CONDUIT™ ATP Inserters

K232014 · Enztec Limited · MAX · Sep 1, 2023 · Orthopedic

Device Facts

Record IDK232014
Device NameCONDUIT™ ATP Inserters
ApplicantEnztec Limited
Product CodeMAX · Orthopedic
Decision DateSep 1, 2023
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3080
Device ClassClass 2

Indications for Use

The CONDUIT™ ATP Inserters are intended to be used with the EIT Cellular Titanium® LLIF Cages. The EIT Cellular Titanium® LLIF Cages with a microscopic roughened surface and micro and nanoscale features are indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S 1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These spinal implants are to be used with autogenous bone graft and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft. EIT Spine LLIF is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.

Device Story

CONDUIT™ ATP Inserters are reusable surgical instruments designed for Lumbar Interbody Fusion surgery via the ATP approach. Instruments interface with EIT Cellular Titanium® LLIF Cages to facilitate cage insertion. Used in operating rooms by surgeons; instruments provide mechanical handling and positioning of spinal implants. Device does not perform automated analysis or provide clinical decision support; it serves as a manual tool for implant delivery. Benefits include enabling precise placement of intervertebral fusion devices during spinal surgery.

Clinical Evidence

Bench testing only. Non-clinical performance testing included impaction endurance testing, formative usability testing, and thread cyclic testing. Results demonstrate performance is sufficient for intended use and equivalent to predicate devices.

Technological Characteristics

Manufactured from Stainless Steel (ASTM F899 or ASTM A693) and Silicone (ESP Class VI). Reusable, non-sterile instruments. Mechanical insertion tools for spinal implants. No software or electronic components.

Indications for Use

Indicated for use with EIT Cellular Titanium LLIF Cages in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD defined as discogenic back pain with disc degeneration confirmed by history and radiography. Patients may have up to Grade 1 spondylolisthesis or retrolisthesis. Requires use with autogenous/allogeneic bone graft and supplemental fixation. Patients must have at least 6 months of prior non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 1, 2023 Enztec Limited % Nathan Wright, MS, RAC Engineer & Regulatory Specialist Empirical Technologies 4628 Northpark Drive Colorado Springs, Colorado 80918 Re: K232014 Trade/Device Name: CONDUIT™ ATP Inserters Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: August 11, 2023 Received: August 11, 2023 Dear Mr. Wright: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {2}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Brent Showalter -S Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use Submission Number (if known) K232014 Device Name CONDUIT™ ATP Inserters Indications for Use (Describe) The CONDUIT™ ATP Inserters are intended to be used with the EIT Cellular Titanium® LLIF Cages. The EIT Cellular Titanium® LLIF Cages with a microscopic roughened surface and micro and nanoscale features are indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S 1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These spinal implants are to be used with autogenous bone graft and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft. EIT Spine LLIF is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # K232014 - 510(K) SUMMARY | Submitter's Name: | Enztec Limited | |--------------------------------|------------------------------------------------------------------------------------------------------------------------------------| | Submitter's Address: | 3/17 Print Place<br>Middleton, Christchurch 8024 | | Submitter's Telephone: | +64 27 829 2440 | | Contact Person: | Nathan Wright MS, RAC<br>Empirical Technologies<br>719-351-0248<br>nwright@empiricaltech.com<br>Image: Empirical Technologies logo | | Date Summary was Prepared: | July 6, 2023 | | Trade or Proprietary Name: | CONDUIT™ ATP Inserters | | Device Classification Name: | Implant insertion tool for intervertebral fusion device, lumbar | | Classification & Regulation #: | Class II per 21 CFR §888.3080 | | Product Code: | MAX | | Classification Panel: | Spinal Devices (DHT6B) | # DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION: The CONDUIT™ ATP Inserters are designed for use during Lumbar Interbody Fusion surgery, specifically using the ATP surgical approach. The inserters have been designed to interface with EIT Cellular Titanium® LLIF Cages (K201605). The reusable instruments are provided non-sterile and made from commonly used orthopedic materials with commonly used manufacturing processes. There are no changes to the implants or other compatible instruments provided with the implants. ### INDICATIONS FOR USE The CONDUIT™ ATP Inserters are intended to be used with the EIT Cellular Titanium® LLIF Cages. The EIT Cellular Titanium® LLIF Cages with a microscopic roughened surface and micro and nano-scale features are indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S 1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These spinal implants are to be used with autogenous bone graft and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft. EIT Spine LLIF is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage. ### TECHNOLOGICAL CHARACTERISTICS The CONDUIT™ ATP Inserters are manufactured from Stainless Steel per ASTM F899 or ASTM A693 and Silicone per ESP Class VI. The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically, the following characteristics are the same between the subject and predicates: - Indications for use ● - Materials of manufacture ● - . Sterility - . Compatibility The differences in instrument handle design for surgeon utility and in handle materials and in the introduction of new instrument options (cobbs and spreaders) not previously part of the CONDUIT™ Inserter set do not {5}------------------------------------------------ introduce any questions for safety and efficacy which were mitigated through the risk assessment and verification and validation testing. #### Predicate Devices | 510k Number | Trade or Proprietary or Model Name | Manufacturer | Predicate Type | |-------------|------------------------------------|---------------|----------------| | K223671 | CONDUIT™ LLIF Straight Inserters | DePuy Synthes | Primary | #### PERFORMANCE DATA Non-clinical testing was conducted to confirm the device performance per intended use including impaction endurance testing, formative usability testing, and thread cyclic testing. The results of this non-clinical testing show that the performance of the CONDUIT™ ATP Inserters is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices. #### CONCLUSION The overall technology characteristics and mechanical performance data lead to the conclusion that the CONDUIT™ ATP Inserters are substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...