EIT Cellular Titanium Cervical Cage, EIT Cellular Titanium ALIF Cage, EIT Cellular Titanium TLIF Cage, EIT Cellular Titanium LLIF Cage, EIT Cellular Titanium T/PLIF Cage
K201605 · Eit Emerging Implant Technologies GmbH · MAX · Jul 15, 2020 · Orthopedic
Device Facts
Record ID
K201605
Device Name
EIT Cellular Titanium Cervical Cage, EIT Cellular Titanium ALIF Cage, EIT Cellular Titanium TLIF Cage, EIT Cellular Titanium LLIF Cage, EIT Cellular Titanium T/PLIF Cage
Applicant
Eit Emerging Implant Technologies GmbH
Product Code
MAX · Orthopedic
Decision Date
Jul 15, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
EIT Cellular Titanium® Cervical Cage The EIT Cellular Titanium® Cervical Cages with a microscopic roughened surface and micro and nano-scale features are intervertebral body fusion devices intended for use for anterior cervical interbody fusion in skeletally mature patients with cervical disc degeneration and/or cervical spinal instability, as confirmed by imaging studies (radiographs, CT, MRI), that results in radiculopathy, myelopathy, and/or pain at multiple contiguous levels from C2 - T1. The EIT Cellular Titanium Cervical Cages are also to be used with supplemental fixation systems that have been cleared for use in the cervical spine. EIT Cellular Titanium Cervical Cages are designed for use with autogenous and/or allogeneic bone graft comprised of cancellous, and/or corticocancellous bone graft to facilitate fusion. EIT Cellular Titanium® ALIF Cage The EIT Cellular Titanium® ALIF Cages with a microscopic roughened surface and micro and nano-scale features in combination with supplemental fixation are indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2 - S1 whose condition requires the use of interbody fusion. Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). Patients must be skeletally mature. Patients should have received 6 months of nonoperative treatment prior to treatment with the devices. EIT Cellular Titanium® TLIF Cage The EIT Cellular Titanium® TLIF Cages with a microscopic roughened surface and micro and nano-scale features in combination with supplemental fixation are indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2 - S1 whose condition requires the use of interbody fusion. Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). Patients must be skeletally mature. Patients should have received 6 months of non-operative treatment prior to treatment with the devices. EIT Cellular Titanium® LLIF Cage The EIT Cellular Titanium® LLIF Cages with a microscopic roughened surface and micro and nano-scale features is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These spinal implants are to be used with autogenous bone graft and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft. EIT Spine LLIF is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage. EIT Cellular Titanium® T/PLIF Cage The EIT Cellular Titanium® T/PLIF Cages with a microscopic roughened surface and micro and nano-scale features in combination with supplemental fixation are indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2 - S1 whose condition requires the use of interbody fusion. Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). Patients must be skeletally mature. Patients should have received 6 months of nonoperative treatment prior to treatment with the devices.
Device Story
3D-printed porous titanium interbody fusion cages; stabilize spinal segments; restore intervertebral height; facilitate fusion. Input: autogenous/allogeneic bone graft packed into hollow cage geometry. Operation: additive manufacturing (Selective Laser Melting) creates solid and porous structures with microscopic roughened surfaces and nano-scale features. Used in cervical (C2-T1) and lumbar (L2-S1) spine procedures; requires supplemental fixation (plates, screws, rods). Surgeon-operated in clinical/OR settings. Output: structural support for spinal fusion. Benefit: promotes fusion in patients with degenerative disc disease or instability.
Clinical Evidence
No clinical data; bench testing only. Substantial equivalence supported by design, material, and performance comparisons to predicate devices.
Technological Characteristics
Material: Ti-6Al-4V ELI (ASTM F3001). Manufacturing: Additive manufacturing (Selective Laser Melting). Features: Porous titanium structure with microscopic roughened surface and nano-scale features. Form factor: Intervertebral cages for cervical and lumbar spine. Connectivity: N/A.
Indications for Use
Indicated for skeletally mature patients requiring intervertebral body fusion for cervical disc degeneration/instability (C2-T1) or lumbar degenerative disc disease (L2-S1) with up to Grade I spondylolisthesis/retrolisthesis. Requires 6 months non-operative treatment. Contraindications include skeletal immaturity.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 15, 2020
EIT Emerging Implant Technologies GmbH % Ms. Christine Cahillane Regulatory Affairs Specialist DePuy Synthes Spine 325 Paramount Dr. Raynham, MA 02767 USA
Re: K201605
Trade/Device Name: EIT Cellular Titanium® Cervical Cage, EIT Cellular Titanium® ALIF Cage, EIT Cellular Titanium® TLIF Cage, EIT Cellular Titanium® LIF Cage, EIT Cellular Titanium® T/PLIF Cage
Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX, ODP Dated: June 12, 2020 Received: June 15, 2020
Dear Ms. Cahillane:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brent Showalter, Ph.D. Acting Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K201605
#### Device Name
EIT Cellular Titanium® Cervical Cage, EIT Cellular Titanium® ALIF Cage, EIT Cellular Titanium® TLIF Cage, EIT Cellular Titanium® LLIF Cage, EIT Cellular Titanium® T/PLIF Cage
Indications for Use (Describe)
EIT Cellular Titanium® Cervical Cage
The EIT Cellular Titanium® Cervical Cages with a microscopic roughened surface and nano-scale features are intervertebral body fusion devices intented for use for anterior cervical interbody firsion in skeletally mature patients with cervical disc degeneration and/or cervical spinal instability, as confirmed by imaging studies (radiographs, CT, MRI), that results in radiculopathy, myelopathy, and/or pain at multiple contiguous levels from C2 - T1. The EIT Cellular Titanium Cervical Cages are also to be used with supplemental fixation systems that have been cleared for use in the cervical spine. EIT Cellular Titanium Cervical Cages are designed for use with autogenous and/or allogeneic bone graft comprised of cancellous, and/or corticocancellous bone graft to facilitate fusion.
#### EIT Cellular Titanium® ALIF Cage
The EIT Cellular Titanium® ALIF Cages with a microscopic roughened surface and micro and nano-scale features in combination with supplemental fixation are indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2 - S1 whose condition reguires the use of interbody fusion. Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may have up to Grade I spondylolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). Patients must be skeletally mature. Patients should have received 6 months of nonoperative treatment with the devices.
#### EIT Cellular Titanium® TLIF Cage
The EIT Cellular Titanium® TLIF Cages with a microscopic roughened surface and micro and nano-scale features in combination with supplemental fixation are indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2 - S1 whose condition requires the use of interbody fusion. Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may have up to Grade I spondylolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). Patients must be skeletally mature. Patients should have received 6 months of non-operative treatment with the devices.
#### EIT Cellular Titanium® LLIF Cage
The EIT Cellular Titanium & LLF Cages with a microscopic roughened surface and micro and nano-scale features are indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two configuous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confinued by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These spinal implants are to be used with autogenous bone graft and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft. ETT Spine LLIF is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment with an intervertebral cage.
#### EIT Cellular Titanium® T/PLIF Cage
The EIT Cellular Titanium® T/PLIF Cages with a microscopic roughened surface and micro and nano-scale features in
{3}------------------------------------------------
combination with supplemental fixation are indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2 - S1 whose condition regures the use of interbody fusion. Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may have up to Grade I spondvlolisthesis at the involved level(s). These patients may have had a previous non-firsion spinal surgery at the involved spinal level(s). Patients must be skeletally mature. Patients should have received 6 months of nonoperative treatment with the devices.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Emerging Implant Technologies (EIT). The logo consists of a geometric design at the top, resembling interconnected hexagons in shades of green, yellow, and orange. Below the geometric design, the letters "EIT" are displayed in a bold, sans-serif font, followed by the words "Emerging Implant Technologies" in a smaller, regular font.
## K201605 - 510(k) Summary
| | K201605 - 510(k) Summary |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| A. Submitter Information | |
| Manufacturer: | EIT Emerging Implant Technologies GmbH<br>Eisenbahnstrasse 84<br>78573 Wurmlingen, Germany<br>Phone: +49 7461 1716900 |
| Contact: | Christine Cahillane<br>DePuy Synthes Spine<br>On behalf of EIT Emerging Implant Technologies GmbH<br>325 Paramount Drive<br>Raynham, MA 02767 |
| Telephone Number: | 508-828-3064 |
| Email Address: | ccahilla@its.jnj.com |
| B. Date Prepared | June 23, 2020 |
| C. Device Name | EIT Cellular Titanium® Cervical Cage<br>EIT Cellular Titanium® ALIF Cage<br>EIT Cellular Titanium® TLIF Cage<br>EIT Cellular Titanium® LLIF Cage<br>EIT Cellular Titanium® T/PLIF Cage |
| Classification: | 21 CFR §888.3080, Intervertebral body fusion device |
| Regulatory Class: | II |
| Product Codes: | ODP - Intervertebral Fusion Device with Bone Graft,<br>Cervical<br>MAX - Intervertebral Fusion Device with Bone Graft,<br>Lumbar |
| D. Predicate Device Names | EIT Cellular Titanium® Cervical Cage (K172888)<br>EIT Cellular Titanium® ALIF Cage (K172888)<br>EIT Cellular Titanium® TLIF Cage (K172888)<br>EIT Cellular Titanium® LLIF Cage (K181644)<br>EIT Cellular Titanium® T/PLIF Cage (K183447) |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for EIT Emerging Implant Technologies. The logo features a stylized molecular structure in yellow and green above the letters "EIT" in bold, black font. Below the letters, the words "Emerging Implant Technologies" are stacked in a smaller, sans-serif font.
### E. Device Description
The EIT Cellular Titanium® Technology platform is a comprehensive portfolio of 3D-printed porous titanium interbody devices intended to stabilize the spinal segment, restore intervertebral height and to facilitate interbody fusion in the cervical (C2-T1) and lumbar spine (L2-S1). Designed to treat cervical and lumbar degenerative disc disease, the platform consists of the Cervical, Transforaminal (TLIF), Lateral (LLIF), Anterior (ALIF) and Transforaminal/Posterior Lumbar (T/PLIF) systems. Each system features a full breadth of sizes. footprints, heights and angles. The devices are intended to be used with supplemental spinal fixation, either applied anterior or posterior (e.g. using posterior pedicle screws, anterior plate system or anterior screw and rod system).
The EIT Cellular Titanium® Cages are made from Ti-6A1-4V ELI conforming to ASTM F3001 with an additive manufacturing process (Selective Laser Melting). The design contains solid structures and porous structures. The hollow geometry of the implants allows them to be packed with autogenous bone graft, and two of the systems allow for the cages to be packed with allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft.
The 3D Printed EIT Cellular Titanium® Cages have a microscopic roughened surface with micro and nano-scale features. The micro and nano features are on all surfaces of the cage, including the superior, inferior, and peripheral surfaces, as well as each member of the internal cell structure.
# F. Intended Uses
### EIT Cellular Titanium® Cervical Cage
The EIT Cellular Titanium® Cervical Cages with a microscopic roughened surface and micro and nano-scale features are intervertebral body fusion devices intended for use for anterior cervical interbody fusion in skeletally mature patients with cervical disc degeneration and/or cervical spinal instability, as confirmed by imaging studies (radiographs, CT, MRI), that results in radiculopathy, myelopathy, and/or pain at multiple contiguous levels from C2 - T1. The EIT Cellular Titanium Cervical Cages are also to be used with supplemental fixation systems that have been cleared for use in the cervical spine. EIT Cellular Titanium Cervical Cages are designed for use with autogenous and/or allogeneic bone graft comprised of cancellous, and/or corticocancellous bone graft to facilitate fusion.
# EIT Cellular Titanium® ALIF Cage
The EIT Cellular Titanium® ALIF Cages with a microscopic roughened surface and micro and nano-scale features in combination with supplemental fixation are indicated for use with
> K201605 510k Summary Page 2 of 4
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for EIT Emerging Implant Technologies. The logo features a geometric design at the top, resembling interconnected hexagons in shades of green, yellow, and orange. Below the geometric design, the letters "EIT" are displayed in a bold, sans-serif font. Underneath "EIT", the words "Emerging Implant Technologies" are written in a smaller, sans-serif font.
autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2 - S1 whose condition requires the use of interbody fusion. Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). Patients must be skeletally mature. Patients should have received 6 months of nonoperative treatment prior to treatment with the devices.
## EIT Cellular Titanium® TLIF Cage
The EIT Cellular Titanium® TLIF Cages with a microscopic roughened surface and micro and nano-scale features in combination with supplemental fixation are indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2 - S1 whose condition requires the use of interbody fusion. Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). Patients must be skeletally mature. Patients should have received 6 months of non-operative treatment prior to treatment with the devices.
### EIT Cellular Titanium® LLIF Cage
The EIT Cellular Titanium® LLIF Cages with a microscopic roughened surface and micro and nano-scale features is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These spinal implants are to be used with autogenous bone graft and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft. EIT Spine LLIF is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.
# EIT Cellular Titanium® T/PLIF Cage
The EIT Cellular Titanium® T/PLIF Cages with a microscopic roughened surface and micro and nano-scale features in combination with supplemental fixation are indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2 - S1 whose condition requires the use of interbody fusion. Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for EIT Emerging Implant Technologies. The logo features a geometric design at the top, resembling interconnected molecular structures, with colors ranging from green to yellow to orange. Below the geometric design, the acronym "EIT" is displayed in bold, dark gray letters, and beneath that, the full name "Emerging Implant Technologies" is written in a smaller, lighter gray font.
spinal surgery at the involved spinal level(s). Patients must be skeletally mature. Patients should have received 6 months of nonoperative treatment prior to treatment with the devices.
### G. Summary of Similarities and Differences in Technological Characteristics, Performance and Intended Use
The fundamental scientific technology of the subject devices remains unchanged from the predicate devices. There is no change to the intended use of the product from the predicate devices. This submission presents an expansion of the device description within the labeling supported by established methods for surface analysis. There has been no change to the design or additive manufacturing process of the EIT Cellular Titanium Cervical or Lumbar Cages previously cleared in K172888, K181644 and K183447.
### H. Materials
The EIT Cellular Titanium Cervical and Lumbar Cages are all constructed from Standard Specification for Additive Manufacturing Titanium-6 Vanadium ELI with Extra Low Interstitials components using full-melt powder bed fusion in conformance with ASTM F3001.
#### I. Performance Testing Summarv
Performance data remains unchanged for the subject devices as compared to the predicate devices.
#### J. Conclusion
The EIT Cellular Titanium Cervical and Lumbar Cages are substantially equivalent to previously cleared devices with respect to indications for use, design, function, materials, and performance.
> K201605 510k Summary Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.