The WMT Modular Shoulder System, humeral head and stem, is intended to be used for total or hemi-arthroplasty. When used for total shoulder arthroplasty, the subject components are to be used with shoulder artill oplaar glenoid components cleared in K920362 or with 3M Neer IITM Shoulder (K895226, SE 11/3/89) compatible glenoid components.
Device Story
Modular shoulder prosthesis system consisting of humeral heads and stems; humeral stems feature proximal flanges (lateral, anterior/posterior, medial) with suture holes for soft tissue attachment; humeral heads available in five sizes. Device implanted via press-fit or cement fixation. Used in total or hemi-arthroplasty procedures to replace damaged humeral joint surfaces. Surgeon-operated in clinical/OR setting. Provides structural replacement for diseased or fractured humeral head; restores joint function; reduces pain. Total shoulder arthroplasty requires compatible glenoid components.
Clinical Evidence
Bench testing only. Mechanical test data demonstrated that the material and subject device design meet the strength requirements of the predicate device.
Technological Characteristics
Humeral stems: titanium alloy (ASTM F 136). Humeral heads: cobalt chrome (ASTM F 1537 or ASTM F 799). Modular design with five head sizes and two stem lengths. Fixation: press-fit or cement. No software or electronic components.
Indications for Use
Indicated for patients with non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, acute humeral head fractures, or cuff tear arthroplasty. Hemi-shoulder replacement indicated for ununited humeral head fractures and avascular necrosis of the humeral head. Revision indicated for failed prior treatments.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
{0}------------------------------------------------
AUG 1 6 2000
Image /page/0/Picture/2 description: The image shows the logo for Wright Medical Technology, Inc. The logo features a stylized "W" on the left, followed by the company name in a serif font. Below the company name is the address "5677 Airline Road, Arlington, TN 38002" and the phone number "901-867-9971".
### 510K Summary
Company: Wright Medical Technology, Inc. 5677 Airline Road Arlington, TN 38002 Contact Person: Lynne Witkowski Date: June 2, 2000 Trade Name: WMT Modular Shoulder System Common Name: Shoulder prosthesis Predicate Device: 3M Modular Shoulder System
### Description/Intended Use:
The WMT Modular Shoulder System includes humeral heads in five sizes. A regular length and long length humeral stem are available in identical diameters. Both humeral stems have four proximal flanges: one lateral, two anterior/posterior and one medial. The lateral flange has three suture holes and the anterior/posterior fins have one suture hole each. Both the regular and long stems are manufactured from titanium alloy (ASTM F 136). The heads are manufactured from cobalt chrome (ASTM F 1537, warm worked or ASTM F 799).
### Intended Use:
The WMT Modular Shoulder System, humeral head and stem, is intended for use as a total or hemi-arthroplasty. When used for total shoulder arthroplasty, the subject components are recommended for use with 3M Neer IITM Shoulder (K895226, SE 11/3/89) compatible glenoid components.
Indications for Use:
- Non-inflammatory degenerative joint disease including osteoarthritis and . avascular necrosis;
- . Rheumatoid arthritis;
- Revision where other devices or treatments have failed; .
- Correction of functional deformity; .
- Treatment of acute fracture of the humeral head unmanageable using . other treatment methods; and
- . Cuff tear arthroplasty.
Hemi-shoulder replacement is also indicated for:
- Ununited humeral head fractures; and .
- Avascular necrosis of the humeral head. .
The humeral stem may be implanted by press-fit or cement fixation.
### Testing Summary:
The WMT Modular Shoulder System was declared substantially equivalent to the predicate device. Mechanical test data demonstrated that the material and subject device design meet the strength requirements of the predicate device.
000069
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines representing wings or feathers.
### Public Health Service
# AUG 1 6 2000
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Lynne Witkowski Regulatory Affairs Associate Wright Medical Technology, Inc. 5677 Airline Road Arlington, Tennessee 38002
Re: K001706
Trade Name: WMT Modular Shoulder System Regulatory Class: II Product Codes: KWS and HSD Dated: May 31, 2000 Received: June 5, 2000
Dear Ms. Witkowski:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 0976, the enactment date of the Medical Device Amendments, or to devices that have been reclassify, in i accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general control provisions of the Act. The general control provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and of doriotics,
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current on one Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{2}------------------------------------------------
# Page 2 - Ms. Lynne Witkowski
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire specific davice for your do ic devices), please contact the Office of Compliance at additionally 607.10 for in This allestions on the promotion and advertising of your device, (201) 594-4637. Rudhionary, for questions at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Information on your responsionales and in the more (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
e R. lochner
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use Statement
KOOITOG 510(k) Number (if known)
# WMT Modular Shoulder System
Device Name
Indications for Use
## Intended Use
The WMT Modular Shoulder System, humeral head and stem, is intended to be used for total or hemi-arthroplasty. When used for total shoulder arthroplasty, the subject components are to be used with shoulder artill oplaar glenoid components cleared in K920362 or with 3M Neer IITM Shoulder (K895226, SE 11/3/89) compatible glenoid components.
# Indications for Use:
- Non-inflammatory degenerative joint disease including . osteoarthritis and avascular necrosis;
- Rheumatoid arthritis; .
- Revision where other devices or treatments have failed; .
- Correction of functional deformity; .
- Treatment of acute fracture of the humeral-head unmanageable . using other treatment methods; and
- Cuff tear arthroplasty. .
Hemi-shoulder replacement is also indicated for:
- Ununited humeral head fractures; and .
- Avascular necrosis of the humeral head. .
The humeral stem may be implanted by press-fit or cement fixation.
# PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| | <div style="border-bottom: 1px solid black;">Dune R. Lochnes.</div> |
| | (Division Sign-Off) |
| | Division of General Restorative Devices |
| | 510(k) Number <div style="border-bottom: 1px solid black;">K001706</div> |
| Prescription Use<br>(per 21 CFR 801.109) <div style="display:inline-block;"><b>✓</b></div> | OR Over-The Counter Use <div style="border-bottom: 1px solid black;"></div> |
000018
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.