CLYDESDALE SPINAL SYSTEM

K132897 · Medtronic Sofamor Danek, Inc. · MAX · Dec 11, 2013 · Orthopedic

Device Facts

Record IDK132897
Device NameCLYDESDALE SPINAL SYSTEM
ApplicantMedtronic Sofamor Danek, Inc.
Product CodeMAX · Orthopedic
Decision DateDec 11, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Intended Use

The CLYDESDALE® Spinal System is designed to be used with autogenous bone graft to facilitate interbody fusion and is intended for use with supplemental fixation systems cleared for use in the lumbar spine. The CLYDESDALE® Spinal System is used for patients diagnosed with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via a minimally invasive lateral approach.

Device Story

The CLYDESDALE® Spinal System is a PEEK intervertebral body fusion cage designed for lumbar interbody fusion. The device features a hollow geometry to be packed with autogenous bone graft to facilitate fusion. It includes tantalum markers for radiographic visualization. The system is implanted via a minimally invasive lateral approach between two lumbar or lumbosacral vertebral bodies to provide support and correction. It is intended for use in conjunction with supplemental fixation systems. The device is used by surgeons in a clinical/OR setting. The system provides structural support to the intervertebral space during the fusion process, potentially benefiting patients by stabilizing the spinal segment and alleviating discogenic back pain.

Clinical Evidence

No clinical testing was required; bench testing only.

Technological Characteristics

PEEK intervertebral cages with tantalum markers. Mechanical testing performed per ASTM F2077 (Static Compression-Shear and Compression-Shear Fatigue).

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1, including those with up to Grade 1 Spondylolisthesis or retrolisthesis, who have failed six months of non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## CLYDESDALE® Spinal System 510(k) Summary December 10, 2013 # DEC 1 1 2013 I. COMPANY: II. CONTACT: Medtronic Sofamor Danek USA, Inc. 1800 Pyramid Place Memphis, Tennessee 38132 Mr. Brad Sheals, MS Senior Regulatory Affairs Specialist Telephone: (901) 396-3133 Fax: (901) 346-9738 #### III. PROPRIETARY TRADE NAME: CLYDESDALE® Spinal System IV. CLASSIFICATION NAME: COMMON NAME: CLASS: PRODUCT CODE: REGULATION NUMBER: Intervertebral Fusion with Bone Graft, Lumbar Intervertebral Body Fusion Device II . MAX 21 CFR 888.3080 #### V. PRODUCT DESCRIPTION: The CLYDESDALE® Spinal System is intended to help provide support in the intervertebral body space during fusion of vertebral bodies in the lumbar spine. This system is intended to be used with supplemental fixation. The CLYDESDALE® Spinal System consists of PEEK cages of various widths and heights, which include tantalum markers. These devices can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. The hollow geometry of the implants allow them to be packed with autogenous bone graft. {1}------------------------------------------------ The purpose of this 510(k) submission includes additional implant sizes and trials for the CLYDESDALE® Spinal System. The subject implants will be offered in 18mm, 22mm, 26mm widths, 0°, 8°, 12°, 18° of lordosis and 8mm, 10mm, 12mm, 16mm. 18mm. 20mm heights, and in the same lengths as the predicate CLYDESDALE® Spinal System, K100175 (S.E. 06/02/2010). ### VI. INDICATIONS FOR USE: The CLYDESDALE® Spinal System is designed to be used with autogenous bone graft to facilitate interbody fusion and is intended for use with supplemental fixation systems cleared for use in the lumbar spine. The CLYDESDALE® Spinal System is used for patients diagnosed with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via a minimally invasive lateral approach. ## VII. SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS: The purpose of this Traditional 510(k) submission is to seek clearance of additional implant sizes and trials into the CLYDESDALE® Spinal System family. The design, materials and fundamental technology are the same as the current CLYDESDALE® Spinal System. ## VIII. IDENTIFICATION OF LEGALLY MARKETED DEVICES USED TO CLAIM SUBSTANTIAL EQUIVALENCE: The design features and indications for use for the subject CLYDESDALE® Spinal System are substantially equivalent to the following predicates: - CLYDESDALE® Spinal System, K100175, (S.E. 06/02/2010) - 트 PERIMETERTM Spinal System, K090353, (S.E. 09/29/2009) - CAPSTONE CONTROLTM Spinal System, K120368 (S.E. 04/09/2012) D - . CAPSTONE® Spinal System K073291, (S.E. 04/24/2008) - 발 CoRoent® System K071795. (S.E. 12/04/2007) {2}------------------------------------------------ #### IX. DISCUSSION OF NON-CLINICAL TESTING: An assessment of the subject CLYDESDALE@ Spinal System was completed in accordance with Medtronic design control processes. Mechanical Testing was conducted according to FDA's guidance document, "Class II Special Controls Guidance Document: Intervertebral Body Fusion Device". For a determination of substantial equivalence, the following nonclinical mechanical tests were performed: | Tests Performed Applicable Standards | | |---------------------------------------|-------------------------------------| | Static Compression- Shear | F2077 (Test Methods for )<br>ASTIM | | Compression- Shear Fatigue | Intervertebral Body Fusion Devices) | The subject implants met the acceptance criteria. Based on the results, the subject implants demonstrated that they are as safe, as effective and perform as well as the predicate device(s). # X. DISCUSSION OF CLINICAL TESTING: No clinical testing was required. # XI. CONCLUSIONS DRAWN FROM THE NON-CLINICAL TESTING: Based on the risk assessment, test results and additional supporting documentation provided in this premarket notification, Medtronic believes the subject CLYDESDALE® Spinal System demonstrates substantial equivalence to the listed predicate devices. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a bird or a stylized human figure. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G6 Silver Spring, MD 20993-0002 December 11, 2013 Medtronic Sofamor Danek USA, Incorporated Brad Sheals, MS Senior Regulatory Affairs Specialist 1800 Pyramid Place Memphis. Tennessee 38132 Re: K132897 Trade/Device Name: CLYDESDALE® Spinal System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: September 13, 2013 Received: September 16, 2013 Dear Mr. Sheals: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {4}------------------------------------------------ #### Page 2 - Brad Sheals, MS forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable. the electronic product radiation control provisions (Sections 331-542 of the Act): 21 CFR 1000-1050. 11 you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, picase note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincercly vours. # Ronald P. Jean -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 510(k) Number (if known): K132897 Device Name: CLYDESDALE® Spinal System Indications for Use: The CLYDESDALE® Spinal System is designed to be used with autogenous bone graft to facilitate interbody fusion and is intended for use with supplemental fixation systems cleared for use in the lumbar spine. The CLYDESDALE® Spinal System is used for patients diagnosed with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via a minimally invasive lateral approach. Prescription Use ਮ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Anton E. Dmitriev, PhD Division of Orthopedio Devices Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...