The LEXUS Cervical Fixation System is indicated for the following: - DDD (neck pain of discogenic origin with degeneration of the disc as confirmed by . patient history and radiographic studies) - . Spondylolisthesis - . Spinal stenosis - . Fracture/dislocation - . Failed previous fusion - . Tumors . The implants are intended to provide stabilization as an adjunct to fusion when used with autogenous bone graft or allograft following the reduction of fractures/dislocations or trauma in the spine. Hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1 -T3) spine. The pedicle screws are limited to placement in T1 -T3 in treating thoracic conditions only. The pedicle screws are not intended to be placed in or treat conditions involving the cervical spine.
Device Story
Posterior cervical-thoracic spinal fixation system; includes polyaxial screws, reduction polyaxial screws, straight/curved rods, set screws, hooks. Used by surgeons in clinical settings to provide spinal stabilization as adjunct to fusion. Implants placed during surgery to treat degenerative disc disease, spondylolisthesis, stenosis, fractures, dislocations, failed fusions, or tumors. Provides mechanical support to promote bone fusion using autogenous bone graft or allograft. Pedicle screws limited to T1-T3 thoracic placement. System components are titanium alloy.
Clinical Evidence
Bench testing only. Performance evaluated per ASTM F1717 standards, including static and dynamic axial compression, static tension, and static torsion testing.
Technological Characteristics
Top-loading, posterior spinal fixation system. Materials: Titanium alloy (Ti-6Al-4V ELI) per ASTM F136. Components: Polyaxial screws, reduction polyaxial screws, straight/curved rods, set screws, hooks. Mechanical testing per ASTM F1717.
Indications for Use
Indicated for patients with DDD, spondylolisthesis, spinal stenosis, fracture/dislocation, failed previous fusion, or tumors requiring stabilization as an adjunct to fusion in the cervical/upper thoracic (C1-T3) spine. Pedicle screws restricted to T1-T3 thoracic use only; not for cervical spine use.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
{0}------------------------------------------------
K120879,
# MAY 2 3 2012
ge if2:
L&K BIOMED
# 510(k) SUMMARY
The following 510(k) summary is being submitted as required by 21 CFR 807.92(a):
| 1. | Submitter: | L&K BIOMED Co., Ltd.<br>#1104, Ace High-end Tower 3 cha, 371-50, Gasan-Dong,<br>Geumcheon-gu, Seoul 153-803 Republic of Korea<br>Phone.82-2-2624-1475<br>FAX.82-2-2624-1477 |
|----|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Contact Person: | KiHyang Kim<br>E-mail : khkim@lnkbiomed.com |
| | Date prepared: | March 23, 2012 |
### 2. Device Identification
| Proprietary Name | LEXUS Cervical Fixation System |
|---------------------|------------------------------------------------------------------------------------------------------|
| Common Name | Spinal Fixation System |
| Product Code | MNI, KWP |
| Regulatory Class | II |
| Classification Name | Pedicle Screw Spinal System (21CFR888.3070)<br>Spinal Interlaminal Fixation Orthosis (21CFR888.3050) |
### 3. Predicate or legally marketed devices which are substantially equivalent
- . Spinal Concept Inc. / Nex-Link Spinal Fixation (K031985)
- . Aesculap Inc. / S4 Spinal Fixation (K050979, K060152)
- Medtronic Sofamor Danek USA Inc. / VERTEX "M Reconstruction System (K003780)
- L&K BIOMED Co.,Ltd. / LEXUS Cervical Fixation System (K103414)
#### 4. Description of the Device
The purpose of this 510(k) submission is to add reduction polyaxial screws and curved rods. The LEXUS Cervical Spinal Fixation System is a top-loading, multiple component, posterior (cervical-thoracic) spinal fixation system which consists of polyaxial screws, reduction poly screws, straight rods, curved rods, set screws, and hooks.
Materials: All products are made of titanium alloy (Ti-6A1-4V ELI/ in conformance with ASTM F136) approved for medical use.
Special 510(K)
Page 16/ 55
{1}------------------------------------------------
page 2 of 2
L&K BIOMED
## 5. Intended use
The LEXUS Cervical Fixation System is indicated for the following:
- DDD (neck pain of discogenic origin with degeneration of the disc as confirmed by . patient history and radiographic studies)
- . Spondylolisthesis
- . Spinal stenosis
- . Fracture/dislocation
- . Failed previous fusion
- . Tumors '
The implants are intended to provide stabilization as an adjunct to fusion when used with autogenous bone graft or allograft following the reduction of fractures/dislocations or trauma in the spine.
Hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1 -T3) spine.
The pedicle screws are limited to placement in T1 -T3 in treating thoracic conditions only. The pedicle screws are not intended to be placed in or treat conditions involving the cervical spine.
# 6. Comparison of the technology characteristics of the device to predicate and legally marketed devices
The LEXUS Cervical Fixation System is considered substantially equivalent to other legally marketed devices. They are similar in design, material, and indications for use and are expected to be equivalent in safety and effectiveness.
## 7. Performance Data
The LEXUS Cervical Fixation System is tested according to the ASTM F1717, specifically, Static and Dynamic Axial Compression, Static Tension and Static Torsion.
## 8. Conclusion
The LEXUS Cervical Fixation System is substantially equivalent to the devices referenced above and is therefore safe and effective for its intended use.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol. The text is in all caps and appears to be a sans-serif font.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
L&K Biomed Co., Ltd. % Ms. Ki Hyang Kim #1104, Ace High-End Tower 3 cha 371-50, Gasan-Dong, Geumcheon-gu Seoul, 153-803, Republic of Korea
Re: K120879
Trade/Device Name: Lexus Cervical Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNI, KWP Dated: May 03, 2012 Received: May 04, 2012
Dear Ms. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
7 3 2012 MAY
{3}------------------------------------------------
Page 2 - Ms. Ki Hyang Kim
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
Sincerely your
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
L&K BIOMEI
# Indications for Use
510(k) Number : K120879
Device Name : LEXUS Cervical Fixation System
Indications for Use :
The LEXUS Cervical Fixation System is indicated for the following:
- DDD (neck pain of discogenic origin with degeneration of the disc as confirmed by patient . history and radiographic studies)
- . Spondylolisthesis
- . Spinal stenosis
- . Fracture/dislocation
- . Failed previous fusion
- . Tumors
The implants are intended to provide stabilization as an adjunct to fusion when used with autogenous bone graft or allograft following the reduction of fractures/dislocations or trauma in the spine.
Hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1 -T3) spine.
The pedicle screws are limited to placement in T1 -T3 in treating thoracic conditions only. The pedicle screws are not intended to be placed in or treat conditions involving the cervical spine.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use __ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Special 510(K)
(Division Sign-Off) Dision of Surgical, Orthopedic, and Restorative Devices
Page 14/ 55
| | 1000 |
|--|------|
|--|------|
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.