SpineFrontier Lumbar Interbody Fusion Device System
K193106 · Spinefrontier, Inc. · MAX · Jun 19, 2020 · Orthopedic
Device Facts
Record ID
K193106
Device Name
SpineFrontier Lumbar Interbody Fusion Device System
Applicant
Spinefrontier, Inc.
Product Code
MAX · Orthopedic
Decision Date
Jun 19, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The SpineFrontier Lumbar Interbody Fusion Device System (Dorado IBC, Dorado PLIFT, Dorado ELIFT, Arena-L, Dorado TILT, Dorado TLIFT, Dorado Wide, and Ursa S-LIFT) is intended for intervertebral body fusion of the spine of skeletally mature patients, using autogenous bone graft to facilitate fusion. The device is indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2-S1. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history or radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. The SpineFrontier Lumbar Interbody Fusion Device System is intended to be used with supplemental spinal fixation system(s) cleared for use in the lumbar spine.
Device Story
Spinal intervertebral body fusion device system; used for lumbar fusion. Implants (Dorado/Arena/Ursa series) inserted into disc space to facilitate fusion using autogenous bone graft. Device features raised surface contours to resist migration. Used in conjunction with supplemental spinal fixation systems. Operated by surgeons in clinical/OR settings. Provides structural support to vertebral bodies during fusion process. Benefits patients by stabilizing spinal segments and addressing discogenic pain.
Clinical Evidence
No clinical data. Engineering rationale provided in lieu of bench testing; modifications determined not to present a worst-case for mechanical testing compared to previous clearance (K142504).
Technological Characteristics
Materials: PEEK Optima LT1 (ASTM F2026), Tantalum markers (ASTM F560), Titanium coating (ASTM F1580). Form factor: Interbody spacers with varying widths, lengths, and heights. Features raised surface contours for fixation. Non-active, mechanical device.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1, including those with up to Grade I spondylolisthesis. Requires six months of failed non-operative treatment. Must be used with supplemental spinal fixation.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
SpineFrontier Lumbar Interbody Fusion Device System (K142504)
Related Devices
K163481 — HALF DOME Posterior Lumbar Interbody System · Astura Medical · May 3, 2017
K203207 — Omnia Medical TiBrid-SA · Omnia Medical, LLC · Dec 23, 2020
K182470 — Plateau Spacer System · Life Spine · Nov 26, 2018
K151310 — Amendia Interbody Fusion Devices · Amendia, Inc. · Jan 6, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 19, 2020
SpineFrontier, Inc. Omkar Joglekar Quality Manager 350 Main St, 2nd Floor Malden, Massachusetts 02148
Re: K193106
Trade/Device Name: SpineFrontier Lumbar Interbody Fusion Device System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: May 22, 2020 Received: May 26, 2020
#### Dear Omkar Joglekar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brent L. Showalter, Ph.D. Acting Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K193106
Device Name SpineFrontier Lumbar Interbody Fusion Device System
Indications for Use (Describe)
The SpineFrontier Lumbar Interbody Fusion Device System (Dorado IBC, Dorado PLIFT, Dorado ELIFT, Arena-L, Dorado TILT, Dorado TLIFT, Dorado Wide, and Ursa S-LIFT) is intended for intervertebral body fusion of the spine of skeletally mature patients, using autogenous bone graft to facilitate fusion. The device is indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2-S1. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s).
Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history or radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.
The SpineFrontier Lumbar Interbody Fusion Device System is intended to be used with supplemental spinal fixation system(s) cleared for use in the lumbar spine.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| <div style="display:inline-block;"><span style="text-decoration: underline;"></span>Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:inline-block;">Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. | |
| <b>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*</b> | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: | |
| Department of Health and Human Services | |
| Food and Drug Administration | |
| Office of Chief Information Officer | |
| Paperwork Reduction Act (PRA) Staff | |
| PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
| FORM FDA 3881 (7/17) | PSC Publishing Services (301) 443-6740 EF |
{3}------------------------------------------------
| Submitter's Name: | SpineFrontier, Inc. |
|----------------------------|----------------------------------------------------------------------|
| Submitter's Address: | 350 Main Street<br>Malden, MA 02148 |
| Submitter's Telephone: | 978.232.3990 Ext-107 |
| Contact Person: | Omkar Joglekar, SpineFrontier, Inc.<br>omkarjoglekar@kicventures.com |
| Date Summary was Prepared: | 05/22/2020 |
| Trade or Proprietary Name: | SpineFrontier Lumbar Interbody Fusion Device System |
| Common or Usual Name: | Intervertebral Fusion Device With Bone Graft, Lumbar |
| Classification: | Class II per 21 CFR §888.3080 |
| Product Code: | MAX |
| Classification Panel: | Division of Orthopedic Devices |
# 510(K) SUMMARY
The purpose of this submission is to expand implant profiles offered for greater lordosis and curvature profile and to replace the titanium coating for the previously cleared implant with a new titanium coating for the SpineFrontier Lumbar Interbody Fusion Device System Dorado P-LIFT, Dorado T-LIFT, and Ursa S-LIFT for implants with a length of 24 mm or longer.
### DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The SpineFrontier Lumbar Interbody Fusion Device System is a spinal intervertebral body fusion device system intended for intervertebral body fusion of the spine of skeletally mature patients, using autogenous bone graft to facilitate fusion. The system is comprised of devices made of PEEK Optima® LT1 or titanium coated PEEK Optima® LT1, with varying widths, lengths, and heights to fit the anatomical needs of patients. The devices have raised contours on the superior and inferior surfaces that will resist device movement following implantation.
### INDICATIONS FOR USE
The SpineFrontier Lumbar Interbody Fusion Device System (Dorado IBC, Dorado PLIFT, Dorado ELIFT, Arena-L, Dorado TLIFT, Dorado TLIFT, Dorado Wide, and Ursa S-LIFT) is intended for intervertebral body fusion of the spine of skeletally mature patients, using autogenous bone graft to facilitate fusion. The device is indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2-S1. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s).
Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history or radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.
The SpineFrontier Lumbar Interbody Fusion Device System is intended to be used with supplemental spinal fixation system(s) cleared for use in the lumbar spine.
{4}------------------------------------------------
# TECHNICAL CHARACTERISTICS
The SpineFrontier Lumbar Interbody Fusion Device System is made from PEEK Optima® LT1 per ASTM F2026 with Tantalum markers per ASTM F560 and Titanium Coating per ASTM F1580. The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically the following characteristics are identical between the subject and predicates:
- Indications for Use
- Materials of manufacture
- Structural support mechanism
- o Specifically, the implant geometry between the vertebral body interfacing teeth geometry
- Principles of operation ●
#### Table 5-1 Predicate Devices
| 510k Number | Trade or Proprietary or Model Name | Manufacturer | Designation |
|-------------|--------------------------------------------------------|---------------------|-------------|
| K142504 | SpineFrontier Lumbar Interbody Fusion<br>Device System | SpineFrontier, Inc. | Primary |
## PERFORMANCE TESTING SUMMARY
In support of this Special 510(k) Device Modification Premarket Notification, engineering rationale was performed in lieu of bench testing. No performance data was required because none of the modifications to the subject from its previous clearance under K142504 present a worst case for mechanical testing to the worst case tested for that clearance.
# CONCLUSION
The modified SpineFrontier Lumbar Interbody Fusion Device System subject is very similar to previously cleared SpineFrontier Lumbar Interbody Fusion Device System (K142504). The subject SpineFrontier Lumbar Interbody Fusion Device System has the same intended uses, indications, technological characteristics, and principles of operation as the predicate SpineFrontier Lumbar Interbody Fusion Device System. The modifications raise no new types of safety or effectiveness questions. The overall technology characteristics and the rationale for not including mechanical performance data lead to the conclusion that the subject SpineFrontier Lumbar Interbody Fusion Device System is substantially equivalent to the predicate.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.