SMART SCREW ACL MODELS 237020, 237025, 237030, 238020, 238025, 238030, 239020, 239025, 239030

K993073 · Bionx Implants, Ltd. · MAI · Dec 6, 1999 · Orthopedic

Device Facts

Record IDK993073
Device NameSMART SCREW ACL MODELS 237020, 237025, 237030, 238020, 238025, 238030, 239020, 239025, 239030
ApplicantBionx Implants, Ltd.
Product CodeMAI · Orthopedic
Decision DateDec 6, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Intended Use

The SmartScrew ACL™ is intended for use in interference fixation of bone-patellar tendon - bone and soft tissue grafts in anterior and posterior cruciate ligament reconstructions.

Device Story

SmartScrew ACL is a fully threaded, cannulated bioabsorbable interference screw used in orthopedic surgery for ACL and PCL reconstructions. It secures bone-patellar tendon-bone or soft tissue grafts within the bone tunnel. The device is manufactured in diameters of 7.0, 8.0, and 9.0mm and lengths of 20, 25, and 30mm. It is intended for use by surgeons in a clinical or hospital setting. By providing secure graft fixation, the screw facilitates ligament reconstruction and healing. The device is made of poly-L/D-lactide copolymer, which is designed to be bioabsorbable, eliminating the need for secondary removal surgery.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Fully threaded, cannulated interference screw with tapered head and tip. Material: poly-L/D-lactide copolymer. Dimensions: 7.0-9.0mm diameter, 20-30mm length. Mechanical fixation principle.

Indications for Use

Indicated for interference fixation of bone-patellar tendon-bone and soft tissue grafts in anterior and posterior cruciate ligament reconstructions. Contraindicated for cases inappropriate for metallic screw fixation, active/potential infection, or patients unable to cooperate (e.g., alcoholism).

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ DEC - 6 1999 # 510(k) Summary Bionx Implants Inc. SmartScrew ACL™ Submitter´s Name, Address, Telephone Number, and Contact Person Bionx Implants, Inc. 1777 Sentry Parkway West Gwynedd Hall, Suite 400 Bluebell, PA 19422 | Contacts: | Gerard S. Carlozzi<br>President and Chief Operating Officer | |-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------| | | Phone: (215) 643-5000 | | | Facsimile: (215) 653-0984 | | | Bionx Implants Ltd. | | | Tuija Annala<br>Regulatory Affairs Assistant | | | P.O.Box 3 | | | FIN-33721 Tampere | | | Finland | | | Phone: 358-3-316 5679 | | | Facsimile: 358-3-316 5688 | | Date prepared: | August 24th, 1999 | | Name of the device: | | | A. Trade or Proprietary Name: | SmartScrew ACLTM | | B. Common Name: | Bioabsorbable Interference Screw | | C. Classification Name: | Biodegradable fixation fastener, bone and soft tissue | | D. Device Product Code: | HWC and MAI | | Predicate Devices: | | | DePuy, Inc. | Phantom PLLA Interference Screw (K955733)<br>Resorbable Interference Screw (K981670)<br>Phantom Resorbable Interference Screw (K982662) | | Linvatec Corporation | BioScrew Fixation System (K933719, K952831, K973758) | | Instrument Makar, | Biologically Quiet Interference Screw (K943249) | | Biomet, Inc. | Arthrotek Interference Screw (K982496) | {1}------------------------------------------------ K095073 2 of 2 | Arthrex, Inc. | The Arthrex Bio-Interference Screw (K971358) | |--------------------------|-----------------------------------------------------------| | Tornier S.A | Phusis Absorbable Interference Screw (K970829) | | Acufex Microsurgical,Inc | Acufex Endofix Absorbable Interference Screw<br>(K954246) | | Smith and Nephew, Inc. | Bioabsorbable Interference Screw (K984320) | | Sulzer Orthopedics, Inc. | Sysorb Interference Screw (983592) | ### Intended Use: The SmartScrew ACL™ is intended for use in interference fixation of bone-patellar tendon - bone and soft tissue grafts in anterior and posterior cruciate ligament reconstructions. ## Device Description: The SmartScrew ACL™ a fully threaded, cannulated interference screw with slightly tapered head and tip. SmartScrew ACL™ is provided with three different diameter, 7.0, 8.0 and 9.0mm and with three lengths, 20, 25 and 30mm and it is made of poly-L/D-lactide copolymer. ### Substantial Equivalence: The SmartScrew ACL™ has the same intended use, principles of operation and technological characteristics than predicate devices. In summary, the SmartScrew ACL™ is, in our opinion, substantially equivalent to the predicate devices. Furthermore, the minor technological differences between the SmartScrew ACL™ and the predicate devices do not raise any new issues of safety or effectiveness. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC -6 1999 Mrs. Tuija Annala Regulatory Affairs Assistant Bionx Implants Limited P.O. Box 3 FIN-33721 Tampere Finland Re: K993073 SmartScrew ACL™ Bioabsorbable Trade Name: Interference Screw Requlatory Class: II Product Codes: MAI and HWC Dated: August 23, 1999 Received: September 14, 1999 Dear Mrs. Annala: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ #### Page 2 - Mrs. Tuija Annala This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Neil RP Ogden Jámes E. Dillard Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### INDICATIONS FOR USE 510(K) Number (if known): _ K993073 Device Name: SmartScrew ACLTM Indications for Use: The SmartScrew ACL™ is intended for use in interference fixation of bone-patellar tendon - bone and soft tissue grafts in anterior and posterior cruciate ligament reconstructions. The SmartScrew ACL™ is not intended for use in and is contraindicated for 1) cruciate ligament reconstructions, which would not be appropriate for fixation with metallic screws and 2) situations, where intra-articular replacement is otherwise contraindicated, e.g., active or potential infection and where patient cooperation cannot be guaranteed (e.g., alcoholism). (Please do not write below this line - continue on another page if needed) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use OR Over-The-Counter Use (Per 21 CFR 801.109) Neo for (Division Sign-Off) Division of General Restorative Devices 510(k) Number _
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...