K121890 · Linvatec Corporation D/B/A Conmed Linvatec · MAI · Aug 22, 2012 · Orthopedic
Device Facts
Record ID
K121890
Device Name
GENESYS PRESSFT SUTURE ANCHOR
Applicant
Linvatec Corporation D/B/A Conmed Linvatec
Product Code
MAI · Orthopedic
Decision Date
Aug 22, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The GENESYS PressFT Suture Anchor is intended to reattach soft tissue to bone in orthopedic surgical procedures. The system may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor system thereby stabilizes the damaged soft tissue, in conjunction with appropriate postoperative immobilization, throughout the healing period. Additional indications include acetabular labral repair and capsular repair in the hip.
Device Story
GENESYS PressFT Suture Anchor assists surgeons in reattaching soft tissue to bone during orthopedic procedures (arthroscopic or open). Device consists of bioabsorbable anchor (96L/4D copolymer + B-TCP) and Hi-Fi ultra-high molecular weight polyethylene sutures. Disposable inserter (stainless steel shaft, polycarbonate handle) used to implant anchor; removed post-implantation. Anchor stabilizes damaged soft tissue (ligaments, tendons, joint capsules) to bone during healing period, supported by postoperative immobilization. Surgeon operates device; output is mechanical fixation of tissue to bone.
Clinical Evidence
Bench testing only. Testing included fixation strength, cyclic loading, insertion, driver strength, biocompatibility, sterilization, shelf-life, and packaging qualifications.
Indicated for patients requiring soft tissue-to-bone reattachment in orthopedic surgical procedures, including arthroscopic or open procedures. Specific applications include repair of ligaments, tendons, joint capsules, acetabular labrum, and hip capsule.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
K200341 — Arthrex FiberTak Suture Anchor · Arthrex, Inc. · Mar 12, 2020
K170501 — CrossFT Knotless Biocomposite Suture Anchor with Disposable Driver · Conmed Corporation · Mar 17, 2017
K192032 — Fixone Biocomposite Small Anchor · Aju Pharm Co., Ltd. · Mar 3, 2020
K041069 — SPIRALOK ANCHOR · Depuy Mitek, A Johnson & Johnson Company · Nov 9, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
# K121990
# 5. 510(k) Summary
AUG 22 2012
In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, ConMed Linvatec is hereby submitting the 510(k) Summary of Safety and Effectiveness for 510(k) Number
Date Prepared: June 26, 2012
### A. Submitter
ConMed Linvatec 11311 Concept Boulevard Largo, Florida 33773-4908 Registration Number: 1017294
### B. Company Contact
Lorna K. Linville Director, Requlatory Affairs Telephone: (727) 399-5396 Fax (727) 399-5264
#### C. Device Name
| Trade Name: | GENESYS PressFT <sup>™</sup> Suture Anchor |
|------------------------|------------------------------------------------|
| Common Name: | Bioabsorbable Suture Anchor |
| Classification Name: | Fastener, Fixation, Biodegradable, Soft tissue |
| Proposed Class/Device: | Class II |
| Product Code: | MAI |
| Regulation: | 21 CFR Part 888.3030 |
## D. Predicate/Legally Marketed Devices
| Device Name: | ConMed Linvatec NANOTM Suture Anchor |
|---------------|--------------------------------------|
| Company Name: | ConMed Linvatec |
| 510(k) #: | K112965 |
ConMed Linvatec CrossFT™-BC Suture Anchor Device Name: Company Name: ConMed Linvatec K101100 510(k) #:
#### E. Device Description
The ConMed Linvatec GENESYS PressFT™ Suture Anchor is a device that is used to assist the surgeon in reattaching soft tissue to bone. The device includes anchors manufactured of 96L/4D copolymer + B-TCP and one (1) or two (2) Hi-Fi® sutures manufactured of ultra-high molecular weight polyethylene. The device is bioabsorbable and is available in two sizes: 1) a 2.5mm diameter x 9.8mm length suture anchor that is
Page 1 of 2
{1}------------------------------------------------
# K121890
utilized with a 2.1mm drill bit, and 2) a 3.0mm diameter x 10.6mm length suture anchor that is utilized with a 2.6mm drill bit. A disposable inserter is included to implant the suture anchor.
The disposable inserter device has a stainless steel shaft with a polycarbonate handle that is provided sterile, for single use and is removed at the end of the repair leaving behind the suture anchor construct. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor system thereby stabilizes the damaged soft tissue, in conjunction with appropriate postoperative immobilization, throughout the healing period. Additional surgical instruments, including drill bits, guides and obturators, are Class I, non-sterile, reusable devices intended for transient use during orthopedic procedures.
## F. Testing
The verification and validation testing of the GENESYS PressFT™ Suture Anchor includes fixation strength, cyclic loading, insertion, driver strength, biocompatibility, sterilization, shelf-life, and packaging qualifications.
#### G. Intended Use / Indications
The GENESYS PressFT™ Suture Anchor is intended to reattach soft tissue to bone in orthopedic surgical procedures. The system may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor system thereby stabilizes the damaged soft tissue, in conjunction with appropriate postoperative immobilization, throughout the healing period. Additional indications include acetabular labral repair and capsular repair in the hip.
#### H. Conclusion
The GENESYS PressFT™ Suture Anchor is substantially equivalent in design, manufacturing materials, intended use, principles of operation, and technical characteristics to the ConMed Linvatec NANO™ Suture Anchor (K112965) and the ConMed Linvatec CrossFT™-BC Suture Anchor (K101100).
Page 2 of 2
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
ConMed Linvatec % Ms. Lorna K. Linville Director, Regulatory Affairs 11311 Concept Boulevard Largo, Florida 33773-4908
AUG 22222012
## Re: K121890
Trade/Device Name: Genesys PressFT Suture Anchor Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories.
Class II
Regulatory Class: Class II Product Code: MAI Dated: July 27, 2012 Received: July 30, 2012
Dear Ms. Linville:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
Page 2 - Ms. Lorna K. Linville
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
,
le
erson
Mark N. Melkerson Director Division of Surgical, Orthopedic & Restorative Devices Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
# 510(k) Number (if known): ≤ / 2 /8 9 6
GENESYS PressFT Suture Anchor Device Name:
Indications for Use:
The GENESYS PressFT Suture Anchor is intended to reattach soft tissue to bone in orthopedic surgical procedures. The system may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor system thereby stabilizes the damaged soft tissue in conjunction with appropriate postoperative immobilization, throughout the healing period. Additional indications include acetabular labral repair and capsular repair in the hip.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Arst
(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number k121890
01:
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.