Leva Spacer System

K151581 · Spine Wave, Inc. · MAX · Jul 9, 2015 · Orthopedic

Device Facts

Record IDK151581
Device NameLeva Spacer System
ApplicantSpine Wave, Inc.
Product CodeMAX · Orthopedic
Decision DateJul 9, 2015
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Leva™ Spacer System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radios. These DDD patients may have up to Grade I spondylolisthesis at the involved level(s). The Leva™ Spacer System is to be used with autogenous bone graft and with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral body fusion device.

Device Story

Leva™ Spacer System is an intervertebral body fusion device for lumbar spine surgery; available in fixed and expandable configurations. Expandable implants are inserted in a collapsed state and expanded in-situ to a predefined height using a dedicated inserter instrument. Implants feature curved, ridged endplates to conform to anatomy and resist expulsion; designed to hold autogenous bone graft. Used by surgeons in clinical settings to facilitate fusion in patients with DDD. Benefits include restoration of disc height and stabilization of the spinal segment when used with supplemental fixation.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing including static/dynamic axial compression, static/dynamic compression shear, and subsidence testing per ASTM F2077 and ASTM F2267, alongside Finite Element Analysis.

Technological Characteristics

Fabricated from unalloyed commercially pure (Class II) titanium per ASTM F67. Available in fixed and expandable configurations. Expandable mechanism is mechanical/in-situ. Features curved, ridged endplates. Non-powered, passive implant.

Indications for Use

Indicated for skeletally mature patients with lumbar degenerative disc disease (DDD) at 1-2 contiguous levels (L2-S1) with/without Grade I spondylolisthesis or retrolisthesis. Requires 6 months prior non-operative treatment. Must be used with autogenous bone graft and supplemental fixation.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a group of people in profile. The logo is black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 9, 2015 Spine Wave, Incorporated Ms. Gail Yaeker-Daunis, RAC, MBA Senior Regulatory Affairs Specialist 3 Enterprise Drive, Suite 210 Shelton, Connecticut 06484 Re: K151581 Trade/Device Name: Leva™ Spacer System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: June 9, 2015 Received: June 11, 2015 Dear Ms. Yaeker-Daunis: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ Page 2 - Ms. Gail Yaeker-Daunis forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. # Lori A. Wiggins -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) #### K151581 Device Name Leva™ Spacer System #### Indications for Use (Describe) The Leva™ Spacer System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radios. These DDD patients may have up to Grade I spondylolisthesis at the involved level(s). The Leva™ Spacer System is to be used with autogenous bone graft and with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral body fusion device. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------|---------------------------------------------------------------------------------| | <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the words "Spine Wave" in a bold, sans-serif font. The words are stacked horizontally, with "Spine" on the left and "Wave" on the right. A blue wave-like graphic is underneath the word "Wave". ## 510(k) Summary Leva™ Spacer System ### 1. Submitter Information | Submitter: | Spine Wave, Inc. | |----------------|----------------------------------------------------------| | Address: | Three Enterprise Drive<br>Suite 210<br>Shelton, CT 06484 | | Telephone: | 203-712-1839 | | Telefax: | 203-944-9493 | | Contact: | Gail Yaeker-Daunis | | Date Prepared: | June 9, 2015 | #### 2. Device Information | Trade Name: | Leva™ Spacer System | |----------------------|------------------------------------------------------| | Common Name: | Intervertebral Body Fusion Device | | Classification: | Class II (special controls) per 21 CFR 888.3080 | | Classification Name: | Intervertebral Fusion Device with Bone Graft, Lumbar | | Product Code: | MAX | #### 3. Purpose of Submission The purpose of this submission is to gain clearance for a line addition to the Leva™ Spacer System, to add additional implant sizes and associated instruments. #### 4. Predicate Device Information The Leva™ Spacer System described in this submission is substantially equivalent to the following predicates: | Primary Predicate Device | Manufacturer | 510(k) No. | |--------------------------|------------------|------------------------| | LevaTM Spacer System | Spine Wave, Inc. | K141980 and<br>K150285 | #### 5. Device Description The Leva™ Spacer System is a lumbar intervertebral body fusion device fabricated from unalloyed commercially pure (Class II) titanium per ASTM F67. The device is available in both expandable and fixed configurations. The expandable device is provided in a collapsed form and is expanded to a predefined height in-situ using the Leva™ Inserter Instrument. Both the fixed and expandable implants are provided in different heights to {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "SPINEWAVE" in bold, black font. A blue wave design is underneath the word. The wave is a simple, curved line that adds a visual element to the logo. accommodate the anatomical needs for a range of patients. The implants are designed to accommodate autogenous bone graft material. The implants have curved endplates to conform to the bony endplates of the patient and ridges on the endplates to resist expulsion. ### 6. Indications for Use The Leva™ Spacer System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The Leva™ Spacer System is to be used with autogenous bone graft and with supplemental fixation. Patients should have at least six (6) months of nonoperative treatment prior to treatment with an intervertebral body fusion device. ## 7. Comparison of Technological Characteristics The substantial equivalence of the Leva™ Spacer System to the predicates is shown by similarity in intended use, indications for use, materials and performance. ## 8. Performance Data The modified implants were compared to constructs previously tested in static and dynamic axial compression (ASTM F2077), static and dynamic compression shear (ASTM F2077), and subsidence (ASTM F2267). Finite Element Analysis and mechanical testing which included oblique compressive shear (ASTM F2077) and subsidence (ASTM F2267) demonstrated substantial equivalence to the predicate devices. #### 9. Conclusion Based on the indications for use, technological characteristics, performance testing and comparison to the predicates, the Leva™ Spacer System has been shown to be substantially equivalent to the predicate devices identified in this submission, and does not present any new issues of safety or effectiveness.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...