<br>Telephone: (901) 396-3133<br>Fax: (901) 346-9738 |
|------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact | Laveeda Leflore<br>Regulatory Affairs Specialist<br>Direct Telephone: 901-399-2953 |
| | Jeff Sprague<br>Sr. Regulatory Affairs Program Manager<br>Direct Telephone: 901-344-1326 |
| Date Prepared | May 15, 2018 |
| Common Name | Kyphon™ HV-R™ Bone Cement |
| Regulatory Class | Kyphon™ HV-R™ Bone Cement |
| Regulation Number | Class II |
| Regulation Name and Device | 21 CFR 888.3027 |
| Product Classification Code | Polymethylmethacrylate (PMMA) bone cement<br>NDN |
| Predicate Devices | 1) Kyphon™ Xpede™ Bone Cement K163032, S.E. 02/27/2017 (Primary<br>Predicate) |
| | 2) Kyphon™ HV-R™ Bone Cement K160983, S.E. 08/24/2016 |
| | The predicate devices have not been subject to a design related recall. |
| Description of Device | Kyphon™ HV-R™ Bone Cement is provided as a two-component system.<br>The powder component consists of a PMMA copolymer<br>(polymethylmethacrylate/ methyl-methacrylate-styrene copolymer) with<br>barium sulfate as a radiopacifier and benzoyl peroxide as an initiator. The<br>liquid component consists of methylmethacrylate monomer, with the<br>addition of hydroquinone as a stabilizer and N,N-dimethyl-p-toluidine as<br>a promoter. The powder and liquid components are mixed prior to use. |
| Indications for Use: | KyphonTM HV-RTM Bone Cement is indicated for the treatment of<br>pathological fractures of the vertebral body due to osteoporosis, cancer, or<br>benign lesions using a cementoplasty (i.e. kyphoplasty or vertebroplasty)<br>procedure. It is also indicated for the fixation of pathological fractures of<br>the sacral vertebral body or ala using sacral vertebroplasty or sacroplasty.<br>Cancer includes multiple myeloma and metastatic lesions, including those<br>arising from breast or lung cancer, or lymphoma. Benign lesions include<br>hemangioma and giant cell tumor. Pathologic fracture may include a<br>symptomatic vertebral body microfracture (as documented by appropriate<br>imaging and/or presence of a lytic lesion) without obvious loss of<br>vertebral body height. |
| Comparison of Technological<br>Characteristics with the<br>Predicate Devices | The subject KyphonTM HV-RTM Bone Cement has the same or similar<br>indications for use, intended use, materials and fundamental scientific<br>technology fundamental scientific technology as the predicates KyphonTM<br>XpedeTM Bone Cement K163032, (S.E. 02/27/2017) and KyphonTM HV-RTM<br>Bone Cement (K160983, S.E. 8/24/2016). The subject device utilizes<br>equivalent implant materials, sterilization methods and bacterial endotoxin<br>testing applying the same 20 EU/ml pyrogen limit specifications utilizing<br>the gel clot test method as the predicate KyphonTM HV-RTM Bone Cement<br>(K160983, S.E. 8/24/2016). |
| Performance Data | Clinical literature data and non-clinical cadaveric testing is provided to<br>support the substantial equivalence of the subject device.<br><br>A retrospective clinical literature review was performed to examine the<br>potential benefits and associated risks of using PMMA bone cement using<br>a sacral vertebroplasty or sacroplasty procedure. The five articles<br>reviewed provided clinical outcomes of 462 patients utilizing both the<br>long-axis and short-axis surgical technique demonstrating the safety and<br>efficacy of PMMA bone cement when used in the sacrum.<br><br>A cadaver study in support of the expanded indication was completed.<br>This study was performed to evaluate the extravasation behavior of the<br>KyphonTM HV-RTM Bone Cement during sacroplasty procedures. The study<br>also defined the surgical procedure steps and imaging needed to minimize<br>the risk of extravasation for the sacroplasty procedure. Both the long-axis<br>and short-axis surgical techniques were evaluated for cement<br>extravasation. The procedure and imaging review was performed by<br>trained physicians. The calculated extravasation rate with the KyphonTM<br>HV-RTM Bone Cement was within the extravasation rate range of the |
| Conclusion | Based on the provided performance data, the subject Kyphon™ HV-R™<br>Bone Cement is substantially equivalent to the primary predicate Kyphon™<br>Xpede™ Bone Cement (K163032, S.E. 02/27/2017) and Kyphon™ HV-R™<br>Bone Cement (K160983, S.E. 08/24/2016). |
{5}------------------------------------------------
{6}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.