Exactech Alteon Modular Dual Mobility (MDM) System

K190890 · Exactech, Inc. · LZO · Oct 22, 2019 · Orthopedic

Device Facts

Record IDK190890
Device NameExactech Alteon Modular Dual Mobility (MDM) System
ApplicantExactech, Inc.
Product CodeLZO · Orthopedic
Decision DateOct 22, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3353
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Exactech Alteon Modular Dual Mobility (MDM) System is a component of the Alteon Acetabular System. The Alteon MDM System is indicated for skeletally mature individuals undergoing surgery for hip replacement due to osteoarthritis, theumatoid arthritis, osteonecrosis or post-traumatic degenerative problems of the hip. The Alteon Modular Dual Mobility System is also potentially indicated for ankylosing spondylitis, congenital hip dysplasia and to restore mobility resulting from previous fusion. The Alteon MDM System is also indicated for use in primary or revision patients at a high risk of hip dislocation. The Alteon Acetabular System is intended for press-fit fixation.

Device Story

Modular hip prosthesis system; provides interface between acetabulum and femoral heads in total hip arthroplasty. Components include CoCr alloy liners and vitamin E-stabilized ultra-high molecular weight polyethylene (UHMWPE) inserts. Used in surgical settings by orthopedic surgeons to restore joint function. System utilizes press-fit fixation. Benefits include restoration of mobility and reduced risk of dislocation in primary or revision patients. Device is provided sterile for single use.

Clinical Evidence

No clinical data. Evidence consists of bench testing and engineering evaluations, including biocompatibility analysis, wear testing, fretting/corrosion testing, impingement testing, lever-out testing, axial push-out testing, torque-out testing, offset pull-out testing, and range of motion analysis. Pyrogen testing conducted per USP <161>, USP <85>, and ANSI/AAMI ST72.

Technological Characteristics

Modular acetabular system components. Materials: CoCr alloy liners and vitamin E-stabilized UHMWPE inserts. Fixation: Press-fit. Sterilization: Sterile, single-use. Compatible with acetabular shells (46-68mm OD) and femoral heads (22mm or 28mm ID). Conforms to recognized performance standards for total hip replacement.

Indications for Use

Indicated for skeletally mature individuals undergoing hip replacement surgery due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative hip problems, ankylosing spondylitis, congenital hip dysplasia, or to restore mobility after previous fusion. Indicated for primary or revision patients at high risk of hip dislocation.

Regulatory Classification

Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ October 22, 2019 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that. Exactech, Inc. Liz Howell Regulatory Specialist 2320 NW 66th Court Gainesville, Florida 32653 Re: K190890 Trade/Device Name: Exactech Alteon Modular Dual Mobility (MDM) System Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO Dated: September 17, 2019 Received: September 19, 2019 Dear Liz Howell: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. FOR Vesa Vuniqi Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) #### K190890 #### Device Name Exactech® Alteon® Modular Dual Mobility (MDM) System Indications for Use (Describe) The Exactech Alteon Modular Dual Mobility (MDM) System is a component of the Alteon Acetabular System. The Alteon MDM System is indicated for skeletally mature individuals undergoing surgery for hip replacement due to osteoarthritis, theumatoid arthritis, osteonecrosis or post-traumatic degenerative problems of the hip. The Alteon Modular Dual Mobility System is also potentially indicated for ankylosing spondylitis, congenital hip dysplasia and to restore mobility resulting from previous fusion. The Alteon MDM System is also indicated for use in primary or revision patients at a high risk of hip dislocation. The Alteon Acetabular System is intended for press-fit fixation. X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## Exactech® Alteon® Modular Dual Mobility (MDM) System Traditional 510(k) - 510(k) Summarv | Company: | Exactech®, Inc<br>2320 NW 66th Court<br>Gainesville, FL 32653 | |----------------------|------------------------------------------------------------------------------------------------------------------------------| | Date: | September 17, 2019 | | Contact Person: | Liz Howell<br>Regulatory Affairs Specialist | | | Phone: (352) 377-1140<br>Fax: (352) 378-2617 | | Proprietary Name: | Exactech® Alteon® Modular Dual Mobility (MDM) System | | Common Name: | Total Hip Prosthesis Acetabular Component | | Classification Name: | 21 CFR 888.3353, Hip Joint Metal/Ceramic/Polymer<br>Semi-Constrained Cemented or Nonporous Uncemented Prosthesis<br>Class II | | Product Code: | LZO | ## Legally Marketed Device to Which Substantial Equivalence Is Claimed: - Corin Trinity™ Dual Mobility System (K170359) ● ## Reference Devices - Exactech® Novation® and AcuMatch® E-HXL Acetabular Liners (K173583) ● - Exactech® Alteon® Acetabular Cups and Acetabular Liners (K182502) ● - Stryker Modular Dual Mobility (MDMTM) Liner and X3® Acetabular Insert (K103233) ● - Biomet G7 Dual Mobility System (K150522), E1TM Avantage™ Head (a.k.a. E1TM ● Active Articulation) (K101336) ## Device Description The Alteon® MDM System provides modular implants for use in total hip arthroplasty (THA) to restore patient joint function. The Exactech® Alteon® Modular Dual Mobility (MDM) System is intended for use in total hip replacement as an interface between the acetabulum and femoral heads. The Alteon® MDM liners and inserts are manufactured from CoCr alloy and Ultra-High Molecular Weight Polyethylene containing vitamin E (alpha tocopherol), respectively. The Alteon® MDM liners and inserts are each available in seven configurations. The Alteon® MDM insert inner diameter (femoral head compatibility) includes 22mm or 28mm, while the Alteon® MDM liners are compatible with acetabular shells with outer diameters 46-68mm. {4}------------------------------------------------ # Exactech® Alteon® Modular Dual Mobility (MDM) System Traditional 510(k) - 510(k) Summary ### Indications for Use The Exactech® Alteon® Modular Dual Mobility (MDM) System is a component of the Alteon® Acetabular System. The Alteon® MDM System is indicated for skeletally mature individuals undergoing surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis or post-traumatic degenerative problems of the hip. The Alteon® Modular Dual Mobility System is also potentially indicated for ankylosing spondylitis, congenital hip dysplasia and to restore mobility resulting from previous fusion. The Alteon® MDM System is also indicated for use in primary or revision patients at a high risk of hip dislocation. The Alteon® Acetabular System is intended for press-fit fixation. #### Summary of Technological Characteristics The rationale for substantial equivalence is based on consideration of the following device use and characteristics: - Indications for Use. The proposed Exactech® Alteon® Modular Dual Mobility System components and the predicate devices have the same or similar indications for use. - Materials. The proposed Exactech® Alteon® Modular Dual Mobility System components and the predicate devices are composed of the same or similar biocompatible materials. - Design Features. The proposed Exactech® Alteon® Modular Dual Mobility System components and the predicate devices have the same or similar design features. - . Dimensions. The proposed Exactech® Alteon® Modular Dual Mobility System components and the predicate devices are dimensionally equivalent or comparable. - Sterilization. The proposed Exactech® Alteon® Modular Dual Mobility System ● components and the predicate devices are provided sterile for single use only. - Performance Requirements. The proposed Exactech® Alteon® Modular Dual Mobility ● System components and the predicate devices conform to recognized performance standards for total hip replacement devices. #### Non-Clinical Testing The following biocompatibility testing/analysis, wear analyses, mechanical testing, and engineering evaluations were performed to demonstrate that the Alteon® MDM System performs as intended and is substantially equivalent to the identified predicate devices: - Biocompatibility Analysis ● - Wear Testing ● - Fretting/Corrosion Testing - Impingement Testing - Lever-Out Testing - Axial Push-Out Testing ● - Torque Out Testing ● - Offset Pull-Out Testing ● - Range of Motion Analysis ● {5}------------------------------------------------ ### Exactech® Alteon® Modular Dual Mobility (MDM) System Traditional 510(k) - 510(k) Summary Pyrogen testing was conducted in accordance with USP <161>, USP <85>, and ANSI/AAMI ST72 to ensure the proposed Alteon® MDM System meets recommended limits per FDA's Guidance Document Submission and Review of Sterility Information in Premarket (510(k)) Submission for Devices Labeled as Sterile. ## Substantial Equivalence Conclusion Based on consideration of indications for use, technological characteristics, biocompatibility of the proposed devices, and results of wear analyses, mechanical testing, and engineering evaluations, the Exactech® Alteon® Modular Dual Mobility System demonstrates substantial equivalence to referenced predicate devices. K190890 Page 3 of 3
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...