Libertas – Taper Uncemented Femoral Stem

K183684 · Maxx Orthopedics, Inc. · LZO · Aug 23, 2019 · Orthopedic

Device Facts

Record IDK183684
Device NameLibertas – Taper Uncemented Femoral Stem
ApplicantMaxx Orthopedics, Inc.
Product CodeLZO · Orthopedic
Decision DateAug 23, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3353
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Libertas™ Hip Replacement System is intended for use in total hip arthroplasty. Total hip arthroplasty is intended to provide patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to fix and support the components. Total hip replacement is indicated for the following conditions: - Non-inflammatory degenerative joint diseases including osteoarthritis, post traumatic arthritis and avascular necrosis. - Rheumatoid arthritis. - Congenital hip dysplasia. - Acute traumatic fracture of the femoral head or neck. - Certain cases of Ankylosis. - Dislocation of the hip. - Correction of functional deformity. - Revision of failed joint reconstruction or treatment. - Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur. NOTE: - The Modular Shell, Uncemented Stem and Taper uncemented stem are intended for press-fit, uncemented use only. - The Cemented stem is intended for cemented use only.

Device Story

Libertas™ Taper uncemented femoral stem is a hip joint prosthesis component for total hip arthroplasty. Device is fabricated from Ti6Al4V-ELI titanium alloy with a commercially pure titanium plasma-sprayed proximal coating. Available in standard and distally reduced designs with varying neck angles (132°/128°). Intended for press-fit, uncemented fixation in the femur. Used by orthopedic surgeons in clinical settings to replace damaged hip joint articulation, aiming to restore patient mobility and reduce pain. Device integrates with previously cleared Libertas femoral heads. Clinical benefit derived from mechanical replacement of the hip joint to address degenerative or traumatic conditions.

Clinical Evidence

Bench testing only. Performance evaluated via proximal fatigue (ISO 7206-6), distal fatigue (ISO 7206-4), range of motion (ISO 21535), axial pull-off (ISO 7206-10), fretting corrosion (ASTM F1875), ceramic head burst/torque testing (ASTM F1820), and Ti coating adhesion (ASTM F1044). Endotoxin testing confirmed manufacturing processes do not introduce contaminants.

Technological Characteristics

Material: Ti6Al4V-ELI titanium alloy; proximal coating: commercially pure Titanium (plasma spray). Design: Tapered uncemented femoral stem, press-fit. Standards: ISO 7206-6, ISO 7206-4, ISO 21535, ISO 7206-10, ASTM F1875, ASTM F1820, ASTM F1044. Sterilization: Not specified (implied standard medical device sterilization).

Indications for Use

Indicated for patients requiring total hip arthroplasty due to non-inflammatory degenerative joint diseases (osteoarthritis, post-traumatic arthritis, avascular necrosis), rheumatoid arthritis, congenital hip dysplasia, acute traumatic femoral head/neck fractures, ankylosis, hip dislocation, functional deformity, revision of failed reconstruction, or nonunion/fractures of the proximal femur, provided sufficient sound bone exists for component fixation.

Regulatory Classification

Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ August 23, 2019 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below that. Maxx Orthopedics Inc. Priscilla Herpai Regulatory Manager 2460 General Armistead Ave, Suite 100 Norristown, Pennsylvania 19403 Re: K183684 Trade/Device Name: Libertas - Taper Uncemented Femoral Stem Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO Dated: July 8, 2019 Received: July 26, 2019 Dear Priscilla Herpai: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, FOR Vesa Vuniqi Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) #### K183684 Device Name Libertas™ - Taper uncemented femoral stem #### Indications for Use (Describe) The Libertas™ Hip Replacement System is intended for use in total hip arthroplasty. Total hip arthroplasty is intended to provide patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to fix and support the components. Total hip replacement is indicated for the following conditions: - · Non-inflammatory degenerative joint diseases including osteoarthritis, post traumatic arthritis and avascular necrosis. - · Rheumatoid arthritis. - · Congenital hip dysplasia. - · Acute traumatic fracture of the femoral head or neck. - · Certain cases of Ankylosis. - Dislocation of the hip. - · Correction of functional deformity. - · Revision of failed joint reconstruction or treatment. - · Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur. #### NOTE: - · The Modular Shell, Uncemented Stem and Taper uncemented stem are intended for press-fit, uncemented use only. - · The Cemented stem is intended for cemented use only. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------|----------------------------------------------------------------------| | <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for Maxx Orthopedics. The word "maxx" is in bold, black letters, with a gray line underneath. Below the line, the word "orthopedics" is written in a smaller, black font. The logo is simple and modern. ## 510(k) Summary This 510(k) Summary is submitted in accordance with the requirements of 21 CFR Part 807, Section 92. ## Applicant Date of Summary: December 27, 2018 Maxx Orthopedics Inc. 2460 General Armistead Ave Suite 100, Norristown, PA 19403 USA | Contact Person | Alternate Contact: | |---------------------------------------|-------------------------------------------------------| | Priscilla Herpai | Gayathri Nair | | Regulatory Manager | Senior Manager – Regulatory Affairs/Quality Assurance | | Maxx Orthopedics, Inc. | Email: gayathri.nair@merillife.com | | Phone: +1 484.342.0092 x 507 | Cell: +91 9909033393 | | Email: priscilla.herpai@maxxortho.com | | ## Device information: | Proprietary Name: | LibertasTM – Taper Uncemented Femoral Stem | |----------------------|---------------------------------------------------------------------------------------------------------------------| | Common / Usual Name: | Hip Joint Prosthesis | | Classification name: | Hip joint metal/ceramic/polymer semi-constrained cemented or<br>nonporous uncemented prosthesis per 21 CFR 888.3353 | | Product Codes: | LZO | | Device Class: | Class II | ## Predicate Devices | Equivalent<br>device category | Manufacturer | Trade name | 510(k) | |-------------------------------|------------------------------------|----------------------------------------------------------------------|----------------------------------------------------------------------------------| | Predicate device | Biomet Manufacturing<br>Corp., USA | Taperloc® Complete Stems<br>Taperloc® Complete Microplasty<br>System | K120030,<br>K110400,<br>K103755,<br>K101086,<br>K062994,<br>K043537,<br>K921301, | {4}------------------------------------------------ | Equivalent<br>device category | Manufacturer | Trade name | 510(k) | |-------------------------------|--------------------------------------|---------------------------------------|---------| | Reference<br>Device | Maxx Orthopedics,<br>Inc, USA | Libertas™ – Hip<br>Replacement System | K180973 | | Reference<br>Device | Meril Healthcare Pvt.<br>Ltd., India | Latitud™ – Hip Replacement<br>System | K172857 | ## Device Description: Libertas™ Hip Replacement System cleared under K180973 consists of a modular acetabular cup system, femoral heads, femoral stems, and related accessories. The purpose of this 510(k) is to add Taper uncemented femoral stem under the Libertas™ Hip device family. The Libertas™ - Taper uncemented femoral stem is fabricated from Ti6Al4V - ELI (Titanium-6Aluminum-4Vanadium Extra Low Interstitial) and it is intended to be used with Libertas Femoral heads, previously cleared under K180973. The proximal portion of stem is coated with commercially pure Titanium with plasma spray method. Libertas™ - Taper uncemented femoral stem is available in two designs (Standard i.e. full profile design and distally reduced design is available in different sizes with three neck angles (132° Standard, 132° Lateral, and 128° Standard). ## Intended use/Indications: The Libertas™ Hip Replacement System is intended for use in total hip arthroplasty. Total hip arthroplasty is intended to provide patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to fix and support the components. Total hip replacement is indicated for the following conditions: - Non-inflammatory degenerative joint diseases including osteoarthritis, post traumatic arthritis . and avascular necrosis. - Rheumatoid arthritis. ● - . Congenital hip dysplasia. - Acute traumatic fracture of the femoral head or neck. ● - Certain cases of Ankylosis. - . Dislocation of the hip. - . Correction of functional deformity. - . Revision of failed joint reconstruction or treatment. - Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur. . NOTE: {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for Maxx Orthopedics. The logo consists of the word "maxx" in bold, black letters, with a gray curved line underneath. Below the line, the word "orthopedics" is written in a smaller, sans-serif font, also in black. - The Modular Shell, Uncemented Stem and Taper uncemented stem are intended for press-fit, . uncemented use only. - . The Cemented stem is intended for cemented use only. ### Comparison of technological characteristics: The subject device is substantially equivalent to the previously cleared predicate devices based on similarities in intended use and technological characteristics, and sterilization method. #### Non clinical Performance data: Non-clinical testing was conducted to evaluate device function/mechanical performance and to demonstrate substantial equivalence. - Proximal fatigue test (ISO 7206-6:2013) - Distal fatigue test (ISO 7206-4:2010) ● - . Range of motion test (ISO 21535-2007/Amd 1:2016) - Axial pull-off test (ISO 7206-10:2003) ● - Fretting corrosion (ASTM F1875-98; Reapproved 2014) ● - Evaluation of Ceramic femoral head includes Burst test, Post fatigue burst test, . Pull-off test (ISO 7206-10:2003) and Torque test (ASTM F1820-13) (Biolox® delta Modular femoral head and Uncemented Stem) - . Ti coating adhesion (shear) test (ASTM F1044-05; Reapproved 2017)e1 ASTM F2003-15, ISO 5834-3 and ISO 5834-2) Endotoxin testing has demonstrated that the manufacturing process does not introduce Endotoxin as a bi-product of the manufacturing and cleaning process. #### Conclusion Based on performance testing results and similarities in intended use, device design/technological characteristics, material, and sterilization method, the Libertas™ - Taper uncemented femoral stem is considered substantially equivalent to the previously cleared predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...