Republic Spine Restore Intervertebral Body Fusion System
Applicant
Republic Spine, LLC
Product Code
MAX · Orthopedic
Decision Date
Sep 12, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Intended Use
When used as a lumbar intervertebral body fusion device, the Republic Spine Restore Interbody Fusion System is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion.
Device Story
Intervertebral body fusion system; consists of spinal cages implanted via anterior, posterior, transforaminal, or lateral surgical approaches. Used by surgeons in clinical settings to treat degenerative disc disease. Device acts as a spacer to facilitate fusion; requires supplemental fixation and autograft. Provides structural support to the lumbar spine. Benefits patient by stabilizing the spinal segment and promoting fusion.
Clinical Evidence
Bench testing only. Mechanical verification testing performed per ASTM F2077 (static/dynamic axial compression) and ASTM F2267 (static subsidence), plus static expulsion testing. Worst-case PEEK device configuration utilized for testing.
Technological Characteristics
Implant components manufactured from titanium alloy (Ti-6Al-4V ELI, ASTM F136) or PEEK (ASTM F2026) with tantalum markers (ASTM F560). Various configurations for different surgical approaches. Mechanical spacer design.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) and up to Grade 1 spondylolisthesis or retrolisthesis at L2-S1, following 6 months of failed non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
Intervertebral Body Fusion Spinal System (K090064)
K200958 — MODULIF-A Anterior Lumbar Interbody Fusion System · Spinal Stability, LLC · Aug 28, 2020
K163481 — HALF DOME Posterior Lumbar Interbody System · Astura Medical · May 3, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right, the text "FDA U.S. FOOD & DRUG ADMINISTRATION" is displayed in blue. The word "FDA" is in a larger font size than the rest of the text.
September 12, 2019
Republic Spine, LLC James Doulgeris Director of Business Development and Quality Systems 350 Camino Gardens Blvd Suite 103 Boca Raton, Florida 33432
Re: K190398
Trade/Device Name: Republic Spine Restore Intervertebral Body Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: July 17, 2019 Received: August 12, 2019
Dear Mr. Doulgeris:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Melissa Hall Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190398
#### Device Name
Republic Spine Restore Intervertebral Body Fusion System
Indications for Use (Describe)
When used as a lumbar intervertebral body fusion device, the Republic Spine Restore Interbody Fusion System is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <svg height="12" width="12"> <rect fill="none" height="12" stroke="black" stroke-width="1" width="12"></rect> <line stroke="black" stroke-width="1" x1="0" x2="12" y1="0" y2="12"></line> <line stroke="black" stroke-width="1" x1="12" x2="0" y1="0" y2="12"></line> </svg> |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <svg height="12" width="12"> <rect fill="none" height="12" stroke="black" stroke-width="1" width="12"></rect> </svg> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K190398: 510(K) Summary
| Submitter's Name: | Republic Spine, LLC |
|----------------------------|-------------------------------------------------------------|
| Submitter's Address: | 2424 N. Federal Hwy Suite 257<br>Boca Raton, FL 33431 |
| Submitter's Telephone: | 561-362-8094 |
| Contact Person: | James Doulgeris james@rspine.com |
| Date Summary was Prepared: | 17 July 2019 |
| Trade or Proprietary Name: | Republic Spine Restore Intervertebral Body<br>Fusion System |
| Common or Usual Name: | Intervertebral Body Fusion System |
| Classification: | Class II per 21 CFR 888.3080 |
| Product Code: | MAX |
| Classification Panel: | Division of Orthopedic Devices |
## Description of the Device Subject to Premarket Notification
The Republic Spine Restore Intervertebral Body Fixation System will be offered in various device configurations based on surgical approach and patient anatomy, and consist of a Republic Spine lumbar intervertebral body fusion device, which may be implanted as a single device via an anterior, posterior, transforaminal or lateral approach.
## Indications for Use
When used as a lumbar intervertebral body fusion device, Republic Spine Restore Intervertebral Body Fusion Spinal System is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion.
## Technological Characteristics
The Republic Spine Restore Intervertebral Body Fusion System is comprised of various device configurations designed to accommodate patient anatomy and provide the surgeon with different surgical approach options.
The Republic Spine Restore Intervertebral Body Fusion System implant components are made of titanium alloy (Ti-6Al-4V ELI) per ASTM F136 or polyether ether ketone (Evonik Vestakeep® i4R) that conforms to ASTM F2026 with tantalum markers (ASTM F560) to assist the surgeon with the proper placement of the device. The subject devices have similar technological characteristics and identical indications s the currently cleared predicate devices.
| 510k Number | Trade or Proprietary or<br>Model Name | Manufacturer | Predicate Type |
|-------------|---------------------------------------------|---------------|----------------|
| K090064 | Intervertebral Body Fusion<br>Spinal System | Eminent Spine | Primary |
## Table 5-1 Predicate Devices
{4}------------------------------------------------
| K112664, K162206 | Zavation IBF System | Zavation LLC | Reference |
|------------------|---------------------------------------------------------------|---------------------|-----------|
| K190889 | Republic Spine Restore<br>Cervical Interbody Fusion<br>System | Republic Spine, LLC | Reference |
Substantial Equivalence Conclusion
The basis of substantial equivalence of the subject device(s) and predicate device(s) is established on the following:
The subject device and the predicate devices have the following similarities:
- Both systems have the same indications for use. The intended patient population and intended use are the same.
- Both systems operate using the same fundamental scientific technology.
- Both systems incorporate the same basic implant design.
- Both systems use the same methods of sterilization.
- Both systems use the same operational principles for the surgical implantation of the interbody cages.
- Both systems are manufactured from the same materials.
The subject device and the predicate devices have the following differences:
- Minor dimensional differences in height, width and depth. ●
## Performance Data
The subject PEEK device is considered worst case and therefore, was utilized during design verification mechanical bench tests to address the design differences between the subject device(s) and the predicate device(s). An engineering analysis consisting of FEA was conducted to determine the worst-case size for testing. The worst case size for the subject device has been tested in the following test modes:
- Static axial compression per ASTM F2077
- . Dynamic axial compression per ASTM F2077
- . Static subsidence per ASTM F2267
- . Static expulsion
Conclusions: The results of the testing demonstrate that both the subject PEEK and Ti devices are substantially equivalent to the predicate.
## Conclusion
Evaluation of the risks and performance data based on the differences between the subject device(s) and predicate(s) does not raise any new issues or concerns related to safety or effectiveness. It is concluded that the subject device(s), is as safe and effective as the predicate device(s) for its intended use and is substantially equivalent to the legally marketed predicate device(s).
> K190398: 510k Summary Page 2 of 2
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.