K171141 · Arthrex, Inc. · MAI · May 4, 2017 · Orthopedic
Device Facts
Record ID
K171141
Device Name
Arthrex SwiveLock Anchors
Applicant
Arthrex, Inc.
Product Code
MAI · Orthopedic
Decision Date
May 4, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Arthrex SwiveLock Anchors are intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the following procedures: Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction. Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair/Tendon Repair and Bunionectomy. Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, and Illiotibial Band Tenodesis. Secondary fixation of ACL/PCL reconstruction or repair (4.75 - 5.5 SwiveLock only). Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction. Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondylitis repair. Hip: Capsular repair, acetabular labral repair.
Device Story
Two-component, knotless suture anchor; consists of eyelet and hollow anchor body; pre-mounted on driver with components separated on shaft. Used by surgeons in OR for soft tissue-to-bone fixation. Device facilitates secure attachment of tendons/ligaments to bone; aids in structural repair of joints. Modification involves eyelet design change from predicate.
Clinical Evidence
Bench testing only. Performance data includes pull-out, compressive load, and insertion mode of failure testing. Bacterial endotoxin testing performed per EP 2.6.14/USP <85> to confirm pyrogen limit compliance.
Technological Characteristics
Two-component knotless suture anchor; metallic bone fixation appliance. Includes eyelet and hollow anchor body. Pre-mounted on driver. Sterilization method not specified.
Indications for Use
Indicated for soft tissue-to-bone fixation in shoulder, foot/ankle, knee, hand/wrist, elbow, and hip orthopedic procedures requiring suture anchoring.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
K192441 — Arthrex SwiveLock Anchors · Arthrex, Inc. · Oct 17, 2019
K180768 — Arthrex SwiveLock Anchors · Arthrex, Inc. · Dec 14, 2018
K193503 — Arthrex SwiveLock Suture Anchor · Arthrex, Inc. · Jan 15, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 4, 2017
Arthrex Inc. Mr. David L. Rogers Project Manager, Regulatory Affairs 1370 Creekside Boulevard Naples, Florida 34108-1945
Re: K171141
Trade/Device Name: Arthrex SwiveLock Anchors Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: MAI, MBI Dated: April 14, 2017 Received: April 18, 2017
Dear Mr. Rogers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K171141
Device Name Arthrex SwiveLock Anchors
#### Indications for Use (Describe)
The Arthrex SwiveLock Anchors are intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the following procedures:
Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair/Tendon Repair and Bunionectomy.
Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, and Illiotibial Band Tenodesis. Secondary fixation of ACL/PCL reconstruction or repair (4.75 - 5.5 SwiveLock only).
Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction.
Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondylitis repair.
Hip: Capsular repair, acetabular labral repair.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary: K171141
| Date Prepared | April 14, 2017 |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter | Arthrex Inc.<br>1370 Creekside Boulevard<br>Naples, FL 34108-1945 |
| Contact Person | David L Rogers<br>Project Manager, Regulatory Affairs<br>1-239-643-5553, ext. 71924<br>david.rogers@arthrex.com |
| Name of Device | Arthrex SwiveLock Anchors |
| Common Name | Suture Anchor |
| Product Code | MAI, MBI |
| Classification Name | 21 CFR 888.3030: Single/multiple component metallic bone fixation appliances<br>and accessories |
| Regulatory Class | II |
| Predicate Device | K151342: Arthrex SwiveLock Anchors |
| Purpose of Submission | This Special 510(k) premarket notification is submitted to obtain clearance for the<br>modification to the eyelet of the Arthrex SwiveLock Suture Anchors cleared under<br>K151342. |
| Device Description | The Arthrex SwiveLock Anchor is a two-component, knotless suture anchor<br>composed of an eyelet and a hollow anchor body. The SwiveLock Anchor is pre-<br>mounted on a driver with the anchor body and eyelet physically separated on the<br>driver shaft. Arthrex suture may also be provided with the device. |
| Indications for Use | The Arthrex SwiveLock Anchors are intended for fixation of suture (soft tissue) to<br>bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the<br>following procedures:<br>Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair,<br>Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair,<br>Capsular Shift or Capsulolabral Reconstruction.Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon<br>Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction,<br>Metatarsal Ligament Repair/Tendon Repair and Bunionectomy.Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament<br>Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, and<br>Illiotibial Band Tenodesis. Secondary fixation of ACL/PCL reconstruction<br>or repair (4.75 - 5.5 SwiveLock only).Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial<br>Collateral Ligament Reconstruction.Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or<br>Radial Collateral Ligament Reconstruction, Lateral Epicondylitis repair.Hip: Capsular repair, acetabular labral repair. |
| Performance Data | Pull-out, compressive load, and insertion mode of failure testing was performed<br>to demonstrate that the SwiveLock Anchors with the proposed modified eyelet<br>are substantially equivalent to the predicate.<br>Bacterial endotoxin per EP 2.6.14/USP <85> was conducted to demonstrate that<br>the device meets pyrogen limit specifications. |
{4}------------------------------------------------
| Conclusion | The Arthrex SwiveLock Anchors are substantially equivalent to the predicate<br>device in which the basic design features and intended uses are the same. Any<br>differences between the proposed device and the predicate device are<br>considered minor and do not raise questions concerning safety or effectiveness. |
|------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Based on the indications for use, technological characteristics, and the summary<br>of data submitted, Arthrex Inc. has determined that the proposed device is<br>substantially equivalent to the currently marketed predicate device. |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.