K122869 · Depuy Mitek, A Johnson & Johnson Company · MAI · Dec 13, 2012 · Orthopedic
Device Facts
Record ID
K122869
Device Name
MILAGRO INTERFERENCE SCREWS
Applicant
Depuy Mitek, A Johnson & Johnson Company
Product Code
MAI · Orthopedic
Decision Date
Dec 13, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The DePuy Mitek MILAGRO BR Interference Screws are designed to attach soft tissues to bone in orthopedic surgical procedures for following indications: Shoulder: Proximal Biceps Tenodesis, Acromio-Clavicular Repair Elbow: Distal Biceps Tenodesis, Ulnar Collateral Ligament repair Knee: Collateral Ligament Repair
Device Story
Milagro Interference Screws are absorbable, tapered, cannulated, threaded fasteners used for interference fixation of soft tissue or bone-tendon grafts. Fabricated from a composite of absorbable Poly (lactide-co-glycolide) polymer and Tricalcium Phosphate (TCP). Used in orthopedic surgical procedures (shoulder, elbow, knee) to attach soft tissue to bone. Operated by surgeons using provided reusable accessories: modular drivers, ratchet handle, guidewires, tap, and notcher. Device provides mechanical fixation during the healing process. Benefit includes secure graft attachment using biocompatible, absorbable material.
Clinical Evidence
Bench testing only. Performance verification included insertion torque, anchor pullout strength at T=0, 3, 6, and 12 weeks of in-vitro physiological aging, and torque to failure testing.
Technological Characteristics
Absorbable, tapered, cannulated, threaded fastener. Material: Poly (lactide-co-glycolide) polymer and Tricalcium Phosphate (TCP) composite. Dimensions: 5mm and 6mm diameters, 12mm length. Accessories: Reusable modular drivers, ratchet handle, guidewires, tap, and notcher. Sterilization: Identical to previously cleared predicates.
Indications for Use
Indicated for patients requiring soft tissue-to-bone attachment in orthopedic procedures, specifically proximal biceps tenodesis, acromio-clavicular repair, distal biceps tenodesis, ulnar collateral ligament repair, and collateral ligament repair.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
K103831 — DEPUY MITEK MILAGRO INTERFERENCE SCREW MODEL 231803, 231804, 231807, 231809 · Depuy Mitek, A Johnson & Johnson Company · Jul 21, 2011
K142948 — BIOSURE REGENESORB interference Screw · Smith & Nephew, Inc. · Jan 13, 2015
K123362 — MILAGRO ADVANCE INTERFERENCE SCREW · Depuy Mitek, A Johnson & Johnson Company · Feb 1, 2013
K202535 — Arthrex FastThread Interference Screw · Arthrex, Inc. · Nov 18, 2020
K060830 — MILAGRO INTERFERENCE SCREW · Depuy Mitek, A Johnson & Johnson Company · Jun 15, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the DePuy Synthes logo. The logo consists of the word "DePuy" in bold black letters, with a black circle to the left of the word. Below the word "DePuy" is the text "Synthes" in a smaller font. To the right of the logo is the tagline "never stop moving" in a smaller font.
5, 6x12mm MILAGRO INTERFERENCE SCREWS
DEC =1-3-2012
#### ATTACHMENT 1
### 5 10(k) SUMMARY - DEPUY MITEK MILAGRO* INTERFERENCE SOREWS
#### SUBMITTER'S NAME AND ADDRESS
DePuy Mitek, Inc. a Johnson & Johnson company 325 Paramount Drive Raynham, MA 02767
#### CONTACT PERSON
Kristine Christo DePuy Mitek Regulatory Affairs
| TELEPHONE | 508-828-3359 |
|---------------|----------------------|
| FACSIMILE | 508-977-6911 |
| E-MAIL | KChristo@its.jnj.com |
| DATE PREPARED | September 18th, 2012 |
#### NAME OF MEDICAL DEVICE
CLASSIFICATION NAME Fastener, Fixation, Biodegradable, Soft Tissue
#### COMMON/USUAL NAME
Bone Anchor
#### PROPRIETARY NAME
DePuy Mitek Milagro® Interference Screws
:
#### SUBSTANTIAL EQUIVALENCE
The proposed 6, 7 and 8 x15mm DePuy Mitek Milagro® Interference Screws are substantially equivalent to the following devices.
- K120589, K103831, K060830, K032717:
Milagro Interference Screws
- · K051726, K041356, K020043:
Arthrex Tenodesis Screws
#### FDA PRODUCT CODE
MAI, HWC
#### DEVICE CLASSIFICATION
This type of fixation screw was originally classified as a Class II medical device by the Orthopedic Review Panel, regulated as 21 CFR 888.3040 Smooth or Threaded Metallic Bone Fixation Fastener.
{1}------------------------------------------------
Image /page/1/Figure/1 description: The image contains the logo for DePuy Synthes, a company specializing in medical devices. The logo features the company name in bold, with the tagline "never stop moving" in smaller font to the right. A black circle is positioned to the left of the company name. The overall image appears to be a cropped section of a larger document or advertisement.
5, 6x12mm MILAGRO INTERFERENCE SCREWS
#### ATTACHMENT 1
Continues ...
# 5 I 0(k) SUMMARY - DEPUY MITEK MILAGRO® INTERFERENCE SOREWS
#### DEVICE DESCRIPTION
The DePuy Mitek Milagro® Interference Screws are absorbable, tapered, cannulated, threaded fasteners for use in interference fixation of soft tissue grafts or bone-tendon grafts. The Interference Screw is made from a composite made of absorbable Poly (lactide-co-glycolide) polymer and Tricalcium Phosphate (TCP).
Modular Drivers, Ratchet Handle, Guidewires, a Tap and Notcher are provided separately as reusable accessories to assist in the placement of the MILAGRO BR Screw.
The proposed DePuy Mitek Milagro® Interference Screws are offered in diameters of 5, 6mm and in the length of 12mm.
#### INDICATIONS FOR USE
The DePuy Mitek MILAGRO BR Interference Screws are designed to attach soft tissues to bone in orthopedic surgical procedures for following indications: Shoulder: Proximal Biceps Tenodesis, Acromio-Clavicular Repair Elbow: Distal Biceps Tenodesis, Ulnar Collateral Ligament repair Knee: Collateral Ligament Repair
#### TECHNOLOGICAL CHARACTERSTICS
The design specifications of the proposed DePuy Mitek Milagro® Interference Screws are substantially equivalent to the existing DePuy Mitek Milagro® Interference Screws cleared under 510(k) K120589, K103831, K060830 and K032717 except that the proposed DePuy Mitek Milagro® Interference Screws are smaller in size. Technological characteristics including design construct, and indications are similar to the predicate devices. There have been no changes to the material used to manufacture the product, packaging, and or to the sterilization processes, sterilization contractors; these are identical to that used for the currently marketed DePuy Mitek Milagro® Interference Screw (previously cleared by the FDA under the premarket notification K032717, K060830, K103831 and K120589).
#### NONCLINICAL TESTING
Product Design Verification activities, such as, Insertion Torque, Anchor Pullout (at T=0, 3, 6 and 12 week in-vitro physiological aging), and Torque to Failure were performed on the implant.
#### SAFETY AND PERFORMANCE
Results of performance and safety testing have demonstrated that the nronosed device is substantially equivalent to the predicate devices.
Based on the indications for use, technological characteristics, and comparison to predicate devices, the proposed Small Size DePuy Mitek Milagro® Interference Screws have been shown to be substantially equivalent to predicate devices under the Federal Food, Drug and Cosmetic Act.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in a bold, sans-serif font. The text is black and the background is white. The image is a close-up of the text.
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is in all caps and appears to be in a sans-serif font.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-002
DePuv Mitek, A Johnson & Johnson Company % Ms. Kristine Christo Regulatory Affairs Manager 325 Paramount Drive Raynham, Massachusetts 02767
Letter dated: December 13, 2012
Re: K122869
Trade/Device Name: 5, 6x12mm Milagro Interference Screws Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MAI, HWC Dated: September 18, 2012 Received: September 19, 2012
Dear Ms. Christo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
#### Page 2 - Ms. Kristine Christo
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the DePuy Synthes logo. The logo includes the company name "DePuy" in a stylized font, along with the tagline "never stop moving". The logo also includes a small graphic element, possibly a stylized representation of a joint or bone structure, which is a common theme for medical device companies specializing in orthopedics.
#### ATTACHMENT 2
# INDICATIONS FOR USE
510(k) Number (if known): K122869
Device Names: 5, 6x12mm Milagro® Interference Screws
Indications for Use:
The DePuy Mitek MILAGRO BR Interference Screws are designed to attach soft tissues to bone in orthopedic surgical procedures for following indications:
Shoulder: Proximal Biceps Tenodesis, Acromio-Clavicular Repair Elbow: Distal Biceps Tenodesis, Ulnar Collateral Ligament repair
Knee: Collateral Ligament Repair
Prescription Use AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
> Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of _ |
# Casey Hanley
For Division of Orthopaedic Devices
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.