K173824 · Shoulder Innovations, Inc. · MBF · Oct 23, 2018 · Orthopedic
Device Facts
Record ID
K173824
Device Name
Humeral Short Stem System
Applicant
Shoulder Innovations, Inc.
Product Code
MBF · Orthopedic
Decision Date
Oct 23, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3670
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Shoulder Innovations Total Shoulder System is intended for use as an orthopedic implant for partial or total shoulder arthroplasty to treat the following: 1. significant disability in degenerative, rheumatoid, or traumatic disease of the glenohumeral joint; 2. avascular necrosis of the humeral head. The assembled humeral component may be used alone for hemiarthroplasty or combined with the glenoid component for total shoulder arthroplasty. The Total Shoulder System components are intended for single use only. The glenoid component is intended for cemented fixation only; the humeral stem may be implanted by press-fit or cement fixation.
Device Story
Humeral Short Stem System is a modular orthopedic implant for shoulder arthroplasty; used in clinical settings by orthopedic surgeons. System comprises collarless titanium alloy (Ti6-4) humeral stems with fins and cobalt-chrome (CoCr) humeral heads. Components connect via Morse-type taper. Stems feature porous coating for press-fit or cemented fixation; designed to prevent subsidence. Device replaces diseased or necrotic humeral head; can be used for hemiarthroplasty or total shoulder arthroplasty when combined with glenoid component. Provides structural support and articulation for shoulder joint; restores joint function and reduces patient disability.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated through non-clinical bench testing, including axial disassembly force testing per ASTM F2009 and fatigue testing of humeral components.
Technological Characteristics
Modular humeral prosthesis. Stems: Titanium Alloy (Ti6-4), collarless, porous-coated, Morse-type female taper. Heads: CoCr, Morse-type male taper. Fixation: Press-fit or cemented. Sterilization: Gamma irradiation. Dimensions: 70mm stem length, 132.5° neck angle, various head diameters/heights.
Indications for Use
Indicated for patients with significant disability due to degenerative, rheumatoid, or traumatic disease of the glenohumeral joint, or avascular necrosis of the humeral head requiring partial or total shoulder arthroplasty.
Regulatory Classification
Identification
A shoulder joint metal/polymer/metal nonconstrained or semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits movement in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral component made of alloys such as cobalt-chromium-molybdenum (Co-Cr-Mo) and titanium-aluminum-vanadium (Ti-6Al-4V) alloys, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, or a combination of an articulating ultra-high molecular weight bearing surface fixed in a metal shell made of alloys such as Co-Cr-Mo and Ti-6Al-4V. The humeral component and glenoid backing have a porous coating made of, in the case of Co-Cr-Mo components, beads of the same alloy or commercially pure titanium powder, and in the case of Ti-6Al-4V components, beads or fibers of commercially pure titanium or Ti-6Al-4V alloy, or commercially pure titanium powder. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance: Shoulder Joint Metal/Polymer/Metal Nonconstrained or Semi-Constrained Porous-Coated Uncemented Prosthesis.”
K241817 — InSet Total Shoulder System · Shoulder Innovations, Inc. · Feb 11, 2025
K193038 — Comprehensive Shoulder System · Biomet Manufacturing Corp · Oct 28, 2020
K102670 — TOTAL SHOULDER SYSTEM · Shoulder Innovations, LLC · Jan 24, 2011
K001706 — WMT MODULAR SHOULDER SYSTEM · Wrightmedicaltechnologyinc · Aug 16, 2000
K210899 — LINK Embrace Shoulder System- Anatomical Configuration · Waldemar Link GmbH & Co. KG · Jun 21, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.
Shoulder Innovations Inc Don Running VP R&D 13827 Port Sheldon Street Holland, Michigan 49424
October 23, 2018
Re: K173824
Trade/Device Name: Humeral Short Stem System Regulation Number: 21 CFR 888.3670 Regulation Name: Shoulder joint metal/polymer/metal nonconstrained or semi-constrained porouscoated uncemented prosthesis Regulatory Class: Class II Product Code: MBF, KWT, KWS, HSD Dated: September 12, 2018 Received: September 14, 2018
Dear Don Running:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Daniel S. Ramsey -S 2018.10.23 16:48:03 -04'00'
For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173824
Device Name Humeral Short Stem System
Indications for Use (Describe)
The Shoulder Innovations Total Shoulder System is intended for use as an orthopedic implant for partial or total shoulder arthroplasty to treat the following:
1. significant disability in degenerative, rheumatoid, or traumatic disease of the glenohumeral joint;
2. avascular necrosis of the humeral head.
The assembled humeral component may be used alone for hemiarthroplasty or combined with the glenoid component for total shoulder arthroplasty.
The Total Shoulder System components are intended for single use only. The glenoid component is intended for cemented fixation only; the humeral stem may be implanted by press-fit or cement fixation.
Type of Use (*Select one or both, as applicable*)
| <span style="font-size:12px"><b>☑</b> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------------------------------------|
| <span style="font-size:12px">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary Pursuant to 21 CFR 807.92c
| Date Prepared: | September 12, 2018 |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted By: | Shoulder Innovations, LLC<br>13827 Port Sheldon St.<br>Holland, MI 49424 |
| Contact: | Don Running<br>VP-R&D<br>574-253-1133<br>don@genesisinnovationgroup.com |
| Proprietary Name: | Humeral Short Stem System |
| Common Name: | Shoulder Prosthesis |
| Classification: | 21 CFR Section 888.3670 – Shoulder joint metal/<br>polymer/metal nonconstrained or semi-constrained<br>porous-coated uncemented prosthesis<br>Product Code: MBF<br>21 CFR Section 888.3650 – Shoulder joint<br>metal/polymer non-constrained cemented<br>prosthesis.<br>Product Code: KWT<br>21 CFR Section 888.3660 – Shoulder joint<br>metal/polymer semi-constrained cemented<br>prosthesis.<br>Product Code: KWS<br>21 CFR Section 888.3690 – Shoulder joint humeral<br>(hemi-shoulder) metallic uncemented prosthesis.<br>Product Code: HSD |
| Substantially<br>Equivalent Device: | K060692, Biomet Comprehensive® Primary Shoulder<br>Stems |
{4}------------------------------------------------
### Device Description:
The Shoulder Innovations Humeral Short Stem consists of modular humeral stems and heads and articulates with the Shoulder Innovations glenoid component (K111596). The humeral stems are collarless and manufactured from Titanium Alloy (Ti6-4) with fins to provide rotational stability. The collarless stems allow the humeral head to prevent stem subsidence. The stems have a female Morse-type taper to interface with the modular humeral heads. The proximal body and fins are coated with a rough, porous coating for un-cemented fixation or for use with bone cement.
The humeral heads are manufactured from CoCr and are available in standard and offset configurations. The heads have a male Morse-type taper to interface with the humeral stems.
#### Intended Use / Indications:
The Shoulder Innovations Total Shoulder System is intended for use as an orthopedic implant for partial or total shoulder arthroplasty to treat the following:
- 1. significant disability in degenerative, rheumatoid, or traumatic disease of the glenohumeral joint;
- 2. avascular necrosis of the humeral head.
The assembled humeral component may be used alone for hemiarthroplasty or combined with the glenoid component for total shoulder arthroplasty.
The Total Shoulder System components are intended for single use only. The glenoid component is intended for cemented fixation only; the humeral stem may be implanted by press-fit or cement fixation.
#### Predicate Device Comparison:
# Humeral Stem Component
| Feature | Biomet<br>Comprehensive<br>Primary Shoulder<br>Stems (K060692) | Proposed Device<br>Shoulder<br>Innovations<br>Humeral Short<br>Stem System | SE? |
|--------------------------|----------------------------------------------------------------|----------------------------------------------------------------------------|-----|
| Distal Stem<br>Diameters | 5.75 mm<br>(Labeled 4mm) | 5.75 mm | Yes |
| Stem<br>Lengths | 72 mm | 70 mm | Yes |
| Stem Geometry | Tapered circular | Tapered circular | Yes |
| Collar | Collarless | Collarless | Yes |
{5}------------------------------------------------
| Feature | Biomet<br>Comprehensive<br>Primary Shoulder<br>Stems (K060692) | Proposed Device<br>Shoulder<br>Innovations<br>Humeral Short<br>Stem System | SE? |
|------------------|----------------------------------------------------------------|----------------------------------------------------------------------------|-----|
| Modular system | Female Morse-type<br>taper | Female Morse-type<br>taper | Yes |
| Bone Integration | Porous Coating | Porous Coating | Yes |
| Neck Angle | 135° | 132.5° | Yes |
| Fixation | Press-fit or<br>Cemented | Press-fit or<br>Cemented | Yes |
| Material | Titanium Alloy | Titanium Alloy | Yes |
| Sterility | Gamma Irradiated | Gamma Irradiated | Yes |
# Humeral Head Component
| Feature | Biomet<br>Comprehensive<br>Primary Shoulder<br>Stems (K060692) | Proposed Device<br>Shoulder<br>Innovations<br>Humeral<br>Short Stem<br>System | SE? |
|-------------------|----------------------------------------------------------------|-------------------------------------------------------------------------------|-----|
| Head<br>Diameters | 38, 42, 46, 50, 54, 58mm | 40mm, 44mm, 48mm, 52mm, and 56mm | Yes |
| Head Heights | 18mm - 27mm | 15mm - 24mm | Yes |
| Configurations | Standard and Offset | Standard and Offset | Yes |
| Geometry | Solid semi-sphere | Solid semi-sphere | Yes |
| Modular system | Male Morse-type<br>taper | Male Morse-type<br>taper | Yes |
| Material | CoCr | CoCr | Yes |
| Sterility | Gamma Irradiated | Gamma Irradiated | Yes |
# Non-Clinical Testing:
Performance testing was performed on the Shoulder Innovations Humeral Stem System. The modular humeral components were tested per ASTM F2009 Standard Test Method for Determining the Axial Disassembly Force of Taper Connections of Modular Prostheses. Fatigue testing was also performed on the humeral components.
#### Clinical Testing:
Clinical testing was not necessary to demonstrate substantial equivalence of the Shoulder Innovations Humeral Short Stem System to the predicate device.
{6}------------------------------------------------
# Summary of Substantial Equivalence:
The results of non-clinical testing and comparative analysis demonstrate that the design, function, intended use, and indications for use of the Shoulder Innovations Humeral Short Stem System is substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.