Meridian Interbody System; WaveForm A Interbody System

K233694 · SeaSpine Orthopedics Corporation · MAX · Jan 12, 2024 · Orthopedic

Device Facts

Record IDK233694
Device NameMeridian Interbody System; WaveForm A Interbody System
ApplicantSeaSpine Orthopedics Corporation
Product CodeMAX · Orthopedic
Decision DateJan 12, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Meridian System with NanoMetalene® surface technology Interbody, when used with or without a Spin Plate, is indicated for use as an adjunct to fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients should be skeletally mature and have had at least six (6) months of non-operative treatment. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The Meridian Interbody is intended for use with supplemental fixation. The Meridian System 2-Hole and 3-Hole No-profile Interbody, when used with Screws and with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed cancellous, cortical, and/or corticocancellous bone. The Meridian System 2-Hole and 3-Hole No-profile Interbody must be used with the maximum number of screws allowed. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation. The Meridian System 4-Hole No-profile Interbody, when used with Screws and with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The Meridian System 4-Hole No-profile Interbody may be used as a standalone device only when at least two Screws are inserted with one inferior and one superior screw trajectory. If the surgeon chooses to use the 4-Hole No-profile Interbody with fewer than two Screws inserted with one inferior and one superior screw trajectory, then additional supplemental fixation system must be used. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation. The Meridian System No-profile Interbody, when used with Inline Fixation Anchors and a No-profile Locking Cover, is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The Meridian System No-profile Interbody with Inline Fixation Anchors is intended for use with supplemental fixation. The Meridian System TruProfile Interbody assembled with the Anterior Plate, when used with Screws, an Anterior Plate Locking Cover, and with or without a Spin Plate, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation. The WaveForm A System Interbody, when used with or without a Spin Plate, is indicated for use as an adjunct to fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients should be skeletally mature and have had at least six (6) months of non-operative treatment. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The WaveForm A System Interbody is intended for use with supplemental fixation. The WaveForm A System 2-Hole and 3-Hole No-profile Interbody, when used with Screws and with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed cancellous, cortical, and/or corticocancellous bone. The WaveForm A System 2-Hole and 3-Hole No-profile Interbody must be used with the maximum number of screws allowed. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation. The WaveForm A System 4-Hole No-profile Interbody, when used with Screws and with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The WaveForm A System 4-Hole No-profile Interbody may be used as a standalone device only when at least two Screws are inserted with one inferior and one superior screw trajectory. If the surgeon chooses to use the 4-Hole No-profile Interbody with fewer than two Screws inserted with one inferior and one superior screw trajectory, then additional supplemental fixation system must be used. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation. The WaveForm A System No-profile Interbody, when used with Inline Fixation Anchors and with a No Profile Locking Cover, is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The WaveForm A System No-profile Interbody with Inline Fixation Anchors is intended for use with supplemental fixation. The WaveForm A System TruProfile Interbody assembled with the Anterior Plate, when used with Screws, an Anterior Plate Locking Cover, and with or without a Spin Plate, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation.

Device Story

Intervertebral body fusion system (Meridian and WaveForm A) for spinal fusion; Meridian uses PEEK with NanoMetalene (titanium) surface; WaveForm A uses additively manufactured titanium alloy; both include central graft window for bone graft; configurations include standalone (with screws/anchors) or adjunct to supplemental fixation; used in surgical setting by orthopedic/neurosurgeons; implants provide structural support to disc space to facilitate fusion; benefits include stabilization of spinal segment and relief of discogenic pain.

Clinical Evidence

No clinical data; substantial equivalence based on non-clinical mechanical testing.

Technological Characteristics

Meridian: PEEK (ASTM F2026) with tantalum (ASTM F560) and Ti-6Al-4V ELI (ASTM F136) markers; NanoMetalene surface (Ti, ASTM F67). WaveForm A: Additively manufactured Ti-6Al-4V ELI (ASTM F3001); optional NaOH surface treatment. Both: Gamma sterilized implants; non-sterile titanium alloy fixation components; steam sterilizable instruments.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1; DDD defined as discogenic back pain with degeneration confirmed by history/radiography; may include up to Grade 1 spondylolisthesis; requires at least 6 months of failed non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ January 12, 2024 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym with the full name of the agency on the right. The seal features a stylized depiction of human figures, while the FDA part includes a blue square with "FDA" in white and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. SeaSpine Orthopedics Corporation Jesse Albright Manager, Regulatory Affairs 5770 Armada Drive Carlsbad, California 92008 # Re: K233694 Trade/Device Name: Meridian Interbody System: WaveForm A Interbody System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX, OVD Dated: November 16, 2023 Received: November 17, 2023 # Dear Jesse Albright: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Katherine D. Kavlock -S for Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K233694 Device Name Meridian Interbody System WaveForm A Interbody System #### Indications for Use (Describe) Meridian Interbody System Interbody Device (IBD) Implants (i.e., interbody implants used alone): The Meridian System with NanoMetalene® surface technology when used with or without a Spin Plate, is indicated for use as an adjunct to fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenc origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients should be skeletally mature and have had at least six (6) months of non-operative treatment. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The Meridian Interbody is intended for use with supplemental fixation. #### No-Profile Interbody Implants with Screws: The Meridian System 2-Hole No-profile Interbody, when used with Screws and with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenc origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed cancellous, cortical, and/or corticocancellous bone. The Meridian System 2-Hole No-profile Interbody must be used with the maximum number of screws allowed. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation. The Meridian System 4-Hole No-profile Interbody, when used with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenc origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The Meridian System 4-Hole No-profile Interbody may be used as a standalone device only when at least two Screws are inserted with one inferior and one superior screw trajectory. If the surgeon chooses to use the 4-Hole No-profile Interbody with fewer than two Screws inserted with one inferior and one superior screw trajectory, then additional supplemental fixation system must be used. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation. No-Profile Interbody Implants with Inline Fixation Anchors: The Meridian System No-profile Interbody, when used with Inline Fixation Anchors and a No-profile Locking Cover, is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogenee bone graft composed of cancellous, cortical, and/or corticocancellous bone. The Meridian System No-profile Interbody with Inline Fixation Anchors is intended for use with supplemental fixation. {3}------------------------------------------------ # TruProfile Interbody Implants: The Meridian System TruProfile Interbody assembled with the Anterior Plate, when used with Screws, an Anterior Plate Locking Cover, and with or without a Spin Plate, is a standalone indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation. #### WaveForm A Interbody System Interbody Device (IBD) Implants (i.e., interbody implants used alone): The WaveForm A System Interbody, when used with or without a Spin Plate, is indicated for use as an adjunct to fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients should be skeletally mature and have had at least six (6) months of non-operative treatment. The interded to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The WaveForm A System Interbody is intended for use with supplemental fixation. #### No-profile Interbody Implants with Screws: The WaveForm A System 2-Hole No-profile Interbody, when used with Screws and with or without a Noprofile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiosraphic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed cancellous, cortical, and/or corticocancellous bone. The WaveForm A System 2-Hole No-profile Interbody must be used with the maximum number of screws allowed. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation. The WaveForm A System 4-Hole No-profile Interbody, when used with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenc origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The WaveForm A System 4-Hole No-profile Interbody may be used as a standalone device only when at least two Screws are inserted with one inferior and one superior screw trajectory. If the surgeon chooses to use the 4-Hole No-profile Interbody with fewer than two Screws inserted with one inferior and one superior screw trajectory, then additional supplemental fixation system must be used. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation. # No-profile Interbody Implants with Inline Fixation Anchors: The WaveForm A System No-profile Interbody, when used with Inline Fixation Anchors and with a No Profile Locking Cover, is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The WaveForm A System No-profile Interbody with Inline Fixation Anchors is intended for use with supplemental fixation. TruProfile Interbody Implants: {4}------------------------------------------------ The WaveForm A System TruProfile Interbody assembled with the Anterior Plate, when used with Screws, an Anterior Plate Locking Cover, and with or without a Spin Plate, is a standalone indicated for use as adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/ or corticocancellous bone. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation. #### Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ # 510(k) Summary | Contact Details | | |-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------| | Applicant Name: | SeaSpine Orthopedics Corporation | | Address: | 5770 Armada Drive, Carlsbad CA | | Phone number: | (760) 216-5176 | | Fax number: | (760) 683-6874 | | Contact person: | Jesse Albright, Manager, Regulatory Affairs | | Date Prepared: | January 9, 2024 | | Device Name | | | Trade Name(s): | Meridian Interbody System<br>WaveForm A Interbody System | | Common Name(s): | Intervertebral Body Fusion Device | | Classification Name(s): | Intervertebral Fusion Device with Bone Graft, Lumbar (21 CFR<br>888.3080), Intervertebral Fusion Device With Integrated Fixation,<br>Lumbar (21 CFR 888.3080) | | Class: | 2 | | Product Code(s): | MAX, OVD | # Legally Marketed Predicate Devices | 510(k) Number | Product Code | Trade Name | Manufacturer | |--------------------------------|--------------|----------------------------------------------|-------------------------------------| | Primary Predicate Device | | | | | K220711 | MAX, OVD | Meridian Interbody System | SeaSpine Orthopedics<br>Corporation | | Additional Predicate Device(s) | | | | | K222732 | MAX, OVD | WaveForm A Interbody System | SeaSpine Orthopedics<br>Corporation | | K121211 | OVD | Vu aPOD Intervertebral Body<br>Fusion Device | SeaSpine Orthopedics<br>Corporation | | K201671 | OVD | A-Link Z | Acuity Surgical Devices, LLC | {6}------------------------------------------------ #### Device Description ### Meridian Interbody System The Meridian Interbody System featuring NanoMetalene® surface technology consists of anterior intervertebral body fusion devices manufactured from polyetheretherketone (PEEK) (per ASTM F2026) with tantalum (per ASTM F560) and titanium alloy (Ti-6Al-4V ELI per ASTM F136) markers for radiographic visualization and NanoMetalene, which is a one-micron thick surface layer of commercially pure titanium (per ASTM F67). NanoMetalene surface technology provides a microscopic roughened surface with nano-scale features. The interbody spacers are available in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy and include a central graft window for receiving autogenous and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The spacers are individually packaged and gamma sterilized. The Meridian Interbody System spacers can be used alone with supplemental fixation or in combination with fixation implants to create the TruProfile and No-profile Interbody configurations. Fixation implants include Anterior Plates, Screws, Inline Fixation Anchors, Anterior Plate Locking Covers, and No-profile Locking Covers, all of which are manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136) and provided non-sterile. The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components. # WaveForm A Interbody System The WaveForm A Interbody System consists of anterior intervertebral body fusion devices additively manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F3001). The interbody spacers are offered in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy as well as with and without a sodium hydroxide (NaOH) surface treatment that provides a microscopic, roughened surface with nanoscale features. The interbody spacers, which include a central graft window for receiving autogenous and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone, are individually packaged and gamma sterilized. The WaveForm A Interbody System spacers can be used alone with supplemental fixation or in combination with fixation implants to create the TruProfile and No-profile Interbody configurations. Fixation implants include Anterior Plates, Screws, Inline Fixation Anchors, Anterior Plate Locking Covers, and No-profile Locking Covers, all of which are manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136) and provided non-sterile. The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that {7}------------------------------------------------ may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components. #### Intended Use/Indications for Use #### Meridian Interbody System #### Interbody Device (IBD) Implants (i.e., interbody implants used alone): The Meridian System with NanoMetalene® surface technology Interbody, when used with or without a Spin Plate, is indicated for use as an adjunct to fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients should be skeletally mature and have had at least six (6) months of non-operative treatment. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The Meridian Interbody is intended for use with supplemental fixation. #### No-Profile Interbody Implants with Screws: The Meridian System 2-Hole No-profile Interbody, when used with Screws and with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed cancellous, cortical, and/or corticocancellous bone. The Meridian System 2-Hole and 3-Hole No-profile Interbody must be used with the maximum number of screws allowed. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation. The Meridian System 4-Hole No-profile Interbody, when used with Screws and with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The Meridian System 4-Hole No-profile Interbody may be used as a standalone device only when at least two Screws are inserted with one inferior and one superior screw trajectory. If the surgeon chooses to use the 4-Hole No-profile Interbody with fewer than two Screws inserted with one inferior and one superior screw trajectory, then additional supplemental fixation system must be used. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation. {8}------------------------------------------------ #### No-Profile Interbody Implants with Inline Fixation Anchors: The Meridian System No-profile Interbody, when used with Inline Fixation Anchors and a Noprofile Locking Cover, is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of nonoperative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The Meridian System No-profile Interbody with Inline Fixation Anchors is intended for use with supplemental fixation. #### TruProfile Interbody Implants: The Meridian System TruProfile Interbody assembled with the Anterior Plate, when used with Screws, an Anterior Plate Locking Cover, and with or without a Spin Plate, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of nonoperative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation. # WaveForm A Interbody System #### Interbody Device (IBD) Implants (i.e., interbody implants used alone): The WaveForm A System Interbody, when used with or without a Spin Plate, is indicated for use as an adjunct to fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients should be skeletally mature and have had at least six (6) months of non-operative treatment. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The WaveForm A System Interbody is intended for use with supplemental fixation. #### No-profile Interbody Implants with Screws: The WaveForm A System 2-Hole No-profile Interbody, when used with Screws and with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may {9}------------------------------------------------ also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed cancellous, cortical, and/or corticocancellous bone. The WaveForm A System 2-Hole and 3-Hole No-profile Interbody must be used with the maximum number of screws allowed. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation. The WaveForm A System 4-Hole No-profile Interbody, when used with Screws and with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The WaveForm A System 4-Hole No-profile Interbody may be used as a standalone device only when at least two Screws are inserted with one inferior and one superior screw trajectory. If the surgeon chooses to use the 4-Hole No-profile Interbody with fewer than two Screws inserted with one inferior and one superior screw trajectory, then additional supplemental fixation system must be used. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation. ## No-profile Interbody Implants with Inline Fixation Anchors: The WaveForm A System No-profile Interbody, when used with Inline Fixation Anchors and with a No Profile Locking Cover, is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of nonoperative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The WaveForm A System No-profile Interbody with Inline Fixation Anchors is intended for use with supplemental fixation. #### TruProfile Interbody Implants: The WaveForm A System TruProfile Interbody assembled with the Anterior Plate, when used with Screws, an Anterior Plate Locking Cover, and with or without a Spin Plate, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of nonoperative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, {10}------------------------------------------------ and/or corticocancellous bone. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation. # Summary of Technological Characteristics The Meridian Interbody System and WaveForm A Interbody System are identical or similar to the cited predicate devices in regard to components, device description, intended use/indications for use, technological characteristics (e.g., operating principle, design, materials, manufacturing, etc.) and performance (mechanical safety). # Non-Clinical Testing The Meridian Interbody System and WaveForm A Interbody System demonstrated substantially equivalent mechanical performance to the predicate systems through static and dynamic axial compression and compression shear (ASTM F2077), subsidence (ASTM F2267), wear evaluation (ASTM F1877), screw pushout, and expulsion. # Clinical Testing Not applicable. The determination of substantial equivalence is not based on an assessment of clinical performance data. # Conclusions The submitted data demonstrates that the Meridian Interbody System and WaveForm A Interbody System are substantially equivalent to the cited legally marketed predicates.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...