The PROFEMUR® RENAISSANCE® Classic Stems are indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; 2. inflammatory degenerative joint disease such as rheumatoid arthritis; 3. correction of functional deformity; and, 4. revision procedures where other treatments or devices have failed. The PROFEMUR® RENAISSANCE® Classic Stems are single use components, intended for use in conjunction with associated ceramic or metal femoral heads as part of uncemented total hip arthroplasty.
Device Story
Monolithic femoral hip stem for uncemented total hip arthroplasty; manufactured from forged titanium alloy (ASTM F620) with titanium plasma spray coating (ASTM F1580). Device provides reduced and standard proximal flare options and neutral/varus neck offsets. Features modified impaction slot oriented 90 degrees from predicate. Implanted by orthopedic surgeons in clinical settings. Functions as structural component in hip replacement; provides mechanical stability for femoral head attachment. Benefits patients via pain relief and improved hip function.
Clinical Evidence
No clinical data provided. Safety and effectiveness supported by nonclinical bench testing, including proximal and distal fatigue testing (ISO 7206-4, 6, 8) and range of motion evaluation (ISO 21535).
Technological Characteristics
Monolithic femoral stem; forged titanium alloy (ASTM F620); titanium plasma spray coating (ASTM F1580). Uncemented prosthesis. Dimensions: 7 reduced flare sizes (10-16), 9 standard flare sizes (10-18). Mechanical fixation via press-fit. No software or electronic components.
Indications for Use
Indicated for skeletally mature patients undergoing total hip arthroplasty for non-inflammatory/inflammatory degenerative joint disease (e.g., osteoarthritis, RA, avascular necrosis, ankylosis, protrusio acetabuli, painful hip dysplasia), functional deformity correction, or revision procedures.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
K141235 — PROFEMUR RENAISSANCE CLASSIC LONG NECK HIP STEMS · Microport Orthopedics, Inc. · Sep 12, 2014
K191632 — PROFEMUR TL2 Stems · Microport Orthopedics, Inc. · Oct 7, 2019
K964218 — PERFECTA PLASMA SPRAY HIP STEM · Wrightmedicaltechnologyinc · Jan 8, 1997
K123688 — PROFEMUR TL CLASSIC HIP STEM · Wrightmedicaltechnologyinc · Feb 8, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
K130984 (pg 1/3)
PROFEMUR® RENAISSANCE® Classic Stems Special 510(k) 510(k) Summary
Image /page/0/Picture/2 description: The image shows the word "WRIGHT." in a bold, sans-serif font. Below the word is a graphic of two angled lines that appear to be overlapping. The text and graphic are black against a white background.
#### 510(k) Summary
# MAY 2 4 2013
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the use of the PROFEMUR® RENAISSANCE® Classic Stems.
(a)(1) Submitted By:
Wright Medical Technology, Inc. 5677 Airline Rd Arlington, TN 38002 (901) 867-4146
Date:
Contact Person:
(a)(2) Proprietary Name of Modified Device:
Common Name:
Classification Name and Reference:
April 03, 2013
Yuan Li. PhD Regulatory Affairs Specialist II ·
PROFEMUR® RENAISSANCE® Classic Stems
Femoral Hip Stem
888.3353 LZO Hip joint metal/ceramic/polymer, cemented or non-porous, uncemented prosthesis Class II
888.3350 JDI Hip joint metal/polymer semiconstrained cemented prosthesis Class II
Subject Product Code and Panel Code:
(a)(3) Predicate Devices:
Orthopedics/87/JDI/LZO
PROFEMUR® RENAISSANCE® Hip Stem, K051995
{1}------------------------------------------------
PROFEMUR® RENAISSANCE® Classic Stems Special 510(k) 510(k) Summary
#### (a)(4) Device Description
The purpose of this submission is to add a monolithic femoral neck option to the PROFEMUR® RENAISSANCE® Hip System. The PROFEMUR® RENAISSANCE® Classic Stems are monolithic stems manufactured from a forged titanium alloy (ASTM F620) and designed for use in uncemented total hip arthroplasty. The PROFEMUR® RENAISSANCE® Classic Stems offer both reduced flare stems (7 sizes, 10-16), and standard flare stems (9 sizes, 10 - 18) and are coated with titanium plasma spray conforming to ASTM F1580, provided by plasma vendors cleared in K123688. The subject device offers two neck offset options, Standard (neutral) and Extended (varus), which are identical to the predicate stems used with modular necks cleared in K100866 and K091423; as well as two proximal flare options, Standard and Reduced. The stem geometry, neck length, and neck offset are designed to create options identical to those available with the predicate modular device.
The subject devices feature one minor additional change in that they have an impaction slot that is oriented 90 degrees from the impaction feature of the predicate device. This is viewed a minor design change, as the rounded slot is not a worse case compared to the predicate design. This modification to the impaction slot was cleared in K123434 and K123688 with similar devices. In addition, these modifications were evaluated by proximal and distal fatigue testing. These modifications do not affect the intended use or alter the fundamental technology of the device.
Instruments used with the system are identical to ones for the predicate system, except for the stem impactor which was reviewed in K123434 and K123688. The subject devices may also be implanted with instruments for alternative surgical approaches, as cleared in K121221 and K122382. Wright has previously provided validation for the cleaning/sterilization of instruments in previous reviews (K122218).
#### (a)(5) Indications for Use
The PROFEMUR® RENAISSANCE® Classic Stems are indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
- 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;
- 2. inflammatory degenerative joint disease such as rheumatoid arthritis;
- 3. correction of functional deformity; and,
- 4. revision procedures where other treatments or devices have failed.
The PROFEMUR® RENAISSANCE® Classic Stems are single use components, intended for use in conjunction with associated ceramic or metal femoral heads as part of uncemented total hip arthroplasty.
The indications for use can be also found in the package insert previously cleared in
Headquarters Wright Medical Technology, Inc. 5677 Airline Road Arlington, TN 38002 901.867.9971 phone www.wmt.com
Tab 007: Page 2 of 3
{2}------------------------------------------------
PROFEMUR® RENAISSANCE® Classic Stems Special 510(k) 510(k) Summary
Image /page/2/Picture/1 description: The image shows the word "WRIGHT." in large, bold, sans-serif font. Above the word "WRIGHT." is the text "K130984 (pg 3/3)". Below the word "WRIGHT." is a graphic of two overlapping shapes.
K123434 and K123688.
# (a)(6) Technological Characteristics of the Device
The indications for use of the PROFEMUR® RENAISSANCE® Classic Stems are identical to those for the predicate device (K051995). The subject devices are made from an identical titanium alloy (ASTM F620) and possess an identical titanium plasma spray coating (ASTM 1580) as the predicate device. The PROFEMUR® RENAISSANCE® Classic Stems are designed to provide identical range of sizes, neck lengths, and neck offsets as those provided by the modular predicate with the exception of featuring a monolithic neck design and a slightly modified impaction slot.
#### (b)(1) Nonclinical Testing
The PROFEMUR® RENAISSANCE® Classic Stems were evaluated by proximal and distal fatigue tests in accordance with ISO 7206-4, 6, and 8 and satisfied the acceptance criteria of each. Range of motion was evaluated in accordance with ISO 21535 and the subject device satisfied the acceptance criteria. The titanium plasma spray coating is applied according to WMT specification and is identical to that used in the predicate device (K051995).
# (b)(2) Clinical Testing
Clinical data was not provided for the subject devices.
#### (b)(3) Conclusions
The indications for use and fundamental scientific technology of the PROFEMUR® RENAISSANCE® Classic Stems are identical to those of the predicate device. The design features and materials of the subject devices are substantially equivalent to those of the predicate devices. The safety and effectiveness of the PROFEMUR® RENAISSANCE® Classic Stems is adequately supported by the substantial equivalence information, materials information, and nonclinical testing data provided within this premarket notification.
Headquarters Arlington, TN 38002 901.867.9971 phone Wright Medical Technology, Inc. 5677 Airline Road www.wmt.com
Tab 007: Page 3 of 3
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes representing human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 24, 2013
Wright Medical Technology, Incorporated % Dr. Yuan Li Regulatory Affairs Specialist II 5677 Airline Road Arlington, Tennessee 38002
Re: K130984
Trade/Device Name: PROFEMUR® RENAISSANCE® Classic Stems Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II Product Code: LZO, JDI Dated: April 25, 2013 Received: April 26, 2013
Dear Dr. Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
# Page 2 - Dr. Yuan Li
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Erin I. Keith
For
Mark N. Melkerson . Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K130984 (pg 1/1)
Device Name: PROFEMUR® RENAISSANCE® Classic Stems
Indications for Use:
The PROFEMUR® RENAISSANCE® Classic Stems are indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
- 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;
- 2. inflammatory degenerative joint disease such as rheumatoid arthritis;
- 3. correction of functional deformity; and,
- 4. revision procedures where other treatments or devices have failed.
The PROFEMUR® RENAISSANCE® Classic Stems are single use components, intended for use in conjunction with associated ceramic or metal femoral heads as part of uncemented total hip arthroplasty.
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Elizabeth - Brank -S
Division of Orthopedic Devices
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.