BERLIN EXTERNAL FIXATOR
Device Facts
| Record ID | K954053 |
|---|---|
| Device Name | BERLIN EXTERNAL FIXATOR |
| Applicant | Encore Orthopedics, Inc. |
| Product Code | LXT · Orthopedic |
| Decision Date | May 16, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3030 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The Berlin External Fixator is a wire connected ring or frame fixator. It can be used as a static fixator or dynamically applying distraction or compression. The Berlin Fixator has been designed as a modular construction system. Through the use of a few individual components, simple static fixation procedures can be achieved. The range of indications includes treatment of acute fractures and reconstructive surgery of all long bones.
Device Story
Modular wire-connected ring/frame external fixator; aluminum and stainless steel construction. Used for static fixation or dynamic distraction/compression of long bones. Operated by orthopedic surgeons in clinical/surgical settings. Provides mechanical stabilization for fracture healing and reconstructive procedures. Output is physical skeletal stabilization. Benefit: fracture management and bone reconstruction.
Clinical Evidence
Bench testing only. Static and dynamic compression testing performed on small, medium, and large diameter ring assemblies. Static load to failure: 600 N (Berlin and Ilizarov). Cyclic load testing at 300 N: Berlin assemblies survived 194,000–253,000 cycles vs. 346,000 cycles for Ilizarov.
Technological Characteristics
Modular ring/frame fixator; materials: aluminum and stainless steel. Mechanical fixation via wires. Static or dynamic distraction/compression capability.
Indications for Use
Indicated for patients requiring treatment of acute fractures and reconstructive surgery of all long bones.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
- Ilizarov device (Smith-Nephew Richards)