Noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis. Rheumatoid arthritis. Correction of functional deformity. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques. Revision of previously failed total hip arthroplasty. Dislocation risks. The E1™ Avantage™ Head is a single-use implant, intended for uncemented applications.
Device Story
Dual mobility acetabular implant system; consists of E1™ Active Articulation Head fitting over femoral modular head, articulating within acetabular metal shell. Used in primary and total revision hip surgeries; intended for uncemented applications. Implanted by orthopedic surgeons in clinical/hospital settings. Device utilizes Vitamin E-infused, crosslinked UHMWPE to prevent oxidative degradation and maintain mechanical strength. Provides two articulating surfaces to improve joint stability and reduce dislocation risk. Single-use device.
Clinical Evidence
No clinical data provided. Evidence based on bench testing, including wear testing, distraction testing of the bipolar portion, range of motion testing, and material characterization (Vitamin E concentration/distribution). Mechanical testing included small punch testing (ASTM F2183), tensile testing (ASTM D638), and accelerated aging (ASTM F2003).
Technological Characteristics
Dual mobility acetabular implant. Material: GUR1020/GUR1050 isostatically compression molded UHMWPE, crosslinked with 100 kGy gamma irradiation under argon, infused with alpha-tocopherol (Vitamin E), and gamma sterilized (25-40 kGy) in Argon. Dimensions: 44-66mm. Uncemented application. Standards: ASTM F2183 (small punch), ASTM D638 (tensile), ASTM F2003 (accelerated aging).
Indications for Use
Indicated for patients with noninflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, non-union, femoral neck/trochanteric fractures, failed total hip arthroplasty, or dislocation risks.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Restoration® ADM System X3® Acetabular Insert-Howmedica Osteonics (K093644)
Tri-Polar System-Biomet, Inc. (Kadi 990)
100 kGy E-Poly™ MaxRom™ Acetabular Liners-Biomet, Inc. (K070364)
100 kGv E-Poly Acetabular Liners, Additional Sizes-Biomet, Inc. (K070399)
RingLoc® 36mm Liners and Modular Femoral Heads-Biomet, Inc. (K032396)
Versafit Cup Double Mobility System-Medacta International (K083116)
E-Poly (Vitamin E) Acetabular Liners-Biomet, Inc. (K050327)
E1™ Antioxidant Infused Technology-Biomet, Inc. (K100048)
Submission Summary (Full Text)
{0}------------------------------------------------
K101336 (pg. 1 of 3)
Image /page/0/Picture/1 description: The image contains the text "BIOMET" in a stylized font at the top. Below that, the text "MANUFACTURING CORP." is printed in a simple, sans-serif font. The text is arranged in two lines, with "BIOMET" above and "MANUFACTURING CORP." below.
JAN 1 4 2011
# 510(k) Summary January 14, 2011 Preparation Date: Applicant/Sponsor: Biomet Manufacturing Corp. 56 East Bell Drive P.O. Box 587 Warsaw. Indiana 46581-0587 Establishment Registration: 1825034 Contact Person: Becky Earl Regulatory Specialist E1™ Avantage™ Head (a.k.a. E1™ Active Articulation) Proprietary Name: Common Name: Artificial Hip Replacement Components -- Acetabular Classification Name: LPH- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented (21 CFR 888.3358) OQG-Prosthesis, hip, semi-constrained, metal/polymer + additive, porous uncemented (888.3358) LZO-Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non-Porous, Uncemented (21 CFR 888.3353) KWY-Prosthesis, Hip, Hemi-Femoral, Metal/Polymer, Cemented (21 CFR 888.3390) Legally Marketed Devices To Which Substantial Equivalence Is Claimed: Restoration® ADM System-Howmedica Oseonics K072020 Restoration® ADM System X3® Acetabular Insert-Howmedica Osteonics K093644 Tri-Polar System-Biomet, Inc. Kadi 990 100 kGy E-Poly™ MaxRom™ Acetabular Liners-Biomet, Inc. K070364 K070399 100 kGv E-Poly Acetabular Liners, Additional Sizes-Biomet, Inc. RingLoc® 36mm Liners and Modular Femoral Heads-Biomet, Inc. K032396 Versafit Cup Double Mobility System-Medacta International K083116 E-Poly (Vitamin E) Acetabular Liners-Biomet, Inc. K050327 E1™ Antioxidant Infused Technology-Biomet, Inc. K100048
Mailing Address: P.O. Box 587 Warsaw, IN 46581-0587 Toll Free: 800.348.9500 Office: 574.267.6639 Main Fax: 574.267.8137 www.biomel.com
Shipping Address: 56 E. Bell Drive Warsaw, IN 46582
{1}------------------------------------------------
K101336 (pg. 2 of 3)
# 510(k) Summary E1™ Active Articulation Biomet Manufacturing Corp.
## Device Description:
The E1 Active Articulation™ belongs to the family of dual mobility acetabular implants; the presence of two articulating surfaces in the same joint device. The E1 Active Articulation™ Head fits over a femoral modular head, which articulates within the E1™ Head. The resultant assembly then articulates within the acetabular metal shell. The E1™ Head is designed to be used with several styles of acetabular shells that have been cleared in previous submissions: Ma Magnum™ (K042037), Magnum™ Tri-Spike (K062995), and Mra 38™ Flared Cups and Non-Flared Cups (K01110). The E1 Active Articulation™ Heads are available in sizes 44-66mm and are manufactured from 100 kGy E1™, which is the same 100 kGy E-Poly™ material cleared in K070364, The claims based on small punch testing for the K070364 E-Poly™ (E1™) material were cleared previously in K100048 and are applicable to the identical E-Poly™ (E1™) subject material. The E1™ Active Articulation is designed for both primary and total revision surgeries, where all device components associated with the wear couple are removed and replaced. The system is intended for uncemented applications.
## FDA CLEARED CLAIMS FOR E1™ ANTIOXIDANT INFUSED TECHNOLOGY2
## Claim 1:
E1™ Antioxidant Infused Technology prevents oxidative degradation of polyethylene. Environmental stress crack testing was conducted by cyclically loading GUR1020 and GUR1050 E1 ™ test specimens in an air atmosphere maintained at 80°C for 5 weeks. Testing was completed per the literature (Nabar, Sean, et al. Transactions of the 54th Annual Meeting of the ORS, Poster No. 1684). E1™ specimens showed no evidence of environmental stress cracking and infrared spectroscopy showed no detectable oxidation in the loaded or unloaded samples (oxidation indices <0.1). E1 samples were machined from either GUR1020 (E1 knee) or GUR1050 (E1 hip) isostatically compression molded UHMWPE crosslinked with 100 kGy gamma irradiation under argon, doped with a-tocopherol, and subsequently gamma sterilized (25-40 kGy) in Argon. Bench testing is not necessarily indicative of clinical performance.
## Claim 2:
E1™ Antioxidant Infused Technology protects polyethylene from oxidation and cracking during environmental stress crack testing. Environmental stress crack testing was conducted by cyclically loading test specimens in an air atmosphere maintained at 80℃ for 5 weeks per the literature (Nabar, Sean, et al. Transactions of the 54th Annual Meeting of the ORS, Poster No. 1684). GUR1050 E1 ™ specimens ran head to head with GUR1050 gamma sterilized (25-40kGy in argon) polyethylene and sequentially crosslinked and annealed polyethylene (GUR 1050 barstock, 33kGy gamma irradiated in air, annealed at 130C in air and repeated for a total dose of 99kGy and machined into final part geometry). GUR1020 E1™ specimens ran head to head with GUR1050 direct compression molded polyethylene that was gamma sterilized (25-40kGy) in argon. The E1™ material tested that showed no evidence of environmental stress cracking or fracture and no detectable oxidation indices <0.1) in the loaded and unloaded samples using infrared spectroscopy. Both gamma sterilized and sequentially crosslinked and annealed polyethylene showed evidence of increased oxidation and cracking or fracture during environmental stress crack testing. E1 samples were machined from either GUR1020 or GUR1050 isostatically compression molded UHMWPE, crosslinked with 100 kGy gamma irradiation under argon, infused with vitamin E, and subsequently gamma sterilized (25-40 kGy) in Argon. Bench testing is not necessarily indicative of clinical performance.
## Claim 3:
E1™ Antioxidant Infused Technology maintains the mechanical strength of conventional UHMWPE under small punch testing. Small punch testing per ASTM F2183 was conducted for the E1 ™ hip material and the E1™ knee material. The E1™ hip material was compared to GUR1050 gamma sterilized in argon isostatic compression molded (ICM) UHMWPE and the E1™ knee material was compared to GUR1050 gamma sterilized (25-40kGy) in argon direct compression molded (DCM) UHMWPE. The ultimate load for the E1™ hip material and the GUR1050 ICM material are 105±5.5N and 75.4±5.3N respectively. The ultimate load for the E1 ™ knee material and the DCM control material are 97.2±6.4N and 86.6±7.5N respectively. The E1™ materials had ultimate loads greater than that of the ICM and DCM control. These differences were statistically significant (p<0.001 for all comparisons). E1 samples were machined
{2}------------------------------------------------
K101336 (pg. 3 of
from either GUR1020 (E1 knee) or GUR1050 (E1 hip) ICM UHMWPE, crosslinked with 100 kGv gamma irradiation under argon, infused with vitamin E, and subsequently gamma sterilized (25-40 kGy) in Argon. Bench testing is not necessarily indicative of clinical performance.
#### Claim 4 :
E1™ Antioxidant Infused Technology maintains mechanical strength after accelerated aging. There was no significant decrease (P>0.05) in ultimate load , ultimate tensile strength, or vield strength after accelerated aging for either the E1™ hip or the E1™ knee material. Ultimate load was measured by small punch testing per ASTM F2183; ultimate tensile strength and yield strength were measured by tensile testing per ASTM D638; Accelerated aging was performed per ASTM F2003 (70°C and 5 alm of oxygen for 14 days). The ultimate load for the E1 ™ knee material before and after accelerated aging was 97.2±6.4N and 100.0±5.0N respectively. The ultimate tensile strength for the E1™ knee material before and after accelerated aging was 45.8±1.6 and 46.1±2.9 MPa respectively. The vield strength for the E1 ™ knee material before and after accelerated aging was 22,640,2 and 22,840,3 MPa resolectively, The ultimate load for the E1 ™ hip material before and after accelerated aging was 105.0±5.5N and 115.0±3.2N respectively. The ultimate tensile strength for the E1™ hip material before and after accelerated aging was 43t3 and 43t2 MPa respectively. The yield strength for the E1™ hip material before and after accelerated aging was 24.2±0.2 and 24.4±0.2 MPa respectively. E1 samples were machined from either GUR1020 (knee material) or GUR1050 (hip material) isostatically compression molded UHMWPE, crosslinked with 100 kGy gamma irradiation under argon, infused with vitamin E. and subsequently gamma sterilized (25-40 kGy) in Argon. Bench testing is not necessarily indicative of clinical performance.
- 1. Cleared through 510(k) K100048
- 2. Note: E1™ Antioxidant Infused Technology may be used interchangeably with any of the following: E1™ Antioxidant Infused Bearings, E1™ Antioxidant Infused Material, E1™ material, E1™ technology, E1 ™ bearings, E1 ™ liners, E1 ™ acetabular liners and E1 ™ tibial bearings.
## Indications for Use:
- 1. Noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis.
- 2. Rheumatoid arthritis.
- 3. Correction of functional deformity.
- 4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques.
- 5. Revision of previously failed total hip arthroplasty.
- 6. Dislocation risks.
The E1™ Active Articulation is a single-use implant, intended for uncemented applications.
## Summary of Technologies:
The E1™ Active Articulation Head has the same technological characteristics as the dual mobility predicates, with the exception that the UHMWPE material used is a 100kGy, Vitamin E infused polyethylene, which was previously cleared in K070364.
#### Non-Clinical Testing:
Wear, distraction testing of the bipolar portion (i.e., distraction of the UHMWPE from the femoral head), and Range of Motion testing were performed on the dual mobility subject device. The concentration and distribution of the Vitamin E in the E1™ Active Articulation Head material was determined. Results of the precinical testing performed are within the range of legally marketed predicates and indicate that the device is functional within its intended use.
#### Clinical Testing:
None provided as a basis for substantial equivalence.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of a bird-like figure with three lines extending from its head.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Biomet Manufacturing Corporation % Ms. Becky Earl Regulatory Specialist P.O. Box 587 Warsaw, Indiana 46581-0587
JAN 1 4 2011 Re: K101336 Trade/Device Name: El™ Avantage™ Head Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, OQG, LZO, KWY Dated: January 10, 2011 Received: January 11, 2011
Dear Ms. Earl:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical
{4}------------------------------------------------
## Page 2 - Ms. Becky Earl
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
A.D. B. 2h.
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
Device Name: E1™ Avantage™ Head_
Indications For Use:
- 1. Noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis.
- 2. Rheumatoid arthritis.
- 3. Correction of functional deformity.
- 4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques.
- 5. Revision of previously failed total hip arthroplasty.
- 6. Dislocation risks.
The E1™ Avantage™ Head is a single-use implant, intended for uncemented applications.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ NO_ (21 CFR 807 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
A. B. Rh f. nxn
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number k101336
Page 1 of 1