MODIFICATION TO MOSS MIAMI SPINAL SYSTEM POLYAXIAL SCREWS

K030383 · Depuyacromed · MNH · Feb 26, 2003 · Orthopedic

Device Facts

Record IDK030383
Device NameMODIFICATION TO MOSS MIAMI SPINAL SYSTEM POLYAXIAL SCREWS
ApplicantDepuyacromed
Product CodeMNH · Orthopedic
Decision DateFeb 26, 2003
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Intended Use

When used as a posterior, noncervical hook, and/or sacral/iliac screw fixation system, or as an anterior, thoracic/lumbar screw fixation system, the Moss Miami Spinal System is intended to treat scoliosis, kyphosis and lordosis, fracture, loss of stability due to tumor, spinal stenosis, spondylolisthesis, a previously failed fusion surgery or degenerative disc disease (i.e., discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies). When used as a pedicle screw fixation system of the noncervical spine in skeletally mature patients, the Moss Miami Spinal System is indicated for deqenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The Moss Miami Spinal System is also indicated for pedicle screw fixation in skeletally mature patients with severe spondylolisthesis (Grades 3 and 4) at the L5 -- S1 vertebral joint, having fusions with autogenous bone graft. with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3 - S1), and for whom the device system is intended to be removed after solid fusion is attained.

Device Story

Moss Miami Spinal System Polyaxial Screws are orthopedic implants designed to stabilize the noncervical spine. The system consists of polyaxial screws and polyaxial reduction screws configured to accept 5.5mm rods. Used by surgeons in clinical settings to provide rigid fixation during spinal fusion procedures. The device acts as a mechanical construct to restore spinal stability, treat deformities, and facilitate bone fusion. It is intended for use with autogenous bone graft; the system is designed for removal after solid fusion is achieved in specific severe spondylolisthesis cases.

Clinical Evidence

No clinical data provided; substantial equivalence is based on design, intended use, and technological characteristics compared to predicate devices.

Technological Characteristics

Polyaxial screw and rod fixation system. Designed to accept 5.5mm rods. Mechanical spinal fixation orthosis. Class II device. Materials and specifications are consistent with previously cleared Moss Miami components.

Indications for Use

Indicated for skeletally mature patients requiring noncervical spinal fixation for scoliosis, kyphosis, lordosis, fracture, spinal tumor, spinal stenosis, spondylolisthesis, degenerative disc disease, or failed fusion. Pedicle screw fixation specifically indicated for degenerative spondylolisthesis with neurologic impairment, dislocation, and severe spondylolisthesis (Grades 3-4) at L5-S1.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KOSO382 FEB 2 6 2003 page of ## 510(k) Summarv IX. SUBMITTER: DePuy AcroMed, Inc. 325 Paramount Drive Raynham, MA 02780 CONTACT PERSON: Lisa A. Gilman DATE PREPARED: January 30, 2003 CLASSIFICATION NAME: Appliance, Fixation, Spinal Interlaminal Orthosis, Spinal Pedicle Fixation PROPRIETARY NAME: Moss Miami Spinal System Polyaxial Screws PREDICATE DEVICES: Moss Miami Spinal System (K953915, K982320, K982511, K982011, K983583, K992168, K011182) DEVICE DESCRIPTION: The Moss Miami Spinal System Polyaxial Screws and Polyaxial Reduction Screws are designed to accept a 5.5mm rod and are available in various sizes. > The Moss Miami Spinal System also contains Class 1 manual surqical instruments and cases that are considered exempt from premarket notification. INTENDED USE: When used as a posterior, noncervical hook, and/or sacral/iliac screw fixation system, or as an anterior, thoracic/lumbar screw fixation system, the Moss Miami Spinal System is intended to treat scoliosis, kyphosis and lordosis, fracture, loss of stability due to tumor, spinal stenosis, spondylolisthesis, a previously failed fusion surgery or degenerative disc disease (i.e., discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies). > When used as a pedicle screw fixation system of the noncervical spine in skeletally mature patients, the Moss Miami Spinal System is indicated for deqenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). {1}------------------------------------------------ Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three lines representing its wings. The eagle's head is formed by the silhouettes of three human faces in profile. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 FEB 2 6 2003 Mr. Frank Maas Regulatory Affairs Director DePuy AcroMed 325 Paramount Drive Raynham, Massachusetts 02767-0350 Re: K030383 Trade/Device Name: Moss Miami Spinal System - Polyaxial Screws Regulation Numbers: 21 CFR 888.3070 and 888.3050 Regulation Names: Pedicle screw spinal system and Spinal interlaminal fixation orthosis Regulatory Class: II Product Codes: MNH, MNI, and KWP Dated: January 31, 2003 Received: February 5, 2003 Dear Mr. Maas: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ Page 2 - Mr. Frank Maas This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Mark n Mellersm Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## III. Indications for Use 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Modified Moss Miami Spinal System Polvaxial Screws Device Name: Indications For Use: When used as a posterior, noncervical hook, and/or sacral/iliac screw fixation system, or as an anterior, thoracic/lumbar screw fixation system, the Moss Miami Spinal System is intended to treat scoliosis, kyphosis and lordosis, fracture, loss of stability due to tumor, spinal stenosis, spondylolisthesis, a previously failed fusion surgery or degenerative disc disease (i.e., discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies). When used as a pedicle screw fixation system of the noncervical spine in skeletally mature patients, the Moss Miami Spinal System is indicated for deqenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The Moss Miami Spinal System is also indicated for pedicle screw fixation in skeletally mature patients with severe spondylolisthesis (Grades 3 and 4) at the L5 -- S1 vertebral joint, having fusions with autogenous bone graft. with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3 - S1), and for whom the device system is intended to be removed after solid fusion is attained. R. Mark N. Milkerum Division Sign Off reral. Restorative ... ological Devices KO30383 510(i.) Namber - (Please do not write below this line - continue on another page if needed) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use: OR Over-The-Counter Use: (Per 21 CFR 801.109) DePuy AcroMed, Inc. Special 510K
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...