SeaSpine Spacer System NM (Hollywood, Hollywood VI, Pacifica, Redondo, Ventura), Vu a•POD-L NanoMetalene, SeaSpine Vu e•POD System, SeaSpine Vu a•POD Prime NanoMetalene Intervertebral, SeaSpine Shoreline ACS - Anterior Cervical Standalone, SeaSpine Cervical Interbody RT System, SeaSpine Cambria System, SeaSpine Regatta Lateral System, SeaSpine Reef TO/TA System, SeaSpine Reef TH System, SeaSpine Meridian System
Applicant
SeaSpine Orthopedics Corporation
Product Code
MAX · Orthopedic
Decision Date
Jul 7, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When used as an intervertebral body fusion device, the SeaSpine Spacer System with NanoMetalene® surface technology is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had six months of non-operative treatment. The device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone and supplemental fixation.
Device Story
Single-use intervertebral fusion devices; PEEK core with NanoMetalene surface technology; tantalum or Ti-6Al-4V ELI markers for radiographic visualization. Devices implanted via anterior or lateral surgical approaches; packed with autogenous or allogeneic bone graft to promote fusion. Used in lumbar or cervical spine to restore disc height and biomechanical integrity; often used with supplemental fixation (pedicle screws, anterior plates). Operated by orthopedic or neurosurgeons in hospital/OR settings. Output is structural support for spinal column; aids in achieving arthrodesis; benefits patients by reducing pain associated with discogenic degeneration.
Clinical Evidence
Bench testing only. Mechanical performance evaluated per ASTM F1978, F1147, F1160, and F1044. Biocompatibility and sterility verified as equivalent to predicate devices.
Technological Characteristics
Materials: PEEK (ASTM F2026), Tantalum (ASTM F560), or Ti-6Al-4V ELI (ASTM F136). Design: Intervertebral spacer with central canal for bone graft. Sterilization: Provided sterile. Connectivity: None (mechanical implant).
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) requiring intervertebral body fusion at one or two contiguous levels (cervical C2-T1 or lumbar L2-S1). DDD defined as discogenic pain with degeneration confirmed by history and radiography. Patients may have up to Grade 1 spondylolisthesis or retrolisthesis. Requires 6 weeks (cervical) or 6 months (lumbar) of non-operative treatment. Contraindications include skeletal immaturity.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
July 7, 2021
SeaSpine Orthopedics Corporation Aly Alvarez Assoc. Manager, Regulatory Affairs 5770 Armada Drive Carlsbad, California 92008
# Re: K210497
Trade/Device Name: SeaSpine Spacer System NM (Hollywood, Hollywood VI, Pacifica, Redondo, Ventura), Vu a•POD-L NanoMetalene, SeaSpine Vu e•POD System, SeaSpine Vu aºPOD Prime NanoMetalene Intervertebral, SeaSpine Shoreline ACS -Anterior Cervical Standalone, SeaSpine Cervical Interbody RT System, SeaSpine Cambria System, SeaSpine Regatta Lateral System, SeaSpine Reef TO/TA System, SeaSpine Reef TH System, SeaSpine Meridian System
Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX, ODP, OVD, OVE, MOP Dated: June 4, 2021 Received: June 7, 2021
Dear Aly Alvarez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
# K210497
#### Device Name
Device Name
SeaSpine Spacer System (Hollywood NanoMetalene, Hollywood VI NanoMetalene, Pacifica NanoMetalene)
#### Indications for Use (Describe)
Indications for Use (Describe)
When used as an intervertebral body fusion device, the SeaSpine Spacer System with NanoMetalene® surface technology with ised as an intel voltonar oody fusion as or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or committed by instoly and radiographies. DDB paxients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had six months of non-operative treatment. The device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone and supplemental fixation.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{3}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K210497
#### Device Name SeaSpine Cambria NanoMetalene System
#### Indications for Use (Describe)
Indications for Use (Describe)
The SeaSpine Cambria NanoMetalene® Surface technology is intended to be used as an I he Seaspille Calliona Nanometate Dystem vo contiguous levels (C3-C7) in skeletally mature patients witho degenerative disc disease (defined as neck pain with discogenio origin with degeneration of history of new operative treatment degenerative discusse (demical spine. Patients should have received at least six weeks of non-operative theatment and radiographic studies) of the cervical sphie. I talended to be implanted via an open, and wed with prior to treatment with the device. Devices are intensed of cancellous, cortical, and/or corticocancellous bone and supplemental fixation, such as an anterior plating system.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the i ne burden time for this ocloolich evisting data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{4}------------------------------------------------
510(k) Number (if known) K210497
#### Device Name
SeaSpine Vu aºPOD-L NanoMetalene Intervertebral Body Fusion Device
#### Indications for Use (Describe)
Indications for Use (Describe)
The SeaSpine Vu aPOD-L NanoMetalene Intervertebral Body Fusion Device with NanoMetalene® surface technology is indicated for use as an adjunct to fusion in patients with degenerative disc disease (DDD) at one or two learly of the device is from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The device is to be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The SeaSpine Vu a POD-L NanoMetalene Intervertebral Body Fusion Device is intended for use with supplemental fixation that is in addition to the integrated buttress spin plate, such as a pedicle screw system or anterior plate. Degenerative disc disease is defined as back pain of discogenic origin with degeneration of the disc allerior plate. Degenerative discuse is assimes should be skeletally mature and have had at least six (6) months of non-operative treatment.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{5}------------------------------------------------
510(k) Number (if known) K210497
Device Name SeaSpine Vu e.POD System
#### Indications for Use (Describe)
Indications for Use (Descine)
When used as an intervertebral body fusion device, the SeaSpine Vu ePOD System with NanoMetalene® surface when used as an mercented or use as adjunct to fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The device is indicated for use with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The SeaSpine Vu e+POD System is intended for use with supplemental fixation. Degenerative disc disease (DDD) is defined as back pain of discogenic origin with degeneration of the disc confirmed by Degencialive dise and issues ( 1) hese patients should be skeletally mature and have had six months of non-operative treatment.
When used as a vertebral body replacement (VBR), the SeaSpine Vu e-POD System is indicated for use in the thoracolumbar spine (TI-LS) to replace a collapsed, damaged, or otherwise unstable vertebral body due to tumor or trauma the arracture). The SeaSpine Vu e-POD System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of fusion for a prolonged period. The device is indicated for use with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The SeaSpine Vu e·POD System is intended for use with supplemental internal spinal fixation.
| Type of Use (Select one or both, as applicable) | | |
|-----------------------------------------------------------------------------------|---------------------------------------------|--|
| > Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. | | |
| *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | |
| rend New Your Comments of Children Comments of Conce | | |
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{6}------------------------------------------------
510(k) Number (if known)
K210497
Device Name SeaSpine Regatta Lateral System
#### Indications for Use (Describe)
Interbody Device (IBD) Implants (i.e., interbody implants used alone):
The SeaSpine Regatta Lateral System with NanoMetalene surface technology is indicated for use as an adjunct to fusion I he Seasonic Regaria Lateral System Mith Autoise disease (DDD, defined as back pain of discogenic origin, with in skelelally mature patients with degencially and radiographic studies). It is intended for use at either one level of the degeneration of this commined by first. The take the treatment of DDD with up to Grade 1 spondylolisthesis at the involved levels in the interior of the interbody spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device.
The SeaSpine Regatta Lateral System is intended for use with supplemental fixation.
#### TruProfile Interbody Implants:
The SeaSpine Regatta Lateral System assembled with the TruProfile Lateral Plate, when used with Screws, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with (DDD) at the of two volumed by history and radiographic studies. The DDD patients may also have up to Grade 1 degencialism of the involved level(s). The interior of the spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device.
The SeaSpine Regatta Lateral System assembled with the 1-hole TruProfile Lateral Plate, when used with Screws, is intended for use with supplemental fixation.
| Type of Use (Select one or both, as applicable) | <span> <span style="padding-right: 20px;"> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{7}------------------------------------------------
510(k) Number (if known) K210497
#### Device Name
SeaSpine Vu a-POD Prime NanoMetalene Intervertebral Body Fusion Device
#### Indications for Use (Describe)
Indications for Use (Describe)
When used with the bone screws, the SeaSpine Vu aªPOD Prime NanoMetalene Intervertebral Body Fusion Device with When used with the bone screws, the Soubpilo for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the interbody spacer component may be allot all to crade i spond from and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone.
When used with the SpinPlate, the SeaSpine Vu a+POD Prime NanoMetalene Intervertebral Body Fusion Device is with used will his opin into, the evalue in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the interbody spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. When used with the SpinPlate, the SeaSpine Vu a POD Prime NanoMetalene Intervertebral Body Fusion Device is intended for use with supplemental fixation.
The SeaSpine Vu aºPOD Prime NanoMetalene Intervertebral Body Fusion used with the bone screws or the bone screws and the SpinPlate, is a stand-alone device. If the SeaSpine Vu a+POD Prime NanoMetalene Intervertebral Body Fusion Device is used only with the SpinPlate, then additional supplemental fixation, which has been cleared by the FDA for use in the lumbar spine, must be used to augment stability. Additionally, implants with hyperlordotic angles of >20° must also be used with additional supplemental fixation (e.g., posterior pedicle screw and rod systems). This device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone.
Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
{8}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the I he burden time for this collection of information to estination the data needed and complete time to review instructions, search existing data sources, gather and his burden estimate or any other aspect and review the collection of information. Ochastions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of of conduct or sponsor, anays a currently valid OMB number."
{9}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K210497
Device Name Shoreline ACS (Anterior Cervical System)
#### Indications for Use (Describe)
Indications for Use (Describe)
The Shoreline ACS (Anterior Cervical System) with NanoMetalene® surface technology are interbody fusion devices I I ill Shoremie Ac s (Anchor Cervical of skeletally mature patients with degenerative disc disease intenced for allerior cervical microody nanon processed pain with degeneration of the disc confirmed by history and radiographic studies. The Shoreline ACS implants are to be used with autogenous bone graft and/or allogeneic bone graft radiographic studies. The Shoremie Not implanted with and implanted via an anterior approach. The device is to be used in patients who have had at least six (6) weeks of non-operative treatment.
When used as a standalone system, Shoreline ACS is intended to be used as an adjunct to spinal fusion procedures at one level (C2-T1) and must be used with the Shoreline ACS bone screw fixation and locking cover.
When used with supplemental fixation, such as anterior cervical plates, the Shoreline Cervical low profile (TruProfile) Interbody Spacer is intended to be used as an adjunct to spinal fusion procedures at one or two levels of the cervical spine (C2-T1).
| AND AND AND INCREAT CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR | | | Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|-------------------------------------------------|--|
| | | | | |
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{10}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
#### K210497
Device Name Shoreline Cervical Interbody RT System
## Indications for Use (Describe)
Indications for Use (Describe)
The Shoreline Cervical Interbody RT System with NanoMetalene® surface technology are interbody fusion devices TV S The Storeme Cervical miterbody (11 by stant with degenerative disc disease (DDD) of the cervical spine (C2-T1) for one mended to tas in skeleadly mature patent. DDD is defined as discogenic pain with degeneration of the disc of two colligabous levels, dopending on the securents should be skeletally mature and have had at least six (6) weeks of non-operative treatment. These devices are to be filled with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone.
When used as a standalone system, the Shoreline Cervical Interbody RT System is intended to be used as an adjunct to within as a standatione system, are simple level (C2-T1) and must be used with the Shoreline ACS bone screw fixation and locking cover.
When used with supplemental fixation, such as anterior cervical plates, the Shoreline Cervical Interbody RT System is intended to be used as an adjunct to spinal fusion procedures at one or two levels of the cervical spine (C3-C7).
When the system is used at two contiguous levels, the Shoreline Cervical Interbody RT System must be used with supplemental fixation.
| Type of Use (Select one or both, as applicable) |
|----------------------------------------------------------------------------------------------------------------------------------------|
| <span> <b> ☑ </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
| <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{11}------------------------------------------------
510(k) Number (if known)
K210497
Device Name SeaSpine Reef TO/TA System
## Indications for Use (Describe)
Indications for Use (Describe)
When used as an intervertebral body fusion device, the SeaSpine Reef TO/TA System with NanoMetalene® surface when used as an merveltedrar our institutes at one or two contiguous levels (L2?S1) in skeletally mature patients technology is mended tor spillar rasion proceanterined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved reiolismes at the involved iven's). These pasis months of non-operative treatment. The device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone and supplemental fixation.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{12}------------------------------------------------
510(k) Number (if known) K210497
Device Name SeaSpine Reef TH System
Indications for Use (Describe)
Indications of Occ (2008/10)
The SeaSpine Reef TH System with NanoMetalene® surface technology is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had six months of nonoperative treatment. The device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone and supplemental fixation.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{13}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K210497
Device Name SeaSpine Meridian System
#### Indications for Use (Describe)
Interbody Device (IBD) Implants (i.e., interbody implants used alone):
Interbody Device (115) Implans (1.5, merood) meet echnology interbody is indicated for use as an adjunct to I lie Seasonic Mendrall System with Nandisease (DDD) at one or two contiguous levels from L2 to S1. These DDD tusion in pations with organetade 1 spondylolisthesis at the involved level(s). The interior of the device is to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The SeaSpine Meridian Interbody is intended for use with supplemental fixation. Degenerative disc disease is defined as I ne ocaspite werendin mith degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment.
#### No-Profile Implants w/ Screws:
No-Frome implants w/ Serews:
The SeaSpine Meridian System No-Profile Interbody, when used with Screws and a No-Profile Locking Cover, is a I he business interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. Hyperlordotic sizes (25 and 30 degrees) are intended for use with supplemental fixation.
### No-Profile Implants w/ Inline Fixation Anchors:
The SeaSpine Meridian System No-Profile Interbody, when used with Inline Fixation Anchors and a No Profile Locking Cover, is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The SeaSpine Meridian No-Profile Implants w/ Inline Fixation Anchors is intended for use with supplemental fixation.
### TruProfile Interbody Implants:
The SeaSpine Meridian System Interbody assembled with the Anterior Plate, when used with Screws and an Anterior Plate Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. Hyperlordotic sizes (25 and 30 degrees) are intended for use with supplemental fixation.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
{14}------------------------------------------------
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the l he burden time for this collection of information is csamated to arelage the data needed and complete time to review the collection. Send comments regarding this burden estimate or any other aspect and review the collection of information. Sogestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{15}------------------------------------------------
# 510(k) Summary
K210497 Page 1 of 8
| Contact Details<br>Applicant Name: | SeaSpine Orthopedics Corporation |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address:<br>Phone number:<br>Fax number: | 5770 Armada Drive, Carlsbad CA<br>(619) 884-4342 (Aly) or (790) 216-5109 (Kate)<br>(760) 683-6874 |
| Contact person:<br>Additional contact: | Aly Alvarez, Assoc. Manager, Regulatory Affairs<br>Kate Hunter, Director, Regulatory Affairs |
| Date Prepared: | February 17, 2020 |
| Device Name<br>Trade Name: | SeaSpine Spacer System – (Hollywood NanoMetalene, Hollywood VI<br>NanoMetalene, Pacifica NanoMetalene, Redondo NanoMetalene, Ventura<br>NanoMetalene),<br>Vu a•Pod L NanoMetalene,<br>SeaSpine Vu e•Pod System,<br>SeaSpine Vu a•Pod Prime NanoMetalene,<br>SeaSpine Regatta Lateral System,<br>SeaSpine Cambria System,<br>SeaSpine Shoreline ACS - Anterior Cervical System,<br>Shoreline Cervical Interbody RT System,<br>SeaSpine Beachside System, and<br>SeaSpine Reef TH System<br>SeaSpine Meridian System |
| Common Name: | Intervertebral body fusion device,<br>Spinal intervertebral body fixation orthosis |
| Classification Name: | Intervertebral body fusion device (21 CFR 888.3080)<br>Spinal intervertebral body fixation orthosis (21 CFR 888.3060) |
| Class: | II |
| Product Code: | MAX (Intervertebral Fusion Device with Bone Graft, Lumbar)<br>ODP (Intervertebral Fusion Device with Bone Graft, Cervical)<br>MQP (Spinal Vertebral Body Replacement Device)<br>OVD (Intervertebral Fusion Device with Integrated Fixation, Lumbar)<br>OVE (Intervertebral Fusion Device with Integrated Fixation, Cervical) |
{16}------------------------------------------------
# K210497 Page 2 of 8
### Legally Marketed Predicate Devices
| 510(k) Number | Product<br>Code | Trade Name | Manufacturer |
|-----------------------|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|
| Primary Predicate | | | |
| K142488,<br>K173260 | MAX,<br>ODP,<br>MQP | SeaSpine Spacer System –<br>(Hollywood NanoMetalene,<br>Hollywood VI NanoMetalene,<br>Pacifica NanoMetalene,<br>Redondo NanoMetalene,<br>Ventura NanoMetalene) | SeaSpine Orthopedics<br>Corporation |
| Additional Predicates | | | |
| K142488 | MAX, ODP | Vu a•POD-L NanoMetalene | SeaSpine Orthopedics<br>Corporation |
| K142488,<br>K173260 | MAX,<br>ODP,<br>MQP | SeaSpine Vu e•POD System | SeaSpine Orthopedics<br>Corporation |
| K173606 | OVD | SeaSpine Vu a•POD Prime<br>NanoMetalene Intervertebral<br>Body Fusion Device | SeaSpine Orthopedics<br>Corporation |
| K190655 | OVD | SeaSpine Shoreline ACS –<br>Anterior Cervical Standalone<br>System | SeaSpine Orthopedics<br>Corporation |
| K171046 | ODP | SeaSpine Cambria System | SeaSpine Orthopedics<br>Corporation |
| K181079 | MAX | SeaSpine Regatta Lateral System | SeaSpine Orthopedics<br>Corporation |
| K183083 | OVE,<br>ODP | Shoreline Cervical Interbody RT<br>System | SeaSpine Orthopedics<br>Corporation |
| K192132 | MAX | SeaSpine Reef TO/TA System<br>(formerly SeaSpine Beachside<br>System) | SeaSpine Orthopedics<br>Corporation |
| K193636 | MAX | SeaSpine Reef TH System | SeaSpine Orthopedics<br>Corporation |
| K200879 | MAX,<br>OVD | SeaSpine Meridian System | SeaSpine Orthopedics<br>Corporation |
# Device Description
The SeaSpine NanoMetalene Systems are single-use intervertebral fusion devices made from polyetheretherketone (PEEK per ASTM F2026) with markers (tantalum per ASTM F560 or Ti-6AI-4V ELI per ASTM F136) for radiographic visualization. The devices have a central canal for receiving autogenous bone graft and or allogenic bone graft, composed of cancellous, cortical, and/or corticocancellous bone prior to implantation. The devices are offered in a variety of sizes
{17}------------------------------------------------
and geometries to accommodate variations in pathology and patient anatomy and are provided sterile-packed.
# Indications for Use
# SeaSpine Spacer System - (Hollywood NanoMetalene, Hollywood VI NanoMetalene, Ventura NanoMetalene, Pacifica NanoMetalene)
When used as an intervertebral body fusion device, the SeaSpine Spacer System with NanoMetalene® surface technology is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had six months of nonoperative treatment. The device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone and supplemental fixation.
## SeaSpine Vu a•POD-L NanoMetalene Intervertebral Body Fusion Device
SeaSpine Vu a•POD-L NanoMetalene Intervertebral Body Fusion Device with The NanoMetalene® surface technology is indicated for use as an adjunct to fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The device is to be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The SeaSpine Vua«POD-L NanoMetalene Intervertebral Body Fusion Device is intended for use with supplemental fixation that is in addition to the integrated buttress spin plate, such as a pedicle screw system or anterior plate. Degenerative disc disease is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment.
### SeaSpine Vu e•POD System
When used as an intervertebral body fusion device, the SeaSpine Vu e•POD System with NanoMetalene® surface technology is indicated for use as an adjunct to fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The device is indicated for use with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The SeaSpine Vu e · POD System is intended for use with supplemental fixation. Degenerative disc disease (DDD) is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.
When used as a vertebral body replacement (VBR), the SeaSpine Vu e • POD System is indicated for use in the thoracolumbar spine (TI-L5) to replace a collapsed, or otherwise unstable K201193 vertebral body due to tumor or trauma (i.e., fracture). The SeaSpine Vu e •POD System
{18}------------------------------------------------
is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of fusion for a prolonged period. The device is indicated for use with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The SeaSpine Vu e•POD System is intended for use with supplemental internal spinal fixation.
## SeaSpine Vu a•POD Prime NanoMetalene Intervertebral Body Fusion Device
When used with the bone screws, the SeaSpine Vu a•POD Prime NanoMetalene Intervertebral Body Fusion Device with NanoMetalene® surface technology is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the interbody spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone.
When used with the SpinPlate, the SeaSpine Vu a•POD Prime NanoMetalene Intervertebral Body Fusion Device is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the interbody spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. When used with the SpinPlate, the SeaSpine Vu a•POD Prime NanoMetalene IntervertebralBody Fusion Device is intended for use with supplemental fixation.
The SeaSpine Vu a•POD Prime NanoMetalene Intervertebral Body Fusion Device, when used with the bone screws or the bone screws and the SpinPlate, is a stand-alone device. If the SeaSpine Vu a•POD Prime NanoMetalene Intervertebral Body Fusion Device is used only with the SpinPlate, then additional supplemental fixation, which has been cleared by the FDA for use in the lumbar spine, must be used to augment stability. Additionally, implants with hyperlordotic angles of >20° must also be used with additional supplemental fixation (e.g., posterior pedicle screw and rod systems). This device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone.
Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device.
## Shoreline ACS (Anterior Cervical System)
The Shoreline ACS (Anterior Cervical System) with NanoMetalene® surface technology are interbody fusion devices intended for anterior cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. The Shoreline ACS implants are to be used with autogenous bone graft and/or allogeneic bone graft
{19}------------------------------------------------
composed of cancellous, cortical, and/or corticocancellous bone and implanted via an anterior approach. The device is to be used in patients who have had at least six (6) weeks of nonoperative treatment.
When used as a standalone system, Shoreline ACS is intended to be used as an adjunct to spinal fusion procedures at one level (C2-T1) and must be used with the Shoreline ACS bone screw fixation and locking cover.
When used with supplemental fixation, such as anterior cervical plates, the Shoreline Cervical low profile (TruProfile) Interbody Spacer is intended to be used as an adjunct to spinal fusion procedures at one or two levels of the cervical spine (C2-T1).
# SeaSpine Cambria System
The SeaSpine Cambria NanoMetalene System with NanoMetalene® surface technology is intended to be used as an adjunct to spinal fusion procedures at one or two contiguous levels (C3-C7) in skeletally mature patients with degenerative disc disease (defined as neck pain with discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Patients should have received at least six weeks of non-operative treatment prior to treatment with the device. Devices are intended to be implanted via an open, anterior approach and used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone and supplemental fixation, such as an anterior plating system.
## SeaSpine Regatta Lateral System
Interbody Device (IBD) Implants (i.e., interbody implants used alone):
The SeaSpine Regatta Lateral System with NanoMetalene® surface technology is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD, defined as back pain of discogenic origin, with degeneration of the disc confirmed by history and radiographic studies). It is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of DDD with up to Grade 1 spondylolisthesis at the involved level(s). The interior of the interbody spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device.
The SeaSpine Regatta Lateral System is intended for use with supplemental fixation.
# TruProfile Interbody Implants:
The SeaSpine Regatta Lateral System assembled with the TruProfile Lateral Plate, when used with Screws, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packedwith autogenous
{20}------------------------------------------------
bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device.
The SeaSpine Regatta Lateral System assembled with the 1-hole TruProfile Lateral Plate, when used with Screws, is intended for use with supplemental fixation.
# Shoreline Cervical Interbody RT System
The Shoreline Cervical Interbody RT System with NanoMetalene® surface technology are interbody fusion devices intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) for one or two contiguous levels, depending on the system. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. These devices are to be filled with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone.
When used as a standalone system, the Shoreline Cervical Interbody RT System is intended to be used as an adjunct to spinal fusion procedures at a single level (C2-T1) and must be used with the Shoreline ACS bone screw fixation and locking cover.
When used with supplemental fixation, such as anterior cervical plates, the Shoreline Cervical Interbody RT System is intended to be used as an adjunct to spinal fusion procedures at one or two levels of the cervical spine (C3-C7).
When the system is used at two contiguous levels, the Shoreline Cervical Interbody RT System must be used with supplemental fixation.
# SeaSpine Reef TO/TA System
When used as an intervertebral body fusion device, the SeaSpine Reef TO/TA System with NanoMetalene® surface technology is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had six months of nonoperative treatment. The device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone and supplemental fixation.
## SeaSpine Reef TH System
The SeaSpine Reef TH System with NanoMetalene® surface technology is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also
{21}------------------------------------------------
have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had six months of nonoperative treatment. The device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone and supplemental fixation.
## SeaSpine Meridian System
Interbody Device (IBD) Implants (i.e., interbody implants used alone):
The SeaSpine Meridian System with NanoMetalene® surface technology interbody is indicated for use as an adjunct to fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the device is to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The SeaSpine Meridian Interbody is intended for use with supplemental fixation. Degenerative disc disease is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment.
## No-Profile Implants w/ Screws:
The SeaSpine Meridian System No-Profile Interbody, when used with Screws and a No-Profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device.Hyperlordotic sizes (25 and 30 degrees) are intended for use with supplemental fixation.
### No-Profile Implants w/ Inline Fixation Anchors:
The SeaSpine Meridian System No-Profile Interbody, when used with Inline Fixation Anchors and a No Profile LockingCover, is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The SeaSpine Meridian No-Profile Implants w/ Inline Fixation Anchors is intended for use with supplemental fixation.
### TruProfile Interbody Implants:
The SeaSpine Meridian System Interbody assembled with the Anterior Plate, when used with Screws and an Anterior Plate Locking Cover, is a standalone interbody implant indicated for use as an adjunct
{22}------------------------------------------------
to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of nonoperative treatment prior to being treated with the device. Hyperlordotic sizes (25 and 30 degrees) are intended for use with supplemental fixation.
# Summary of Technological Characteristics
The SeaSpine NanoMetalene Systems are identical to the cited predicate devices. They have identical technological characteristics to the cited predicate devices in regard to components, device description, intended use/indications for use, technological characteristics (operating principle, design, materials, sterility, etc.) and performance (mechanical safety).
# Non-Clinical Testing
The SeaSpine NanoMetalene Systems have demonstrated equivalent mechanical performance to the predicate systems in accordance with requirements outlined in ASTM F1978, ASTM F1147, ASTM F1160 and ASTM F1044. In addition, the SeaSpine NanoMetalene Systems have equivalent biocompatibility and sterility to that of their predicate counterparts.
# Conclusions
The submitted data demonstrates that the subject SeaSpine NanoMetalene Systems have been shown to be substantially equivalent to legally marketed predicate devices for their intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.