DM Converter - TiN coated
K211891 · Medacta International S.A. · LZO · Jul 22, 2022 · Orthopedic
Device Facts
| Record ID | K211891 |
| Device Name | DM Converter - TiN coated |
| Applicant | Medacta International S.A. |
| Product Code | LZO · Orthopedic |
| Decision Date | Jul 22, 2022 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3353 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Medacta DM Converter is designed to be used in combination with the Mpact and Versafitcup CC Trio family cementless cups in total hip arthroplasty in primary or revision surgery. Total hip Arthroplasty is indicated in the following cases: · Severely painful and/or disabled joint as a result of arthritis, rheumatoid polyarthritis or congenital hip dysplasia • Avascular necrosis of the femoral head · Acute traumatic fracture of the femoral head or neck · Failure of previous hip surgery: joint reconstruction, internal fixation, arthroplasty, hip resurfacing replacement or total hip arthroplasty
Device Story
DM Converter - TiN coated is a modular acetabular component for total hip arthroplasty; replaces acetabulum in primary or revision surgery. Device consists of highly polished High Nitrogen Stainless Steel (ISO 5832-9) liner with external Titanium Nitride (TiN) coating. Inserted into acetabular shell (Mpact or Versafitcup CC Trio family); couples with mobile UHMWPE or HighCross (highly crosslinked UHMWPE) liners. Used in clinical settings by orthopedic surgeons. Provides articulation surface for femoral head. Benefits include restoration of joint function and pain relief. Device is a line extension of K131458, offering additional sizes and lateralization features.
Clinical Evidence
No clinical studies were conducted. Substantial equivalence is supported by non-clinical bench testing, including stability (ASTM F1820-13), fretting corrosion, range of motion (ROM) evaluation, wear testing (ISO 14242-1), abrasion resistance (ASTM 1978-18), and coating adhesion (ISO 20502).
Technological Characteristics
Materials: High Nitrogen Stainless Steel (ISO 5832-9) with external TiN coating. Modular acetabular liner design. Connectivity: None. Sterilization: Not specified. Software: None. Standards: ASTM F1820-13 (stability), ISO 14242-1 (wear), ASTM 1978-18 (abrasion), ISO 20502 (adhesion).
Indications for Use
Indicated for patients requiring total hip arthroplasty (primary or revision) due to painful/disabled joint from arthritis, rheumatoid polyarthritis, congenital hip dysplasia, avascular necrosis of femoral head, acute traumatic fracture of femoral head/neck, or failure of previous hip surgery (reconstruction, fixation, arthroplasty, resurfacing).
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
Predicate Devices
- Mpact DM Converter (K131458)
Reference Devices
- Mpact Double Mobility System (K143453)
- Versafitcup DM (K083116, K092265)
- OMNI TiN Coated Apex Knee System (K191765)
- Biocore9 Humeral Resurfacing System (K193122)
Related Devices
- K103518 — TRINITY ACETABULAR SYSTEM · Corin USA · Mar 9, 2011
- K071784 — DEPUY PINNACLE WITH GRIPTON ACETABULAR CUPS · DePuy Orthopaedics, Inc. · Jul 25, 2007
- K131458 — MPACT DM LINER CONVERTER · Medacta International · Oct 1, 2013
- K102423 — DEPUY PINNACLE ALTRX ACETABULAR LINERS · DePuy Orthopaedics, Inc. · Oct 29, 2010
- K103721 — MPACT ACETABULAR SYSTEM · Medacta International · Mar 21, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
July 22, 2022
Medacta International S.A. % Chris Lussier Senior Director, Quality and Regulatory Medacta USA 3973 Delp Street Memphis, Tennessee 38118
Re: K211891
Trade/Device Name: DM Converter - TiN coated Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II Product Code: LZO, MEH Dated: June 24, 2022 Received: June 27, 2022
Dear Chris Lussier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
Device Name DM Converter - TiN coated
| <b>Indications for Use (Describe)</b> | |
|---------------------------------------|--|
|---------------------------------------|--|
The Medacta DM Converter is designed to be used in combination with the Mpact and Versafitcup CC Trio family cementless cups in total hip arthroplasty in primary or revision surgery.
Total hip Arthroplasty is indicated in the following cases:
· Severely painful and/or disabled joint as a result of arthritis, rheumatoid polyarthritis or congenital hip dysplasia
• Avascular necrosis of the femoral head
· Acute traumatic fracture of the femoral head or neck
· Failure of previous hip surgery: joint reconstruction, internal fixation, arthroplasty, hip resurfacing replacement or total hip arthroplasty
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| × | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### I. Submitter
Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
Contact Person: Stefano Baj, Regulatory and Compliance Director, Medacta International SA Applicant Correspondent: Chris Lussier, Senior Director of Quality and Regulatory, Medacta USA Date Prepared: June 17, 2021 Date Revised: July 22, 2022
#### II. Device
| Device Proprietary Name: | DM Converter – TiN coated |
|--------------------------|--------------------------------------------------------------|
| Common or Usual Name: | Total Hip Prosthesis |
| Classification Name: | Hip joint metal/ceramic/polymer semi-constrained cemented or |
| | nonporous uncemented prosthesis |
| Primary Product Code: | LZO |
| Secondary Product Code: | MEH |
| Regulation Number: | 21 CFR 888.3353 |
| Device Classification: | II |
#### III. Predicate Device
Substantial equivalence is claimed to the following device:
- > Mpact DM Converter, K131458, Medacta International SA
In addition, the following reference devices are cited within the submission:
- A Mpact Double Mobility System, K143453, Medacta International SA
- > Versafitcup DM, K083116 and K092265, Medacta International SA
- OMNI TiN Coated Apex Knee™ System, K191765, OMNI Life Science, Inc. A
- > Biocore9 Humeral Resurfacing System, K193122, Biocore9, LLC.
#### IV. Device Description
The purpose of this submission is to gain clearance for the new DM Converter - TiN coated Implants and to extend the shell's compatibility of the already cleared DM Converter (K131458). Specifically, both the new DM Converter - TiN coated and the already cleared DM Converter (K131458) can be
{4}------------------------------------------------
coupled with Mpact (K103721, K122641, K132879 and K171966) and Versafitcup CC Trio Family shells (K103352 and K121911).
The DM Converter - TiN coated is a line extension to the DM Converter (K131458) to provide a larger product offering.
The DM Converter - TiN coated is a component of a total hip joint prosthesis that is used to replace the acetabulum in both primary or revision surgery. It is a highly polished metal liner made of High Nitrogen Stainless Steel according to ISO 5832-9 with a TiN coating on the external surface.
The DM Converter - TiN coated is inserted into the acetabular shell and can be coupled with a mobile liner: Medacta Double Mobility UHMWPE (ISO 5834-2 Type 1) liners (K083116 and K131458) or HighCross (highly crosslinked UHMWPE) liners (K092265 and K131458).
The DM Converter – TiN coated is available in 6 six sizes with a nomenclature that identifies the cup liner size (first letter) and the double mobility liner size (final three letters): D/DMB, E/DME, G/DMF, J/DMH and K/DML.
#### V. Indications for Use
The Medacta DM Converter is designed to be used in combination with the Mpact and Versafitcup CC Trio family cementless cups in total hip arthroplasty in primary or revision surgery.
Total hip Arthroplasty is indicated in the following cases:
- Severely painful and/or disabled joint as a result of arthrosis, traumatic arthritis, rheumatoid . polyarthritis or congenital hip dysplasia
- Avascular necrosis of the femoral head ●
- Acute traumatic fracture of the femoral head or neck
- Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, partial hip ● arthroplasty, hip resurfacing replacement or total hip arthroplasty
#### VI. Comparison of Technological Characteristics
The DM Converter - TiN coated and the predicate device, Mpact DM converter (K131458), share the following characteristics:
- sizes; ●
- indications for use ●
- shape and design (except for the lateralization and groove for coating limit); ●
- modular connection fixation;
- biocompatibility;
- device usage;
- sterility;
- shelf life; and ●
- packaging.
{5}------------------------------------------------
The subject DM Converter - TiN coated differs from the predicate device, Mpact DM converter (K131458), with respect to:
- lateralization of both articulating surface and edge;
- groove for coating limit; and ●
- . material and coating.
### Discussion
Medacta International SA has not made any change to the indications for use, device usage, device sizes biocompatibility, sterility, shelf life, and packaging of the subject devices respect to the predicate devices.
#### VII. Performance Data
Based on the risk analysis, design validation and performance testing were conducted to written protocols. The following validation and tests are being provided in support of the substantial equivalence determination:
### Non-Clinical Studies
- DESIGN VALIDATION
- O DM Converter TiN coated, Validation Workshop and Evaluation forms
- PERFORMANCE TESTING ●
- DM Converter TiN coated Stability test according to ASTM F1820-13 Standard Test o Method For Determining the Forces for Disassembly of Modular Acetabular Devices
- DM Converter TiN coated Fretting corrosion test o
- DM Converter TiN coated Jumping distance and ROM evaluation rationale o
- DM Converter TiN coated Wear test according to ISO 14242-1 Implants for surgery о Wear of total hip-joint prostheses - Part 1: Loading and displacement parameters for weartesting machines and corresponding environmental conditions for test [Including AMENDMENT 1 (2018)]
- o Taber test according to ASTM 1978-18 Standard Test Method for Measuring Abrasion Resistance of Metallic Thermal Spray Coatings by Using the Taber Abraser
- Adhesive shear strength through scratch testing to ISO 20502 Fine ceramics о (advanced ceramics, advanced technical ceramics) - Determination of adhesion of ceramic coatings by scratch testing
- Double Mobility Converter Implants Compatibility Extension Rationale O
- PYROGENICITY
- o Bacterial endotoxin test (LAL test) according to European Pharmacopoeia §2.6.14 (which is equivalent to USP chapter <85>)
- 0 Pyrogen test according to USP chapter <151> for pyrogenicity determination
- o The subject devices are not labeled as non-pyrogenic or pyrogen free.
- BIOCOMPATIBILITY assessment ●
{6}------------------------------------------------
K211891 Page 4 of 4
- SHELF-LIFE evaluation •
## Clinical Studies:
- . No clinical studies were conducted.
# VIII. Conclusion
The information provided above supports that the DM Converter – TiN coated is substantially equivalent to the predicate device.