FIDJI-VERTEBRAL BODY REPLACEMENT TYPE 1, 2, 3 AND 4
K042714 · Spine Next America Corporation · MQP · Dec 29, 2004 · Orthopedic
Device Facts
Record ID
K042714
Device Name
FIDJI-VERTEBRAL BODY REPLACEMENT TYPE 1, 2, 3 AND 4
Applicant
Spine Next America Corporation
Product Code
MQP · Orthopedic
Decision Date
Dec 29, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3060
Device Class
Class 2
Attributes
Therapeutic
Intended Use
FIDJI types 1, 2, 3, and 4 are vertebral body replacement devices that are intended for use in the thoracic and/or thoracolumbar spine (T1-L5) to replace a collapsed, damaged or unstable vertebral body resected or excised (i.e., partial or total vertebrectomy procedures) due to tumor or trauma (i.e., fracture). The device is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of fusion for a prolonged period. These devices are intended to be used with bone graft and supplemental fixation, such as the Spinal Concepts, Inc. InCompass System.
Device Story
FIDJI Types 1, 2, 3, and 4 are hollow, tapered vertebral body replacement blocks; machined from PEEK OPTIMA (polyetheretherketone); feature angled teeth on cephalad and caudal surfaces for stability; include tantalum inserts for radiographic visualization. Used in thoracic/thoracolumbar spine (T1-L5) to replace collapsed, damaged, or unstable vertebral bodies following vertebrectomy. Implanted by surgeons; requires supplemental fixation (e.g., InCompass System) and bone graft. Restores biomechanical integrity of spinal column. Mechanical testing confirms functional support under physiologic loads.
Clinical Evidence
Bench testing only. Mechanical testing demonstrated that the devices exhibit the functional requirements to support their use as vertebral body replacements under normal physiologic loads in the spine.
Technological Characteristics
Materials: PEEK OPTIMA (ASTM F2026). Features: Hollow, tapered blocks with angled teeth on cephalad/caudal surfaces; tantalum inserts for radiographic marking. Energy source: None (mechanical implant). Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for patients requiring vertebral body replacement in the thoracic or thoracolumbar spine (T1-L5) following partial or total vertebrectomy due to tumor or trauma (fracture).
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
K042713 — FIDJI-VERTEBRAL BODY REPLACEMENT TYPE 5 · Spine Next America Corporation · Dec 29, 2004
K101302 — ALEUTIAN IBF SYSTEM · K2m, Inc. · Jul 22, 2010
K033702 — BLACKSTONE PEEK VERTEBRAL BODY REPLACEMENT SYSTEM · Blackstone Medical, Inc. · Mar 31, 2004
K062132 — AVS PL PEEK SPACER SYSTEM · Stryker Spine · Aug 16, 2006
K050449 — QUANTUM VERTEBRAL BODY REPLACEMENT · Quantum Orthopedics, Inc. · Jul 7, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(K) Summary
| Submitted By: | Lisa Peterson<br>Regulatory Affairs Manager<br>Spinal Concepts, Inc.<br>5301 Riata Park Court, Bldg. F<br>Austin, TX 78727<br>512-918-2700<br>December 8, 2004 |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name: | Spinal Concepts Inc. FIDJI® Types 1,2,3, and |
| Classification Name: | Vertebral Body Replacement |
| Product Code: | MOP |
Predicate Device: FIDJI types 1,2,3, and 4 are substantially equivalent to the Synthes Vertebral Spacer (K011037), EBI CAS Spine Spacer (K042268), and Medtronic Sofamor Danek Vertestack (K031780).
Device Description: The FIDJI Vertebral Body Replacement Types 1, 2, 3, and 4 devices are hollow blocks that are machined from PEEK OPTIMA® (polyetheretherketone, ASTM F2026); the blocks are tapered to aid in maintaining lordosis of the spine following implantation. The devices have angled teeth on the cephalad and caudal surfaces. The devices include tantalum inserts that serve as a location and orientation markers for radiographs.
Intended Use: FIDJI types 1,2,3, and 4 are vertebral body replacement devices that are intended for use in the thoracic and/or thoracolumbar spine (T1 - L5) to replace a collapsed, damaged or unstable vertebral body resected or excised (i.e., partial or total vertebrectomy procedures) due to tumor or trauma (i.e., fracture). These devices are designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period. These devices are intended to be used with bone graft and supplemental fixation, such as the Spinal Concepts, Inc. InCompass System.
Mechanical Testing: Mechanical testing demonstrated that FIDJI types 1,2,3, and 4 exhibit the functional requirements to support their use as vertebral body replacements under normal physiologic loads in the spine.
Page 1 of 1
{1}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight, composed of three curved lines.
Public Health Service
DEC 2 9 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Lisa Peterson Regulatory Affairs Manager Spinal Concepts, Inc. 5301 Riata Park Court, Building. F Austin, Texas 78727
Re: K042714
Trade/Device Name: FIDJI Vertebral Body Replacement Types 1, 2, 3, and 4 Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: II Product Code: MQP Dated: December 3, 2004 Received: December 6, 2004
Dear Ms. Peterson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 - Mr. Justin Eggleton
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Mark A. Milken
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Ko42714
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K042714
Device Name:
Spinal Concepts, Inc. FIDJI Vertebral Body Replacement Types 1, 2, 3, and 4
Indications for Use:
FIDJI types 1, 2, 3, and 4 are vertebral body replacement devices that are intended for use in the thoracic and/or thoracolumbar spine (T1-L5) to replace a collapsed, damaged or unstable vertebral body resected or excised (i.e., partial or total vertebrectomy procedures) due to tumor or trauma (i.e., fracture). The device is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of fusion for a prolonged period. These devices are intended to be used with bone graft and supplemental fixation, such as the Spinal Concepts, Inc. InCompass System.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
Mark A. Millerson
Division of General, Restorative, and Neurological Devices
**510(k) Number** K042714
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches