K230394 · Teknimed · MAI · May 12, 2023 · Orthopedic
Device Facts
Record ID
K230394
Device Name
EUROSCREW NG; EUROSCREW TCP NG
Applicant
Teknimed
Product Code
MAI · Orthopedic
Decision Date
May 12, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Intended Use
EUROSCREW® NG and EUROSCREW® TCP NG are intended for attaching ligaments or tendons to bone in orthopaedic surgical procedures.
Device Story
EUROSCREW® NG and EUROSCREW® TCP NG are bioabsorbable interference screws used for fixation of bone-tendon-bone and soft tissue grafts during knee ACL reconstruction. Devices are implanted by surgeons in a clinical setting to secure ligaments or tendons to bone. Screws provide mechanical fixation during the healing process. Performance is verified through torsion, torque, and pull-out testing. Devices are supplied sterile for single-patient use.
Clinical Evidence
Bench testing only. Performance testing included torsion, torque (insertion), and pull-out testing per ASTM F 2502, and in vitro degradation per ISO 13781. Biocompatibility evaluated per ISO 10993-1. Sterilization validated per ISO 11135-1 and ISO 10993-7.
Technological Characteristics
Bioabsorbable interference screws. Materials: Poly Lactic Acid (PLA) for EUROSCREW® NG; Tri calcium phosphate (beta-TCP) and PLA for EUROSCREW® TCP NG. Compliant with ASTM F2502. Sterilization: Ethylene Oxide (EtO) to SAL 10^-6. Form factor: Cannulated screw, various diameters (6-11mm) and lengths (20-35mm).
Indications for Use
Indicated for knee anterior cruciate ligament (ACL) reconstruction in ligamentoplasty procedures. Intended for use by qualified healthcare professionals (surgeons) in a surgical setting.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
Teknimed % Barry E Sands President and Founder RQMIS, Inc. 110 Haverhill Rd., Suite 524 Amesbury, Massachusetts 01913
Re: K230394
Trade/Device Name: Euroscrew® NG; Euroscrew® TCP NG Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: MAI, Dated: February 14, 2023 Received: February 14, 2023
Dear Barry Sands:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Yu-chieh Chiu - Digitally signed by Yu-chieh
Chiu-s Chiu -S Date: 2023.05.12 13:43:23 S -04'00
Yu-Chieh Chiu, Ph.D. Acting Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230394
Device Name EUROSCREW® NG EUROSCREW® TCP NG
Indications for Use (Describe)
EUROSCREW® NG and EUROSCREW® NG TCP screws are indicated in knee anterior cruciate ligament reconstruction in ligamentoplasty procedures.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(K) SUMMARY
#### Submitter information
| Date Prepared per 21 CFR<br>807.92(a)(1) | May 11, 2023 |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------|
| Applicant/Manufacturer | Teknimed<br>11-12 rue Apollo<br>Z.I. Montredon<br>L'Union<br>31240, France |
| Primary Contact | Barry Sands<br>President and Founder, RQMIS<br>Phone: (978) 358-7307<br>Email: regulatorysubmissions@rqmis.com |
| Device information | |
| Trade Name of Subject Device | EUROSCREW® NG & EUROSCREW® TCP NG |
| Common Name | Fastener, Fixation, Biodegradable, Soft Tissue |
| Device class | Class II |
MAI
#### Device Description
Product Code
Regulation Number
Predicate Device:
Teknimed's EUROSCREW® NG and EUROSCREW® TCP NG are bioabsorbable interference screws used in ligament surgery specifically designed for fixation of both bone-tendon-bone and soft tissue grafts. The screws are suitable and designed for ligamentoplasty surgical procedures specifically for knee anterior cruciate ligament reconstructions. Both are implantable screw devices in long term contact with bone and tissue grafts. Teknimed's EUROSCREW® NG and
888.3030 Single/multiple component metallic bone
Fixation appliances and accessories
LIGAFIX Interference Screw (K122228)
{4}------------------------------------------------
EUROSCREW® TCP NG are implant devices that are utilized in a surgical setting with a qualified healthcare professional (i.e., surgeons).
### Indications for Use
EUROSCREW® NG and EUROSCREW® TCP NG screws are indicated in knee anterior cruciate ligament reconstruction in ligamentoplasty procedures.
#### Intended Use
EUROSCREW® NG and EUROSCREW® TCP NG are intended for attaching ligaments or tendons to bone in orthopaedic surgical procedures.
### Intended Use/Indication for Use - Comparison of the Subject Devices with Predicate Device
The subject devices, EUROSCREW® NG and TCP NG, have equivalent intended use, indications for use, material composition to the predicate device, LIGAFIX Interference Screw (K122228).
# Technological Characteristics - Comparison with the Predicate Device, LIGAFIX Interference Screw (K122228)
The technological characteristics comparison demonstrated that the subject devices, EUROSCREW® NG and EUROSCREW® TCP NG, are equivalent to the previously cleared predicate device in terms of intended uses, designs, materials, and operational principles.
#### Basis of Substantial Equivalence
The substantial equivalence of both subject devices was determined per the FDA guidance document, "The 510(k) program: evaluating substantial equivalence in premarket Notification [510(k]]." Comparison with the predicate device (K12228) does not raise new questions of safety and effectiveness. The analysis of the technological characteristics, which include design, material composition and other device features, as defined in section 513(i)(1)(B) of the FD&C Act and 21 CFR 807.100(b)(2) (ii)(A), demonstrated that EUROSCREW® NG and EUROSCREW® TCP NG are substantially equivalent to the identified predicate device, LIGAFIX Interference Screw (K122228).
### Sterilization
EUROSCREW® NG and TCP NG will be supplied sterile. EUROSCREW® NG and EUROSCREW® TCP NG are delivered in double sterile packaging. Both subject devices are sterilized by ethylene oxide and are provided sterile for single patient use. The sterilizations are performed per ISO 11135-1 and ISO 10993-7 FDA recognized consensus standards. The sterilization validation demonstrated that both subject devices are effective at all points of load and achieve a sterility assurance level (SAL) of 10-6.
{5}------------------------------------------------
## Biocompatibility
Biocompatibility testing was performed on both Teknimed subject devices, and the test results demonstrated that EUROSCREW® NG and EUROSCREW® TCP NG are biocompatible. Biocompatibility of both devices were evaluated according to the FDA guidance document entitled, Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing within a Risk Management Process" issued on September 4, 2020. Based on this FDA guidance document, both the subject devices are considered implants which come in permanent contact (>30 days) with bone and tissue of the patient's body.
### Performance Testing
Both the subject devices, EUROSCREW® NG and EUROSCREW® TCP NG, were tested for torsion, torque (insertion) and pull-out as per ASTM F 2502, and for In Vitro Degradation as per ISO 13781. The company also performed Insertion Validation and Association screwdriver-screw tests to assess screw insertion criteria. Performance test reports confirmed that the difference between the predicate (LIGAFIX Interference Screw (K122228)) and subject devices, EUROSCREW® NG and TCP NG, do not raise any new questions of safety or effectiveness. Therefore, both subject devices are substantially equivalent to the predicate device identified throughout this submission.
| Device<br>Characteristics | Subject Devices | Predicate Device | Substantial<br>Equivalence<br>Discussion |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Products | EUROSCREW® NG<br>and TCP NG | LIGAFIX Interference Screw<br>(30/70) | N/A |
| Applicant | Teknimed | SBM | N/A |
| 510(k)<br>Number | K230394 | K122228 | N/A |
| Classification | Class II | Class II | Identical |
| Classification<br>Name | 888.3030<br>Single/multiple<br>component<br>metallic bone<br>Fixation appliances<br>and accessories | 888.3030 Single/multiple<br>component metallic bone<br>Fixation appliances and<br>accessories | Identical |
| Product Code | MAI | MAI | Identical |
| Indications for use | EUROSCREW® NG<br>and EUROSCREW®<br>TCP NG screws are<br>indicated in knee<br>anterior cruciate<br>ligament<br>reconstruction in<br>ligamentoplasty<br>procedures. | LIGAFIX/ComposiTCP is a<br>cannulated; sterile, single<br>use, resorbable interference<br>bone screw made of a<br>mixture of tri calcium<br>phosphate (beta-TCP) and<br>Poly Lactic Acid (PLA)<br>designed for' the<br>interference fixation of<br>grafts in anterior cruciate<br>ligament reconstruction. | Equivalent<br>Both Subject devices<br>and the predicate<br>device have equivalent<br>Indications for use. |
| Sterile | Ethylene Oxide<br>(EtO) | Unknown | Equivalent<br>Both Subject devices<br>meet ISO 10993-7<br>specification. The<br>Sterilization tests'<br>results for both subject<br>devices passed, and<br>reports are attached in<br>the original 510(K)<br>body. |
| Material | Euroscrew NG: Poly<br>Lactic Acid (PLA)<br>&<br>Euroscrew TCP NG<br>Tri calcium<br>phosphate (beta-<br>TCP) and Poly Lactic<br>Acid (PLA) | Tri calcium phosphate<br>(beta-TCP) and Poly Lactic<br>Acid (PLA) | Equivalent<br>Both subject devices<br>and the predicate<br>devices have<br>equivalent materials.<br>Both Subject devices<br>and the predicate<br>devices meet ASTM<br>F2502 specifications. |
| Available sizes | Ø 6 mm L 20 mm<br>Ø 7 mm L 24 mm<br>Ø 8 mm L 24 mm<br>Ø 9 mm L 24 mm<br>Ø 7 mm L 30 mm<br>Ø 8 mm L 30 mm<br>Ø 9 mm L 30 mm<br>Ø 10 mm L 30 mm<br>Ø 11 mm L 35 mm | Flat head<br>Ø 10 mm - L 30 mm<br>Ø 10 mm - L 35 mm<br>Ø 11 mm - L 30 mm<br>Ø 11 mm - L 35 mm<br>Round head<br>Ø 6 mm - L 20 mm<br>Ø 7 mm - L 20 mm<br>Ø 7 mm - L 25 mm<br>Ø 7 mm - L 30 mm<br>Ø 8 mm - L 20 mm<br>Ø 8 mm - L 25 mm<br>Ø 8 mm - L 30 mm | Equivalent<br>Size range is<br>equivalent to the<br>Predicate device. |
| | | | |
| | | Ø 8 mm - L 35 mm | |
| | | Ø 9 mm - L 20 mm | |
| | | Ø 9 mm - L 25 mm | |
| | | Ø 9 mm - L 30 mm | |
| | | Ø 9 mm - L 35 mm | |
| | | Ø 10 mm - L 25 mm | |
| | | Ø 10 mm - L 30 mm | |
| | | Ø 10 mm - L 33 mm | |
| | | Ø 10 mm - L 35 mm | |
| How Supplied | Resorbable<br>interference bone<br>screw, Single use<br>only | Resorbable<br>interference bone screw,<br>Single use only | Identical |
| Biocompatible | ISO 10993-1 | Biocompatible | Equivalent<br>Both Subject and<br>predicate devices<br>testing<br>protocols/reports for<br>the screws meet ISO<br>10993 specification. All<br>tests' results for the<br>subject device passed<br>and reports are<br>attached in the original<br>510(K) body |
Table 1: Substantial Equivalence Comparison Table between EUROSCREW®NG and TCP NG (Subject Devices) with LIGAFIX® Interference Screw (K122228)
{6}------------------------------------------------
{7}------------------------------------------------
### Substantial Equivalence Conclusion
EUROSCREW® NG and EUROSCREW® TCP NG have equivalent intended use, material comparison, design, operational principle and equivalent technological characteristics as the predicate device (K122228). The non-clinical performance test reports support the safety, effectiveness and performance of the subject devices and demonstrate that any difference in technological characteristics and material composition do not rase any new questions of safety and effectiveness. Therefore, EUROSCREW® NG and EUROSCREW® TCP NG are substantially equivalent to the identified predicate device, LIGAFIX Interference Screw (K122228).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.