Parallax® Acrylic Resin Cartridge with TRACERS-Ta is intended for the fixation of pathological fractures of the vertebral body using vertebroplasty or kyphoplasty procedures. Painful vertebral compression fractures may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (metastatic cancers, myeloma).
Device Story
Parallax Acrylic Resin Cartridge with TRACERS-Ta is an opacified polymethylmethacrylate (PMMA) bone cement; used for fixation of pathological vertebral body fractures during vertebroplasty or kyphoplasty procedures. Device functions as a bone filler/stabilizer; provides structural support to fractured vertebrae. Used in clinical settings by physicians; delivered via cartridge system. Benefits include stabilization of painful fractures caused by osteoporosis or lesions; reduction of pain; restoration of vertebral integrity.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and intended use compared to predicate devices.
Technological Characteristics
Opacified polymethylmethacrylate (PMMA) bone cement. Contains TRACERS-Ta (tantalum) as a radiopaque agent. Delivered via cartridge system. Class II device per 21 CFR 888.3027.
Indications for Use
Indicated for fixation of pathological vertebral body fractures via vertebroplasty or kyphoplasty. Patient population includes individuals with painful vertebral compression fractures due to osteoporosis, benign lesions (hemangioma), or malignant lesions (metastatic cancers, myeloma).
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
Predicate Devices
Parallax® Acrylic Resin with TRACERS® -Ta (K050864)
Reference Devices
Parallax Acrylic Resin Cartridge with TRACERS (K053180)
Related Devices
K053180 — PARALLAX ACRYLIC RESIN CARTRIDGE WITH TRACERS · Arthrocare Corp. · Jan 25, 2006
K042947 — PARALLAX ACRYLIC RESIN WITH TRACERS, MODEL KP-PAR-003 · Arthrocare Corp. · Dec 27, 2004
K050864 — PARALLAX ACRYLIC RESIN WITH TRACERS TA · Arthrocare Corp. · Jun 17, 2005
K122175 — MENDEC SPINE HV, MENDEC SPINE HV SYSTEM · Tecres S.P.A. · Mar 28, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
K060425
Image /page/0/Picture/1 description: The image shows the logo for ArthroCare Corporation. The logo consists of a stylized graphic to the left of the company name. The graphic is composed of several curved lines that converge to form a circular shape. The text "ArthroCare" is in a bold, sans-serif font, with "Corporation" in a smaller font size below it.
# 510(k) Summary
JUL 1 2 2006
### ArthroCare Corporation Parallax® Acrylic Resin Cartridge with TRACERS-Ta
| General Information | |
|------------------------------------|-----------------------------------------------------------------------|
| Manufacturer: | ArthroCare, Corporation<br>680 Vaqueros Avenue<br>Sunnyvale, CA 94085 |
| Establishment Registration Number: | 2951580 |
| Contact Person: | Valerie Defiesta-Ng<br>Director, Regulatory Affairs |
| Date Prepared: | February 15, 2006 |
Device Description
| Classification: | PMMA Bone Cement: Class II per 21 CFR<br>888.3027 |
|----------------------|------------------------------------------------------------------|
| Trade Name: | Parallax <sup>®</sup> Acrylic Resin Cartridge with<br>TRACERS-Ta |
| Device Code: | NDN |
| Generic/Common Name: | Bone Cement |
Predicate Device
| Parallax <sup>®</sup> Acrylic Resin with TRACERS <sup>®</sup> -Ta | K050864 |
|-------------------------------------------------------------------|---------|
|-------------------------------------------------------------------|---------|
#### Parallax® Acrylic Resin with TRACERS®-Ta
Parallax® Acrylic Resin Cartridge with TRACERS® K050864 K053180
### Intended Use
Parallax® Acrylic Resin Cartridge with TRACERS-Ta is intended for the fixation of pathological fractures of the vertebral body using vertebroplasty or kyphoplasty procedures. Painful vertebral compression fractures may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (metastatic cancers, myeloma).
### Product Description
Parallax® Acrylic Resin Cartridge with TRACERS-Ta is an opacified polymethylmethacrylate bone cement.
### Substantial Equivalence
The subject device is substantially equivalent to the ArthroCare Parallax Acrylic Resin with TRACERS-Ta (K050864) and the Parallax Acrylic Resin Cartridge with TRACERS (K053180).
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the Department of Health and Human Services, USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three curved lines representing its wings or feathers.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 1 2 2006
Arthrocare Corporation, Inc. % Ms. Valerie Defiesta-Ng Director of Regulatory Affairs 680 Vaqueros Avenue Sunnyvale, California 94085-3523
Re: K060425
Trade/Device Name: Parallax Acrylic Resin Cartridge with Tracers - Ta Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: NDN Dated: June 1, 2006 Received: June 2, 2006
Dear Ms. Defiesta-Ng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 – Ms. Valerie Defiesta-Ng
This letter will allow you to begin marketing your device as described in your Section 510(k) I ma letter notification. The FDA finding of substantial equivalence of your device to a legally premaince noticate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please in you attine office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Herbert Lemmercus
Mark N. Melkerson
Director
Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use Statement
510(k) Number (if known):
Device Name:
Parallax® Acrylic Resin Cartridge with TRACERS-Ta
Indications for Use:
Parallax® Acrylic Resin Cartridge with TRACERS-Ta is intended for the fixation of pathological fractures of the vertebral body using vertebroplasty or kyphoplasty procedures. Painful vertebral compression fractures may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (metastatic cancers, myeloma).
| Prescription Use (Part 21 CFR 801 Subpart D) | X | AND/OR | Over-The-Counter Use (21 CFR 801 Subpart C) | |
|----------------------------------------------|---|--------|---------------------------------------------|--|
|----------------------------------------------|---|--------|---------------------------------------------|--|
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDHH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative, and Neurological Devices
| 510(k) Number | K060425 | xii |
|---------------|---------|-----|
|---------------|---------|-----|
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.