GRYPHON BR ANCHOR WITH PERMACORD, GRYPHON BR DS ANCHOR WITH PERMACORD,GRYPHON PEEK ANCHOR WITH PERMACORD, GRYPHON PEEK

K141259 · Depuy Mitek, A Johnson & Johnson Company · MAI · Jun 10, 2014 · Orthopedic

Device Facts

Record IDK141259
Device NameGRYPHON BR ANCHOR WITH PERMACORD, GRYPHON BR DS ANCHOR WITH PERMACORD,GRYPHON PEEK ANCHOR WITH PERMACORD, GRYPHON PEEK
ApplicantDepuy Mitek, A Johnson & Johnson Company
Product CodeMAI · Orthopedic
Decision DateJun 10, 2014
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Intended Use

The GRYPHON BR Anchor is intended for: Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction Hip: Capsular Repair, Acetabular Labral Repair The GRYPHON PEEK Anchor is intended for: Shoulder: Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction Foot/Ankle: Lateral Stabilization, Medial Stabilization Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis Elbow: Ulnar or Radial Collateral Ligament Reconstruction Hip: Capsular Repair, Acetabular Labral Repair

Device Story

Suture anchor system for soft tissue-to-bone fixation; preloaded on disposable inserter assembly. Available in absorbable BR or non-absorbable PEEK materials with non-absorbable Permacord suture. Used by surgeons in clinical settings for orthopedic repairs (shoulder, knee, hip, elbow, foot/ankle). Device functions as mechanical fastener to secure suture to bone; provides stable fixation to facilitate tissue healing. Supplied sterile for single use.

Clinical Evidence

No clinical data. Bench testing only, including pull-out testing and in-vitro assessments, demonstrated performance equivalent to predicate devices.

Technological Characteristics

Materials: Absorbable BR or non-absorbable PEEK. Suture: Permacord (non-absorbable). Form factor: Suture anchor preloaded on disposable inserter. Sterilization: Sterile, single-use.

Indications for Use

Indicated for soft tissue-to-bone fixation in shoulder, foot/ankle, knee, elbow, and hip procedures, including rotator cuff, Bankart, SLAP, biceps tenodesis, ligament, and capsular repairs.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # JUN 1 0 2014 ・ : # SECTION 2 - 510(k) SUMMARY l Gryphon™ Anchor with Permacord™ | Submitter's<br>Name and<br>Address | DePuy Mitek<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767<br>Date Prepared: May 14, 2014 | |------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person | Julie Vafides<br>Regulatory Affairs Specialist<br>DePuy Mitek, Inc.<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767, USA<br>Telephone: 508-977-6645<br>Facsimile: 508-977-6911<br>e-mail: jvafides@its.jnj.com | | Name of<br>Medical Device | Proprietary Name: a) Gryphon™ BR Anchor with Permacord™<br>b) Gryphon™ PEEK Anchor with Permacord™<br>Classification Name: a) Single/multiple component metallic bone fixation appliances and accessories<br>b) Smooth or threaded metallic bone fixation fasteners<br>Common Name: Suture Anchor | | Substantial<br>Equivalence | The Gryphon™ Anchor with Permacord™ is substantially equivalent to:<br>■ K090124, K100012 Gryphon™ P BR Anchor with Orthocord®<br>■ K103712 Gryphon™ PEEK Anchor with Orthocord®<br>■ K133794 Healix Advance™ Anchor with Permacord™ | | Device<br>Classification | ➤ Gryphon™ BR Anchors with Permacord™ are classified as:<br>Single/multiple component metallic bone fixation appliances and accessories, classified as Class II, product code MAI, regulated under 21 CFR 888.3030.<br>➤ Gryphon™ PEEK Anchors with Permacord™ are classified as:<br>Smooth or threaded metallic bone fixation fasteners, classified as Class II, product code MBI, regulated under 21 CFR 888.3040. | {1}------------------------------------------------ | Device<br>Description | The Gryphon™ Anchor with Permacord™ is a suture anchor preloaded on a disposable<br>inserter assembly intended for fixation of suture to bone. Gryphon Anchors are<br>available in absorbable BR and non-absorbable PEEK materials. Permacord suture is<br>non-absorbable. The Gryphon Anchor with Permacord suture is supplied sterile and is<br>for single use only. | |----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Technological<br>Characteristics | The proposed Gryphon Anchors with Permacord suture have the same anchor<br>materials and design as predicate Gryphon Anchors (K090124, K100012, K103712).<br>The proposed device principal operation is the same as predicate Gryphon Anchors<br>(K090124, K100012, K103712). The Permacord suture is the same suture as<br>referenced in predicate Healix Advance ™ Anchor with Permacord™ (K133794). | | Indications for<br>Use | <b>The GRYPHON BR Anchor</b> is intended for: | | | Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps<br>Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair,<br>Capsular Shift or Capsulolabral Reconstruction | | | Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair | | | Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair,<br>Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis | | | Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament<br>Reconstruction | | | Hip: Capsular Repair, Acetabular Labral Repair | | | <b>The GRYPHON PEEK Anchor</b> is intended for: | | | Shoulder: Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-<br>Clavicular Separation Repair, Deltoid Repair, Capsular Shift or<br>Capsulolabral Reconstruction | | | Foot/Ankle: Lateral Stabilization, Medial Stabilization | | | Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair,<br>Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis | | | Elbow: Ulnar or Radial Collateral Ligament Reconstruction | | | Hip: Capsular Repair, Acetabular Labral Repair | 2 of 3 / {2}------------------------------------------------ . . | Non clinical<br>Testing | Verification activities were performed on the implant and / or its predicates. Testin<br>assessments include pull out testing and in-vitro testing. | |---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Safety and<br>Performance | Results of performance testing have demonstrated that the proposed devices are<br>suitable for their intended use.<br><br>Based on similarities in the indications for use, technological characteristics, and<br>performance in comparison to the predicate devices, the proposed Gryphon Anchor<br>with Permacord suture has shown to be substantially equivalent to the predicate<br>devices under the Federal Food, Drug and Cosmetic Act. | . : {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 10, 2014 DePuy Mitek, Incorporated Ms. Julie Vafides Regulatory Affairs Specialist 325 Paramount Drive Raynham, Massachusetts 02767 Re: K141259 Trade/Device Name: Gryphon™ BR and PEEK with Permacord™ Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: MAI, MBI Dated: May 14, 2014 Received: May 15, 2014 Dear Ms. Vafides: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device- {4}------------------------------------------------ ### Page 2 - Ms. Julie Vafides related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. .. : . Sincerely yours. ## Lori A. Wiggins for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use 510(k) Number (if known) K141259 #### Device Name Gryphon BR and PEEK with Permacord Indications for Use (Describe) The GRYPHON BR Anchor is intended for: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Sepair, Shoulder: Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair Foot/Ankle: Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Posterior Oblique Ligament Repair, Iliotibial - Band Tenodesis - Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction - Hip: Capsular Repair, Acetabular Labral Repair The GRYPHON PEEK Anchor is intended for: Shoulder: Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Sepair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction Foot/Ankle: Lateral Stabilization, Medial Stabilization Knee: Medial Collateral Ligament Repair. Lateral Ligament Repair. Posterior Oblique Ligament Repair. Nigilbial Band Tenodesis - Elbow: Ulnar or Radial Collateral Ligament Reconstruction - Hip: Capsular Repair, Acetabular Labral Repair Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. CHANGE ON FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR CONSTITU Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Casey L. Hanley, Ph.D.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...